SOURCE GUIDE TUBE WITH LOCKING MECHANISM FOR 113 MM NEEDLES, GAMMAMEDPLUS
K141336 · Varian Medical Systems, Inc. · JAQ · Aug 6, 2014 · Radiology
Device Facts
Record ID
K141336
Device Name
SOURCE GUIDE TUBE WITH LOCKING MECHANISM FOR 113 MM NEEDLES, GAMMAMEDPLUS
Applicant
Varian Medical Systems, Inc.
Product Code
JAQ · Radiology
Decision Date
Aug 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GammaMedplus Source Guide Tubes are intended to connect between the GammaMedplus Remote Afterloader system and its range of applicators. This connection creates a conduit for the source wire to travel through and allows the radioactive isotopes to be positioned within the patient's tumor site
Device Story
GammaMedplus Source Guide Tubes serve as mechanical conduits in brachytherapy; they connect GammaMedplus Remote Afterloader systems to various applicators (probes, needles, interstitial tubes). The device provides a protected path for the radioactive source wire to travel from the afterloader to the patient's tumor site. Used in hospital brachytherapy treatment rooms by trained clinical staff (radiation oncologists, physicists, technicians). The device is a passive accessory; it does not process data or provide clinical outputs. It facilitates precise positioning of radioactive isotopes as defined by the treatment plan, enabling targeted radiotherapy while protecting the source wire and ensuring mechanical compatibility between the afterloader and the applicator.
Clinical Evidence
No clinical data included. Bench testing only. Testing confirmed functional compatibility with specified afterloaders/applicators, mechanical durability over the device lifetime, and material resistance to radiation exposure. Usability assessed per IEC 62366:2007.
Technological Characteristics
Non-sterile, passive mechanical accessory. Materials: FEP, stainless steel, PEEK, titanium, ABS/PP. Dimensions vary by applicator compatibility (113 mm to 500 mm). No energy source, software, or electronic components. Not in direct contact with body tissue.
Indications for Use
Indicated for use in brachytherapy procedures to connect remote afterloader systems to applicators for radioactive isotope delivery to tumor sites. No specific patient population limitations; usage restricted by maximal radiation dose to individual organs. Intended for use by trained radiation oncologists, physicians, radiologists, dosimetrists, medical physicists, and radiology technicians.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized representation of human figures.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 6, 2014
Varian Medical Systems, Inc. % Mr. Peter J. Coronado Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
Re: K141336
Trade/Device Name: GammaMedplus Source Guide Tubes Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ Dated: April 11, 2014 Received: May 23, 2014
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K141336
Device Name GammaMedplus Source Guide Tubes
#### Indications for Use (Describe)
The GammaMedplus Source Guide Tubes are intended to connect between the GammaMedplus Remote Afterloader system and its range of applicators. This connection creates a conduit for the source wire to travel through and allows the radioactive isotopes to be positioned within the patient's tumor site
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Michael D'Hara
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## PREMARKET NOTIFICATION
## 510(k) Summary
## GammaMedplus Source Guide Tubes
## As required by 21 CFR 807.92
| Submitter's Name: | Varian Medical Systems<br>3100 Hansen Way, m/s E-110<br>Palo Alto CA94304 | |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|--|
| | Contact Name: Peter J. Coronado | |
| | Phone: 650/424.6230<br>Fax: 650/ 646.9200 | |
| | e-mail: submissions.support@varian.com | |
| | Date: 15th May 2014 | |
| | | |
| Proprietary Name: | GammaMedplus Source Guide Tubes and PDR Quick Connectors | |
| Classification Name: | Remote controlled radionuclide applicator system<br>21CFR892.5700 | |
| | Class II | |
| Common/Usual Name: | Source Guide Tubes | |
| Product code: | JAQ | |
| Predicate Devices: | K113766 VariSource Transfer Guide Tube sets | |
| Device Description: | Source Guide Tubes are Brachytherapy applicator accessories. | |
| | Brachytherapy is a form of radiotherapy using Gamma rays from a | |
| | radioactive source placed at locations close to or within a tumor<br>or other treatment area to a predefined treatment plan. | |
| | They are designed to provide a path for the dummy and source | |
| | wire from the Afterloader to the Applicator. The applicator end of | |
| | a Source Guide Tube can vary in design to accommodate a range | |
| | of Applicators. | |
| | They are intended to be used by trained and qualified personnel | |
| | such as Radiation Oncologists, Physicians, Radiologists,<br>Dosimetrists, Medical Physicists, and Nurses/MTRAs/Radiology | |
| | Technicians/Radiographers in a hospital environment. | |
| Indications for Use: | The GammaMedplus Source Guide Tubes are intended to connect | |
