K151501 · Custom Orthopaedic Solutions, Inc. · LLZ · Oct 2, 2015 · Radiology
Device Facts
Record ID
K151501
Device Name
OrthoVis Web Portal
Applicant
Custom Orthopaedic Solutions, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The OrthoVis Web Portal is intended for use as a software interface and for the transfer of imaging information from a medical scanner such as a CT scanner. It is also intended as pre-operative software for simulating implant placement and surgical treatment options. The OrthoVis Web Portal is intended for use with the Glenoid Intelligent Reusable Instrument system and with the OrthoVis Preoperative Plan. It is also intended for use with the Arthrex and DePuy shoulder replacement implant systems listed below. - Arthrex Univers Apex - Arthrex Univers II - Arthrex Univers Revers - DePuy Global AP - DePuy Global StepTech - DePuy Delta Xtend Reverse The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended for use to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant.
Device Story
OrthoVis Web Portal acts as a secure software interface for transferring medical images (CT scans) between surgeons and Custom Orthopaedic Solutions (COS) technicians; facilitates preoperative planning for total shoulder replacement. Input: medical images uploaded by surgeon; Output: 3D preoperative plan for surgeon review/approval. Workflow: surgeon uploads images via web portal; COS technicians download images via SFTP to create plans; plans uploaded back to portal; surgeon reviews/interacts with 3D plan via web browser; surgeon approves or requests adjustments; final approved plan downloaded by surgeon. Replaces manual processes (CD/DVD mailers, phone calls, screen sharing). Benefits: streamlined communication, secure data transfer, improved preoperative planning efficiency. Used in clinical settings by surgeons and technicians.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical testing: software verification and validation, dimensional validation, and simulated use comparison. Automated unit tests, integration testing (WebDriver), and manual code inspection verified UI functionality, constraint enforcement, and error handling.
Technological Characteristics
Web-based software portal; utilizes HTTPS/SSH for secure network communication and SFTP for file transfer. Employs an encrypted database for case storage. Uses the same image processing/display code as cleared OrthoVis Shoulder software (K123122, K133367). Operates via standard web browsers (Chrome, Firefox).
Indications for Use
Indicated for use as a software interface for transferring medical imaging (e.g., CT scans) and as pre-operative software for simulating/evaluating implant placement and surgical treatment options for patients undergoing total shoulder replacement using specified Arthrex or DePuy systems.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Custom Orthopaedics Solutions, Incorporated Mr. Keith Grafmeyer Product Engineer 10000 Cedar Avenue Cleveland, Ohio 44106
October 2, 2015
Re: K151501
Trade/Device Name: OrthoVis Web Portal Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ, KWS, PHX Dated: May 26, 2015 Received: June 4, 2015
Dear Mr. Grafmeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151501
Device Name OrthoVis Web Portal
#### Indications for Use (Describe)
The OrthoVis Web Portal is intended for use as a software interface and for the transfer of imaging information from a medical scanner such as a CT scanner. It is also intended as pre-operative software for simulating implant placement and surgical treatment options. The OrthoVis Web Portal is intended for use with the Glenoid Intelligent Reusable Instrument system and with the OrthoVis Preoperative Plan. It is also intended for use with the Arthrex and DePuy shoulder replacement implant systems listed below.
