Patients with aneurysms (2-22 mm) not located on crossing vessels
Not applicable for this study
Segmentation success rate (average 73%)
Indications for Use
Allura 3D-RA Rel. 6: The Allura 3D-RA Rel. 6 is intended to assist physicians in analyzing two dimensional X-ray images by creating three dimensional views from sets of two dimensional images created during rotational angiographic runs. The Allura 3D-RA Rel. 6 is intended to assist in the diagnosis and treatment of endovascular diseases, for example, stenosis, aneurysms, and arteriovenous malformations. The Allura 3D-RA Rel. 6 also supports measurement of lesion dimensions and anatomical distances. 3D Roadmap Rel. 1: 3D Roadmap Rel. 1 is an extension of Allura 3D-RA, which provides image guidance by superimposing live fluoroscopic images on a 3D reconstruction of the vessel anatomy to assist in catheter maneuvering. MR-CT Roadmap Rel. 1: MR-CT Roadmap Rel. 1 is an extension of Allura 3D-RA, which provides image guidance by superimposing live fluoroscopic images on a 3D reconstruction of the vessel anatomy to assist in catheter maneuvering.
Device Story
Software suite (Allura 3D-RA Rel. 6, 3D Roadmap Rel. 1, MR-CT Roadmap Rel. 1) used with Allura X-ray systems. Allura 3D-RA generates 3D reconstructions from 2D rotational angiographic X-ray images to visualize vascular/bone anatomy and measure dimensions. 3D Roadmap and MR-CT Roadmap provide live 3D image guidance by overlaying live 2D fluoroscopic images onto 3D reconstructions (from 3D-RA or previously acquired MRA/CTA data). Used by physicians in clinical settings during endovascular procedures to assist in catheter navigation and diagnosis. Output displayed to physician to aid real-time decision-making and intervention accuracy.
Clinical Evidence
Retrospective in-house validation study using 3DRA clinical datasets with aneurysms. Performance validated by three experts (including an interventional neuroradiologist). Aneurysm tool success rate for segmentation was 73% (95% CI: 60%, 100%) for aneurysms 2-22 mm not located on crossing vessels.
Technological Characteristics
Software-based interventional tools. Operates on Allura X-ray system platform. Provides 3D reconstruction, visualization, and overlay guidance. No specific hardware materials or energy sources defined as these are software applications.
Indications for Use
Indicated for physicians to assist in the diagnosis and treatment of endovascular diseases (e.g., stenosis, aneurysms, arteriovenous malformations) and to provide image guidance for catheter maneuvering during endovascular interventions in any anatomy, including cerebral, abdominal, and peripheral vasculature.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K121772
Page 1 of 4
# MAR 2 1 2013
## 510(k) Summary of Safety and Effectiveness
3D Interventional tools software medical device
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR § 807.92.
Date prepared: March, 1 2013
## ldentification of manufacturer
| Company: | Philips Medical Systems Nederland B.V |
|----------------------|--------------------------------------------------|
| Address: | Veenpluis 4-6,<br>5684-PC, Best, The Netherlands |
| Registration number: | 3003768277 |
#### Identification of U.S. designated agent
| Company: | Philips Medical Systems |
|----------------------|-----------------------------------------------------------------|
| Address: | 22100 Bothell Everett Highway<br>Bothell, WA 98021-8431, U.S.A. |
| Registration number: | 1217116 |
#### Identification of official correspondent
| Name: | Liselotte Kornmann, PhD<br>Regulatory Affairs Manager |
|----------|--------------------------------------------------------------------------------------------|
| Address: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6,<br>5684-PC, Best, The Netherlands |
| Phone: | +31 6 11 621238 |
| Fax: | +31 40 2769100 |
| E-mail: | Liselotte.Kornmann@philips.com. |
#### Device identification
Regulation description: Regulation number:
Trade name: Device name:
Product code:
Class:
Allura 3D-RA, 3D Roadmap, MR-CT Roadmap Allura 3D-RA. Rel. 6, 3D Roadmap Rel. 1, MR-CT Roadmap Rel. 1 Image-intensified fluoroscopic x-ray system 21CFR 892.1650 OWB, JAK, LLZ
#### Legally marketed devices
| Trade names: | Allura 3D-RA Rel. 4.2 |
|-----------------|----------------------------|
| Manufacturer: | Philips Medical Systems |
| 510(k) numbers: | K040254 - Feb 19, 2004 |
| Trade names: | Innova Vision Applications |
| Manufacturer: | GE Medical Systems |
| 510(k) numbers: | K092639 - Dec 2, 2009 |
{1}------------------------------------------------
K121772
### Device description
The device description of the 3D Interventional Tools software medical devices that are subject of this 510(k) submission are as follows:
### Allura 3D-RA Rel. 6 software medical device (Allura 3D-RA Rel. 6)
The Allura 3D-RA Rel. 6 is a software product (Interventional Tool) intended to provide high-speed and high resolution 3D visualization of vessels and bones anatomy. Allura 3D-RA is intended to be used in combination with an Allura X-ray system.
