GLENOID INTELLIGENT REUSABLE INSTRUMENT SYSTEM (GLENOID IRIS)
K142072 · Custom Orthopaedic Solutions, Inc. · KWS · Nov 24, 2014 · Orthopedic
Device Facts
Record ID
K142072
Device Name
GLENOID INTELLIGENT REUSABLE INSTRUMENT SYSTEM (GLENOID IRIS)
Applicant
Custom Orthopaedic Solutions, Inc.
Product Code
KWS · Orthopedic
Decision Date
Nov 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Indications for Use
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant. The Glenoid IRIS is indicated for use in planning and placing the central glenoid guide pin for the DePuy Anchor Peg Glenoid (APG) component of the DePuy AP Shoulder System, the DePuy Global StepTech Glenoid component, or the DePuy Delta Xtend Reverse Shoulder metaglene component as an alternative to the standard instruments provided for placing the guide pin with these implant systems. The indications for use of the DePuy shoulder systems with which the Glenoid IRIS is intended to be used are the same as those described in the labeling for these shoulder systems except that the Glenoid IRIS is not intended for use in hemi-shoulder replacement.
Device Story
Patient-specific manual instrument system; facilitates preoperative planning and intraoperative placement of central glenoid guide pin for total shoulder arthroplasty. System includes SmartBone models (photopolymer), reusable stainless steel IRI legs, and SmartBone mount. Used by orthopedic surgeons in clinical settings. Surgeon uses preoperative plan to guide pin placement; device acts as physical template/guide to ensure accurate trajectory relative to patient anatomy. Benefits include improved surgical precision and alignment during glenoid preparation. Reusable components require on-site steam sterilization.
Clinical Evidence
No clinical data. Bench testing included biocompatibility (cytotoxicity, irritation, sensitization), shipping/distribution testing at temperature extremes (-112°F to 140°F), and a simulated-use artificial cadaver study. Cadaver study compared mean version and inclination deviation between original and modified systems, demonstrating equivalent performance.
Technological Characteristics
Manual surgical instrument system. Materials: Stainless steel (ISO 16061 compliant) and photopolymer. Components: SmartBone models, IRI legs, SmartBone mount, and instrument tray. Sterilization: Steam sterilization (on-site). Connectivity: None (manual system).
Indications for Use
Indicated for patients undergoing total shoulder replacement surgery for osteoarthritis, traumatic arthritis, rheumatoid arthritis, proximal humerus fracture-dislocation, or revision of failed primary components. Also indicated for patients with severe arthropathy or failed joint replacement with grossly deficient rotator cuff (Delta Xtend). Contraindicated for hemi-shoulder replacement.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Glenoid Intelligent Reusable Instrument System (Glenoid IRIS) (K123122)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, featuring three human profiles in silhouette.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 24, 2014
Custom Orthopaedic Solutions, Incorporated Justin Baker, Ph.D. Regulatory and Quality Manager 10000 Cedar Avenue Cleveland, Ohio 44106
Re: K142072
Trade/Device Name: Glenoid Intelligent Reusable Instrument System (Glenoid Iris) Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KWS. PHX Dated: October 21, 2014 Received: October 27, 2014
Dear Dr. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K142072
#### Device Name
Glenoid Intelligent Reusable Instrument System (Glenoid IRIS)
#### Indications for Use (Describe)
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant.
The Glenoid IRIS is indicated for use in planning and placing the central glenoid guide pin for the DePuy Global AP Shoulder glenoid component, Global Shoulder StepTech Anchor Peg glenoid component, or Delta Xtend Reverse Shoulder metaglene component as an alternative to the standard instruments provided for placing the guide pin with these implant systems.
The indications for use of the DePuy shoulder systems with which the Glenoid IRIS is intended to be used are as those described in the labeling for these shoulder systems except that the Glenoid IRIS is not intended for use in hemishoulder replacement. The Indications for Use of these DePuy shoulder systems are:
DePuy Global AP Shoulder System and Global StepTech Anchor Peg Glenoid
The DePuy Global AP Shoulder System and Global StepTech Anchor Peg Glenoid are indicated for use in total shoulder replacement surgery for patients suffering from:
· A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis.
· Fracture-dislocation of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience dictates that alternative methods of treatment are unsatisfactory.
· Other difficult clinical problems where shoulder surgery arthrodesis or resection arthroplasty are not acceptable (e.g., revision of a failed primary component).
The Global AP Glenoid and StepTech glenoid components are indicated only for use with bone cement.
DePuy Delta Xtend Reverse Shoulder System
· The Delta Xtend Reverse Shoulder prosthesis is intended for use as total shoulder replacement.
· The Delta Xtend Reverse Shoulder prosthesis is indicated for use in a grossly efficient rotator cuff joint with severe arthropathy or a previous failed joint replacement with a grossly deficient rotator cuff joint.
• The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.
· In cases of bone defects in the proximal humerus, the monoblock implant should be used and then only in cases where the residual bone permits firm fixation of this implant.
