K151640 · Enraf-Nonius, B.V. · ITH · Oct 16, 2015 · Physical Medicine
Device Facts
Record ID
K151640
Device Name
ELTRAC 471
Applicant
Enraf-Nonius, B.V.
Product Code
ITH · Physical Medicine
Decision Date
Oct 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5900
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ELTRAC 471 Traction device, with its accessories, is intended to provide relief for the management of pain and symptoms from a variety of pressures on muscular or skeletal structures. The ELTRAC 471 Traction device may be used to treat pain and symptoms associated with the following conditions: herniated discs, bulging or protruding intervertebral discs, degenerative disease, posterior facet syndrome, acute facet problems, radicular pain, prolapsed discs, spinal root impingement, hypomobility, degenerative joint disease, facet syndrome, compression fractures, joint pain, and discogenic pain.
Device Story
Eltrac 471 is a powered traction device for clinical use by physiotherapists and rehabilitation professionals. It applies controlled mechanical forces to the spine to alleviate pain from musculoskeletal pressures. The device consists of a motor housed in a plastic unit with a touch-screen interface for parameter entry and protocol storage. Forces are transferred to the patient via cords and accessories. The system supports lumbar (face up) and cervical (lying) traction. Safety features include a patient-held stop switch that overrides software to release force, hardware-based force limitations, and a remote control for decreasing force. The device operates on 100-240V AC power. By providing controlled traction, it aims to reduce symptoms associated with spinal conditions. Healthcare providers use the interface to program treatment parameters, while the patient maintains control over immediate cessation of therapy via the stop switch.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including verification and validation of performance and functionality according to ISO 14971, IEC 60601-1, IEC 60601-1-2, and IEC 62304 standards.
Technological Characteristics
Powered traction equipment; plastic housing; motor-driven force application via cords/accessories; 100-240V 50/60Hz power; touch-screen interface; software-controlled force speed; hardware-based patient stop switch; electronic and mechanical force limiters. Standards: ISO 14971 (risk management), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software lifecycle).
Indications for Use
Indicated for patients requiring relief from pain and symptoms associated with herniated/prolapsed/bulging discs, degenerative disc/joint disease, posterior facet syndrome, acute facet problems, radicular pain, spinal root impingement, hypomobility, compression fractures, joint pain, and discogenic pain.
Regulatory Classification
Identification
Powered traction equipment consists of powered devices intended for medical purposes for use in conjunction with traction accessories, such as belts and harnesses, to exert therapeutic pulling forces on the patient's body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing the concept of health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2015
Enraf-Nonius B.V. % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K151640
Trade/Device Name: Eltrac 471 Regulation Number: 21 CFR 890.5900 Regulation Name: Power Traction Equipment Regulatory Class: Class II Product Code: ITH Dated: September 16, 2015 Received: September 17, 2015
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151640
Device Name Eltrac 471
#### Indications for Use (Describe)
The ELTRAC 471 Traction device, with its accessories, is intended to provide relief for the management of pain and symptoms from a variety of pressures on muscular or skeletal structures. The ELTRAC 471 Traction device may be used to treat pain and symptoms associated with the following conditions: hermiated discs, bulging or protruding intervertebral discs, degenerative disease, posterior facet syndrome, acute facet problems, radicular pain, prolapsed discs, spinal root impingement, hypomobility, degenerative joint disease, facet syndrome, compression fractures, joint pain, and discogenic pain.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| □ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for ENRAF NONIUS. The logo consists of the words "ENRAF NONIUS" stacked on top of each other, with a geometric shape above the words and a solid black shape below the words. Below the logo is the text "Partner for Life".