The Allura 3D-RA Rel. 6 generates a 3D reconstruction from 2D X-ray images to visualize vascular anatomy, and helps to identify vascular pathologies from a single contrast-enhanced rotational angiogram. It can be used during anv angiography procedure and can cover any anatomical area including cerebral, abdominal, and peripheral. It also allows for 3D visualization of the spine.
#### 3D Roadmap Rel. 1 software medical device (3D Roadmap Rel. 1)
The 3D Roadmap Rel. 1 is a software product intended (Interventional Tool) to assist the physician during complex interventions by providing live 3D image guidance for navigating endovascular devices through vascular structures anywhere in the body. 3D Roadmap is intended to be used in combination with an Allura X-ray system.
The 3D Roadmap Rel. 1 overlays live 2D fluoroscopic images on a 3D reconstruction of the vessel tree processed by Allura 3D-RA Rel. 6 and therefore, assist the physician in catheter maneuvering. It can be used during any endovascular intervention and can cover any anatomy, including cerebral, abdominal, and peripheral vasculature.
### MR-CT Roadmap Rel. 1 software medical device (MR-CT Roadmap Rel. 1)
The MR-CT Roadmap Rel.1 is a software product intended to provide live 3D image guidance for navigating endovascular devices through vascular structures anywhere in the body, reusing segmented MRA or CTA data that has been acquired previously. The MR-CT Roadmap Rel.1 is intended to be used in combination with an Allura X-ray system.
The MR-CT Roadmap Rel.1 overlays live 2D fluoroscopic images on a previously acquired MRA or CTA volume, which is registered by the Allura 3D-RA Rel. 6 with the X-ray system using a rotational scan and therefore, assist the physician in catheter maneuvering. It can be used during any endovascular intervention and can cover any anatomy, including for example cerebral, abdominal, and peripheral vasculature.
{2}------------------------------------------------
K12/772
Page 3 of 4
### Indications for Use:
The indications for use of the 3D Interventional Tools software medical devices are as follows:
#### Allura 3D-RA Rel. 6
- The Allura 3D-RA Rel. 6 is intended to assist physicians in analyzing two . dimensional X-ray images by creating three dimensional views from sets of two dimensional images created during rotational angiographic runs.
- The Allura 3D-RA Rel. 6 is intended to assist in the diagnosis and treatment of . endovascular diseases, for example, stenosis, aneurysms, and arteriovenous malformations.
- The Allura 3D-RA Rel. 6 also supports measurement of lesion dimensions and . anatomical distances.
#### 3D Roadmap Rel.1
- 3D Roadmap Rel. 1 is an extension of Allura 3D-RA. which provides image . quidance by superimposing live fluoroscopic images on a 3D reconstruction of the vessel anatomy to assist in catheter maneuvering.
- The Allura 3D-RA Rel. 6 is intended to assist in the diagnosis and treatment of . endovascular diseases, for example, stenosis, aneurysms, and arteriovenous malformations.
#### MR-CT Roadmap Rel.1
- MR-CT Roadmap Rel. 1 is an extension of Allura 3D-RA, which provides image . guidance by superimposing live fluoroscopic images on a 3D reconstruction of the vessel anatomy to assist in catheter maneuvering.
