The IMPROVE BLADDER SUPPORT is an over-the-counter device intended for the temporary management of stress urinary incontinence (SUI) in women.
Device Story
Single-use vaginal pessary for temporary management of stress urinary incontinence (SUI); self-inserted by user via applicator similar to menstrual tampon. Device consists of flexible silicone rubber core covered with non-woven mesh and removal string. Provides tension-free mid-urethral support; worn for maximum 8 hours in 24-hour period. User removes and discards device after use. Intended for over-the-counter (OTC) use. Benefits include temporary SUI management without prescription. Safety and efficacy supported by clinical studies on vaginal microflora and labeling comprehension/self-selection studies.
Clinical Evidence
Clinical evidence includes a multicenter, prospective, randomized, controlled, cross-over study (n=45) evaluating vaginal microflora changes over 30 days of use versus pad control; results showed no clinically significant microflora changes. A labeling comprehension and selection study (n=45, divided into low literacy, general, and elder populations) demonstrated 92% accuracy in user self-selection and 96% accuracy in understanding instructions. A limited R&D clinical evaluation confirmed safety and efficacy comparable to the predicate. Adverse events were mild, temporary, and resolved without intervention.
Technological Characteristics
Single-use vaginal pessary. Materials: medical-grade silicone rubber core, non-woven mesh cover, removal string. Dimensions/shape modified from predicate for comfort. Biocompatibility testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization/irritation), ISO 10993-11 (systemic toxicity), and ISO 10993-6 (implantation). Mechanical testing: insertion force, retraction, string detachment, and removal string integrity.
Indications for Use
Indicated for temporary management of stress urinary incontinence (SUI) in women.
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
{0}------------------------------------------------
## 510(k) SUMMARY ConTIPI LTD's IMPROVE BLADDER SUPPORT
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
ConTIPI Ltd. 2 Alon Ha'Tavor St. Caesarea Industrial Park 38900 POB 3135, Israel
Phone: 972-4-6276066 Facsimile: 972-4-6276068 Contact Person: Tsvia Erlich, VP Regulatory and clinical affairs AUG 2 7 2013
August 26, 2013 Date Prepared:
### Name of Device and Name/Address of Sponsor
IMPROVE BLADDER SUPPORT ConTIPI Ltd. 2 Alon Ha'Tavor St. Caesarea Industrial Park 38900 POB 3135, Israel
#### Common or Usual Name
Intra-vaginal stress incontinence device
#### Classification Name
Vaginal pessary
#### Predicate Devices
ConTIPI Ltd. TIPI (K060526)
#### Intended Use / Indications for Use
The IMPROVE BLADDER SUPPORT is an over-the-counter device intended for the temporary management of stress urinary incontinence (SUI) in women.
#### Device Description and Technological Characteristics
The IMPROVE BLADDER SUPPORT device is a single use vaginal pessary, inserted into the vagina, by the user herself, using an applicator similar to a menstrual tampon applicator. The device comprises of a flexible silicone rubber core, covered with a non-woven mesh, and a
{1}------------------------------------------------
removal string. Once inserted into the vagina, the core and cover of the device provides tension free mid-urethral support for the temporary management of stress urinary incontinence (SUI). After a maximum of 8 hours of use in a 24 hours period, the device should be removed from the vagina using its pull string and discarded.
### Comparison to Predicate Device
The IMPROVE BLADDER SUPPORT is a modification to its predicate, the cleared ConTIPI TIPI vaginal pessary device, and is substantially similar to its predicate device. While the predicate TIPI device was cleared for prescription use, the IMPROVE BLADDER SUPPORT is indicated for over-the-counter (OTC) use.
The designs of the subject and predicate versions of the device are substantially similar, with only minor modifications to the device materials, sizes and shape. Both the IMPROVE BLADDER SUPPORT and the predicate consist of a flexible silicone rubber core covered with mesh, an applicator and removal string and have very similar dimensions, materials, and performance. Minor changes have been made to IMPROVE BLADDER SUPPORT materials and shape as follows: replacing the core material with a similar medical silicon material; change in the cover mesh and removal string materials: slight change to core edges; a color has been added to the applicator; applicator and core shape and dimensions were slightly changed. These changes were made to improve manufacturability and promote patient comfort.
Additionally, the labeling materials of the IMPROVE BLADDER SUPPORT and the predicate are substantially the same but the labeling for the IMPROVE BLADDER SUPPORT was updated with self-selection information and is designed to conform to FDA's guidance on labeling for OTC devices.
#### Performance Data
The following testing was performed to confirm that the subject device met its specifications and demonstrates substantial equivalence to the identified predicate device:
- Design verification testing to confirm that the subject device met its specifications and was . substantially equivalent to the specifications of the predicate device. This testing included the following:
- Force required to push a device out of the applicator (insertion force) ্
- Retraction during device removal o
- String detachment force ു
- Removal string integrity 0
- . Bench testing to evaluate the force exerted on the vaginal wall in comparison to the predicate device
- Subject device's potential contribution to the development to S. aureus and TSST-1 when . used as intended
{2}------------------------------------------------
| Test | Standard | Conclusion |
|-------------------------------------|--------------------------------|--------------------------------------------------------------|
| Cytotoxicity | ISO 10993-5:2009, 2nd edition | Non-cytotoxic |
| Sensitization | ISO 10993-10:2010,3rd edition | Non-sensitizing |
| Vaginal Irritation | ISO 10993-10:2010, 3rd edition | Non-irritating |
| Acute Systemic<br>Toxicity | ISO 10993-11:2006, 2nd edition | No evidence of<br>systemic toxicity |
| Local Effects after<br>Implantation | ISO 10993-6:2007, 2nd edition | No evidence of tissue<br>reactions relevant to<br>actual use |
- Biological Evaluation: The subject device has undergone the following biocompatibility tests: .
