K151863 · Precision Spine, Inc. · PEK · Aug 7, 2015 · Orthopedic
Device Facts
Record ID
K151863
Device Name
Precision Spine Interspinous Plate System
Applicant
Precision Spine, Inc.
Product Code
PEK · Orthopedic
Decision Date
Aug 7, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reli SP Spinous Plating System of Precision Spine, Inc. is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1) of skeletally mature patients. It is intended for single level plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (i.e. fracture or dislocation), spondylolisthesis, and/or tumor. It is not intended for stand-alone use.
Device Story
Reli SP Spinous Plating System is a posterior, non-pedicle supplemental fixation device for the non-cervical spine (T1-S1). System comprises male and female titanium plates with teeth to interface with spinous process bone. Plates are placed on both sides of the spinous process and compressed together to achieve fixation. Used by surgeons in clinical settings to facilitate spinal fusion in patients with degenerative disc disease, trauma, spondylolisthesis, or tumor. Device is not for stand-alone use; requires supplemental fixation. Expanded size range provided to accommodate various patient anatomies.
Clinical Evidence
No clinical data. Bench testing only (analysis of equivalence to predicates).
Technological Characteristics
Manufactured from medical grade Titanium (Ti 6Al-4V) per ASTM F136. Interspinous plates range in total height from 28-55mm. Non-sterile; requires sterilization. Mechanical fixation via compression of male/female plates onto spinous processes.
Indications for Use
Indicated for skeletally mature patients requiring supplemental posterior, non-pedicle fixation in the non-cervical spine (T1-S1) for single-level fusion. Conditions include degenerative disc disease, trauma (fracture/dislocation), spondylolisthesis, and tumor. Not for stand-alone use.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2015
Precision Spine, Incorporated % Mr. Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Testing, Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K151863
Trade/Device Name: Reli SP Spinous Plating System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: July 7, 2015 Received: July 8, 2015
Dear Mr. Maxwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Kenneth C. Maxwell II
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known)
K151863
Device Name
Reli SP Spinous Plating System
Indications for Use (Describe)
The Reli SP Spinous Plating System of Precision Spine, Inc. is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1) of skeletally mature patients. It is intended for single level plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (i.e. fracture or dislocation), spondylolisthesis, and/or tumor. It is not intended for stand-alone use.
Type of Use (Select one or both, as applicable)
□ Over-The-Counter Use (21 CFR 801 Subpart C) 区 Prescription Use (Part 21 CFR 801 Subpart D) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 5.510(K) SUMMARY
| Submitter's Name: | Precision Spine |
|----------------------------|------------------------------------------------------------------------|
| Submitter's Address: | 2050 Executive Drive<br>Pearl, MS 39208 |
| Submitter's Telephone: | 973-455-7150 |
| Contact Person: | Kenneth C Maxwell II<br>Empirical Testing Corp.<br>904.392.7576 |
| Date Summary was Prepared: | 06 Aug 2015 |
| Trade or Proprietary Name: | Reli SP Spinous Plating System |
| Common or Usual Name: | Spinous Process Plate |
| Classification: | Class II per 21 CFR §888.3050<br>Spinal interlaminal fixation orthosis |
| Product Code: | PEK |
| Classification Panel: | 87 Orthopedics |
#### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The Precision Spine Reli SP Spinous Plating System consists of an ISP female plate and an ISP male plate for posterior fixation of the spine in order to achieve fusion. The ISP female plate and an ISP male plate are available in multiple sizes to accommodate various patient anatomies. The ISP female plate and an ISP male plate feature teeth to interface with the bone of the spinous processes. The ISP male plate is passed through the insert such that, in their final position, the ISP female plate and an ISP male plate surround the spinous processes on both sides, and fixation is achieved via compression of the two components onto the spinous processes. The purpose of this submission is to offer an expanded size range for the Reli SP Spinous Plating System.
## INDICATIONS FOR USE
The Reli SP Spinous Plating System of Precision Spine, Inc. is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1) of skeletally mature patients. It is intended for single level plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (i.e. fracture or dislocation), spondylolisthesis, and/or tumor. It is not intended for stand-alone use.
#### TECHNICAL CHARACTERISTICS
The Precision Spine Reli SP Spinous Plating System is manufactured from medical grade Titanium (Ti 6Al-4V) per ASTM F136. The implants are provided non-sterile with instructions for sterilization. The interspinous plates are designed in total heights of 28-55mm.
{5}------------------------------------------------
Reli SP Spinous Plating System is made from material that conforms to ASTM F136. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:
- Indications for Use
- Materials of manufacture
- Structural support mechanism ●
#### Table 5-1 Predicate Devices
| 510k Number | Trade or Proprietary or Model Name | Manufacturer | Type |
|-------------|------------------------------------|-----------------|------------|
| K142378 | Interspinous Plate System | Precision Spine | Primary |
| K130438 | Axle Interspinous Fusion System | Spine-X | Additional |
### PERFORMANCE TESTING SUMMARY
Precision Spine has not conducted mechanical testing as part of this Special 510(k) Device Modification Premarket Notification. An analysis was performed to determine the subject implants were substantially equivalent to the predicate devices, therefore no mechanical testing was conducted as part of this submission.
### CONCLUSION
The subject modified is the Reli SP Spinous Plating System and is very similar to previously cleared Reli SP Spinous Plating System. The subject Reli SP Spinous Plating System has the same intended uses, indications, technological characteristics, and principles of operation as the predicate devices. The modifications raise no new types of safety or effectiveness questions. The overall technology characteristics lead to the conclusion that the Reli SP Spinous Plating System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.