The implants are intended to support normal bone healing for osteotomies, fractures, and reconstructions. The RECON system consists of various system components and it is indicated for the treatment of fracture fixation, joint decompression and fusion, reconstruction or arthrodeses of small bones. The system can be used in both adult and pediatric patients. Additional information is provided in the corresponding surgical techniques.
Device Story
RECON system comprises metallic plates, screws, and washers for internal bone fixation; supports bone healing in osteotomies, fractures, and reconstructions. Implants available in various sizes to match bone geometry; used by surgeons in clinical settings. Device provides mechanical stabilization of bone segments; facilitates fusion or reconstruction. No electronic or software components; purely mechanical orthopedic fixation system.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and the long-standing clinical history of similar predicate devices.
Technological Characteristics
Materials: Titanium alloy Ti-6Al-4V (ASTM F136) or C.P. Titanium (ASTM F67). Components: Plates, screws, and washers. Mechanical properties: Fatigue strength, bending stiffness, and axial pullout performance verified. Biocompatibility: Tested per ISO 10993-1. Sterilization: Non-sterile (implants), reusable (instruments).
Indications for Use
Indicated for fracture fixation, joint decompression, fusion, reconstruction, or arthrodeses of small bones in adult and pediatric patients.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, and Herbert Mini Bone Screw (K143165)
Universal Locking Plate System 2.7, Plates and Screws (K063303)
Universal Locking Plate System 3.5, Plates and Screws (K060710)
Normed Titanium Osteotomy Plating System (K022325)
Normed Distal Radius Reconstruction System (K022231)
Extremity Fixation Implants for Osteosynthesis (K111678)
Submission Summary (Full Text)
{0}------------------------------------------------
August 12, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Normed Medizin-Technik GmbH Mr. Arne Briest RA/QA Manager Ulrichstrasse 7 D-78532 DE Tuttlingen Germany
Re: K151407 Trade/Device Name: RECON System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: June 16, 2015 Received: June 18, 2015
Dear Mr. Arne Briest:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
#### Page 2 - Mr. Arne Briest
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|-----------------------------------------|--|
| Food and Drug Administration | |
### Indications for Use
510(k) Number (if known)
K151407
page 1 of 1
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017
See PRA Statement below.
Device Name RECON system
Indications for Use (Describe)
The implants are intended to support normal bone healing for osteotomies, fractures, and reconstructions.
The RECON system consists of various system components and it is indicated for the treatment of fracture fixation, joint decompression and fusion, reconstruction or arthrodeses of small bones. The system can be used in both adult and pediatric patients. Additional information is provided in the corresponding surgical techniques.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a logo for NORMED, a Zimmer company. The logo features the text "NORMED" in a bold, sans-serif font, with a globe graphic incorporated into the letter "O". Above the text is the tagline "The specialist for small bones". Below the text is the text "A ZIMMER COMPANY" with a stylized Z to the left.
# 510(k) Summary
9611091
#### Submission Sponsor and Correspondent 1.
Normed Medizin-Technik GmbH Ulrichstrasse 7 D-78532 Tuttlingen Germany
| Phone: | + 49 7461 93 43 0 |
|----------|--------------------|
| Fax: | + 49 7461 93 43 20 |
| Contact: | Mr. Arne Briest |
FDA Establishment Registration #:
#### 2. Date Prepared
Date Prepared: July 22, 2015
#### 3. Device Identification
Trade/Proprietary Name: Regulation Name:
Classification Regulation Product Code:
Device Class: Classification Panel
#### RECON system
Single/Multiple Component Metallic Bone Fixation Appliances and Accessories 21CFR 888.3030 HRS HWC HTN Class II Orthopedic
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a logo for NORMED, a Zimmer company. The logo features the word "NORMED" in a bold, stylized font, with a globe graphic incorporated into the first letter. Above the logo, the text "The specialist for small bones" is written in a smaller font. Below the logo, the text "A ZIMMER COMPANY" is written in a smaller font.
