The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner. The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended for use by healthcare professionals. The Expression MR400 MRI Patient Monitoring System (Model MR400) provides monitoring for the following vital sign parameters: ECG, pulse oximetry (SpO2), non-invasive blood pressure (NIBP), and optionally, invasive blood pressure (IBP), carbon dioxide (CO2) and respiration rate, anesthetic agents, nitrous oxide (N2O), oxygen (02), and/or temperature.
Device Story
Multi-parameter patient monitoring system for MRI suites; monitors ECG, SpO2, NIBP, IBP, CO2, respiration, anesthetic agents, O2, N2O, and temperature. Inputs include physiological signals via wireless ECG/SpO2 modules and wired sensors (cuffs, probes, cannulas). System processes signals to display waveforms, numeric values, and alarms on a cart-mounted touchscreen. Provides synchronization signals for MRI scanners. Operated by healthcare professionals in MRI magnet rooms/suites. Output enables real-time patient status assessment and clinical decision-making during imaging. Benefits include continuous patient safety monitoring in high-field MRI environments.
Clinical Evidence
No clinical data was required. Substantial equivalence is supported by non-clinical verification, validation, and system-level performance testing, including hazard analysis and comparison against predicate specifications.
Technological Characteristics
Multi-parameter monitor with touchscreen cart. Wireless ECG/SpO2 modules (IP21 rated). Power: AC or internal battery. Parameters: ECG, SpO2, NIBP, IBP, CO2, anesthetic agents, O2, N2O, temperature. Connectivity: 10 user-selectable wireless networks. Software-based processing of physiological signals.
Indications for Use
Indicated for patients of all sizes (neonatal, infant, pediatric, adult, large adult) undergoing MRI procedures requiring vital sign monitoring and scanner synchronization. Intended for use by healthcare professionals.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
MRI Patient Monitoring System (Model 865214) (K124061)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 23, 2015
Invivo Corporation c/o Christine Trefethen Regulatory Manager Philips Medical Systems 3000 Minuteman Rd. Andover, MA 01801
Re: K152330
Trade/Device Name: Expression MR400 MRI Patient Monitoring System (Model MR400) Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: November 24, 2015 Received: November 25, 2015
Dear Christine Trefethen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152330
Page 1 of 1
Device Name
Expression MR400 MRI Patient Monitoring System (Model MR400)
Indications for Use (Describe)
The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.
The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended for use by healthcare professionals.
The Expression MR400 MRI Patient Monitoring System (Model MR400) provides monitoring for the following vital sign parameters: ECG, pulse oximetry (SpO2), non-invasive blood pressure (NIBP), and optionally, invasive blood pressure (IBP), carbon dioxide (CO2) and respiration rate, anesthetic agents, nitrous oxide (N2O), oxygen (02), and/or temperature.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10px;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 10px;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Invivo logo. The logo consists of a purple circle with the letter "i" inside, followed by the word "Invivo" in purple. Below the word "Invivo" is the text "A Philips Company" in gray.