· Delta Xtend hemi-shoulder replacement is also indicated for hemi-arthroplasty if the glenoid is fractured intraoperatively.
· The metaglene component is HA coated and is intended for cementless use with the addition of screws for fixation. All other components are for cemented use only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
| 510(k) SPONSOR / MANUFACTURER: | Custom Orthopaedic Solutions, Inc.<br>A subsidiary of Cleveland Clinic<br>10000 Cedar Avenue<br>Cleveland, Ohio 44106 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------|
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Justin J. Baker, PhD, RAC<br>Regulatory & Quality Manager<br>Custom Orthopaedic Solutions<br>10000 Cedar Avenue<br>Cleveland, OH 44106<br>Tel: (216) 445-2164<br>Email: jbaker@customorthopaedics.com |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
DATE PREPARED: November 22, 2014
TRADE NAME: Glenoid Intelligent Reusable Instrument System (Glenoid IRIS)
COMMON NAMES:
Total shoulder replacement instruments
| Product | Product Code | Regulation and<br>Classification Name | Device Class |
|------------------------------------------------------------------------|--------------|--------------------------------------------------------------------------------|--------------|
| Glenoid Intelligent<br>Reusable<br>Instrument System<br>(Glenoid IRIS) | KWS, PHX | 21 CFR 888.3660<br>Shoulder, semi-<br>constrained metal /<br>polymer, cemented | II |
#### PREDICATE DEVICES:
Glenoid Intelligent Reusable Instrument System (Glenoid IRIS)(K123122)
# DEVICE DESCRIPTION:
The Glenoid Intelligent Reusable Instrument System described in this submission is identical to the Glenoid IRIS cleared in K123122, except for the following modifications:
- In K123122, the IRI legs were plastic, single use, and were shipped sterile in quantities of 2 of each recommended leg length. In this submission, the IRI legs are stainless steel, reusable, and are shipped non-sterile in the Glenoid IRIS instrument
{5}------------------------------------------------
tray/carrier/kit for steam sterilization prior to use at the hospital where the procedure will be taking place. Each Glenoid IRIS kit contains a full set of reusable IRI legs.
- The SmartBone instruments (SmartBone-Pin Trajectory, SmartBone-Reamed, SmartBone-Fixation Feature Prep, and SmartBone-Implant) are made of a different photopolymer from K123122.
-The SmartBone instruments are shipped non-sterile and are sterilized on-site where the procedure will be taking place.
-The IRI wrench was removed from the Glenoid IRIS system and the IRI cap was modified to facilitate tightening and loosening the IRI instrument by hand. The Glenoid IRIS instrument tray was made to include a slot that could be used to replace the disassembly feature of the nubs on the IRI wrench.
-The SmartBone—Pin Trajectory instrument is pre-marked where the recommended IRI legs contact the SmartBone instrument. The marker is a non-toxic marker that is compatible with steam, EO & Sterrad Sterilization. We performed testing to determine how great the risk of the marking from the marker bleeding and or transferring when wiped with a polar and non-polar solvent. The study indicated that the marker was compatible with steam sterilization and did not undergo bleeding. Likewise, the marker did not transfer to a white rag when rubbed with a polar or non-polar solvent.
-The SmartBone mount is made of stainless-steel, is reusable, and is shipped nonsterile as part of the Glenoid IRIS system in a kit/carrier. When the SmartBone is placed on the SmartBone mount, the glenoid is positioned such that vertical is the plane of the scapula. When the SmartBone is set on its base without the SmartBone Mount, the guide pin hole is vertical for ease in setting the IRI.
-The Glenoid IRIS system includes a case/tray carrier for ease in shipping, storing, carrying, and organizing the Glenoid IRI device, the Glenoid IRI legs, the SmartBone instrument (or a place to put the SmartBone instrument, when it arrives after having been shipped separately), and the SmartBone Mount.
-Minor dimensional and aesthetic marking changes to the IRI components
## INTENDED USE AND INDICATIONS:
The intended use and indications for use of the modified Glenoid Intelligent Reusable Instrument System remain unchanged from that cleared in K123122.
{6}------------------------------------------------
The Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is a patient specific manual instrument system intended to facilitate preoperative planning and intraoperative placement of the central glenoid guide pin used in the preparation of the glenoid in total shoulder systems that utilize a central guide pin for preparing the glenoid to receive the glenoid implant.
The Glenoid IRIS is indicated for use in planning and placing the central glenoid guide pin for the DePuy Anchor Peg Glenoid (APG) component of the DePuy AP Shoulder System, the DePuy Global StepTech Glenoid component, or the DePuy Delta Xtend Reverse Shoulder metaglene component as an alternative to the standard instruments provided for placing the guide pin with these implant systems.
The indications for use of the DePuy shoulder systems with which the Glenoid IRIS is intended to be used are the same as those described in the labeling for these shoulder systems except that the Glenoid IRIS is not intended for use in hemi-shoulder replacement. The Indications for Use of these DePuy shoulder systems are:
## DePuy Global APTM Shoulder System and Global StepTech™ Anchor Peg Glenoid
The DePuy Global AP™ Shoulder System and Global StepTech™ Anchor Peg Glenoid are indicated for use in total shoulder replacement surgery for patients suffering from:
- A severely painful and/or disabled joint resulting from osteoarthritis, ● traumatic arthritis or rheumatoid arthritis.
