K151657 · Varian Medical Systems, Inc. · JAQ · Aug 7, 2015 · Radiology
Device Facts
Record ID
K151657
Device Name
Shielded Applicator Set, Cervical Stop
Applicant
Varian Medical Systems, Inc.
Product Code
JAQ · Radiology
Decision Date
Aug 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Shielded Applicator Set is intended for cancer treatment of the vagina, vaginal stump or rectum where partial shielding is required using HDR or PDR Brachytherapy. The Cervical Stop is indicated for use with Varian metallic intrauterine tandems/probes to provide a physical stop for depth of insertion of these applicators within the uterus during HDR or PDR brachytherapy.
Device Story
Shielded Applicator Set and Cervical Stop are accessories for Varian afterloader systems used in HDR/PDR brachytherapy. Applicator set includes cylinders, probes, and tungsten alloy shielding to deliver radiation to vaginal/rectal tumors while providing partial shielding. Cervical stop is a stainless steel/PEEK/titanium component providing a physical depth stop for intrauterine tandems/probes. Used in radiation oncology departments by clinicians. Devices are non-sterile, reusable, and require sterilization between uses. Output is physical positioning of radioactive sources; clinical benefit is precise tumor irradiation while sparing healthy tissue.
Clinical Evidence
No clinical data. Bench testing only. Testing confirmed correct function with afterloaders, mechanical durability, positional accuracy of radioactive sources, material radiation resistance, and sterilization efficacy. Usability assessed per IEC 62366:2007.
Technological Characteristics
Materials: Tungsten alloy (shielding), stainless steel (probes/screws), PMMA (cylinders), PEEK/Titanium (cervical stop), silicone (cleaning caps). Principle: Intracavitary radiation delivery with physical shielding. Dimensions: Cylinders 20-35mm diameter, 140mm length; probes 3mm diameter, 320mm length. Sterilization: Steam (probes/screws), PMMA parts non-sterile (use with medical condom). Connectivity: Mechanical interface with Varian afterloaders.
Indications for Use
Indicated for cancer treatment of the vagina, vaginal stump, or rectum (Shielded Applicator Set) and for use with Varian metallic intrauterine tandems/probes to provide a physical stop for insertion depth during HDR or PDR brachytherapy (Cervical Stop). Used in female and male patients.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2015
Varian Medical Systems, Inc. % Mr. Peter J. Coronado Director, Global Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304
Re: K151657
Trade/Device Name: Shielded Applicator Set-GM11004380. Cervical Stop-GM11000670 Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ Dated: June 17, 2015 Received: June 19, 2015
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151657
### Device Name
GM11004380 - Shielded Applicator Set GM11000670 - Cervical Stop
#### Indications for Use (Describe)
The Shielded Applicator Set is intended for cancer treatment of the vaginal stump or rectum where partial shielding is required using HDR or PDR Brachytherapy.
The Cervical Stop is indicated for use with Varian metallic intrauterine tandems/probes to provide a physical stop for depth of insertion of these applicators within the uterus during HDR or PDR brachytherapy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Varian Medical Systems, Inc. 3100 Hansen Way Palo Alto, CA 94304-1038 USA Tel +1 650 493 4000 www.varian.com
# Premarket Notification [510(k)] Summary
## GM11004380 Shielded Applicator Set GM11000670 Cervical Stop
The following information is provided following the format of 21 CFR 807.92(c).