| | between the GammaMedplus Remote Afterloader system and its | |
| | range of applicators. This connection creates a conduit for the | |
{4}------------------------------------------------
### source wire to travel through and allows the radioactive isotopes to be positioned within the patient´s tumor site.
Technological Characteristics:
| Predicate Device<br>Clearance Number: | K113766 | N/A |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Compatible<br>Afterloader | VariSource Remote HDR Afterloader<br>system | GammaMed afterloaders |
| Compatible<br>Applicators | All applicators and needles that are<br>compatible with VariSource series<br>afterloaders | All applicators and needles that are<br>compatible with GammaMed series<br>afterloaders |
| Intended use | The VariSource Reusable transfer<br>Guide Tubes are intended to connect<br>between the VariSource Remote<br>HDR Afterloader system and its<br>range of applicators. This connection<br>creates a conduit for the source wire<br>to travel through and allows<br>radioisotopes to be positioned within<br>the patient's tumour site. | The GammaMedplus Source Guide<br>Tubes are intended to connect<br>between the GammaMedplus<br>Remote Afterloader system and its<br>range of applicators. This connection<br>creates a conduit for the source wire<br>to travel through and allows the<br>radioactive isotopes to be<br>positioned within the patient's<br>tumour site. |
| Indications for Use | Same as above | Same as above |
| Target population | There is no limitation to patient<br>population. Limitations are on the<br>maximal radiation dose applicable to<br>the individual organs. The equipment<br>is designed to be used by a trained<br>and qualified Radiation Oncologist,<br>Physician, Radiologist, Dosimetrist,<br>Medical Physicist, and Nurse / MTRA<br>/ Radiology Technician /<br>Radiographer | There is no limitation to patient<br>population. Limitations are on the<br>maximal radiation dose applicable to<br>the individual organs. The<br>equipment is designed to be used by<br>a trained and qualified Radiation<br>Oncologist, Physician, Radiologist,<br>Dosimetrist, Medical Physicist, and<br>Nurse / MTRA / Radiology<br>Technician / Radiographer. |
| Design | Transfer Guide Tubes are designed to<br>provide a secure connection<br>between the VariSource afterloader<br>and applicators, needles or catheters<br>along which the source wire travels<br>during Brachytherapy. | The Source Guide Tubes and PDR<br>Quick Connectors were developed<br>to connect between the<br>GammaMed Afterloader systems to<br>applicators with an internal length<br>of:<br>113 mm, 180 mm, 200 mm, 250 mm<br>257 mm, 320 mm, 500 mm,<br>285 mm (Third party applicators),<br>187 mm (Adjustable Luer) |
| Materials | ABS, FEP, PEEK, Titanium, stainless<br>steel, ABS/PP.<br>Not in direct contact with body<br>tissue. | FEP, Stainless Steel, PEEK, Titanium<br>and ABS/PP.<br>Not in direct contact with body<br>tissue. |
| Packing | Individual | Individual |
| Sterility | Non sterile | Non sterile |
| Sterilization method | N/A | N/A |
| Biocompatibility | No body contact | No body contact |
| Mechanical safety | Protection caps | Protection caps |
| Anatomical sites | NA | NA |
| Compatibility with<br>the environment and<br>other devices | Compatible with a range of<br>applicators, e.g. probes, needles,<br>interstitial tubes | Compatible with a range of<br>applicators, e.g. probes, needles,<br>interstitial tubes |
| Where used | Brachytherapy treatment room | Brachytherapy treatment room |
{5}------------------------------------------------
#### Non Clinical Tests
Bench Testing has been performed to demonstrate that
- the device functions correctly with the specified afterloader and applicators
- . the device can withstand the number of cycles of use that it will experience in its lifetime;
- . the device is constructed of materials that are not significantly affected by the radiation to which they are exposed in the lifetime of the product;
Usability was assessed to the requirements of IEC 62366:2007. Results of Bench Testing showed conformance to applicable requirements and specifications
Clinical Tests No clinical tests have been included in this pre-market submission.
Conclusions All the tests that were performed met the applied pass criteria. Varian considers to be safe and effective and to perform as well or better than the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.