For the temporary management of stress urinary incontinence (SUI) in women
Device Story
Single-use vaginal pessary; inserted by user via tampon-like applicator; composed of high-density polyethylene core, nonwoven overwrap, and withdrawal cord. Device provides mechanical support to anterior vaginal wall to increase pressure on urethra; manages stress urinary incontinence. Used in home setting; patient self-administered. Daily use up to 12 hours in 24-hour period. Device removed via withdrawal cord and disposed of in trash. Clinical benefit includes reduction in SUI episode frequency and pad weight gain; improves quality of life for women with SUI.
Clinical Evidence
Two clinical studies performed. 1) 3-month randomized, 2-arm safety study (n=not specified) showed device was well-tolerated, minimal effect on vaginal wall, no serious AEs. 2) 4-week single-arm, open-label efficacy study demonstrated statistically significant and clinically meaningful improvement in SUI episode frequency and pad weight gain. Study confirmed women could appropriately self-identify SUI symptoms and self-administer device.
Technological Characteristics
Single-use vaginal pessary. Core: High-density polyethylene. Overwrap: Nonwoven. Withdrawal cord: Tampon-style. Dimensions/form factor: Tampon-like applicator. Biocompatibility per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization/irritation), and ISO 10993-6 (implantation). Mechanical testing per ASTM D2256 (string tensile) and AS 2869-2008 (anchor strength).
Indications for Use
Indicated for women (18+) with stress urinary incontinence (SUI) symptoms for temporary management.
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other.
Re:
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2015
Procter & Gamble Lenore Faulhaber Regulatory Affairs Manager 6110 Center Hill Ave. Cincinnati, OH 45224
> K151413 Trade/Device Name: Always/Tampax bladder supports Regulation Number: 21 CFR 884.3575 Regulation Name: Vaginal Pessary Regulatory Class: Class II Product Code: HHW Dated: November 19, 2015 Received: November 20, 2015
Dear Lenore Faulhaber,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151413
Device Name Always/Tampax bladder supports
Indications for Use (Describe) For the temporary management of stress urinary incontinence in women
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K151413
#### Submitter
The Procter & Gamble Company 6110 Center Hill Avenue Cincinnati, OH 45224
#### Contact Person
Lenore Faulhaber, Ph.D., M.B.A. Regulatory Affairs Manager (513) 634-2466
Date prepared December 11, 2015
Common Name Intravaginal stress incontinence device
Classification Name Vaginal Pessary (21 CFR 884.3575, Product Code HHW)
Proprietary Name Always/Tampax bladder supports
#### Indications for Use/ Intended Use
For the temporary management of stress urinary incontinence (SUI) in women
#### Predicate Device
Improve Bladder Supports/TIPI OTC K131198; Manufactured by ConTIPI Ltd.
#### Device Description
The device is a single use vaginal pessary, inserted into the vagina, by the user, using a tampon-like applicator. The device is composed of a high density polyethylene core, a nonwoven overwrap, and a withdrawal cord. The overwrap helps to facilitate ease of insertion and removal. The device is removed by a withdrawal cord made from tampon withdrawal cord material. Upon removal, the device is disposed of in a trash receptacle. The pessary is a single-use device. It is designed for daily use up to 12 hours in 24 hours. The device is available in one size and is sold over-the-counter.
{4}------------------------------------------------
| Comparison Element | 510(k) Device | Predicate Tipi OTC [K131198] |
|---------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Classification Name | Pessary, Vaginal | Pessary, Vaginal |
| Manufacturer | Procter & Gamble | ConTIPI, Ltd. |
| Indications for Use | For the temporary management<br>of stress urinary incontinence<br>(SUI) in women. | For the temporary management<br>of stress urinary incontinence<br>(SUI) in women. |
| Intended Population | Any women (18 and older) who<br>have SUI symptoms | Women 21 years of age or older<br>who have SUI |
| Mode of Action | Increases pressure through the<br>anterior vaginal wall onto the<br>urethra | Tension free, mid-urethral<br>support |
| Wear Time | Up to 12 hours in a 24 hour<br>period | Up to 8 hours in a 24 hour<br>period |
| Insertion | Via a tampon-like applicator | Via a tampon-like applicator |
| Removal | Tampon-like withdrawal cord | Tampon-like withdrawal cord |
| Single Use? | Yes | Yes |
| Pessary Core Design | Shaped High Density<br>Polyethylene core covered by a<br>non-woven overwrap | Soft flexible silicone rubber,<br>multiple armed cores covered by<br>non-woven mesh |
| Sizing | Available in one size | Available in multiple sizes |
| Overwrap | Nonwoven | Nonwoven |
| Withdrawal Cord | Removal string | Removal string |
| Applicator | Tampon-like applicator | Tampon-like applicator |
#### Technological Characteristics of Subject Device Compared to Predicate
#### Summary of Non-Clinical Tests
Design Verification Assessment
#### Pessary Applicator Force
A quantitative method to measure pessary applicator expulsion using a semi-automated tester to measure tension and compression.