| Submitter's Name: | Varian Medical Systems, Inc.<br>3100 Hansen Way E-110<br>Palo Alto, CA 94304 |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| | Contact Name: Peter J. Coronado<br>Phone: 650.424.6320<br>Fax: 650.646.9200 |
| | |
| | Date: 17 June 2015 |
| Proprietary Name: | Shielded Applicator Set - GM11004380<br>Cervical Stop- GM11000670 |
| Classification Name: | Remote controlled radionuclide applicator system<br>21 CFR 892.5700, Class II<br>Product Code: JAQ |
| | |
| Common/Usual Name: | Remote controlled radionuclide applicator system |
| | |
| Predicate Devices: | K033371 Intracavitary Brachytherapy Applicators<br>(Primary predicate: GM11004380 – Shielded Applicator Set<br>Secondary predicate: 1100404 – Cervix Applicator Set, Cervical Stopper) |
| Device Description: | The applicator set in this submission is designed to be used with all Varian<br>afterloaders to deliver high dose rate (HDR) and pulsed-dose-rate (PDR)<br>brachytherapy treatment for gynecological and rectal applications. The<br>intracavitary applicator will be used for the treatment of cancerous tumors, and<br>are designed to be inserted into a body cavity.<br><br>The cervical stop is a stainless steel component that is placed on the applicator<br>probe during HDR and PDR brachytherapy treatment of the uterus. It provides a<br>physical stop for depth of insertion. |
| Intended Use Statement -<br>Shielded Applicator Set | The Shielded Applicator Set is intended for cancer treatment of the vagina, vaginal<br>stump or rectum where partial shielding is required using HDR or PDR<br>Brachytherapy. |
| | |
| Intended Use Statement -<br>Cervical Stop | The Cervical Stop is indicated for use with Varian metallic intrauterine<br>tandems/probes to provide a physical stop for depth of insertion of these<br>applicators within the uterus during HDR or PDR brachytherapy. |
| | |
| | |
{4}------------------------------------------------
| Indications for Use<br>Statement –<br>Shielded Applicator Set | The Shielded Applicator Set is intended for cancer treatment of the vagina, vaginal<br>stump or rectum where partial shielding is required using HDR or PDR<br>Brachytherapy. |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>Statement<br>– Cervical Stop | The Cervical Stop is indicated for use with Varian metallic intrauterine<br>tandems/probes to provide a physical stop for depth of insertion of these<br>applicators within the uterus during HDR or PDR brachytherapy. |
{5}------------------------------------------------
## Technological Characteristics:
| | Shielded Applicator Set (K033371) | Modified Shielded Applicator Set |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Compatible<br>Afterloaders | VariSource Afterloader Systems | VariSource iX™<br>VariSource 200™<br>GammaMedplus iX™<br>GammaMedplus™ |
| Intended use | Shielded applicator is developed to treat<br>cancer of the vagina or rectum where<br>partial shielding is required. The max<br>implantation time for this applicator is 2<br>days | The Shielded Applicator Set is intended for<br>cancer treatment of the vagina, vaginal<br>stump or rectum where partial shielding is<br>required using HDR or PDR Brachytherapy. |
| Indications for Use | The applicator will be used with the<br>VariSource High Dose Rate Afterloaders to<br>deliver brachytherapy treatment for<br>gynecological and rectal applications. The<br>applicator will be used in medical<br>intracavitary for treatment of cancerous<br>tumors. | The Shielded Applicator Set is intended for<br>cancer treatment of the vagina, vaginal<br>stump or rectum where partial shielding is<br>required using HDR or PDR Brachytherapy. |
| Design | Cylinder:<br>- Ø 20, 23, 26, 30, 35mm<br>- length 140mm<br>Applicator probe:<br>- Ø 3.0mm, straight<br>- Length 320mm<br>Tungston Alloy Shielding:<br>- 90°, 180°<br>- Length 122.5mm<br>Marking Screw (shielded)<br>0°, 90°, 2 x 90°, 180°, 270°<br>Clamping Nut for marking screw<br>Clamping screw for probe, 3mm<br>Plexiglass Filler Piece, 90°<br>Yellow Cleaning Cap | Cylinder:<br>- Ø 20, 23, 26, 30, 35mm<br>- length 140mm<br>Applicator probe:<br>- Ø 3.0mm, straight<br>- Length 320mm<br>Tungston Alloy Shielding:<br>- 90°, 180°<br>- Length 122.5mm<br>Marking Screw (shielded)<br>0°, 90°, 2 x 90°, 180°, 270°<br>Clamping Nut for marking screw<br>Clamping screw for probe, 3mm<br>Plexiglass Filler Piece, 90°<br>Blue Cleaning Cap |
| Materials | - Shielding - Tungsten Alloy<br>- Marking Screw, Clamping nut/Screw<br>and applicator probe - Stainless Steel<br>- Cylinders - PMMA<br>- Cleaning cap- Silicone | - Shielding - Tungsten Alloy<br>- Marking Screw, Clamping nut/Screw and<br>applicator probe - Stainless Steel<br>- Cylinders - PMMA<br>- Cleaning cap- Silicone |
| Guide Tubes | - Source guide tube with locking<br>mechanism for 320mm applicators<br>GM plus<br>- VariSource Transfer Guide Tubes | - Source guide tube with locking<br>mechanism for 320mm applicators GM<br>plus (K141336)<br>ClickFit transfer guide tube, set of 4 for<br>i<br>VS 200 (K113766) |