- · Arthrex Univers Apex
- · Arthrex Univers II
- · Arthrex Univers Revers
- · DePuy Global AP
- DePuy Global StepTech
- · DePuy Delta Xtend Reverse
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended for use to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k)
Summary
| 510(k) SPONSOR / MANUFACTURER: | Custom Orthopaedic Solutions, Inc.<br>10000 Cedar Avenue<br>Cleveland, Ohio<br>44106 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Keith Grafmeyer<br>Product Engineer<br>Custom Orthopaedic Solutions<br>10000 Cedar Ave.<br>Cleveland, OH 44106<br>Tele: (216) 445 -3403<br>Email: kgrafmeyer@customorthopaedics.com |
| TRADE NAME: | OrthoVis Web Portal |
| DATE PREPARED: | 26-May -2015 |
| DATE REVISED: | 08-Aug-2015 |
| COMMON NAMES: | -Image processing system and preoperative software for<br>simulating / evaluating implant placement and surgical<br>treatment options<br>-Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer<br>Cemented<br>-Shoulder Prosthesis, Reverse Configuration |
| Product | Product Code | Regulation and Classification Name | Device Class |
|---------------------------------------------------|--------------|-------------------------------------------------------------------|--------------|
| OrthoVis Web Portal | LLZ | 21 CFR 892.2050<br>Picture Archiving and<br>Communications System | II |
| Secondary Product Names and Codes | | | |
| OrthoVis Web Portal -<br>Accesory to Glenoid IRIS | KWS | 21 CFR 888.3660 | II |
| OrthoVis Web Portal -<br>Accesory to Glenoid IRIS | PHX | 21 CFR 888.3660 | II |
{4}------------------------------------------------
#### Predicate Devices:
K133367: OrthoVis Preoperative Plan
K142072: Glenoid Intelligent Reusable Instrument System (Glenoid IRIS)
#### DEVICE DESCRIPTION
The OrthoVis Web Portal is composed of software intended for use to facilitate upload of medical images and preoperative planning and plan approval of placement and orientation of Arthrex and DePuy total shoulder replacement systems. Each surgeon user's uploaded images are grouped into cases and associated with that user's profile. Uploaded images can be downloaded from the portal by COS technicians and used to create preoperative plans (see 510(k) K13367 and K151568). The user is then able to login to the OrthoVis Web Portal to review the preoperative plan and either approve or modify the location and/or orientation of the shoulder replacement component and then approved plan is then available to download by COS technicians for further preoperative planning production (see 510(k) K123122 and K142072) and for download by the user/surgeon.
#### INTENDED USE AND INDICATIONS FOR USE
The OrthoVis Web Portal is intended for use as a software interface and for the transfer of imaging information from a medical scanner such as a CT scanner. It is also intended as pre-operative software for simulating / evaluating implant placement and surgical treatment options. The OrthoVis Web Portal is intended for use with the Glenoid Intelligent Reusable Instrument system and with the OrthoVis Preoperative Plan. It is also intended for use with the Arthrex and DePuy shoulder replacement implant systems listed below.
- Arthrex Univers Apex
- · Arthrex Univers II
- Arthrex Univers Revers
- DePuy Global AP
- DePuy Global StepTech
- DePuy Delta Xtend Reverse
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended for use to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant.
#### BASIS OF SUBSTANTIAL EQUIVALENCE
The OrthoVis Web Portal has nearly the same intended use and indications for use as the OrthoVis Preoperative Plan (K133367) and the Glenoid Intelligent Reusable Instrument System (Glenoid IRIS) (K142072). The main difference between the original OrthoVis software and the OrthoVis Web Portal is that the Web Portal facilitates the communication between the COS planning technician and surgeon.
#### Technological Comparison to Predicate
The device described here is intended to collect images and preferences from a surgeon relative to a case, move those images to an in-house workstation for processing, and to allow those plans to be displayed for a surgeon to either request adjustments to the plan or approve the plan, followed by distribution of the final plan materials.
These functions were previously done by a largely manual process involving secure email and physical mail. This table compares the new technology to the old process, describing the use of the web portal, SFTP transfer, and email to do these functions.
{5}------------------------------------------------
(Note that email is only used for indication of status change, and does not ever contain PHI.)