- � Shelf life study intended to establish that the IMPROVE BLADDER SUPPORT (subject device) retains its properties, in terms of microbiological load and performance, throughout its stated shelf life
In all instances the subject device met its specifications and demonstrated substantial equivalence to the predicate device.
## Clinical Testing
Clinical testing was necessary to support the change from prescription use to over-the-counter use of this device due to concerns that this change could potentially affect the safety profile of the device. To reduce the level of new clinical testing that was required to support clearance of the OTC device, the clinical study used to demonstrate safety and effectiveness of the predicate TIP! device was leveraged along with the new bench studies described above and a limited R&D supporting clinical evaluation. In addition, the change from prescription to OTC use required a demonstration that women could appropriately understand the contraindications for use (i.e., who would and would not be appropriate users of the device) and comprehend the instructions for use. The following studies were conducted to address these concerns:
- . A limited R&D supporting clinical evaluation to evaluate device safety and efficacy in comparison to the predicate device was performed. The subject device was evaluated in a limited patient population in a clinic setting. The results from the subject device were compared to those obtained using the predicate device and no device from the same patient population. The results showed substantially equivalent performance between the subject and predicate devices.
All adverse events were considered to be minimal, temporary and resolved without medical treatment and no sequelae.
- . A multicenter, prospective, randomized, controlled, cross-over study to assess changes in vaginal microflora while using the IMPROVE BLADDER SUPPORT in the temporary
{3}------------------------------------------------
management of SUI in females was conducted. A total of 45 subjects completed the study. All screened patients were randomized between two groups. One group used the pessary for the first 30 days (or one menstrual cycle) followed by pads (control) for the second 30 days (or second menstrual cycle). The other group used the pads first then the pessary. A clinical evaluation of vaginal microflora was completed before and after each phase. The results demonstrate no clinically significant change in vaginal microflora occurred following the use of the IMPROVE BLADDER SUPPORT and no meaningful change in microflora compared to the control (pad only). No serious adverse events were observed. The data also show that there were few adverse events which were generally mild, anticipated and resolved without medical intervention. The most frequently observed device related adverse effects included discomfort, spotting, bleeding and urinary tract infection. Subanalyses were conducted by race, menopausal status and hormone replacement therapy (HRT) use. The populations of African American. Hispanic and HRT users were too small to draw meaningful conclusions about them, but do not appear to show adverse effects of use.
- . Following several preliminary studies to evaluate and refine the labeling, a selection and labeling comprehension study was performed. The objectives were to assess the ability of users to read the labeling and determine the appropriate user of the product and whether key concepts in the labeling were understood. The patient population was divided into three subgroups comprising a low literacy population, general population and elder population. There were 15 women in each group. In the selection study, participants were required to read the exterior packaging and were then questioned by a moderator who posed different clinical scenarios to determine whether the participant could appropriately determine whether the woman in each scenario should use the product. The labeling comprehension study required that the participant read the instructions for use. The moderator then asked the participant questions to determine the participant's understanding of key points. On average, the participants in the selection study were able to correctly identify the appropriateness of the device 92% of the time. The participants in the labeling comprehension study were able to provide a correct answer 96% of the time. In both studies, the lowest performance was from the low literacy group; however, their results were still acceptable as they did not pose any additional risk to the user.
#### Substantial Equivalence
The IMPROVE BLADDER SUPPORT is as safe and effective as the legally cleared predicate device (K060526-ConTIPI TIPI vaginal pessary). The IMPROVE BLADDER SUPPORT has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. The IMPROVE BLADDER SUPPORT is indicated for overthe-counter use, however, this modification in the indications for use and the labeling does not alter the therapeutic effect of the predicate device. The IMPROVE BLADDER SUPPORT is substantially equivalent to the identified predicate device in terms of overall design, principles of operation, components and configuration, and the technological characteristics. The subject device incorporates minor differences in materials, sizes and shape to improve manufacturability and promote user comfort. Comparative bench testing and clinical data demonstrates substantially equivalent performance of the IMPROVE BLADDER SUPPORT and the predicate device. The minor technological differences between the IMPROVE BLADDER SUPPORT and
{4}------------------------------------------------
the predicate device raise no new issues of safety or effectiveness. .Thus, the IMPROVE BLADDER SUPPORT is substantially equivalent.
:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal for the Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Public Health Service
Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 27, 2013
ConTIPI Ltd. % Jonathan Kahan Partner Hogan Lovells US, LLP 555 Thirteenth St., NW Washington, DC 20004
Re: K131198
Trade/Device Name: IMPROVE BLADDER SUPPORT Regulation Number: 21 CFR 884.3575 Regulation Name: Vaginal Pessary Regulatory Class: Class II Product Code: HHW Dated: July 29, 2013 Received: July 29, 2013
Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 - Jonathan Kahan
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K131198
Device Name: IMPROVE BLADDER SUPPORT
Indications for Use:
The IMPROVE BLADDER SUPPORT is an over-the-counter device intended for the temporary management of stress urinary incontinence (SUI) in women.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _ _ of __ ]
Herbert P. Lerner -S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K131198 510(k) Number
1000 - 037551/000001 - 5107773 v1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.