VISUAL-ICE CRYOABLATION SYSTEM

K123865 · Galil Medical, Inc. · GEH · Feb 6, 2013 · General, Plastic Surgery

Device Facts

Record IDK123865
Device NameVISUAL-ICE CRYOABLATION SYSTEM
ApplicantGalil Medical, Inc.
Product CodeGEH · General, Plastic Surgery
Decision DateFeb 6, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Visual-ICE™ Cryoablation System is intended for cryoablative destruction of tissue during surgical procedures. The Visual-ICE System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The Visual-ICE Cryoablation System has the following specific indications: Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH") Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention) Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.) Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia) General surgery (palliation of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma) ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth) Thoracic surgery (ablation of arrhythmic cardiac tissue cancerous lesions) Proctology (ablation of benign or malignant growths of the anus or rectum, and hemomhoids)

Device Story

Mobile, computer-controlled cryoablation console; uses Joule-Thomson effect with high-pressure argon gas for tissue freezing and helium gas or internal heating elements for active thawing. Operated by clinicians in surgical settings. System features touch screen interface for procedure monitoring and control. Software revision 1.2.2 introduces UI enhancements, including modified organ map display, updated timer capacity, programmable freeze-thaw cycles for all channels, and adjusted gas/pressure warning thresholds. Device provides controlled thermal energy to cryoablation needles to destroy targeted tissue; benefits include minimally invasive treatment of tumors and lesions.

Clinical Evidence

Bench testing only. No clinical data required or provided. Software verification testing conducted per approved protocol; all tests passed.

Technological Characteristics

Mobile console; Joule-Thomson effect cooling; argon/helium gas delivery; closed-tip cryoablation needles; touch screen interface; computer-controlled; software-based monitoring and control.

Indications for Use

Indicated for patients requiring cryoablative tissue destruction in general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, proctology, and urology. Specific conditions include prostate cancer, BPH, malignant/benign tumors, skin lesions, warts, gynecological dysplasia, hemorrhoids, and arrhythmic cardiac tissue.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ G**GALILMEDICAL** Leading Through Innovation K123865 Page 1 of 2 # Section 5. 510(k) Summary #### FEB 6 2013 # 510(k) SUMMARY TEMPLATE A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92. Submitter: Galil Medical Inc. 4364 Round Lake Road Arden Hills, MN 55112 USA | Company Contact Person: | Amy E. McKinney<br>Director, Regulatory Affairs | |--------------------------------|------------------------------------------------------------| | Phone:<br>Fax: | Galil Medical Inc.<br>651-287-5096 | | Email: | 651-287-5097<br>amy.mckinney@galilmedical.com | | Device Name: | Visual-ICE® Cryoablation System, Software Revision 1.2.2 | | Device Classification<br>Name: | Cryosurgical unit and accessories (GEH)<br>21 CFR 878.4350 | | Predicate Device: | Visual-ICE® Cryoablation System (K113860) | ## Device Description: The Visual-ICE Cryoablation System is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by circulating. helium gas through the needles or, alternatively, by the use of Galil Medical i-Thaw technology in which a heating element inside the cryoablation needle can be energized to cause thawing. This Special 510(k) is being submitted to modify the software with a variety of changes to enhance usability of the system. These changes are primary being made to the user interface and the service and maintenance screens. Basic functionality of the system remains unchanged, however, the following functional modifications have been made: - . Organ map display has been modified to allow users to place MTS needles on the Organ Мар, - When all channels are locked and linked together the Test All button icon state identifies the . state for all channels. - The software defaults all needles to the first 1-wire needle that was attached. . Page 10 of 46 {1}------------------------------------------------ K123865 Image /page/1/Picture/1 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter "G" with a small circle inside, followed by the text "GALILMEDICAL" in bold, sans-serif font. Below the company name is the tagline "Leading Through Innovation" in a smaller font size. Page 2 of (2) - User interface can display up to three large timers instead of two. . - Freeze-thaw cycles can now be programmed for all channels using the All button, . - The software wams the user of low gas when the remaining gas time reaches 10 minutes . instead of 5 minutes, - "Stick" intensity has been changed from 20% to 5%, . - Helium low-pressure warning was reduced to 1800 PSI from 2100 PSI. . # Intended Use: The Visual-ICE™ Cryoablation System is intended for cryoablative destruction of tissue during surgical procedures. The Visual-ICE System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The Visual-ICE Cryoablation System has the following specific indications: - Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH") - . Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention) - . Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.) - . Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia) - . General surgery (palliation of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma) - . ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth) - Thoracic surgery (ablation of arrhythmic cardiac tissue cancerous lesions) . - . Proctology (ablation of benign or malignant growths of the anus or rectum, and hemomhoids) # Summary of Performance Data and Substantial Equivalence: The Visual-ICE System and software were evaluated in accordance with Galil Medical's risk management plan. No new risks were identified based on the changes incorporated into Software Revision 1.2.2. A complete software verification test was conducted based on the originally approved protocol and test methods. Software Revision 1.2.2 passed all verification testing. Additional validation testing was not required based on the changes made to the software. No changes were made to the Visual-ICE System hardware. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three curved shapes that resemble a person embracing or being embraced. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 6, 2013 Galil Medical, Incorporated % Ms. Amy McKinney Director, Regulatory Affairs 6518 Tamarind Sky Lane Fulshear, Texas 77441 Re: K123865 Trade/Device Name: Visual-ICE Cryoablation System, Software Revision 1.2.2 Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: January 08, 2013 Received: January 11, 2013 Dear Ms. McKinney: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Amy McKinney forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, For # Peter D. Rumm -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement # 510(k) Number (if known): K123865 Device Name: Visual-ICE Cryoablation System, Software Revision 1.2.2 ### Indications For Use: The Visual-ICE™ Cryoablation System is intended for cryoablative destruction of tissue during surgical procedures. The Visual-ICE System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery, ENT gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The Visual-ICE Cryoablation System has the following specific indications: - Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH") - . Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention) - Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. . Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the evelid or canthus area, ulcerated basal cell turnors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemangiomas, perianal condylomata, and palliation of tumors of the skin.) - Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia) . - General surgery (palliation of the rectum, hemorrhoids, anal fissures, pilonidal . cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma) - ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth) . - Thoracic surgery (ablation of arrhythmic cardiac tissue cancerous lesions) . - Proctology (ablation of benign or malignant growths of the anus or rectum, and . hemorrhoids) Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {5}------------------------------------------------ Long H. Chen, bays, bays, bays, bays, bays, cause and many, chen - S anger H. Chen, bays FDA outeredde, co-long H. Chen S, 0.9.2342.19200300.100.1.1=1300369056 Oate: 2013.02.06 13:32:18 -05T00 -S ing the state of the state of the state of the state of the state of the formation of the formation of the finisters. And the formation of the finish of the finish and the fi (Division Sign-Off) Division of Surgical Devices 510(k) Number ___K123865_ ・
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...