Masimo Root Vital Signs Monitoring System and Accessories
Applicant
Masimo Corporation
Product Code
MWI · Cardiovascular
Decision Date
Feb 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Masimo Root Monitoring System is indicated for use by healthcare professionals for the monitoring of multiple physiological parameters in healthcare environments. The Root Monitoring System, when used with the optional ISA module, is not intended to be used in road ambulances. The Masimo Root Monitoring System can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station). The optional Masimo Radical-7 Pulse CO-Oximeter and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical-7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices. The optional Masimo Radius-7 Wearable Pulse Oximeter and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and/or respiratory rate (RRa). The Masimo Radius-7 Wearable Pulse Oximeter and accessories are indicated for use with adult, and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, and hospital-type facilities. The optional ISA product family consists of three types of sidestream gas analyzers (ISA CO2, ISA AX+ and ISA OR+), intended to be connected to other medical backboard devices for monitoring of breath rate and the following breathing gases: ISA CO2: CO2 ISA AX+: CO2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane ISA OR+: CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane ISA CO2, ISA AX+ and ISA OR+ are intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The intended environment is the operating suite, intensive care unit and patient room. ISA CO2 is also intended to be used in road ambulances. The intended patient population is adult, pediatric and infant patients. The optional SEDLine Sedation Monitor is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The optional temperature module is indicated to measure temperature (oral, adult axillary, pediatric axillary, and rectal) of adult and pediatric patients. The device is intended to be used by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider. The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure in healthcare environments. The NIBP module is designed to measure blood pressure for patient population described in the following table: Newborn (neonate) Birth to 1 month of age; Infant 1 month to 2 years of age; Child 2 to 12 years of age; Adolescent 12-21 years of age; Adult 21 years of age and older.
Device Story
Root Monitoring System acts as central user interface/display for multiple physiological modules; inputs include SpO2, pulse rate, SpCO, SpMet, SpHb, RRa (via Radical-7/Radius-7), breathing gases (via ISA), EEG/PSI (via Sedline), temperature, and NIBP. Device transforms raw module data into visual/audible alarms and trend displays; operates via touchscreen; communicates with network systems (e.g., Patient SafetyNet) via wired/wireless interfaces. Used in hospitals, OR, ICU, and clinical labs by clinicians/medically qualified personnel. Output allows real-time patient status assessment; facilitates clinical decision-making regarding anesthesia, sedation, and vital sign stability. Benefits include centralized monitoring and integrated data management.
Clinical Evidence
No clinical data. Bench testing only, including alarm testing per IEC60601-1-8, usability testing per FDA Human Factors guidance, and software verification.
Technological Characteristics
Multifunctional patient monitor with LCD touchscreen. Powered by AC or internal rechargeable battery. Connectivity: Ethernet, Wi-Fi (802.11 a/b/g), Bluetooth 2.0, USB, SD card, nurse call. Interfaces: MOC-9, temperature probe port, NIBP port. Integrated modules for temperature and NIBP. Software-based alarm management compliant with IEC60601-1-8.
Indications for Use
Indicated for monitoring physiological parameters in healthcare environments. Patient population includes neonates, infants, children, adolescents, and adults. Specific modules indicated for anesthesia, recovery, respiratory care, and sedation monitoring. Contraindicated for use of ISA module in road ambulances.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 12, 2016
Masimo Corporation Ms. Marguerite Thomlinson Sr. Director, Regulatory Affairs 52 Discovery Irvine, California 92618
Re: K151644
Trade/Device Name: Masimo Root Vital Signs Monitoring System and Accessories Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, DXN, DQA, FLL Dated: January 13, 2016 Received: January 15, 2016
Dear Ms. Thomlinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Athel Stei
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Section 4. Indications for Use Statement
# Indications for Use
510(k) Number:
Device Name: Masimo Root Monitoring System and Accessories
The Masimo Root Monitoring System is indicated for use by healthcare professionals for the monitoring of multiple physiological parameters in healthcare environments. The Root Monitoring System, when used with the optional ISA module, is not intended to be used in road ambulances.
The Masimo Root Monitoring System can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).
The optional Masimo Radical-7 Pulse CO-Oximeter and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical-7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices.
The optional Masimo Radius-7 Wearable Pulse Oximeter and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and/or respiratory rate (RRa). The Masimo Radius-7 Wearable Pulse Oximeter and accessories are indicated for use with adult, and pediatric
| Prescription Use X | AND/OR Over The Counter Use |
|-----------------------------|--------------------------------|
| (Part 21 CFR 801 Subpart D) | (Part 21 CFR 801 Subpart D) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### Section 4. Indications for Use Statement
patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, and hospital-type facilities.