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
### SPECIAL 510(K) SUMMARY
# SUBMITTER INFORMATION
| Establishment Name:<br>Establishment Address: | Invivo Corporation<br>12151 Research Parkway<br>Orlando, FL 32826 USA |
|-----------------------------------------------|------------------------------------------------------------------------------------------------|
| Company Phone: | (407) 252-0414 |
| Company Fax: | (407) 249-2022 |
| Person to contact<br>regarding questions: | Rusty Kelly<br>Sr. Manager Quality and Regulatory<br>(407)-252-0414<br>Rusty.Kelly@philips.com |
| Establishment<br>Registration Number: | 1051786 |
## DEVICE IDENTIFICATION
Date Summary Prepared:
| Trade name: | Expression MR400 MRI Patient Monitoring<br>System (Model MR400) |
|----------------------|-------------------------------------------------------------------------------------------------------|
| Common name: | MRI patient monitoring system |
| Classification name: | Cardiac monitor (including cardiotachometer and<br>rate alarm)<br>(21 CFR 870.2300, Product Code MWI) |
August 11, 2015
{4}------------------------------------------------
#### IDENTIFICATION OF LEGALLY MARKETED PREDICATE DEVICE
The Expression MR400 MRI Patient Monitoring System also referred to as the Expression MR400 or MR400, is substantially equivalent to the following predicate device:
| Predicate Device | Manufacturer | 510(k) No. | Clearance Date |
|-------------------------------------------------|--------------------|------------|----------------|
| MRI Patient Monitoring System<br>(Model 865214) | Invivo Corporation | K124061 | Feb 22, 2013 |
#### DEVICE DESCRIPTION
The Expression MR400 is a multi-parameter patient monitoring system used to monitor the vital signs of patients in an MRI magnet room and throughout an MRI suite. The MR400's software simultaneously processes and displays multiple parameters, waveforms, measurement numeric values and alarms. All patient information is provided on a wheeled cart which consists of a patient monitor with a touch screen display. The device is powered by either AC line power or its internal battery. The standard and optional parameters which can be monitored by the MR400 include:
- Electrocardiogram (ECG)
- Heart rate (HR, ECG and SPO2 derived) ●
- Blood oxygen saturation/pulse oximetry (SPO2)
- End-tidal and fractional inspired CO2 (EtCO2 and FiCO2) ●
- Invasive blood pressure (IBP) (P1 and P2) ●
- Anesthetic agents (AGENT) ●
- o Desflurane (DES)
- o Enflurane (ENF)
- o Halothane (HAL)
- o Isoflurane (ISO)
- o Sevoflurane (SEV)
- Fractional inspired O2 (FiO2), and end-tidal and fractional inspired N2O (EtN2O ● and FiN2O)
- Temperature (TEMP) ●
- Non-invasive blood pressure (NIBP)
- . Respiration rate (CO2-derived and/or bellows-derived)
{5}------------------------------------------------
The MR400 consists of the following key components:
- Cart with touchscreen display ●
- Wireless ECG module .
- Wireless SpO2 module ●
- Batteries for both the Cart and Wireless Modules ●
- Battery charger ●
- Power cord ●
Accessories to the MR400 are offered to accommodate various patient sizes-neonatal, infant, pediatric, adult, and large adult. Key accessories include:
- . ECG lead cables and electrodes
- SPO2 probes and clips ●
- NIBP cuffs and hoses ●
- IBP transducer
- Temperature probe and sheath ●
- Cannulas ●
- Water traps ●
- Chest pneumograph ●
- Expression IP5 (optional secondary display for control room w/printer) ●
### INDICATIONS FOR USE
The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.
The Expression MR400 MRI Patient Monitoring System (Model MR400) is intended for use by healthcare professionals.
The Expression MR400 MRI Patient Monitoring System (Model MR400) provides monitoring for the following vital sign parameters: ECG, pulse oximetry (SpO2), non-invasive blood pressure (NIBP), and optionally, invasive blood pressure (IBP), carbon dioxide (CO2) and respiration rate, anesthetic agents, nitrous oxide (N2O), oxygen (02), and/or temperature.
#### TECHNOLOGICAL CHARACTERISTICS
The modified device has the same fundamental scientific technology as the predicate device.
