K151812 · Philips Medical Systems · MHX · Jan 26, 2016 · Cardiovascular
Device Facts
Record ID
K151812
Device Name
PATIENT MONITORS VARIOUS MODELS
Applicant
Philips Medical Systems
Product Code
MHX · Cardiovascular
Decision Date
Jan 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Efficia CM10, CM12, CM100, CM120 and CM150 patient monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The Efficia CM10, CM12, CM100, CM120 and CM150 patient monitors are intended for monitoring, analysis, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Device Story
Efficia CM10/12/100/120/150 are multi-parameter patient monitors for healthcare environments. Devices accept inputs via sensors for ECG, respiration (impedance/acoustic/capnography), NBP, SpO2/SpHb/SpCO, temperature, IBP, CO2, and cardiac output. Monitors process signals using established Philips and third-party (Masimo, Respironics) algorithms to perform arrhythmia detection, ST-segment analysis, and vital sign calculation. Outputs are displayed on 10-15 inch screens, providing real-time monitoring, recording, and alarming. Operated by healthcare professionals in clinical settings or during intra-facility transport. Output informs clinical decision-making regarding patient status and intervention. Benefits include continuous physiological surveillance and timely alerts for abnormal conditions.
Clinical Evidence
No clinical trials were conducted. Substantial equivalence is supported by bench testing, performance verification, and usability studies. Devices comply with IEC 60601-1, IEC 60601-2-27, ISO 80601-2-55, ISO 80601-2-56, ISO 80601-2-61, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62366, IEC 62304, IEC 80601-2-30, IEC 60601-2-34, IEC 60601-2-49, and ANSI/AAMI EC-13.
Technological Characteristics
Multi-parameter monitors with 10, 12, or 15-inch displays. Powered by AC with secondary battery. Materials identical to primary predicate. Algorithms: Philips proprietary (ECG, NBP, IBP, Temp), Masimo (SpO2, SpHb, SpCO, acoustic respiration), Respironics (mainstream CO2), and Philips ST/AR (arrhythmia/ST-segment). Connectivity includes standard patient monitoring interfaces. Compliant with IEC 60601 series, ISO 80601 series, and ANSI/AAMI EC-13.
Indications for Use
Indicated for healthcare professionals to monitor, analyze, record, and alarm physiological parameters in adult, pediatric, and neonatal patients within healthcare facilities, including intra-facility transport. Parameters include ECG, Respiration (Impedance, RRa, awRR), NBP, SpO2, SpHb, SpCO, Temperature, CO2, IBP, Cardiac Output, Arrhythmia, and ST-segment analysis. Contraindicated for transport outside the healthcare facility.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Philips SureSigns VM Series Patient Monitors (K123900)
Masimo Rainbow SET® Radical 7R CO-Oximeter and Accessories (K100428)
Philips ST/AR ST and Arrhythmia Software, Release K.O (K101521)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The text is centered horizontally. The background is white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2016
Philips Medical Systems Mary Kruitwagen OA and Regulatory Affairs 3000 Minuteman Road Andover, MA 01810-1099
Re: K151812
Trade/Device Name: Philips Efficia CM10 863301, Philips Efficia CM12 863303, Philips Efficia CM100 863300. Philips Efficia CM120 863302, Philips Efficia CM150 863304 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, MLD, DSJ, DSK, DXN, DRT, DQA, DSA, FLL, CCK, BZQ Dated: December 16, 2015 Received: December 21, 2015
Dear Mary Kruitwagen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
[signature]
for Pram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration | |
| <b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
| 510(k) Number (if known) | K151812 |
| Device Name | Philips Efficia CM10 863301, Philips Efficia CM12 863303, Philips Efficia CM100 863300, Philips Efficia CM120 863302<br>Philips Efficia CM150 863304 |
| Indications for Use (Describe) | The Efficia CM10, CM12, CM100, CM120 and CM150 patient monitors are indicated for use by health care<br>professionals whenever there is a need for monitoring the physiological parameters of patients. The Efficia CM10,<br>CM12, CM100, CM120 and CM150 patient monitors are intended for monitoring, analysis, recording and alarming<br>of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the<br>monitor may be used in transport situations within a healthcare facility. |