### 1. Date Prepared:
2015-09-09
#### 2. Owner:
Anne-Marie van Maurik, O.A. Manager Enraf-Nonius B.V. Vareseweg 127 3047 AT Rotterdam The Netherlands Phone: +31102030680 Email: Anne-Marie.van.Maurik@enraf-nonius.nl
#### 3. Contact person:
Amy Fowler DuVal & Associates, P.A. 825 Nicollet Mall, Suite 1820 Minneapolis, MN 55402 United States Phone: 612-338-7170, ext. 4 Email: fowler@duvalfdalaw.com
#### 4. Trade name:
Trade name: Common/Usual Name: Classification Name: Product code:
ELTRAC 471 Traction Device 890.5900 Powered traction equipment ITH
### 5. Predicate device:
The Triton/Tru-Trac/TX/Triton DTS Traction devices provide traction and mobilization of skeletal structures and skeletal muscles. They were cleared for marketing in K053223 on February 24th, 2006, under Product Code ITH
### 6. Device Description:
The Eltrac 471 traction device applies forces to the spine to alleviate pain caused by a variety of pressure on muscular or skeletal structures. It exerts defined forces on the human body to alleviate pain and utilizes computer technology for programming of different kinds of traction.
This device is intended for use by professional users only, such as qualified personnel in physiotherapy, rehabilitation and adjacent areas. The device is designed with a plastic housing that holds the motor and has a full color touch screen interface on which the professional user can enter various parameters by touch. The forces from the motor are transferred to the patient by cords and connected accessories. Sequential programming is possible for programming combinations of traction which can be stored as favorites for later use.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for ENRAF NONIUS. The logo consists of the words "ENRAF NONIUS" stacked on top of each other in a bold, sans-serif font. Above the words is a geometric shape that resembles a stylized hexagon with three horizontal lines above it. Below the words is the tagline "Partner for Life" in a smaller, italicized font.
The device has an on/off switch to power a medical power supply for 100-240V ~ 50/60Hz, and consumes a maximum of 50VA. The device has been designed for the standard US electric power sources and has a 100-240 VAC 50/60 Hz power supply.
To protect the patient from potential hazards, multiple mitigations are designed into the hardware and software. There are electronic safeguards of force and mechanical safety for maximum force limitation. Additional hardware mitigations are applied to protect the patient from excessive force. If no software is running, or if no patient stop is connected, the hardware is designed to lower the force. The patient stop must be held by the patient during therapy. The patient can press it to stop therapy immediately. If it is pressed, the force will lower to release the patient from the device. The patient stop is controlled in the hardware, and any software will be overruled if it is pressed. A remote control can be used as an option for decreasing the force delivered to the patient during treatment. As a safety feature, the remote control cannot increase the force.
#### 7. Intended Use:
The ELTRAC 471 Traction device, with its accessories, is intended to provide relief for the management of pain and symptoms from a variety of pressures on muscular or skeletal structures. The ELTRAC 471 Traction device may be used to treat pain and symptoms associated with the following conditions: herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet syndrome, acute facet problems, radicular pain, prolapsed discs, spinal root impingement, hypomobility, degenerative joint disease, facet syndrome, compression fractures, joint pain, and discogenic pain.
#### 8. Summary of Technological Characteristics of the Eltrac 471 compared to predicate device:
The intended uses of the Eltrac 471 and the predicate device are essentially still the same even though their indications for use statements have slightly different wording.
The technological characteristics of the Eltrac 471 and the predicate device are largely the same.
Both devices have an electrical input of 100-240V, 50/60Hz. Both devices employ software with touch-screen control to adjust settings, store treatment protocols, and provide reference information. Both devices utilize a software controlled motor to deliver force via a cord and accessories to apply traction to the patient.
Both devices are designed with safety/warning features so that cervical traction treatment decisions are made with care. In addition, treatment times
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for ENRAF NONIUS. The logo is black and white and features the company name in bold, uppercase letters. Above the company name is a geometric shape that resembles a stylized building or structure. Below the company name is the tagline "Partner for Life" in a smaller, italicized font.
are limited and the patient is able to stop the treatment at any time with a switch.
