Galil Medical's IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures. The IcePearl 2.1 CX and lceFORCE 2.1 CX Cryoablation Needles are indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, oncology, proctology, and urology. The systems are designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
Device Story
Cryoablation needles (2.1 mm) convert high-pressure gas into extreme cold for tissue destruction or warm gas for thawing; used in surgical settings (OR/clinic) by physicians. Needles feature sharp cutting tips, color-coded handles, and gas connectors; available in straight or 90° configurations. Operated via Visual-ICE Cryoablation System; software (v1.4.0) manages gas flow, monitors needle temperature, and provides real-time power feedback. Physician positions needle under imaging guidance (e.g., CT); system output (temperature/cycle status) informs ablation progress. Benefits include precise tissue necrosis for tumors, lesions, and palliative care.
Clinical Evidence
Bench testing only. Performance verified through dimensional testing, functional testing, freezing performance, and in vivo needle track ablation (characterizing depth of tissue necrosis). Testing confirmed needles meet defined specifications and perform equivalently to predicates when used with Visual-ICE Software Revision 1.4.0.
Technological Characteristics
Sterile, single-use 2.1 mm cryoablation needles. Principle: Joule-Thomson effect (high-pressure gas expansion). Features: reconfigured heat exchanger in handle, vacuum-insulated handle tubing. Connectivity: connects to Visual-ICE Cryoablation System. Software: v1.4.0. Sterilization: not specified.
Indications for Use
Indicated for cryoablative destruction of tissue in urology (prostate cancer, BPH), oncology (malignant/benign tumors, palliative), dermatology (skin cancers, warts, lesions, angiomas, keratoses), gynecology (malignant neoplasia, benign dysplasia), general surgery (rectal tumors, hemorrhoids, fibroadenoma), ENT (oral tumors, leukoplakia), thoracic surgery (excluding cardiac), and proctology (anal/rectal growths).
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 28, 2015
Galil Medical Ltd. % Ms. Lynne Davies Galil Medical Incorporated 4364 Round Lake Road Arden Hills, Minnesota 55112
Re: K152133
Trade/Device Name: IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: July 31, 2015 Received: July 28, 2015
Dear Ms. Davies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{1}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter G with a green dot inside, followed by the text "GALIL MEDICAL" in a bold, sans-serif font. Below the company name is the tagline "Leading Through Innovation" in a smaller, lighter font.
# K152133
# Indications for Use Statement
510(k) Number (if known): K152133
#### Device Name: IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles
#### Indications For Use:
Galil Medical's IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures. The IcePearl 2.1 CX and lceFORCE 2.1 CX Cryoablation Needles are indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, oncology, proctology, and urology. The systems are designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
The Galil Medical IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles have the following specific indications:
- . Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")
- . Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention)
- . Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cvsts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.)
- . Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia)
- . General surgery (palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma)
- . ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth)
- Thoracic surgery (with the exception of cardiac tissue)
- . Proctology (ablation of benign or malignant growths of the anus or rectum, and hemorrhoids)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter "G" with a green dot inside, followed by the text "GALIL MEDICAL" in a bold, sans-serif font. Below the logo is the tagline "Leading Through Innovation" in a smaller, lighter font.
# Section 5. 510(k) Summary
# 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Galil Medical Ltd.<br>Tavor 1 Building<br>Shaar Yokneam<br>Yokneam Industrial Park 20692<br>Israel |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person: | Ms. Lynne A Davies<br>Sr. Manager, Regulatory Affairs<br>Galil Medical Ltd.<br>Phone: 651-287-5098<br>Fax: 651-287-5099<br>Email: lynne.davies@galilmedical.com |
| Alternate Contact Person: | Ms. Amy E McKinney<br>VP, Regulatory Affairs and Quality<br>Galil Medical Ltd.<br>Phone: 651-287-5096<br>Fax: 651-287-5097<br>Email: amy.mckinney@galilmedical.com |
| Device Name: | IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation<br>Needles |
| Device Classification Name: | Cryosurgical unit and accessories (GEH)<br>21 CFR 878.4350 |
| Predicate Devices / Reference<br>510(k): | IceRod CX Cryoablation Needles (K121251 and<br>K140584)<br>IceEDGE 2.4 Cryoablation Needle (K111859)<br>Visual-ICE Cryoablation System (K113860, K123865,<br>K143564) |
| Date of Preparation: | July 28, 2015 |
# Device Description:
Galil Medical's IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles are sterile, single use, disposable components, when used in conjunction with Galil Medical's Visual-ICE Cryoablation System Software Revision 1.4.0, to perform cryoablative destruction of tissue. The needles are intended to convert high-pressure gas to either a very cold freezing application or to a warm thawing application. The IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles are disposable 2.1 mm needles that have a sharp cutting tip, a color-coded handle, a gas tube, and a connector. Both IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles offer handles configured as straight and in a 90° angled configuration (to aid positioning of the needle within the CT imaging system gantry).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter G with a green dot inside, followed by the text "GALILMEDICAL" in a bold, sans-serif font. Below the logo is the tagline "Leading Through Innovation" in a smaller font.