| Function | Predicates | Subject |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Surgeon authentication | Phone conversation. | Secure portal login with hashed<br>password. |
| Confidential communication | Postal mail, phone call, or secure email. | 1. HTTPS/SSH network<br>communication for all web portal and<br>surgeon upload/downloads<br>2. STFP for local office<br>uploads/downloads.<br>3. Email for status change notification<br>(e.g. plan ready, etc.) |
| Create case | Phone conversation, file on local<br>computer. | Web portal case creation page. |
| Store case | Local files | Encrypted database on server |
| Submit images | CD/DVD in postal mail, files put on<br>local server | 1. Files uploaded via HTTP web portal<br>upload page to protected directory.<br>2. Local office is notified that files are<br>uploaded via case status change.<br>3. Files are then downloaded via SFTP<br>secure protocol from network server<br>to local server. |
| Create surgical plan | Planning software on local<br>workstation. | Planning software on local<br>workstation. |
| Preview surgical plan with surgeon | Surgeon uses commercial screen<br>sharing software (i.e. PC Anywhere) to<br>view local workstation screen during<br>phone conversation. | 1. Files uploaded by office to web<br>portal protected directory via SFTP<br>secure protocol.<br>2. Web portal software imports plan<br>information into web database, and<br>imports images into web image store.<br>3. Web portal notifies surgeon of plan<br>availability via case status change<br>email.<br>4. Surgeon logs into portal.<br>5. Surgeon views plan interactively in<br>3D on plan view page. |
| Surgeon requests adjustments to plan | During phone conversation, surgeon<br>verbally requests adjustments and sees<br>changes on shared screen. | 1. Surgeon uses web controls to change<br>3D image on plan view page.<br>2. Surgeon indicates on plan view page<br>via web controls that changes are<br>requested.<br>3. Local office is notified via email of<br>status change. |
| Surgeon approves plan | During phone conversation or secure<br>email, surgeon verbally approves plan. | 1. Surgeon views plan on plan view<br>page.<br>2. Surgeon indicates on plan via web<br>controls that the plan is approved.<br>3. Local office is notified via email of<br>status change. |
| Function | Predicates | Subject |
| Surgeon gets final copy of approved<br>plan | The Approved Preoperative Plan is<br>created in OrthoVis and sent in a PDF<br>file via secure email. | 1. Local office prepares PDF file on<br>OrthoVis and uploads via STFP secure<br>protocol.<br>2. Surgeon is notified via email of<br>status change.<br>3. Surgeon logs into web portal.<br>4. Surgeon downloads PDF containing<br>final approved plan. |
| Surgeon is reminded of plan | Reminder phone call. | Periodic generation of reminder emails<br>about status change until download<br>occurs. |
{6}------------------------------------------------
#### NON-CLINICAL TESTING
The following testing was performed to demonstrate substantial equivalency of the OrthoVis Web Portal to the predicate device
- Software verification and validation
- Dimensional validation
- Simulated Use Comparison between OrthoVis Web Portal Process and CD Mailer/Go-To-Meeting/Email Process
#### Software Verification and Validation
Software Verification and Validation was performed in order to establish substantial equivalence to the predicate device.
#### Dimensional Validation
Dimensional validation of the OrthoVis system was performed (see Exhibit A). The same code responsible for reading and display of CT scans is used in the cleared OrthoVis Shoulder software (K123122, and K133367) as is used in the OrthoVis Web Portal software.
#### Simulated Use Comparison between OrthoVis Web Portal Process and CD Mailer/Go-To-Meeting/Email Process
Surgeons with experience in using the OrthoVis Preoperative Plan via the CD Mailer/Go-To-Meeting/Email process were asked to compare that experience with the OrthoVis Web Portal Process and fill out a survey based on the two experiences.
#### CLINICAL TESTING
Clinical testing was not necessary to determine substantial equivalence between the OrthoVis Web Portal and the predicate device.
#### NON-CLINICAL TESTING SUMMARY AND CONCLUSION
The testing done for the web portal consisted of three parts:
- The first part consisted of automated unit tests. -
- -The second part used an automatic integration-testing tool to test the Web Portal user interface.
- -The third part manually tested remaining errors.
The sum of our automated and manual tests described below provide comprehensive coverage of the application requirements, and proved that our device is as safe and effective as the predicate device.
We tested each constraint and protection using an automated testing framework that created test cases that followed the constraint rules to verify correct results, and by creating test cases that violate these
{7}------------------------------------------------
rules to verify that the constraints and error messages were working correctly.
Operational behavior of the user interface (those part of the requirements that could be observed by a user of the application) was tested using the WebDriver testing tool (which is now a W3C standard) to drive the user interface in a large number of scenarios. By "robotically" operating the user interface, the application was tested exactly as if a user (using Chrome or Firefox) was executing a series of tests. Automated tests were made from prepared data to test each requirement. By using rigorous automation, tests can be repeated many times when moving to each specific environment, to test servers or to run regression tests when features are added.
Since some errors are difficult to mitigate with automated testing, we verified, by manual code inspection, that error handlers are in place and operational. We completed manual testing that was executed as usual during each testing phase.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.