| | Parameter (patient types): ECG (Adult, Pediatric, Neonatal); Respiration - Impedance (Adult, Pediatric, Neonatal);<br>Respiration - RRa (Adult); Respiration - awRR (Adult, Pediatric, Neonatal); NBP (Adult, Pediatric, Neonatal); SpO2<br>(Adult, Pediatric, Neonatal); SpHb (Adult, Pediatric); SpCO (Adult, Pediatric); Temperature (Adult, Pediatric, Neonatal);<br>CO2 (Adult, Pediatric, Neonatal); IBP (Adult, Pediatric, Neonatal); Cardiac Output (Adult, Pediatric); Arrhythmia<br>(Cardiotach) (Adult, Pediatric, Neonatal); Arrhythmia (Basic) (Adult, Pediatric); Arrhythmia (Advanced) (Adult,<br>Pediatric); ST segment Analysis (Adult); |
| | Contraindications: Not for transport outside the healthcare facility |
| Type of Use (Select one or both, as applicable) | |
| | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
# Efficia CM10, CM12, CM100, CM120 and CM150 Patient Monitors 510(k) Summary
| Preparation Date | September 18, 2015 |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Type: | Traditional 510(k) |
| Regulatory Information<br>Classification | Physiological Monitor, Patient Monitor |
| Common/ Usual Name | Multi-parameter Patient monitor |
| Trade Name | Philips Efficia |
| Proprietary Name | Efficia CM10<br>Efficia CM12<br>Efficia CM100<br>Efficia CM120<br>Efficia CM150 |
| Models (with Reference numbers) | CM10 (863301)<br>CM12 (863303)<br>CM100 (863300)<br>CM120 (863302)<br>CM150 (863304) |
| Device Sponsor | Philips Medical Systems,<br>3000 Minuteman Rd,<br>Andover, MA, USA, 01810-1099 |
| Device Owner | Philips Medical Systems,<br>3000 Minuteman Rd,<br>Andover, MA, USA 01810-1099 |
| Establishment Registration number: | 1218950 |
| Sponsor/Manufacturer/Owner/Operator: | 1217116 |
| Contact | Mary Kruitwagen<br>Philips Medical Systems,<br>3000 Minuteman Rd,<br>Andover, MA, USA, 01810-1099<br>Phone: (978) 659-4932<br>Fax: 978-659-7323<br>mary.kruitwagen@philips.com |
{4}------------------------------------------------
Peter Schipelliti, Philips Medical Systems, 3000 Minuteman Rd, Andover, MA 01810-1099 USA Email: peter.schipelliti@philips.com Phone: (978) 659-4744 Fax: 978-659-7323
#### Device Classification:
The product codes and individual classification monographs applicable to the subject devices are listed below:
| Device Panel | Classification | ProCode | Description |
|-----------------------------------------|----------------|---------|-----------------------------------------------------------------------|
| Cardiovascular | 870.1025, II | MHX | Monitor, Physiological, Patient (with arrhythmia detection or alarms) |
| Cardiovascular | 870.1025, II | MLD | Monitor, St Segment With Alarm |
| Cardiovascular | 870.1110, II | DSJ | Alarm, Blood Pressure |
| Cardiovascular | 870.1110, II | DSK | Computer, Blood Pressure |
| Cardiovascular | 870.1130, II | DXN | System, Measurement, Blood Pressure, Non-Invasive |
| Cardiovascular | 870.2300, II | DRT | Monitor, Cardiac (incl. Cardiotachometer & Rate Alarm) |
| Cardiovascular | 870.2700, II | DQA | Oximeter |
| Cardiovascular | 870.2900, I | DSA | Cable, Transducer and Electrode, incl. Patient connector |
| Cardiovascular | 890.1175, II | IKD | cable, electrode |
| General Hospital and<br>Personal Use | 880.2910, II | FLL | Thermometer, Electronic, Clinical |
| Anesthesiology &<br>Respiratory Therapy | 868.1400, II | CCK | Analyzer, Gas, |
| Anesthesiology &<br>Respiratory Therapy | 868.2375, II | BZQ | Monitor, Breathing Frequency |
#### Device Description
The subject Philips Efficia CM10, CM120 and CM150 patient monitors are a family of physiological vital signs monitors. The monitors come with a variety of display sized (10 inch, 12 inch or 15 inch) and with a variety of measurements, some optional. The monitors primarily run on AC power with a secondary battery. The exterior material is the same as the primary predicate devices.
#### Fundamental Scientific Technology:
The subject Efficia CM10, CM120 and CM150 patient monitors include the following parameters (some optional): ECG, Respiration, NBP, SpO, Temperature, Invasive Blood Pressure, Cardiac Output, Arrhythmia Analysis, ST-Segment monitoring, and CO2 .monitoring.
- . The ECG algorithm is a Philips algorithm which is well established in the marketplace and is used in the primary predicate devices (K123900).