- The Allura 3D-RA Rel. 6 is intended to assist in the diagnosis and treatment of . endovascular diseases, for example, stenosis, aneurysms, and arteriovenous malformations.
### Technological characteristics
The technological characteristics of the 3D Interventional Tools software medical devices are as follows:
The Allura 3D-RA Rel. 6, 3D Roadmap Rel. 1 and MR-CT Roadmap Rel. 1 software medical device applications are extensions of the currently marketed and predicate Allura 3D-RA rel 4.2 with new functionality that are built re-using the same technology. They are provided on the hosting software functionality platform of the currently marketed and predicate Allura 3D-RA Rel 4.2. Please note that upon receiving 510(k) clearance (K121296) from FDA for the independent hosting software functionality platform, Interventional Workspot, which is currently pending with FDA, the Allura 3D-RA Rel. 6, 3D Roadmap Rel. 1 and MR-CT Roadmap Rel. 1, will be provided as separate software medical devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a handwritten string of characters, "K121772". The characters are written in a cursive style, with the "K" being the largest and most distinct character. Below the string of characters, the text "Page 4 of 4" is printed in a smaller font size.
### Summary of testing
Verification and validation tests were performed with reqards to Intended Use. Claims, Requirement specifications and Risk management requirements. The test results demonstrate that the Allura 3D-RA Rel. 6, 3D Roadmap Rel. 1 and MR-CT Roadmap Rel. 1 software medical devices comply with all the requirements, including international recognized standards as detailed in this premarket submission and met the acceptance criteria.
This premarket submission included an in-house validation study to assess the performance of the aneurysm tool in segmentation of the aneurysm. In this retrospective study performed on 3DRA clinical datasets with aneurysms, the performance of the tool was validated by three experts; one of the experts was an interventional neuro radiologist. The aneurysm tool can be used successfully for aneurysms between 2 and 22 mm that are not located on crossing vessels. According to the validation study, the average success rate of segmentation is 73% with a 95% confidence interval of (60%, 100%].
### Conclusion:
The 3D Interventional Tools software medical devices are substantially equivalent to the currently marketed and predicate devices, based on the similar design functionality, indications for use and software requirements.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is smaller than the caduceus symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2013
Philips Medical Systems Nederland B.V. % Liselotte Kornmann, Ph.D. Regulatory Affairs Manager Veenpluis 4-6 5684-PC, BEST THE NETHERLANDS
Re: K121772
Trade/Device Name: Allura 3D-RA Rel. 6, 3D Roadmap Rel. 1, MR-CT Roadmap Rel. 1 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAK, LLZ Dated: February 28, 2013 Received: March 05, 2013
Dear Dr. Kornmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Dr. Kornmann
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K121772
Device Name: 3D Roadmap Rel. 1
Indications For Use:
The 3D Roadmap Rel. 1 is an extension of Allura 3D-RA, which provides image guidance by superimposing live fluoroscopic images on a 3D reconstruction of the vessel anatomy to assist in catheter maneuvering.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Division of Radiological Health 510(k): K121772
Page 1 of 1
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K121772
Device Name: Allura 3D-RA Rel. 6
Indications For Use:
The Allura 3D-RA Rel. 6 is intended to assist physicians in analyzing two dimensional X-ray images by creating three dimensional views from sets of two dimensional images created during rotational angiographic runs.
The Allura 3D-RA Rel. 6 is intended to assist in the diagnosis and treatment of endovascular diseases, for example, stenosis, aneurysms, and arteriovenous malformations.
The Allura 3D-RA Rel.6 also supports measurement of lesion dimensions and anatomical distances.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Division of Radiological Health 510(k): K121772
Page 1 of 1
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known): K121772
Device Name: MR-CT Roadmap Rel. 1
Indications For Use:
The MR-CT Roadmap Rel. 1 is an extension of Allura 3D-RA, which provides image guidance by superimposing live fluoroscopic images on a 3D reconstruction of the vessel anatomy to assist in catheter maneuvering.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OlR)
್ತು
Division of Radiological Health 510(k): K121772
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.