K053223 · Chattanooga Group · ITH · Feb 24, 2006 · Physical Medicine
Device Facts
Record ID
K053223
Device Name
TRITON/TRU-TRAC/TX/TRITON DTS TRACTION
Applicant
Chattanooga Group
Product Code
ITH · Physical Medicine
Decision Date
Feb 24, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Triton/ Tru-Trac/ TX/ Triton DTS Traction devices provide traction and mobilization of skeletal structures and skeletal muscles. The Triton/ Tru-Trac/ TX/ Triton DTS Traction devices provide a treatment in static, intermittent, progressive and cycling distraction forces to relieve pressures on micentifies that may be causing pain of skeletal or muscular origin (cervical, thoracic, lumbar, hip, wrist, shoulder). Therapeutic distraction can be applied in a variety of programmable patterns, cycles and functions. The Triton/ Tru-Trac/ TX/ Triton DTS Traction devices with the optional EMG (a.k.a. sEMG) biofeedback feature may be used to relieve peripheral radiation/sciatica and nain associated with: Protruding discs, Spinal root impingement, Bulging discs, Hypomobility, Herniated discs, Degenerative joint disease, Degenerative disc disease, Facet syndrome, Posterior facet syndrome, Compressions fractures, Acute facet problems, Joint pain, Radicular pain, Discogenic pain, Prolapsed discs. EMG (a.k.a. sEMG) Determination of the activation magnitude and timing of muscles for: a) retraining of muscle activation b) coordination of muscle activation Determination of the force produced by muscle for control and maintenance of muscle contractions and Relaxation muscle training Muscle re-education
Device Story
Powered traction device for hospitals, physical therapy, and chiropractic clinics; operated by licensed practitioners. Inputs include clinician-set traction parameters (force, time, cycle patterns) and sEMG signals for biofeedback. Device consists of electric traction unit, display, traction rope, and patient stop switch. Transforms inputs into static, intermittent, progressive, or cyclic distraction forces applied to patient via traction belts. EMG feature detects muscle activation/relaxation; clinician sets threshold via visual display; traction unit adjusts based on these thresholds. Output is mechanical distraction force and visual biofeedback data. Benefits include relief of pressure on skeletal/muscular structures causing pain and facilitation of muscle re-education/coordination.
Clinical Evidence
Bench testing only. Compliance with UL 60601-1 (safety) and IEC 60601-1-2 (electromagnetic compatibility) standards provided. No clinical data included.
Technological Characteristics
Powered traction equipment; electric traction unit with display, rope, and patient stop switch. Includes optional sEMG biofeedback. Standards: UL 60601-1 (2003), IEC 60601-1-2 (2001).
Indications for Use
Indicated for patients requiring traction/mobilization of skeletal structures/muscles to relieve pain of skeletal/muscular origin (cervical, thoracic, lumbar, hip, wrist, shoulder) associated with disc issues, spinal impingement, hypomobility, degenerative joint/disc disease, facet syndrome, compression fractures, and radicular/discogenic pain. EMG biofeedback indicated for muscle retraining, coordination, force control, and relaxation training.
Regulatory Classification
Identification
Powered traction equipment consists of powered devices intended for medical purposes for use in conjunction with traction accessories, such as belts and harnesses, to exert therapeutic pulling forces on the patient's body.
{0}------------------------------------------------
Page 1 of 3
| 510(k) Summary - K053223 | FEB 2 4 2006 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | Chattanooga Group,<br>A Division of Encore Medical, L.P.<br>4717 Adams Road<br>Hixson, TN 37343 |
| CONTACT PERSON: | Michael Treas<br>Manager of Regulatory Affairs<br>Phone: (423) 870-2281 ext. 7207<br>Fax: (423) 870-7404 |
| DATE PREPARED: | February 22, 2006 |
| DEVICE TRADE NAME: | Triton/ Tru-Trac/ TX/ Triton DTS Traction |
| COMMON/ USUAL NAME: | Powered Traction Equipment |
| CLASSIFICATION | Class II (Product Code ITH) |
| PREDICATE DEVICES | K051938 - Triton/ Tru-Trac/ TX Traction,<br>K844385 - Eskotek EST Trac 401 Traction,<br>K992733 - Akron ATP9,<br>K053503 - VAX-D Genesis System,<br>K031862 - SpineRx-LDM,<br>K001361 - Jilco,<br>K030060 - SpineMed,<br>K972487 - EMG Retrainer,<br>K881416 - MyoTrac |
#### DEVICE DESCRIPTION:
The Chattanooga Group powered traction device consists of an electric traction treatment unit, and may be sold with the following: EMG (a.k.a. sEMG) biofeedback feature, Pull-Pattern intermittent traction preset feature, patient traction treatment table device, positioning bolster accessories, patient traction belt accessories, high volt pulsed electrotherapy device and miscellaneous traction devices and/ or accessories. The device is intended to be used in hospitals, physical therapy clinics and chiropractic clinics for treatment in static, intermittent, progressive, regressive and cyclic, distraction forces to relieve pressures on structures that may be causing pain of skeletal or muscular origin (cervical, thoracic, lumbar, hip, wrist, shoulder). The device is restricted to sale by or on the order of a licensed physician or licensed practitioner.
The primary features of the Triton/ Tru-Trac/ TX/ Triton DTS Traction hardware are the traction unit (a.k.a. traction head), display screen, traction rope, mounting assembly and patient stop treatment switch.