{6}------------------------------------------------
| Technological<br>Characteristic | Predicate: MRI Patient<br>Monitoring System<br>(K124061) | Modified Device: Expression<br>MR400 MRI Patient<br>Monitoring System | Comparison |
|--------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Cart | | | |
| Display | Detachable, located on<br>top front of display | Permanently installed, located on<br>the rear of the display | Substantially<br>Equivalent |
| Expression IP5 | Software is compatible<br>with the predicate device | Software was updated to be<br>compatible with the predicate<br>device and the Expression<br>MR400 | Substantially<br>Equivalent |
| Printer | Integrated on side of<br>display | No printer permanently installed<br>on display. | Substantially<br>Equivalent |
| Alarms Set Up,<br>Printer Set Up and<br>Monitor Set Up | Separate key on the<br>keypad which is visible to<br>the operator at all times | Collapsed within one Setup key<br>on the virtual keypad that is<br>visible to the operator at all times | Substantially<br>Equivalent |
| Power Supply | External | Integrated | Substantially<br>Equivalent |
| Internal Batteries | Four User-replaceable | Two Permanently installed and<br>two User-replaceable | Substantially<br>Equivalent |
| Alarms | Vital Sign Alarms | Added user adjustable alarms for<br>extreme bradycardia, extreme<br>tachycardia, apnea and<br>desaturation | Substantially<br>Equivalent |
| Non-Invasive Blood Pressure | | | |
| NIBP Connector | Connector ports are for<br>unique predicate<br>accessories | Accommodate locking<br>connectors unique to the<br>modified device accessories | Substantially<br>Equivalent |
| NIBP Hose | 3 meters in length | 5 meters in length | Substantially<br>Equivalent |
| NIBP Data<br>Collection &<br>Measurement | Performed in the<br>processing unit within the<br>cart. | Performed in the processing unit<br>within the cart using the<br>picoNIBP OEM Module | Substantially<br>Equivalent |
| CO2 | | | |
| CO2 Connector | Connector ports are for<br>unique predicate<br>accessories | Accommodate locking<br>connectors unique to the<br>modified device accessories | Substantially<br>Equivalent |
| CO2 Collection &<br>Measurement<br>CO2 cannula | Completed using<br>Excelitas CO2 Sensor<br>22 ft. in length | Completed using LoFloC5 Co2<br>Sensor<br>17ft. in length | Substantially<br>Equivalent |
| ECG | | | |
| ECG Cable Cover | Wire bundle enclosed in a<br>silicone foam trunk | Flat wire assembly enclosed in<br>Azote foam | Substantially<br>Equivalent |
| Wireless ECG<br>Module- communication | 5 user-selectable<br>networks | 10 user-selectable networks.<br>(only 5 per module, either 1-5 or<br>6-10) | Substantially<br>Equivalent |
| Wireless ECG<br>Module- battery | Operates on one (1) 3.7<br>VDC user-replaceable<br>lithium battery | Operates on two (2) 3.7 VDC<br>user-replaceable lithium batteries | Substantially<br>Equivalent |
| Wireless ECG<br>Module- housing | Plastic covering, single<br>battery slot, Ingress<br>Protection IPX1 | Plastic housing made of same<br>materials as predicate, supports<br>two batteries, has eject buttons<br>for batteries, improved Ingress<br>Protection rating of IP21 | Substantially<br>Equivalent |
| SPO2<br>Wireless SPO2<br>Module- communication | 5 user-selectable<br>networks | 10 user-selectable networks.<br>(only 5 per module, either 1-5 or<br>6-10) | Substantially<br>Equivalent |
| Wireless SPO2<br>Module- housing | Plastic covering, single<br>battery slot, Ingress<br>Protection IPX1 | Plastic housing made of same<br>materials as predicate, has eject<br>button for battery, improved<br>Ingress Protection rating of IP21 | Substantially<br>Equivalent |
{7}------------------------------------------------
#### SUMMARY OF NON-CLINICAL PERFORMANCE DATA
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and the specifications of the modified device. Results of the non-clinical testing demonstrate that the modified device operates as intended within the performance specifications and is substantially equivalent to the cleared device. The results do not raise issues regarding the safety and effectiveness of the device. Clinical data was not required to substantiate claims of safety and effectiveness. The modified device, the Expression MR400 MRI Patient Monitoring System, described in this submission is substantially equivalent to the cleared device, the MRI Patient Monitoring System (Model 865214).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.