The optional ISA product family consists of three types of sidestream gas analyzers (ISA CO2. ISA AX+ and ISA OR+), intended to be connected to other medical backboard devices for monitoring of breath rate and the following breathing gases:
ISA CO2: CO2
ISA AX+: CO2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane
ISA OR+: CO2. O2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane
ISA CO2. ISA AX+ and ISA OR+ are intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The intended environment is the operating suite, intensive care unit and patient room. ISA CO2 is also intended to be used in road ambulances. The intended patient population is adult, pediatric and infant patients.
The optional SEDLine Sedation Monitor is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents.
The optional temperature module is indicated to measure temperature (oral, adult axillary, pediatric axillary, and rectal) of adult and pediatric patients. The device is intended to be used by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over The Counter Use (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
#### Indications for Use Statement Section 4.
The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure in healthcare environments. The NIBP module is designed to measure blood pressure for patient population described in the following table:
| Patient Population | Approximate Age Range |
|--------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
| Adult | 21 years of age and older |
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over The Counter Use (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{5}------------------------------------------------
# Section 5. 510(k) Summary
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>Phone: (949) 297-7683<br>FAX: (949) 297-7592 | |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date: | June 17, 2015 | |
| Contact: | Marguerite Thomlinson<br>Senior Director, Regulatory Affairs | |
| Trade Name: | Masimo Root Monitoring System and Accessories | |
| Common Name: | Patient Monitor | |
| Classification Regulation/<br>Product Code: | 21 CFR 878.2300, Class II/MWI<br>21 CFR 862.3220, Class II/JKS<br>21 CFR 868.1400, Class II/CCK<br>21 CFR 868.2375, Class II/BZQ<br>21 CFR 870.2700, Class II/DQA<br>21 CFR 870.2710, Class II/DPZ<br>21 CFR 882.1320, Class II/GXY<br>21 CFR 882.1400, Class II/GWQ<br>21 CFR 882.1400, Class II/OLT<br>21 CFR 882.1400, Class II/OLW<br>21 CFR 882.1400, Class II/OMC<br>21 CFR 882.1400, Class II/ORT<br>21 CFR 870.1130, Class II/DXN<br>21 CFR 880.2910, Class II/FLL | |
| Establishment Registration<br>Number: | 2031172 | |
| Reason for Premarket<br>Notification: | Device modification and new indications for use | |
| Predicate Devices: | K142394 – Masimo Root Monitoring System<br>K101680 - Welch Allyn Spot Ultra Vital Signs/Welch Allyn<br>Spot Vital Signs LXi<br>K090989 - Zoll R Series with NIBP & EtCO2 Options | |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. | |
{6}------------------------------------------------
# Device Description - Disease/Conditions that Device Diagnose, Treat, Prevent, Cure or Mitigate, Including Patient Population
The Root Monitoring System (Root) is a multifunctional device that monitors vital signs of patients from neonates to adults. Parameters monitored by Root include non-invasive functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), respiratory rate (RRa), inspired/expired gases during anesthesia, recovery and respiratory care, state of the brain by real-time data acquisition and processing of EEG signals, temperature, non-invasive blood pressure (NIBP) and Patient State Index (PSI) which is an EEG variable that is related to the effect of anesthetic agents.
# Explanation of Why Differences in Indication Statement Are Not Critical to Intended Use, and Why Difference Do Not Affect Safety and Effectiveness of Device When Used as Labeled
Root is intended to be used with the previously FDA cleared measurement technologies for the modules of:
- External Masimo Radical-7 Pulse CO-Oximeter (Radical-7 module), with cleared . technologies of SpO2, pulse rate, SpCO, SpMet, SpHb and RRa monitoring per K110028.
- . External Masimo Radius-7 Pulse Oximeter (Radius-7 module), with cleared technologies of SpO2, pulse rate and RRa monitoring per K110028.
- . External ISA-Infrared Sidestream Gas Analyzer (ISA module), with cleared technologies of breathing gases and respiratory rate monitoring per K103604.
- External Sedline Sedation Monitor with Frontal PSI and SEDTrace EEG Electrode Set (Sedline module), with cleared technologies of EEG and PSI monitoring per K051874.
- . Internal temperature module, with cleared temperature measurement technologies (the module was integrated in the Welch Allyn Spot Ultra Vital Signs/Welch Allyn Spot Vital Signs LXi per K101680).