- Respiration rate can be measured through impedance technology, from capnography or via acoustic respiration. The Impedance respiration algorithm is the same algorithm as used in the primary predicate devices (K123900).
- Respiration from capnography, shown as awRR, is either from Microstream from the Oridion algorithm or main stream CO2.
- The Sidestream method is the same as in the primary predicate (K123900). The main stream CO2 including O respiration is from Respironics.
- The algorithm of the respirations from the mainstream CO₂ is from Respironics M2105A sensor (K042601 O and K050762) and is used unchanged in the subject devices.
- The acoustic respirations (awRR) measurement is achieved via the Masimo rainbow Set using their O algorithm unchanged and previously cleared (K100428).
{5}------------------------------------------------
- The NBP algorithm is proprietary Philips algorithm and is the same algorithm as in the primary predicate device (K123900).
- The subject devices offer two different SpO3 algorithms. The default SpO3 algorithm is a Philips proprietary algorithm and is unchanged from the primary predicate device (K123900). Optionally, the user can select the Masimo Set or Masimo rainbow Set. This option uses Masimo's algorithm unchanged from its predicate (K100428). SpHb and SpCO are part of Masimo's algorithm and are implemented unchanged from the Masimo predicate.
- . The temperature measurement is similar algorithm of primary predicate (K123900).
- The invasive blood pressure algorithm is a Philips algorithm which is unchanged from that in the primary predicate device (K123900).
- . The arrhythmia algorithm and the ST-Segment analysis algorithm used the Philips ST/AR algorithm and incorporate a subset of the STAR algorithm as cleared by K101521.
Performance data supports that the subject devices performance to same or better level of safety and effectiveness as the predicate devices. Clinical user preference and usability studies support that the customer requirements are met. The subject devices are in compliance with the following standards:
| IEC 60601-1 | IEC 60601-2-27 | ISO 80601-2-55 |
|------------------------------|----------------|-----------------|
| IEC 60601-1-2 | IEC 80601-2-30 | ISO 80601-2-56 |
| IEC 60601-1-6 | IEC 60601-2-34 | ISO 80601-2-61 |
| IEC 62366 included IEC 62304 | IEC 60601-2-49 | ANSI/AAMI EC-13 |
| IEC 60601-1-8 | | |
The predicates for the subject devices are:
| K123900 | 03/19/2013 | Philips SureSigns VM Series Patient Monitors |
|--------------------------|------------|-----------------------------------------------------------------------------|
| K050762 | 04/22/2005 | Philips MP20, MP30, MP40, MP50, MP60, MP70 MP90 IntellliVue Patient Monitor |
| K042601 | 11/19/2004 | Respironics CAPNOSTAT 5 CO2 sensor |
| K100428 | 07/09/2010 | Masimo Rainbow SET® Radical 7R CO-Oximeter and Accessories |
| K101521 | 07/25/2010 | Philips ST/AR ST and Arrhythmia Software, Release K.O |
| Primary 510(k) Predicate | | |
Philips SureSigns VM Series Patient Monitors
#### K123900 03/19/2013
## Indications for Use:
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The intended uses are for monitoring, analysis, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
| | Patient Types | | |
|-------------------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| ECG | √ | √ | √ |
| Respiration - Impedance | √ | √ | √ |
| Respiration - RRa | √ | | |
{6}------------------------------------------------
| | Patient Types | | |
|-------------------------|---------------|-----------|----------|
| Parameter | Adult | Pediatric | Neonatal |
| Respiration - awRR | √ | √ | √ |
| NBP | √ | √ | √ |
| SpO2 | √ | √ | √ |
| SpHb | √ | √ | |
| SpCO | √ | √ | |
| Temperature | √ | √ | √ |
| CO2 | √ | √ | √ |
| IBP | √ | √ | √ |
| Cardiac Output | √ | √ | |
| Arrhythmia (Cardiotach) | √ | √ | √ |
| Arrhythmia (Basic) | √ | √ | |
| Arrhythmia (Advanced) | √ | √ | |
| ST segment Analysis | √ | | |
Contraindications: Not for transport outside the healthcare facility
### Conclusion
The subject Efficia CM10, CM120 and CM150 patient monitors utilize the same fundamental scientific technology as the predicate devices on the market. The energy source is electrical with secondary battery back like the primary predicate. The subject Efficia CM10, CM120 and CM150 patient monitors use the same algorithms and have the similar performance to their predicates. They have the same or better level of safety and essential performance as compared to their predicate devices when reviewing the data of the device history file. It is our assessment that the subject devices demonstrate that they are substantially equivalent to the identified predicates and have the same level of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.