| Optional accessories | - X-ray marker for 320mm applicators, 0.9mm, uncoded, GammaMedPlus<br>- X-ray marker probe, for 320mm applicators, uncoded, GammaMed 12i(t)<br>- VariSource X-ray marker wire<br>- Leak stop button<br>- VariSource Ruler<br>- Flexible fixation device for gynecological applicators<br>- Universal applicator clamping device (third party product)<br>- Source guide tube support, GammaMed units only | - X-ray marker BV, up to 320 mm applicators, uncoded, GammaMedPlus<br>- Leak stop button<br>- Measurement Ruler<br>- Measurement Marker Wire VS Measurement<br>- Measurement Marker Clip<br>- Flexible fixation device for gynecological applicators<br>- Universal applicator clamping device (third party product)<br>- Source guide tube support, GammaMed units only<br>- Length gauge, GM plus<br>- Cervical Stop<br>- Allen Wrench |
| Packing | Individual | Individual |
| Sterility | Non sterile | Non sterile |
| Sterilization method | Steam sterilization (applicator probe, tungsten shielding and marking screws)<br>EO Sterilization for PMMA parts | Steam sterilization (applicator probe and marking screws)<br>No sterilization of PMMA parts (use sterile medical condom) |
| | Use of yellow Cleaning CAP (GM110003260) | Use of dark blue Cleaning CAP (GM11010770) |
| Biocompatibility | Fully biocompatible only when used in combination with a sterile medical condom. | Fully biocompatible only when used in combination with a sterile medical condom. |
| Anatomical sites | The Shielded Applicator Set is used on female (vaginal and rectal) and male (rectal) patients. | The Shielded Applicator Set is used on female (vaginal and rectal) and male (rectal) patients.<br>Cervical Stop |
| Compatibility with the environment and other devices | N/A | Cervical Stop |
| Where used | The Shielded Applicator Set is intended to be used in a radiation therapy or radiation oncology department in a hospital environment. | The Shielded Applicator Set is intended to be used in a radiation therapy or radiation oncology department in a hospital environment. |
{6}------------------------------------------------
## Cervical Stop – GM11000670
| | Cervical Stopper<br>(part of K033371 – Cervix Applicator Set) | Cervical Stop |
|------------------------|---------------------------------------------------------------|----------------------------------|
| Compatible Applicators | Cervix Applicator Set | All Varian metallic intrauterine |
{7}------------------------------------------------
| | | tandems/probes |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Cervical Stopper is indicated for use<br>with the Cervix Applicator Set to provide a<br>physical stop for depth of insertion of<br>these applicators within the uterus during<br>HDR or PDR brachytherapy. | The Cervical Stop is indicated for use with<br>Varian metallic intrauterine<br>tandems/probes to provide a physical<br>stop for depth of insertion of these<br>applicators within the uterus during HDR<br>or PDR brachytherapy. |
| Indications for Use | The Cervical Stopper is indicated for use<br>with the Cervix Applicator to provide a<br>physical stop for depth of insertion of<br>these applicators within the uterus during<br>HDR or PDR brachytherapy. | The Cervical Stop is indicated for use with<br>Varian metallic intrauterine<br>tandems/probes to provide a physical<br>stop for depth of insertion of these<br>applicators within the uterus during HDR<br>or PDR brachytherapy. |
| Material | PEEK/Titanium | PEEK/Titanium |
| Sterility | Non sterile | Non sterile |
| Reusable | Yes | Yes |
| Design | Ø 16mm flattened to 10mm<br>Thickness 5.8mm | Ø 16mm flattened to 10mm<br>Thickness 5.8mm |
## Non Clinical Tests
Bench Testing has been performed to demonstrate that
- the applicator functions correctly with the specified afterloaders; .
- the devices can withstand the number of cycles of use that it will experience in its lifetime;
- the applicator enables the radioactive source to be located to the accuracy required
- the devices are constructed of materials that are not significantly affected by the radiation to which they • are exposed in the lifetime of the products;
- . the device components may be sterilized effectively (as appropriate)
- . the devices can be used and sterilized for the specified number of times
- the positional accuracy of the source within the applicator is adequate. ●
Usability was assessed to the requirements of IEC 62366:2007.
Results of Bench Testing showed conformance to applicable requirements and specifications
Clinical Tests No clinical tests have been included in this pre-market submission.
Conclusions All the tests that were performed met the applied pass criteria. Varian considers to be safe and effective and to perform as well or better than the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.