- Internal non-invasive blood pressure (NIBP) module with cleared NIBP measurement . technologies (the module was integrated in Zoll R Series per K090989).
Root is also intended to be used as a user interface to facilitate access control and monitoring device functions and to connect system networks such as the Patient SafetyNet (K071047).
# Device Description - General Description from Labeling, Including Explanation of How Device Functions, Scientific Concepts that Form Basis For the Device
Root displays patient monitoring information from the connected modules. Visual alarms are shown on the Root display and audible alarms are generated through the Root internal speaker. The user accesses the connected modules' monitoring functions, using the Root
{7}------------------------------------------------
display. When the module is disconnected from Root, the monitoring information from the module is no longer displayed on Root.
Data from connected modules, including patient monitoring data, can be communicated to network systems. Root also functions as a pass-through means for communicating information between connected devices and network systems.
The subject device is the same as the primary predicate (K142394). The main difference is that the subject device has been modified to integrate internal modules for temperature and NIBP measurement technologies.
### Device Description - Significant Physical and Performance Characteristics of the Device
The significant physical characteristics for Root include an LCD touchscreen for patient monitoring. The instrument can be powered by AC or by its internal rechargeable battery. The approximate size and weight of the instrument are 11' x 10.5'' x 5.5'' (27.9 cm x 26.7 cm x 14 cm) and approximately 8 pounds.
The device specifications are shown below for the general functions of the subject device, Root.
| FEATURE | SPECIFICATION |
|---------------------------------------|-----------------------------------------------------|
| Display | Color LCD touchscreen |
| External Module | Parameter |
| Radical-7 | Parameters per K110028 |
| Radius-7 | Parameters per K110028 for SpO2, pulse rate and RRa |
| ISA | Parameters per K103604 |
| Sedline | Parameters per K051874 |
| Internal Module | |
| Temperature | Temperature parameters per K101680 |
| Non-Invasive Blood<br>Pressure (NIBP) | Arterial blood pressure parameters per K090989 |
| General | |
| Visual/audible alarm | IEC60601-1-8 compliant |
| Storage/recording | Trend/data storage |
| Electrical | |
| AC Power | 100-240 volt, 47-63 Hz |
| Battery | Rechargeable battery |
| Interface | |
| Root and Device/Module<br>Connection | Wired/docking interface |
| | Wireless interface |
| | MOC-9 interface |
| | Temperature Probe Port |
| | NIBP Port (Nib) |
| | Iris interface |
| | Nurse call interface |
| | USB interface<br>SD card interface |
| Network Connectivity | Ethernet<br>Wi-Fi, 802.11 a/b/g; Bluetooth 2.0 |
{8}------------------------------------------------
| FEATURE | SPECIFICATION |
|-----------------------|--------------------------------------------|
| Mechanical | |
| Dimensions | 11 x 10.5 x 5.5 inch (27.9 x 26.7 x 14 cm) |
| Weight | 3.37 kg |
| Environmental | |
| Operating Temperature | 32 to 122°F (0-50°C) |
| Storage Temperature | -40 to 158°F (-40 to 70°C) |
| Humidity | 10-95% non-condensing humidity |
| Mode of Operation | |
| Mode of Operation | Continuous |
### Intended Use
The Masimo Root Monitoring System is indicated for use by healthcare professionals for the monitoring of multiple physiological parameters in healthcare environments.
Root also serves as a convenient alternative user interface to integrate modules to provide health care professionals the ability to access, control and monitor measurement technologies (within the respective modules) that have been previously cleared by the FDA. Root does not affect the intended use for the cleared measurement technologies with which it is intended to function. Additionally, Root is intended to communicate with network systems.
### Indications For Use
The Masimo Root Monitoring System is indicated for use by healthcare professionals for the monitoring of multiple physiological parameters in healthcare environments. The Root Monitoring System, when used with the optional ISA module, is not intended to be used in road ambulances.
The Masimo Root Monitoring System can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).
The optional Masimo Radical-7 Pulse CO-Oximeter and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical-7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices.
The optional Masimo Radius-7 Wearable Pulse Oximeter and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial
{9}------------------------------------------------
hemoglobin (SpO2), pulse rate, and/or respiratory rate (RRa). The Masimo Radius-7 Wearable Pulse Oximeter and accessories are indicated for use with adult, and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, and hospital-type facilities.