#### Retraction during Device Removal
Not applicable
#### Pessary Withdrawal Cord Anchor Strength
This procedure determines the maximum force required to break or detach the pessary withdrawal cord. Reference: Australian / New Zealand Standard (AS 2869-2008)
#### Removal String Integrity/ String Tensile Testing
Specified string properties (for example, Tensile Strength) are confirmed to meet finished product specifications. Reference Method: ASTM D2256 (1998)
{5}------------------------------------------------
## Stiffness of Core/ Dimension Verification
Specified core properties (dimensions, weight and resin material properties) are confirmed to meet finished product specifications.
| Test | Standard | P&G Device | FDA<br>Recognition<br>Number |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------------------------------|
| Cytotoxicity | ISO 10993-5:2009: Biological<br>evaluation of medical devices,<br>Part 5: Tests for in vitro<br>cytotoxicity. 2009-06-01. | Non-cytotoxic | 2-153 |
| Sensitization | ISO 10993-10:2010: Biological<br>evaluation of medical devices,<br>Part 10: Tests for irritation and<br>skin sensitization. 2010-08-01. | Non-sensitizing | 2-174 |
| Vaginal Irritation | ISO 10993-10:2010: Biological<br>evaluation of medical devices,<br>Part 10: Tests for irritation and<br>skin sensitization. 2010-08-01. | Non-irritating | 2-174 |
| Genotoxicity | N/A | Risk Analysis<br>Approach | N/A |
| Acute Systemic<br>Toxicity | N/A | Risk Analysis<br>Approach | N/A |
| Subchronic<br>Systemic Toxicity | N/A | Risk Analysis<br>Approach | N/A |
| Chronic Systemic<br>Toxicity | N/A | Risk Analysis<br>Approach | N/A |
| Reproductive/<br>Developmental<br>Toxicity | N/A | Risk Analysis<br>Approach | N/A |
| Carcinogenicity | N/A | Risk Analysis<br>Approach | N/A |
| Local Effects after<br>Implantation | ISO 10993-6:2007: Biological<br>evaluation of medical devices,<br>Part 6: Tests for local effects<br>after implantation. 2007-04-15. | No evidence of<br>microscopic or<br>macroscopic<br>tissue reaction | 2-177 |
## Biocompatibility Assessment
## Evaluation of Toxic Shock Syndrome Risk
The pessary was evaluated in vitro to determine if it enhances the growth of Staphylococcus aureus, increases the production of Toxic Shock Syndrome Toxin-1 (TSST-1), or alters the growth of normal vaginal microflora. Under the test conditions, the pessary device did not promote or inhibit either bacterial growth or TSST-1 toxin production.
{6}------------------------------------------------
#### Label Comprehension and Self-Selection
The labeling was assessed in a combined self-selection and label comprehension study. The study objectives were to evaluate self-selection and comprehension of targeted communication objectives from the external package labeling and Instructions for Use (IFU) that were different from the predicate device (K131198). Three cohorts were tested: 1) A population of women who may use product based on personal medical history, ages 18 and older; 2) A population of women who may use product based on personal medical history, ages 18 and older with low literacy; 3) A population of women who may use product based on personal medical history, ages 65 and older.
#### Summary of Clinical Tests
Two clinical studies were performed:
- Safety study: A randomized, 2-arm (pad control), single-blind (vaginal examiner), multi-center, 3-month safety study, in which safety was assessed by adverse events and physical examination, visual vaginal examinations, clean catch urine collection and vaginal swabbing at baseline, week 2, week 4, and week 12.
This study showed that the device was well-tolerated, with no serious AEs. The majority of the AEs that were observed were 'mild' and spontaneously resolved. The device showed minimal effect on the vaginal wall and surrounding structures, which is a particularly important finding for postmenopausal women who represented over 70% of the study population. This study shows that the device was well tolerated in, and used appropriately by, women with self-identified SUI.
- . Efficacy and safety study: A single-arm, open label, multi-center, 4-week efficacy study (2-week baseline followed by a 2-week treatment period) that assessed the change from baseline for both pad weight gain and SUI episode frequency. A clinically meaningful improvement as a composite of the symptoms that are most relevant to the improved QoL for women was evaluated: 1) SUI episode frequency and 2) pad weight gain. Safety of the pessary was assessed by adverse events, clean catch urine collection and vaginal swabbing.
This study demonstrated that the single-use, disposable pessary was effective for women with SUI, and was safe and comfortable. The data show a statistically significant and clinically meaningful improvement in the symptoms that were most important to each woman, whether it was a reduction in SUI episode frequency, a reduction in pad load, or both. The study also showed that women were able to appropriately self-identify symptoms of SUI and appropriately self- administer the device, which is important for an OTC product.
## Conclusion
The differences between Always/Tampax bladder supports and its predicate device do not introduce a new intended use and do not raise new issues of safety and effectiveness. Non
{7}------------------------------------------------
clinical and clinical testing has demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended. Therefore, Always/Tampax bladder supports are substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.