K053223 Summary Page 1 of 3
{1}------------------------------------------------
K053223
Page 2 of 3
The EMG biofeedback feature is used to detect muscle activation and relaxation for Frie ENC oferous traction therapy. The treating clinician determines the minimum muscle activation or minimum muscle relaxation threshold set point. This is done by utilizing the visual display. The intermittent traction reacts to these threshold settings and administers prescribed traction treatment. The Pull Pattern feature is a preset of intermittent traction settings, less the setting of traction force, less setting of traction time. Force and time parameters are determined by the treating clinician. STATEMENT OF TECHHICAI. CHARACTERISTICS COMPARISON
A comparison of device features demonstrate that the Triton/ Tru-Trac/ TX/ Triton DTS Traction device is substantially equivalent to the following devices:
| Predicate device name and product code | 510(k) number |
|-------------------------------------------------------------------------------------------|---------------|
| Triton/ Tru-Trac/ TX Traction<br>ITH, Powered Traction Equipment | K051938 |
| Eskotek EST Trac 401 Traction<br>(a.k.a. Tru-Trac 401)<br>ITH, Powered Traction Equipment | K844385 |
| Akron ATP9<br>Traction Device<br>ITH Powered Traction Equipment | K992733 |
| VAX-D Genesis System<br>Traction Device<br>ITH, Powered Traction Equipment | K053503 |
| SpineRx-LDM<br>Traction Device<br>ITH, Powered Traction Equipment | K031862 |
| Jilco<br>Traction Device<br>ITH, Powered Traction Equipment | K001361 |
| SpineMed<br>Traction Device<br>ITH, Powered Traction Equipment | K030060 |
| EMG Retrainer<br>HCC, Biofeedback | K972487 |
| MyoTrac<br>HCC, Biofeedback | K881416 |
{2}------------------------------------------------
K05322}
Page 3 of 3
## Declarations of Conformity
The Triton/ Tru-Trac/ TX/ Triton DTS Traction devices is in compliance with the following FDA recognized Consensus Standards:
UL 60601-1: 2003, Standards for Medical Equipment Part 1: General Requirements for Safety, 1st Edition
IEC 60601-1-2: 2001, Medical Electrical Equipment Part 1 - 2: General requirements for Safety - Collateral Standard, Electromagnetic Compatibility --Requirements and Tests, 2nd Edition
## Truthful and Accurate Statement
A statement attesting to the truthfulness and accuracy of the information contained in this submission is attached in Section 3.
### Further Information
In the event that additional information is required, please contact:
Michael Treas Manager of Regulatory Affairs Chattanooga Group, A Division of Encore Medical, L.P. 4717 Adams Road Hixson, TN 37377 U.S.A. Telephone: (423) 870-2281 ext. 7207 Fax: (423) 870-7404 E-mail: Michael Treas(@ChattGroup.com
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an abstract design of an eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The seal is black and white and appears to be a simplified or stylized version of the official HHS seal.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 4 2006
Chattanooga Group c/o Mr. Michael Treas Manager of Regulatory Affairs 4717 Adams Road Hixson, Tennessee 37343
Re: K053223
Trade/Device Name: Triton/Tru-Trac/TX/Triton DTS Traction Regulation Number: 21 CFR 890.5900 Regulation Name: Powered traction equipment Regulatory Class: II Product Code: ITH Dated: February 7, 2006 Received: February 9, 2006
Dear Mr. Treas:
We have reviewed your Section 510(k) premarket notification of intent to market the device wo nare reveal above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass based in to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Errth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Treas
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA linding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
dell
Mark N. Melkerson, M.S. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): _ K 05 32 23 Device Name:
Indications for Use:
The Triton/ Tru-Trac/ TX/ Triton DTS Traction devices provide traction and mobilization of skeletal structures and skeletal muscles.
The Triton/ Tru-Trac/ TX/ Triton DTS Traction devices provide a treatment in static, intermittent, progressive and cycling distraction forces to relieve pressures on micentifies that may be causing pain of skeletal or muscular origin (cervical, thoracic, lumbar, hip, wrist, shoulder). Therapeutic distraction can be applied in a variety of programmable patterns, cycles and functions.
The Triton/ Tru-Trac/ TX/ Triton DTS Traction devices with the optional EMG (a.k.a. sEMG) biofeedback feature may be used to relieve peripheral radiation/sciatica and nain associated with:
| ➤ Protruding discs | ➤ Spinal root impingement |
|-----------------------------|------------------------------|
| ➤ Bulging discs | ➤ Hypomobility |
| ➤ Herniated discs | ➤ Degenerative joint disease |
| ➤ Degenerative disc disease | ➤ Facet syndrome |
| ➤ Posterior facet syndrome | ➤ Compressions fractures |
| ➤ Acute facet problems | ➤ Joint pain |
| ➤ Radicular pain | ➤ Discogenic pain |
| ➤ Prolapsed discs | |
EMG (a.k.a. sEMG)
Determination of the activation magnitude and timing of muscles for:
a) retraining of muscle activation
- b) coordination of muscle activation
Determination of the force produced by muscle for control and maintenance of muscle contractions
and the comments of the comments of the comments of the comments of the comments of the comments of
Relaxation muscle training
Muscle re-education
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
---------------------------------------------------------e Evaluation (ODE) Concurrence of CDRH, Qff
## Q.Y.M.D.L.
(Division of City Costs)
(Division Signe Off) Division of General, Restorative.
# and Neurological Devices
K A5 327 7
210/1.\ KT____LL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.