The optional ISA product family consists of three types of sidestream gas analyzers (ISA CO2, ISA AX+ and ISA OR+), intended to be connected to other medical backboard devices for monitoring of breath rate and the following breathing gases:
ISA CO2: CO2
ISA AX+: CO>, NoO. Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane
ISA OR+: CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane
ISA CO2, ISA AX+ and ISA OR+ are intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The intended environment is the operating suite, intensive care unit and patient room. ISA CO2 is also intended to be used in road ambulances. The intended patient population is adult, pediatric and infant patients.
The optional SEDLine Sedation Monitor is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents.
The optional temperature module is indicated to measure temperature (oral, adult axillary, pediatric axillary, and rectal) of adult and pediatric patients. The device is intended to be used by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.
The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure in healthcare environments. The NIBP module is designed to measure blood pressure for patient population described in the following table:
| Patient Population | Approximate Age Range |
|--------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
| Adult | 21 years of age and older |
{10}------------------------------------------------
# Technological Characteristics
#### Principle of Operation
Root functions as a user interface that allows access, control and monitoring from the connected modules (external and internal).
Data from connected modules, including patient monitoring data, can be communicated to network systems. Root also functions as a pass-through means for communicating information between connected devices and network systems.
### Mechanism of Action for Achieving the Intended Effect
Root communicates with connected modules and displays the modules' patient monitoring information on the Root display. The healthcare provider controls the functions of each module using the Root touchscreen display. Visual alarms are shown on the Root display and audible alarms are generated through the Root internal speaker.
By connecting external modules or devices to Root, data can be communicated between Root and network systems via wired or wireless connection. Information from network systems can be shown on the Root display for viewing and notification purposes.
Once use is complete, the user then turns the power "off" for Root.
# Summary of Technological Characteristics of Subject Device Compared to Predicate Device
Similarities and Differences between the Subject Device, Root and the Primary Predicate Device, Root (K142394)
The subject device and the primary predicate have the following key similarities:
- Both have the same intended use as a patient monitoring device ●
- Both can function with the same external Radical-7, Radius-7, ISA and/or . Sedline modules;
- . Both have the same principle of operation and the mechanism of action for achieving the intended effect; and
- Both can communicate with a network system such as the Patient SafetyNet . (K071047) through wired or wireless connection.
The subject device and the primary predicate have the following key differences:
{11}------------------------------------------------
- . The subject device includes an internal temperature module; whereas the predicate does not; and
- . The subject device includes an internal non-invasive blood pressure (NIBP) module; whereas the predicate does not and
- The patient applied parts for the temperature and NIBP modules are connected ● to the Root; whereas the predicate does not have this feature.
Similarities and Differences between the Subject Device, Root and the Predicate Device, Welch Allyn Spot Ultra Vital Signs/Welch Allyn Spot Vital Signs LXi (K101680)
The subject device and the predicate have the following key similarities:
- Both have the same integrated temperature measurement module;
- Both have the same intended use and similar indications for use (IFU) for the ● temperature measurement and
- Both include the same temperature probes as accessories.
The subject device and the predicate have the following key differences:
- . The subject device and predicate have different IFU for their respective functions except for the IFU for temperature measurement and
- . Both have different user interfaces.
# Similarities and Differences between the Subject Device, Root and the Predicate Device, Zoll R Series with NIBP & EtCO2 Options (K090989)
The subject device and the predicate have the following key similarities:
- . Both have the same integrated NIBP measurement module;
- Both have the same intended use and similar indications for use for the NIBP ● measurement and
- Both include the same blood pressure cuffs and patient hoses as accessories. ●
The subject device and the predicate have the following key differences:
- . The subject device and predicate have different IFU for their respective functions except for the IFU for NIBP measurement and
- . Both have different user interfaces.
### Non-clinical Testing
Bench testing was performed on the subject device. The following non-clinical testing was performed in accordance with Masimo design control requirements and quality system to demonstrate substantial equivalence of the subject device to its predicates:
{12}------------------------------------------------
- Alarm testing per IEC60601-1-8
- . Usability testing per FDA Human Factors and Usability Draft Guidance
- Software verification per FDA Software Guidance ●
The results demonstrate that all requirements and performance specifications were satisfied, and that the subject device is substantially equivalent to the predicate.
### Clinical Testing
Not applicable. Clinical performance testing was not performed with the subject device, Root, to support substantial equivalence.
#### Conclusion
It is concluded that the subject device, Masimo Root Monitoring System, is substantially equivalent to its predicates with respect to safety and effectiveness, based on the nonclinical tests discussed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.