AXLE INTERSPINOUS FUSION SYSTEM

K130438 · X-Spine Systems, Inc. · PEK · Jun 27, 2013 · Orthopedic

Device Facts

Record IDK130438
Device NameAXLE INTERSPINOUS FUSION SYSTEM
ApplicantX-Spine Systems, Inc.
Product CodePEK · Orthopedic
Decision DateJun 27, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Axle Interspinous Fusion System is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1 inclusive). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. The Axle Interspinous Fusion System is intended for use with bone graft material, and not for stand-alone use.

Device Story

The Axle Interspinous Fusion System is a posterior, non-pedicle supplemental fixation device designed to stabilize the spinous processes to facilitate fusion. It consists of titanium alloy (Ti6Al4V) and PEEK (Polyetheretherketone) plates and inserts. The device is implanted by a surgeon in a clinical setting to provide mechanical support to the spine at a single level (T1-S1). It is not a stand-alone device and must be used in conjunction with bone graft material. By providing supplemental stabilization, the system aids in the fusion process for patients suffering from degenerative disc disease, spondylolisthesis, trauma, or tumors. The components are provided non-sterile and must be steam autoclaved by the healthcare facility prior to use.

Clinical Evidence

Bench testing only. Testing included static compression bending, static torsion, and fatigue compression bending per ASTM F1717, as well as spike pullout and dissociation tests.

Technological Characteristics

Materials: Titanium alloy (Ti6Al4V) per ASTM F136 and PEEK Optima LT-1 per ASTM F2026. Form factor: Plates and inserts for spinous process fixation. Energy source: None (mechanical). Sterilization: Steam autoclave by user. Connectivity: None.

Indications for Use

Indicated for patients requiring supplemental spinal fusion at a single level in the non-cervical spine (T1-S1) due to degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), or tumor. Not for stand-alone use; must be used with bone graft material.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) Summary #### ADMINISTRATIVE INFORMATION ### JUN 2 7 2013 | Manufacturer Name: | X-spine Systems, Inc.<br>452 Alexandersville Rd.<br>Miamisburg, OH 45342 | |-------------------------|--------------------------------------------------------------------------| | Official Contact: | David Kirschman, M.D.<br>Chief Medical Officer | | Date Prepared: | February 20, 2013 | | DEVICE NAME | | | Trade/Proprietary Name: | Axle™ Interspinous Fusion System | | Common Name: | Spinous Process Plate | | Classification Name: | Spinal Interlaminal Fixation Orthosis | | Product Code: | PEK | | Classification: | §888.3050 | | Device Class: | Class II | #### ESTABLISHMENT REGISTRATION NUMBER The X-spine Systems, Inc. establishment registration number is 3005031160. The owner/ operator number for X-spine Systems, Inc. is 9063903. #### INTENDED USE The Axle Interspinous Fusion System is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1 inclusive). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. The Axle Interspinous Fusion System is intended for use with bone graft material, and not for stand-alone use. #### DEVICE DESCRIPTION The Axle Interspinous Fusion System consists of plates and inserts of various sizes that are used to provide supplemental stabilization of the spinous processes to support fusion. The svstem components can be assembled in a variety of configurations so that adaptations can be made to take into account pathology and individual patient anatomy. The systems components are manufactured using standard manufacturing processes of medical grade Titanium alloy (Ti6Al4V) that complies with ASTM F136 - Standard {1}------------------------------------------------ Specifications for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications and Invibio PEEK Optima LT-1 in accordance with ASTM F2026 -Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications. The implant components are provided clean and non-sterile. These devices are supplied in a rigid sterilization tray and are to be sterilized by a healthcare professional using a Steam Autoclaye in accordance with the instructions for use provided by X-spine Systems Inc., as well as the instructions provided by the Autoclave manufacturer. #### EQUIVALENCE TO MARKETED PRODUCT X-spine Systems. Inc. has submitted information to demonstrate that, for the purposes of FDA's requlation of medical devices, the Axle Interspinous Fusion System is substantially equivalent to the predicate device based on a comparison including the following characteristics: - FDA Product Code . - . Intended Uses - . Surgical Approach - Anatomical Region . - . Implant Materials - . Product Features #### PREDICATE DEVICES - . X-spine Systems, Inc. - Axle Interspinous Fusion System (K101471) - . X-spine Systems. Inc. - Axle PEEK Interspinous Fusion Svstem (K112592) - Medtronic-Sofamor-Danek-CD Horizon Spinal System Spinous Process Plate . (K091445) - . Lanx, LLC - Aspen Spinous Process Fusion Plate (K071877) #### PERFORMANCE COMPARISON The implant components were tested using the following standards: ASTM F1717 - Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model - . Static Compression Bending - . Static Torsion - Fatigue Compression Bending . In addition to the above standard testing, spike pullout and dissociation tests were performed as part of this submission. There are no cited standards for these tests. The modified device is used to treat the same indications for use, utilizes the same scientific and operational principles, and is manufactured using the same manufacturing practices from identical materials as the parent device. In conclusion, X-spine Systems, Inc. has submitted information to demonstrate that, for the purposes of FDA's requlation of medical devices, the Axle Interspinous Fusion System modifications substantially meet the performance criteria established by the cleared parent device. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 27, 2013 X-Spine Systems, Incorporated % David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342 Re: K130438 Trade/Device Name: Axle™ Interspinous Fusion System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: May 30, 2013 Received: May 31, 2013 Dear Dr. Kirschman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ #### Page 2 - David Kirschman. M.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Erin-Keith For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Special 510(k): Device Modification Axle Interspinous Fusion System #### Indications for Use 510(k) Number (if known): K130438 Device Name: Axle™ Interspinous Fusion System Indications for Use: The Axle Interspinous Fusion System is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1 inclusive). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. The Axle Interspinous Fusion System is intended for use with bone graft material, and not for stand-alone use. Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ## Ronald P. Jean -S (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130438 X Page 15 of 47
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...