The Eltrac 471 contains a subset of the traction body positions that the predicate device allows (i.e., the Eltrac 471 does not add anything new or more in that regard). The Eltrac 471 allows for only Lumbar (Face up) or Cervical (Lying) traction, whereas the predicate device allows Lumbar (Face up), Lumbar (Face down) or Cervical (Lying). This difference with the Eltrac only means that the professional user cannot treat patients for face down lumbar traction. This difference is not raising any new technological characteristic with the Eltrac 471.
There are negligible differences between the Eltrac 471 and the predicate device in regards to traction force and base force. The Eltrac 471 and the predicate device are substantially equivalent in this regard.
The difference in the maximum treatment time between the Eltrac 471 to the predicate device also does not add any new or additional risks. The Eltrac 471 allows for 150 minutes total treatment time, whereas the predicate device allows 165 minutes. (For both devices, the normal treatment time will generally be around 20-30 minutes.) Further, total allowable treatment time can always be extended by the professional user using either the Eltrac 471 or the predicate device, as both allow the professional user to restart or extend a treatment time. Therefore, the difference in maximum treatment time between the Eltrac 471 and the predicate device does not add any new or additional risks to the predicate device (i.e., the Eltrac 471 is substantially equivalent to the predicate device in this regard).
The Eltrac 471 is also substantially equivalent to the predicate device in regards to force speed. The Eltrac 471 has a defined force speed, whereas the predicate device has an unknown force speed because of variables that influence the force speed, which are mostly the level of force and the elasticity of the patient. The predicate device's level of force varies because the higher the force, the more power is dissipated into heat, which results in a lower force speed (N/s). The elasticity of the patient is also a variable, through which a force set point is reached faster (i.e., higher force speed) when a patient has low elasticity compared to a patient with higher elasticity. The Eltrac 471 has a controlled force speed which is independent of the patient's elasticity; this aspect would not render the Eltrac 471 to not be substantially equivalent to the predicate device.
The Traction Force Hold Time and Traction Base Force Rest Time appear to be different, but the professional user is responsible for setting acceptable values for both the Eltrac 471 and the predicate device. When a very long Hold Time or Rest Time is set, it will become a Static Force. Both traction types are valid and can be used safely in both devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for ENRAF NONIUS. The logo consists of the words "ENRAF NONIUS" stacked on top of each other, with a geometric shape above the words and a horizontal line below the words. Below the line are the words "Partner for Life".
The differences between the Eltrac 471 and the identified predicate do not negatively affect the safety and effectiveness of the Eltrac 471 for the proposed indications for use, and most importantly, the Eltrac 471 is substantially equivalent to the predicate device despite these differences.
#### 9. Non-clinical Performance:
Verification and validation tests were executed by Enraf-Nonius, B.V. to establish the performance and functionality of ELTRAC 471. Several standards were utilized:
| - | ISO 14971 | Medical Devices – Application of risk management to<br>medical devices |
|---|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| - | IEC 60601-1 | Medical electrical equipment – Part 1: General<br>requirements for basic safety and essential<br>performance |
| - | IEC 60601-1-2 | Medical electrical equipment – Part 1-2: General<br>requirements for basic safety and essential<br>performance – Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests |
| - | IEC 62304 | Medical Device Software – Software Life Cycle<br>Process |
### 10. Clinical Performance:
Not applicable. This device does not require clinical testing for demonstration of substantial equivalence and safety/efficacy.
#### 11. Conclusion:
The Eltrac 471 traction device has the same intended use and technological characteristics as the predicate device. The testing done according to the company's internal standards and industry technical standards demonstrate that the Eltrac 471 meets the requirements for electrical and EMC safety, and meets the same safety and efficacy level of the predicate device.
In conclusion, the Eltrac 471 traction device is substantially equivalent to the Triton/Tru-Trac/TX/Triton DTS traction device (K053223).
**Enraf-Nonius B.V.**
Vareseweg 127
3047 AT Rotterdam
P.O. Box 12080
3004 GB Rotterdam
The Netherlands
+31-(0)10- 203 06 00
= +31-(0)10- 203 06 99
info@enraf-nonius.nl
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.