# K152133
The IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles were designed to provide the same cryoablation functionality with similar iceball shapes on a 2.1 mm shaft as compared to Galil Medical's predicate needles, IceRod CX (1.5 mm) needle and IceEDGE 2.4 mm needle.
The IcePearl 2.1 CX and IceFORCE 2.1 CX Crvoablation Needles differ from the predicate devices in that the reconfigured heat exchanger has been relocated from the needle shaft to the needle handle and an additional vacuum insulated tube has been incorporated into the handle. The changes were made to ensure optimal freezing performance utilizing a 2.1 mm shaft.
The table below provides a summary comparison of the submitted devices compared to the predicate devices.
| Description of Submitted Device: | Comments related to Predicates: |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IcePearl 2.1 CX and IceFORCE 2.1<br>CX Cryoablation Needles | IceRod CX Cryoablation Needles (K121251, K140584); and<br>IceEDGE 2.4 (K111859) |
| Design and Construction | |
| Needle Tip | Same as predicate |
| Needle Shaft | Shaft size (diameter) is in-between the shaft sizes of the predicates |
| Gas Pathway Tubing | Same as predicate |
| Handle | Same as the predicates with the exception of a larger diameter resulting<br>from the incorporation of:<br>• the reconfigured heat exchanger located in the proximal end of the handle toward the connector versus being in the needle shaft; and<br>• the addition of a vacuum insulation tube |
| Needle Connector | Same as predicate |
| Performance and Function | |
| Freezing/Thawing Technology | Same as predicate |
| Function | Same as predicate |
| Freezing Parameters | Same as predicate |
| Thaw Parameters | Same as predicate |
| Track Ablation | Same as predicate |
| Indications for Use | Same as predicate |
In summary, the submitted IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles have the same technology, principle of operation and indications for use as the predicate devices.
This Special 510(k) also contains changes to the Visual-ICE System software denoted as Software Release 1.4.0. Software Release 1.4.0 is an updated version of the Visual-ICE software previously cleared. There are no changes to the system hardware required as a result of the changes being made in the software. The following modifications were incorporated based on the added software support of the proposed IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles as well as customer and sales feedback to enhance the robustness of the software programming.
| User Interface<br>Changes | Updated the needle selection dialog to display marketing names and to default to the<br>currently selected needle type. |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Service and | Added features for service personnel to diagnose and troubleshoot issues in the field. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter "G" with a green dot inside, followed by the text "GALIL MEDICAL" in a bold, sans-serif font. Below the logo, the tagline "Leading Through Innovation" is written in a smaller font.
K152133
| Maintenance<br>Changes | These changes include:<br>Ability to access and troubleshoot to needle memory parameters from diagnostic screen Ability to export multiple log files to USB Stick Modified the Visual ICE message to accommodate for 2 year service intervals. |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional<br>Changes | Added support for needles that have been calibrated with heater wire temperature information during production. This allows for a lower needle shaft temperature display. Added support for the 2.1 CX Cryoablation Needles to the software. Added reporting of timeline when channel is disconnected. Modified the software to display lowest temperature during freezing and to display temperatures in the MTS graphs without first having to test a needle. |
| User Interface and<br>Functional<br>Changes | Added support and dialog messages related to real time feedback of power consumption by needles in i-Thaw, FastThaw, and cautery modes. The big timers on the system are now easier to activate from different screens and display needle temperature. Streamlined the channel controls by removing the redundant advanced features on the freeze and thaw dialogs. Added the ability to FastThaw directly from the channel controls instead of having to filter through multiple menu options. Added the ability to save and load multiple programmed cycles. |
# Intended Use:
Galil Medical's IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures. The IcePearl 2.1 CX and lceFORCE 2.1 CX Cryoablation Needles are indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, oncology, and urology. The systems are designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
The Galil Medical IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles have the following specific indications:
- Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate . Hyperplasia "BPH")
- . Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention)
- . Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.)
- Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia)
- . General surgery (palliation of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma)
- . ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth)
- . Thoracic surgery (with the exception of cardiac tissue)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Galil Medical. The logo features a stylized letter "G" with a green dot inside, followed by the text "GALILMEDICAL" in a bold, sans-serif font. Below the logo, the tagline "Leading Through Innovation" is displayed in a smaller, lighter font.
- Proctology (ablation of benign or malignant growths of the anus or rectum, and ● hemorrhoids)
#### Summary of Performance Data and Substantial Equivalence:
The IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles use the same technology and have the same intended use and method of operation as compared to the predicate devices.
Performance testing was conducted on the IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles with Visual-ICE Software Revision 1.4.0 to verify safety and performance characteristics and to establish substantial equivalence. Testing was conducted according to protocols based on international standards and in-house requirements and included dimensional testing, functional testing, freezing performance, and in vivo needle track ablation, based on depth of tissue necrosis characterization. Test results demonstrated that the IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles when used with Visual-ICE Software Revision 1.4.0 meet defined specifications and do not raise any new safety or effectiveness issues.
#### Conclusion:
The information and data provided in this Special 510(k) Notification establish that the IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles when used with Visual-ICE Software Revision 1.4.0 are substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.