K151598 · Philips Medical Systems Nederland B.V. · OWB · Aug 17, 2015 · Radiology
Device Facts
Record ID
K151598
Device Name
VesselNavigator Rel. 1.0
Applicant
Philips Medical Systems Nederland B.V.
Product Code
OWB · Radiology
Decision Date
Aug 17, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
VesselNavigator provides image guidance by superimposing live fluoroscopic images on a 3D volume of the vessel anatomy to assist in catheter maneuvering and device placement. VesselNavigator is intended to assist in the treatment of endovascular diseases during procedures such as (but not limited to) AAA, TAA, carotid stenting, iliac interventions.
Device Story
VesselNavigator is an interventional software tool used with Philips Interventional X-ray systems; assists clinicians during endovascular procedures. Input: previously acquired 3D diagnostic images (CT/MR) and live 2D fluoroscopic X-ray signals. Operation: registers 3D volume to X-ray system; overlays live 2D fluoroscopy onto 3D roadmap; automatically adjusts 3D volume based on gantry/table movements. Features: vessel segmentation (user-editable), landmark placement, distance measurement, and view angle storage. Output: real-time 3D/2D fused visualization on interventional workstation. Used in clinical settings by physicians to navigate catheters/devices through vasculature. Benefits: improved anatomical visualization and guidance during complex endovascular interventions.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, software verification/validation, and usability testing.
Technological Characteristics
Software-based interventional tool running on Philips Interventional Workspot. Integrates with X-ray systems via DICOM. Features: 3D volume registration, automated gantry/table tracking, vessel segmentation, and measurement tools. Complies with IEC 62304, IEC 62366, ISO 14971, and NEMA PS 3.1-3.20.
Indications for Use
Indicated for patients undergoing endovascular interventions (e.g., AAA, TAA, carotid stenting, iliac interventions) requiring image guidance for catheter maneuvering and device placement. Excludes coronary and intracranial vessels.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 17, 2015
Philips Medical Systems Nederland B.V. % Liselotte Kornmann, Ph.D. Regulatory Affairs Manager Veenpluis 4-6 5684 PC Best THE NETHERLANDS
Re: K151598
Trade/Device Name: VesselNavigator Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, LLZ Dated: May 18, 2015 Received: June 12, 2015
Dear Dr. Kornmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K151598
Device Name
VesselNavigator
Indications for Use (Describe)
VesselNavigator provides image guidance by superimposing live fluoroscopic images on a 3D volume of the vessel anatomy to assist in catheter maneuvering and device placement.
VesselNavigator is intended to assist in the treatment of endovascular diseases during procedures such as (but not limited to) AAA, TAA, carotid stenting, iliac interventions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
{4}------------------------------------------------
{5}------------------------------------------------
# 510(k) summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | June 9, 2015 | |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277 | |
| Contact Person: | Ms. Liselotte Kornmann, PhD<br>Regulatory Affairs Manager<br>Phone: +31 611621238<br>E-mail: Liselotte.Kornmann@philips.com | |
| Device: | Trade Name: | VesselNavigator |
| | Device Name: | VesselNavigator |
| | Classification Name: | Image-intensified fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Primary Product Code:<br>Secondary Product Code: | OWB (Interventional x-ray system)<br>LLZ (system, image processing, radiological), |
| Primary Predicate<br>Device: | Trade Name: | MR-CT Roadmap Rel. 1 |
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K121772 (March 21, 2013) |
| | Classification Regulation: | 21 CFR, Part 892.1650 |
| | Classification Name: | Image-intensified fluoroscopic x-ray system |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | OWB (primary), JAK, LLZ (secondary) |
| Reference Device: | Trade Name: | HeartNavigator Release 2 |
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K140138 (June 10, 2014) |
| | Classification Regulation: | 21 CFR, Part 892.1650 |
| | Classification Name: | Image-intensified fluoroscopic x-ray system |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | OWB (primary), LLZ (secondary) |
{6}------------------------------------------------
#### Device description: VesselNavigator is a software product (Interventional Tool) intended to assist in the treatment of endovascular diseases during an endovascular intervention procedure. VesselNavigator is intended to be used in combination with a Philips Interventional X-ray system. VesselNavigator can be used during any endovascular intervention and covers all vascular anatomy except coronaries and intracranial vessels.