#### 4. Legally Marketed Predicate Device
K143165 - Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, and Herbert Mini Bone Screw
K063303 - Universal Locking Plate System 2.7, Plates and Screws, manufactured by Zimmer Inc., cleared November 22, 2006
K060710 - Universal Locking Plate System 3.5, Plates and Screws manufactured by Zimmer Inc., cleared April 26, 2006
K022325 - Normed Titanium Osteotomy Plating System, manufactured by Normed Medizin-Technik GmbH, cleared September 9, 2009
K022231 - Normed Distal Radius Reconstruction System, manufactured by Normed Medizin-Technik GmbH, cleared September 13, 2002
K143066 - Zimmer Plates and Screws System (ZPS) - Non-sterile ZPS Plate Line Extensions, Sterile/Non-sterile ZPS Screws and Washers manufactured by Zimmer Inc., cleared November 28, 2014
K090675 - VLP Medial Column Fusion Plate - manufactured by Smith & Nephew Inc., cleared June 4. 2009
K121651 - Ortholoc 3DI Midfoot/Flatfoot System - manufactured by Wright Medical Technology, Inc. cleared October 12, 2012
K141784 - UltOS Plating System - manufactured by Ortho Solutions Limited, cleared July 23, 2014
K111678 - Ortho Solutions - Extremity Fixation Implants for Osteosynthesis- manufactured by Ortho Solutions Limited, cleared February 7, 2012
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for NORMED, a Zimmer company. The logo is black and white, with the word "NORMED" in large, bold letters. Above the word "NORMED" is the text "The specialist for small bones". Below the word "NORMED" is the text "A ZIMMER COMPANY" in a smaller font.
#### 5. Device Description
The RECON system is a plate and screw system intended for internal fracture fixation. The system consists of the following plates:
- Reconstruction Plates 3.5, ●
- Medialis Plates 3.5, ●
- Universal Plates 3.5
- Universal Plates 3.5, H-shaped .
- USG Plates 3.5 ●
- Interpositioning Arthrodesis Plates ●
- Lapidus Plates ●
- . Dwyer Plates
screws and washers of different sizes and designs.
The plates, screws, and washers are either made of titanium alloy Ti-6AI-4V (ASTM F136) or commercially pure Titanium (C.P. Titanium; ASTM F67).
The implants are offered in various sizes to accommodate the variations of bone size and geometry. The implants are provided non-sterile and single-use only. The instruments are non-sterile and reusable.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for NORMED, a Zimmer company. The logo is black and white and features the word "NORMED" in large, bold letters. Above the word "NORMED" is the text "The specialist for small bones". Below the word "NORMED" is the text "A ZIMMER COMPANY".
#### 6. Indications for Use
The implants are intended to support normal bone healing for osteotomies, fractures, and reconstructions.
The RECON system consists of various system components and it is indicated for the treatment of fracture fixation, joint decompression and fusion, reconstruction or arthrodeses of small bones. The system can be used in both adult and pediatric patients. Additional information is provided in the corresponding surgical techniques.
#### 7. Substantial Equivalence Discussion
The RECON system has the same intended use, similar performance characteristics, is manufactured from similar materials and is similar in design to the predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image is a logo for NORMED, a Zimmer company. The logo features the text "NORMED" in a bold, sans-serif font, with the "O" replaced by a globe graphic. Above the logo, the text "The specialist for small bones" is written in a smaller font. Below the logo, the text "A ZIMMER COMPANY" is written in a smaller font, accompanied by the Zimmer logo.
#### 8. Non-Clinical Performance Data
- -Biocompatibility - Biocompatibility testing on the plates, screws, and washer material was conducted and evaluated per ISO 10993-1. All testing passed
- -Cross sectional engineering analysis of the plates, proof load, bending strength, bending stiffness, equivalent bending stiffness and fatigue strength of the RECON System -plates and the predicate devices, the Zimmer Universal Locking System (ULS) Plates and Screws (K063303 and K060710) , resulted in similar mechanical properties and performance. The subject and predicate devices are substantially equivalent.
- -Self-tapping performance, driving and removal torque, to failure and axial pullout of the RECON System -screws and the predicate devices, the Zimmer Universal Locking System (ULS) Plates and Screws (K063303 and K060710) and Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, and Herbert Mini Bone Screw (K143165), resulted in similar mechanical properties and performance. The subject and predicate devices are substantially equivalent.
- -The subject washers are substantially equivalent to the washers cleared in K143066
#### 9. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device.
#### 10. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the RECON system and the predicate devices do not raise any questions regarding its safety and effectiveness.
Performance testing and compliance with voluntary standards, demonstrate that the RECON system are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for NORMED, a Zimmer company. The logo is black and white and features the word "NORMED" in a stylized font. Above the word "NORMED" is the text "The specialist for small bones". Below the word "NORMED" is the text "A ZIMMER COMPANY".
The RECON system is determined to be substantially equivalent to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.