K152014 · Pmbs, LLC · KCT · Oct 20, 2015 · General Hospital
Device Facts
Record ID
K152014
Device Name
MTS300 Multiple Tray Sterilization System
Applicant
Pmbs, LLC
Product Code
KCT · General Hospital
Decision Date
Oct 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The MTS300 System is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterlization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the MTS300 Transfer Cart, MTS300 filters and integrity locks. The MTS300 System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray. The MTS300 System was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm. Use only uncovered, perforated or wire mesh general delivery trays within the MTS300 System.
Device Story
MTS300 Multiple Tray Sterilization System; stainless steel sterilization cabinet and adjustable transfer cart. Used by healthcare providers to enclose medical devices for steam sterilization. Inputs: medical instruments placed in uncovered, perforated, or wire mesh trays. Process: cabinet loaded onto transfer cart, deployed into prevacuum steam sterilizer (270°F, 4-minute cycle), followed by 30-minute drying. System uses disposable filters and integrity locks with sterilization indicator dots to maintain sterility for 30 days. Optional STEAMPlus tray record card/integrator may be used. Output: sterilized, protected instrument trays ready for clinical use. Benefits: enables organized, sterile storage and transport of surgical instruments; maintains sterility post-processing.
Clinical Evidence
Bench testing only. Sterilization efficacy verification performed per ANSI/AAMI ST77:2013 and FDA guidance for sterilization packaging. Stability demonstrated via engineering analysis and usability study. No clinical data.
Technological Characteristics
16-gauge stainless steel cabinet; adjustable transfer cart; disposable heavy-duty sterilization wrap filters; integrity locks with indicator dots. Sterilization via prevacuum steam (270°F). Compatible with all uncovered, perforated, or wire mesh general delivery trays. Dimensions/volume-to-vent ratio specified. Mechanical containment device; no software or electronic components.
Indications for Use
Indicated for enclosing medical devices for sterilization by healthcare providers; maintains sterility for up to 30 days. Compatible with prevacuum steam sterilizers (270°F, 4-minute exposure). Validated for rigid instruments with lumens ≥3.8mm ID and ≤370mm length. Requires use of MTS300 Transfer Cart, filters, and integrity locks.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, formed by three stylized human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 20, 2015
PMBS, LLC c/o Ms. Annette Hillring President Hillring & Associates, Inc. 3012 St. Charles Dr. Tampa, FL 33618
Re: K152014
Trade/Device Name: MTS300 Multiple Tray Sterilization System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: September 16, 2015 Received: September 17, 2015
### Dear Ms. Hillring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Hillring
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K152014
Device Name MTS300 Multiple Tray Sterilization System
### Indications for Use (Describe)
The MTS300 System is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterlization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the MTS300 Transfer Cart, MTS300 filters and integrity locks.
The MTS300 System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray.
The MTS300 System was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm.
Use only uncovered, perforated or wire mesh general delivery trays within the MTS300 System.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared | September 16, 2015 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Michele E. Mauzerall<br>PMBS, LLC<br>40 Milltown Road<br>Stockton, New Jersey 08559<br>Phone: 908-300-7093<br>Email: mem.pmbs@gmail.com |
| Device | <ul><li>Trade Name: MTS300 Multiple Tray Sterilization System</li><li>Common Name: Sterilization Container and Accessories</li><li>Classification Name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories</li><li>Product Code: KCT</li><li>Classification Regulation: 21 CFR 880.6850 Sterilization wrap, Class II</li></ul> |
| Predicate Device | AmMed Surgical Equipment, LLC, S.C.O.R.E.S. Units, Self Contained Operating Room Equipment Sterilization Containers cleared via Premarket Notification 510(k) K110898 on July 26, 2012 |
| Device Description | The MTS300 Multiple Tray Sterilization System (also referred to as the MTS300 System) includes the stainless steel sterilization cabinet and adjustable transfer cart. Single use filters and integrity locks with sterilization indicator dots are used with the system for each sterilization cycle. An optional STEAMPlus <sup>TM</sup> tray record card with STEAMPlus sterilization integrator (SPS Medical) may be utilized by the healthcare provider. The MTS System is intended to sterilize a maximum weight of 300 pounds of devices and instruments for a single patient surgery. The loaded cabinet is deployed from the adjustable transfer cart and processed in a pre-vacuum steam sterilizer and then dried for 30 minutes. Sterility is maintained for up to 30 days. |
| Indications for Use | The MTS300 System is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The |
Continued on next page
{4}------------------------------------------------
| Indications for<br>Use, continued | unit must be used with the MTS300 Transfer Cart, MTS300 filters and<br>integrity locks. | |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The MTS300 System is intended to be used in prevacuum steam sterilizers<br>with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no<br>more than 3 trays per shelf or 25 lbs. per tray. | |
| | The MTS300 System was tested and validated with rigid instruments<br>containing lumens with an inner diameter of 3.8mm and an overall length of<br>370mm. Do not use with instruments containing lumens with an inner<br>diameter smaller than 3.8mm and an overall length longer than 370mm. | |
| | Use only uncovered, perforated or wire mesh general delivery trays within the<br>MTS300 System. | |
| Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate<br>Device | There are no new technological characteristics associated with the MTS300<br>Multiple Tray Sterilization System as compared to the predicate device, the<br>S.C.O.R.E.S. Units, Self Contained Operating Room Equipment Sterilization<br>Container. A comparison of the technological characteristics of the subject<br>device, the MTS300 Multiple Tray Sterilization System, to the predicate<br>device, the S.C.O.R.E.S. Units, Self Contained Operating Room Equipment<br>Sterilization Container is provided in the table on the following page. | |
| Characteristic | Subject Device:<br>MTS300 Multiple Tray Sterilization System<br>("MTS300 System")<br>K152015 | Predicate Device:<br>S.C.O.R.E.S. Units, Self Contained Operating<br>Room Equipment Sterilization Container<br>K110898 |
| Indications for Use | The MTS300 System is indicated for enclosing other<br>medical devices that are to be sterilized by a healthcare<br>provider. It is intended to allow sterilization of the<br>enclosed materials and maintain sterility for up to 30 days<br>until used. The unit must be used with the MTS300<br>Transfer Cart, MTS300 filters and integrity locks. | The SCORES Sterilization Container is indicated for<br>enclosing other medical devices that are to be sterilized<br>by a healthcare provider. It is intended to allow<br>sterilization of the enclosed materials and maintain<br>sterility for up to 30 days until used. The unit must be<br>used with the SCORES Transfer Cart, SCORES filters<br>and integrity locks. |
| | The MTS300 System is intended to be used in prevacuum<br>steam sterilizers with a prevacuum cycle of 270°F and<br>exposure time of 4 minutes. Use no more than 3 trays per<br>shelf or 25 lbs. per tray. | The unit is intended to be used in prevacuum steam<br>sterilizers with a prevacuum cycle of 270°F and exposure<br>time of 4 minutes. Use no more than 3 trays per shelf or<br>25 lbs. per tray. |
| | The MTS300 System was tested and validated with rigid<br>instruments containing lumens with an inner diameter of<br>3.8mm and an overall length of 370mm. Do not use with<br>instruments containing lumens with an inner diameter<br>smaller than 3.8mm and an overall length longer than<br>370mm. | The SCORES Unit was tested and validated with rigid<br>instruments containing lumens with an inner diameter of<br>3.8mm and an overall length of 370mm. Do not use with<br>instruments containing lumens with an inner diameter<br>smaller than 3.8mm and an overall length longer than<br>370mm. |
| | Use only uncovered, perforated or wire mesh general<br>delivery trays within the MTS300 System. | Use only uncovered, perforated or wire mesh general<br>delivery trays within the SCORES Sterilization<br>Container. |
| Characteristic | Subject Device:<br>MTS300 Multiple Tray Sterilization System<br>("MTS300 System")<br>K152015 | Predicate Device:<br>S.C.O.R.E.S. Units, Self Contained Operating<br>Room Equipment Sterilization Container<br>K110898 |
| Sterilization<br>Parameters | Prevacuum cycle of 270°F and exposure time of 4<br>minutes | Prevacuum cycle of 270°F and exposure time of 4<br>minutes |
| Drying Time | 30 minutes | 30 minutes |
| Sterility | 30 days | 30 days |
| Maintenance | | |
| Cabinet Material | 16 gauge stainless steel | 12 and 14 gauge stainless steel |
| Weight w/ Shelves | 212 lbs. | 214 lbs. |
| Filter Material | Heavy duty sterilization wrap by SPS Medical | Heavy duty sterilization wrap by SPS Medical |
| Volume-to-Vent<br>Ratio | 129.131 | 192.22 |
| Deployment into<br>Sterilization<br>Chamber | Adjustable transfer cart | Adjustable transfer cart |
| Recommended<br>Sterilization Trays | All manufacturers' trays - uncovered, perforated or<br>wire mesh general delivery trays | All manufacturers' trays - uncovered, perforated or<br>wire mesh general delivery trays |
| Recommended<br>Sterilizers | All makes and models with dimensions compatible<br>with the MTS System | Three Steris models |
| Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate<br>Device,<br>continued | Modifications introduced into the MTS300 Multiple Tray Sterilization<br>System do not affect the technological characteristics. The cabinet material<br>remains stainless steel and a filtration system utilizing disposable filters of the<br>same material from the same supplier (with only dimensional changes) is the<br>method by which sterilization is achieved and maintained. The cabinet<br>continues to be deployed into the sterilization chamber with an adjustable<br>transfer cart. The sterilization parameters, drying time and sterility<br>maintenance specifications are identical to the predicate device along with the<br>indications for use. | |
| Performance<br>Data | Nonclinical data was generated to ensure the MTS300 Multiple Tray<br>Sterilization System continues to meet the intended use. Sterilization efficacy<br>verification studies were performed in accordance with the requirements of<br>ANSI/AAMI ST77:2013 Containment devices for reusable medical device<br>sterilization and the recommendations of Premarket Notification [510(k)]<br>Submissions for Medical Sterilization Packaging Systems in Healthcare<br>Facilities, Draft Guidance for Industry and FDA, March 7, 2002, and<br>included all of the studies performed on the predicate device. Stability of the<br>system (cabinet on transfer cart) was demonstrated by an engineering analysis<br>and within a usability study. The usability study demonstrated that the<br>modified cabinet and adjustable transfer cart will continue to meet user needs<br>and intended use. | |
| Conclusions | In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR<br>807, and based upon the information and scientifically valid data provided in<br>this premarket notification, PMBS, LLC, concludes that the subject device,<br>the MTS300 Multiple Tray Sterilization System is as safe, as effective and<br>performs as well as the predicate device, the S.C.O.R.E.S. Units, Self<br>Contained Operating Room Equipment Sterilization Container (K110898). | |
{5}------------------------------------------------
omparison of Technological Characteristics with the Predicate Device, continu
{6}------------------------------------------------
omparison of Technological Characteristics with the Predicate Device, continu
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.