It provides live 3D image guidance for navigating endovascular devices through intended vascular structures in the body, reusing previously acquired diagnostic 3D images. After registration, the 3D volume can be used as a 3D roadmap for navigation; live 2D fluoroscopic images will be overlaid on the 3D volume. In addition, VesselNavigator provides tools to segment the relevant vasculature in the 3D volume (where the end-user is able to edit the segmentation results), place landmarks for easy recognition of key anatomical points of interests, and store and recall of preferred view angles.
Indications for Use: VesselNavigator provides image guidance by superimposing live fluoroscopic images on a 3D volume of the vessel anatomy to assist in catheter maneuvering and device placement.
> VesselNavigator is intended to assist in the treatment of endovascular diseases during procedures such as (but not limited to) AAA, TAA, carotid stenting, iliac interventions.
The Indications for Use statement for VesselNavigator is not identical to the predicate device; compared to MR-CT Roadmap, VesselNavigator is intended to assist in the treatment of endovascular diseases during procedures. However, this difference does not alter the intended use of the device nor does it affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate device have the same intended use, by superimposing live fluoroscopic images on a 3D volume of the vessel anatomy and are accessories to the Allura Xper X-ray system (K133292), which is an Interventional X-ray imaging system.
{7}------------------------------------------------
Technological VesselNavigator employs comparable technology as its predicate device MR-CT characteristics: Roadmap:
- Both tools run on a separate interventional tools workstation (Philips ● Interventional Workspot, K121296).
- . Both tools provide basic viewing operations to manipulate image data.
- . Both tools have comparable snapshot functionality.
- Both tools provide an overlay of live 2D fluoroscopy images on a 3D . reconstruction (3D volume) of the vessel tree acquired from a previous MR or CT scan; in this way the 3D volume serves as a roadmap to provide additional information for navigating endovascular devices through vascular structures of the patient.
- Both tools support registration of 3D volume with the x-ray system.
- . Both tools support manual correction of the registration.
- Both tools provide dynamic update to changes of the position of the X-rav equipment; the 3D volume is automatically adjusted to any gantry changes and any lateral or longitudinal table movements.
- Both tools have functionality to store roadmaps.
- . Both tools can be controlled from the tableside.
The technological differences between VesselNavigator and its predicate device MR-CT Roadmap are noted below
- VesselNavigator provides functionality to measure distances. This function ● is identical to the reference device HeartNavigator Release 2.
- . VesselNavigator provides functionality to segment the relevant vasculature in the 3D volume (where the end-user is able to edit the segmentation results), place landmarks for easy recognition of key anatomical points of interests, and store and recall of preferred view angles.
- . VesselNavigator tools to remove the table and segment and remove specific bone structures from the view
As these differences are considered low risk (only providing further support to the clinicians in performing interventions) and the functionalities were verified and validated with equivalent methods, these differences do not raise new questions on safety or effectiveness.
Therefore, the VesselNavigator is substantially equivalent to the currently marketed predicate device MR-CT Roadmap in terms of technological characteristics.
{8}------------------------------------------------
Summary of Nonclinical Performance Data:
Non-clinical performance testing has been performed on VesselNavigator and demonstrates compliance with the following International and FDA-recognized consensus standards and FDA guidance document:
- IEC 62304 Medical device software Software life cycle processes (Ed. 1.0, ● 2006),
- . IEC 62366 Medical devices - Application of usability engineering to medical devices (Ed. 1.0, 2007),
- ISO 14971 Medical devices - Application of risk management to medical devices (Ed. 2.0. 2007).
- . NEMA PS 3.1-3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2011), and
- Guidance for Industry and FDA Staff Guidance for the Content of . Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005, document number 337).
Software verification testing has been performed to cover the main system level requirements as well as the identified hazard mitigations. Software validation testing included testing of the vessel segmentation tool, the intended use and commercial claims, and usability testing with representative intended users. All of these tests were used to support substantial equivalence of the subject device.
The test results in this 510(k) premarket notification demonstrate that VesselNavigator:
- complies with the aforementioned international and FDA-recognized consensus standards and FDA guidance document, and
- . meets the acceptance criteria and is adequate for its intended use.
Therefore, VesselNavigator is substantially equivalent to the currently marketed predicate device MR-CT Roadmap in terms of safety and effectiveness.
Summary of Clinical The subject of this premarket submission, VesselNavigator, did not require Performance Data: clinical studies to support substantial equivalence.
Substantial The VesselNavigator software medical device is substantially equivalent to the Equivalence currently marketed predicate device MR-CT Roadmap in terms of design Conclusion: features, fundamental scientific technology, indications for use, and safety and effectiveness. The (non-)clinical performance tests provided in this 510(k) premarket notification demonstrates that the proposed VesselNavigator is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.