- . Fracture-dislocation of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience dictates that alternative methods of treatment are unsatisfactory.
- Other difficult clinical problems where shoulder surgery arthrodesis or . resection arthroplasty are not acceptable (e.g., revision of a failed primary component).
The Global AP™ Glenoid and StepTech™ glenoid components are indicated only for use with bone cement.
## DePuy Delta Xtend™ Reverse Shoulder System
- . The Delta Xtend™ Reverse Shoulder prosthesis is intended for use as total shoulder or hemi-shoulder replacement.
- The Delta Xtend™ Reverse Shoulder prosthesis is indicated for use in a grossly ● efficient rotator cuff joint with severe arthropathy or a previous failed joint replacement with a grossly deficient rotator cuff joint.
- . The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.
{7}------------------------------------------------
- In cases of bone defects in the proximal humerus, the monoblock implant should be used and then only in cases where the residual bone permits firm fixation of this implant.
- Delta Xtend™ hemi-shoulder replacement is also indicated for hemi-arthroplasty . if the glenoid is fractured intraoperatively.
- The metaglene component is HA coated and is intended for cementless use with . the addition of screws for fixation. All other components are for cemented use only.
| Prescription Use _____________________________________________________________________________________________________________________________________________________________ | | Over-The-Counter Use |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | AND/OR | (21 CFR 801 Subpart C) |
# BASIS OF SUBSTANTIAL EQUIVALENCE:
The modified Glenoid Intelligent Reusable Instrument System ("Glenoid IRIS") is substantially equivalent to the Glenoid Intelligent Reusable Instrument System described in K123122 as it is the same system with minor dimensional modifications, modifications to whether the product is reusable and shipped sterile, and modifications to the material from which the instruments are manufactured. The convenience of a system carrier/kit/tray has been included as a part of the Glenoid IRIS system in this submission. The IRI wrench has been removed from the modified system. The modified Glenoid Intelligent Reusable Instrument System functions in the same way as the original Glenoid IRIS. The indications for use for the Glenoid Intelligent Reusable System with these modifications have not changed from K123122.
## Non-Clinical Testing
The following testing was performed to demonstrate substantial equivalency of the Glenoid IRIS to the predicate device in K123122.
- Drawings comparison (see list of changes in Device Description above) ●
- Biocompatibility testing of new photopolymer printed SmartBones ●
- Shipping/Distribution Testing of new photopolymer printed/packaged . SmartBones
- Artificial Cadaver Surgeon Evaluation (Validation)
## Clinical Testing
Clinical testing was not necessary to determine substantial equivalence between the modified Glenoid IRIS and the predicate Glenoid IRIS (K123122).
## Drawings Comparison
Changes in the IRI drawings were minor and are listed in the Device Description above.
## Biocompatibility
As the SmartBone models are considered an external communicating device that contacts
{8}------------------------------------------------
bone/tissue/dentin for a limited duration, cytotoxicity and intracutaneous irritation and sensitization testing was performed. Testing was performed on steam sterilized samples of the SmartBones in the new photopolymer. All tests passed, indicating that the new photopolymer is biocompatible. The original Glenoid IRIS submission likewise used biocompatible materials, and so this modification in the SmartBone material is still substantially equivalent to the original Glenoid IRIS (K123122).
The modified Glenoid IRIS changes the IRI legs from a biocompatible acetal copolymer to a stainless steel that is in conformance with ISO 16061: Instrumentation for use in association with non-active surgical implants - General requirements. In this regard, the change in IRI leg material does not alter the biocompatibility of the modified Glenoid IRIS and the original and modified Glenoid IRIS systems remain substantially equivalent.
#### Shipping Distribution Testing of New Photopolymer Printed/Packaged SmartBones
The original Glenoid IRIS (K123122) tested packaged SmartBones at temperature extremes of 140°F and -20°F for 48 hours. The modified Glenoid IRIS tested the new SmartBone material with packaged SmartBones in temperature extremes of 140°F and -112°F for 120 hours. The new SmartBone material successfully passed the shipping/distribution temperature tests and drop testing, and in this regard despite the SmartBone material change, remains substantially equivalent to the original Glenoid IRIS submission (K123122).
#### Artificial Cadaver Surgeon Evaluation
A simulated use, artificial cadaver comparison test between the original and modified Glenoid IRIS was performed. The original Glenoid IRIS was used on an artificial training cadaver of a patient with severe glenoid retroversion five times. The modified Glenoid IRIS was also used on this same patient five times. The mean version and inclination of the original and modified Glenoid IRIS was compared between each other and to the amount of deviation from the planned guide pin location. The modified Glenoid IRIS was shown to be substantially equivalent in performance to the original Glenoid IRIS (K123122).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.