The SCORES® Sterilization Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SCORES® transfer cart, SCORES® filters, and integrity locks. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 27.0° and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray. The SCORES® Unit was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm. Use only uncovered, perforated or wire mesh general delivery trays within the SCORES® Sterilization Container.
Device Story
S.C.O.R.E.S. Unit is a stainless steel sterilization container and delivery system; holds multiple instrument trays for single surgical cases. System includes cabinet with internal shelves, filtration system for steam penetration, and adjustable transport cart. Used in clinical settings; operated by healthcare staff. Loaded unit transported via cart into prevacuum steam autoclave; processed in standard cycle; dried in chamber. Maintains sterility of contents for up to 30 days post-sterilization. Eliminates need for individual tray wrapping. Benefits include streamlined instrument delivery to surgical suite and maintained sterile integrity.
Clinical Evidence
Bench testing only. Validation performed using a hospital steam sterilizer (Amsco Model 3043) with prevacuum steam half-cycles. Testing confirmed successful sterilization and drying of contents. Sterile maintenance validated for up to 30 days. Validated for specific lumen dimensions (≥3.8mm ID, ≤370mm length) and weight limits (max 25 lbs/tray).
Technological Characteristics
Stainless steel sterilization cabinet with filtration system. Wheeled form factor for transport via adjustable cart. Operates in prevacuum steam sterilizers. Mechanical design; no electronic or software components.
Indications for Use
Indicated for healthcare providers to enclose medical devices for steam sterilization and maintenance of sterility for up to 30 days. Compatible with prevacuum steam sterilizers (270°F, 4-minute exposure). Limits: max 3 trays/shelf, max 25 lbs/tray. Validated for rigid instruments with lumens ≥3.8mm ID and ≤370mm length. Requires specific SCORES® accessories (cart, filters, locks).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Allegiance Genesis Container System
Reference Devices
Amsco (Steris) Model 3043
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 26 2012
# 510(k) Summary Pursuant to 21 CFR 807.92
## K110898
## Date of preparation: 5 July 2012
- Submitted By: 1.
AmMed Surgical Equipment, LLC 3801 Corporex Park Drive, Suite 120 Tampa, Florida 33619
2. Contact: David C. Furr FDC Services, LLC 8708 Capehart Cove Austin, Texas 78733 512-906-9654
3. Product: S.C.O.R.E.S Units. Self Contained Operating Room Equipment Sterilization Containers Regulation: 880.6850. Sterilization Wrap Class II Product Code: KCT
#### 4. Common/Trade Name:
S.C.O.R.E.S Units
Sterilization Wrap/Container
#### Description:
The S.C.O.R.E.S. Unit (self contained operating room equipment sterilization unit) is a sterilization container and delivery system capable of holding multiple instrument sterilization trays. The S.C.O.R.E.S. container can be loaded with multiple surgical instrument trays needed for a single surgery, steam sterilized, dried in the chamber drying cycle, removed from the autoclave on a transport cart, stored for up to 30 days, and delivered to the surgical suite for opening and use of the contents. The system includes the S.C.O.R.E.S. cabinet with wheels for loading into sterilizers, internal shelves for placement of sterilization trays, a filter system to allow steam penetration and maintain sterility, and an adjustable transport cart to allow for movement of the loaded system.
{1}------------------------------------------------
Instrument trays within the S.C.O.R.E.S Unit do not need to be placed in sterilization containers or wrapped to maintain sterile integrity as the filtered Unit does so. Once opened in the surgical suite and the contents are to be used immediately. The subject device has been extensively tested, and verified to perform in a safe and effective manner.
## Intended Use:
The SCORES® Sterilization Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SCORES® transfer cart, SCORES® filters, and integrity locks.
The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 27.0° and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray.
The SCORES® Unit was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm.
Use only uncovered, perforated or wire mesh general delivery trays within the SCORES® Sterilization Container.
# Technological Characteristics:
The S.C.O.R.E.S. Unit is a sterilization cabinet constructed from stainless steel and provided with a filtration system.
The Unit can be loaded with all of the instrumentation needed for a surgical case. Instruments are placed into trays, which need not be wrapped. Individual trays of instruments should weigh no more than 251bs. and up to 12 trays can be arranged on the shelves of the S.C.O.R.E.S. Unit. The loaded unit has a filtration system to allow for steam penetration.
{2}------------------------------------------------
The wheeled Unit is transported on an adjustable cart, which can be adjusted to the height of a prevacuum autoclave chamber. The sealed, loaded unit can be rolled into the chamber and processed in a normal cycle.
The S.C.O.R.E.S. Unit has been validated in accordance with relevant industry standards using a hospital steam sterilizer. Validation was done with prevacuum steam half cycles and drying was also accomplished. Validation was done with an Amsco (Steris) Model 3043.
The product has also been shown to be capable of maintaining sterility of the contents for up to 30 days.
## Substantial Equivalence:
The S.C.O.R.E.S. Unit has been determined to be substantially equivalent to the Allegiance Genesis Container System. For prevacuum sterilization applications, the predicate device has similar technical features, indications for use, and the safety and effectiveness of the devices is equivalent.
# Conclusions:
The predicate devices and the AmMed Surgical S.C.O.R.E.S Unit container system share similar indications, technology, and application. Although the S.C.O.R.E.S. unit holds more equipment, the device is essentially equivalent to the predicate device products in key areas of performance that affect safety and effectiveness.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a single snake entwined around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 2 6 2012
Ammed Surgical Equipment, LLC C/O Mr. David C. Furr FDC Services, LLC 8708 Capehart Cove Austin, Texas 78733
Re: K110898
Trade/Device Name: Scores Unit Sterilization Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: July 5, 2012 Received: July 13, 2012
Dear Mr. Furr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Furr
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
# Indications for Use Statement
K110898 . 510(k) Number:
Device Name: SCORES® Unit Sterilization Container
Indications for Use:
The SCORES® Sterilization Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SCORES® transfer cart, SCORES® filters, and integrity locks.
The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270° and exposure time of 4 minutes. Use no more than 3 travs per shelf or 25 lbs. per tray.
The SCORES® Unit was tested and validated with rigid instruments containing lumens with an inner diameter of 3.8mm and an overall length of 370mm. Do not use with instruments containing lumens with an inner diameter smaller than 3.8mm and an overall length longer than 370mm.
Use only uncovered, perforated or wire mesh general delivery trays within the SCORES® Sterilization Container.
| Prescription Use | |
|-------------------|--|
| (per CFR 801.109) | |
or
| Over-the-counter use | X |
|----------------------|---|
|----------------------|---|
Concurrence of CDRH
E. Mitchell F. O'Quinn - Williams
Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, Deneral Devices Division of Arcel. Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.