Vitrea Software Toshiba Package is an application package developed for use on Vitrea, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea has the following additional indications: The CT/XA Cerebral Artery Morphological Analysis application is intended to facilitate the extraction and segmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio. The MR Wall Motion Tracking application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and regional myocardial function that is used for patients with suspected heart disease.
Device Story
Vitrea Software Toshiba Package (VSTP-001A) is a post-processing software suite for medical imaging workstations. It processes CT, XA, and MR image data to assist physicians in specialized clinical analysis. The CT/XA Cerebral Artery Morphological Analysis module performs semi-automatic segmentation of user-identified cerebral aneurysms and calculates size and aspect ratio. The MR Wall Motion Tracking module performs cardiac functional analysis, including global/regional myocardial function, wall motion, wall thickness, and strain analysis. Physicians use these outputs as an adjunct to diagnosis to assess heart disease or cerebral vascular morphology. The software is intended for clinical use by physicians to aid in clinical decision-making and patient management.
Clinical Evidence
Clinical evaluations were conducted to demonstrate that the CT/XA Cerebral Artery Morphological Analysis and MR Wall Motion Tracking applications performed as intended. Results confirmed that the CT/XA application was comparable to manual measurements and segmentations, and the MR Wall Motion Tracking contour tracking process met the required success ratio. Bench testing using numerical phantoms was also performed to verify the accuracy of aneurysm extraction, measurement calculations, and cardiac function/strain analysis.
Technological Characteristics
Software-based medical image processing package. Operates on Vitrea workstation. Implements semi-automatic segmentation (seed-point based) and automated measurement algorithms. Complies with IEC 62304 (software lifecycle) and IEC 62366 (usability). Developed under ISO 13485 quality systems. Moderate level of concern software.
Indications for Use
Indicated for patients with suspected heart disease requiring cardiac functional analysis and patients requiring cerebral artery aneurysm measurement and size/aspect ratio assessment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around the perimeter. In the center is a stylized caduceus symbol, which is often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 13. 2015
Toshiba Medical Systems Corporation % Mr. Orlando Tadeo Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K151091
Trade/Device Name: Vitrea Software Toshiba Package, VSTP-001 A Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 22, 2015 Received: April 23, 2015
Dear Mr. Tadeo:
This letter corrects our substantially equivalent letter of July 13, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael O'Hara
For
Robert Ochs, Ph.D Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
k151091
Device Name
Vitrea Software Toshiba Package, VSTP-001A
#### Indications for Use (Describe)
Vitrea Software Toshiba Package is an application package developed for use on Vitrea, a medical image processing software, which includes the following post-processing software applications.
CT/XA Cerebral Artery Morphological Analysis This software is intended to facilitate the extraction and seqmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio.
#### MR Wall Motion Tracking
This application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and reqional myocardial function that is used for patients with suspected heart disease.
Type of Use (Select one or both, as applicable)
ك Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
TOSHIBA AMERICA MEDICAL SYSTEMS, INC. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000
## 510(k) SUMMARY
- 1. SUBMITTER'S NAME: Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
#### 2. OFFICIAL CORRESPONDENT: Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
- 3. ESTABLISHMENT REGISTRATION: 9614698
## 4. CONTACT PERSON:
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
## 5. Date Prepared:
April 30, 2015 (Updated 07/07/2015)
## 6. TRADE NAME(S):
Vitrea Software Toshiba Package, VSTP-001A
- 7. COMMON NAME:
Radiological Image Processing Software
## 8. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.2050, Picture Archiving and Communications System)
## 9. PRODUCT CODE / DESCRIPTION: 90LLZ / Picture Archiving and Communications System
# 10. PERFORMANCE STANDARD:
None
{4}------------------------------------------------
# 11. PREDICATE DEVICES:
| Device Name | Vitrea, Version 7.0<br>Medical Image<br>Processing Software | MRT-301/A5, Vantage<br>Titan 3T, v2.0<br>(Reference Device) | Aquilion ONE Vision,<br>V7.0<br>(Reference Device) |
|----------------------------|-------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------|
| Application<br>Name | | Cardiac Analysis<br>Application<br>For Superconducting<br>MRI System | 4D Cerebral Artery<br>Morphological Analysis |
| Marketed By | Vital Images, Inc. | Toshiba America<br>Medical Systems | Toshiba America<br>Medical Systems |
| 510(k) Clearance<br>Number | K150258 | K120487 | K142465 |
| Clearance Date | March 5, 2015 | May 23, 2012 | March 10, 2015 |
| Product Code | LLZ, Picture Archiving<br>and Communications<br>System | LNH, Magnetic<br>Resonance Diagnostic<br>Device | JAK, Computed<br>Tomography X-Ray<br>System |
## 12. REASON FOR SUBMISSION:
Adding existing software applications to an imaging workstation (Vitrea; Product Code LLZ). MR Wall Motion Tracking includes an improved Contour Tracking function which allows for heart wall strain analysis. CT/XA Cerebral Artery Morphological Analysis allows for the import of volume data from an interventional system in addition to the cleared CT volume analysis.
## 13. DEVICE DESCRIPTION:
Vitrea Software Toshiba Package, VSTP-001A, an application package developed for use on Vitrea, a medical image processing software, marketed by Vital Images. Inc. Vitrea Software Toshiba Package, VSTP-001A, includes two post processing applications, CT/XA Cerebral Artery Morphological Analysis and MR Wall Motion Tracking, which use brain and cardiac image data, obtained from CT/XA/MR systems, to assist physicians in performing specialized measurements and analysis.
## 14. INDICATIONS FOR USE:
Vitrea Software Toshiba Package is an application package developed for use on Vitrea®, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea has the following additional indications:
The CT/XA Cerebral Artery Morphological Analysis application is intended to facilitate the extraction and segmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio.
The MR Wall Motion Tracking application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and regional myocardial function that is used for patients with suspected heart disease.
{5}------------------------------------------------
# 15. SUBSTANTIAL EQUIVALENCE:
The software applications included in the Vitrea Software Toshiba Package, VSTP-001A, perform in a manner similar to and are intended for the same use as the predicate devices. These applications include modifications to the originally cleared devices, 4D Cerebral Artery Morphological Analysis (K142465) and Cardiac Analysis Application For Superconducting MRI System (K120487), in order to be available on the Vitrea, Medical Image Processing Software (K150258). The subject and predicate device applications are all post processing applications used to aid physicians with performing specialized measurement and analysis of CT, XA and MR image data.
Complete comparison tables are included in this submission. See below for a brief summary of technological characteristics of the software applications:
| Item | CT/XA Cerebral Artery<br>Morphological Analysis | 4D Cerebral Artery<br>Morphological Analysis for<br>Aquilion ONE |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------|
| 510(k) Clearance | Subject Device | K142465 |
| Type of Input Data | One volume acquired by X-ray<br>CT system or interventional<br>angiography system | Multiple volumes acquired by X-<br>ray CT system |
| Anatomical Region | Cerebral Artery | Cerebral Artery |
| Segmentation<br>(Extraction Of<br>Aneurysm-Shaped<br>Region) | Semi-Automatic Segmentation<br>With One-Clicked Seed Point | Semi-Automatic Segmentation<br>with One-Clicked Seed Point |
| Measurement<br>(Distance, Angle,<br>Length) | Automatic | Automatic |
| ltem | MR Wall Motion Tracking | Cardiac Analysis Application<br>For Superconducting MRI<br>System |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance | Subject Device | K120487 |
| Anatomical Region | Cardiac | Cardiac |
| Myocardium<br>Contour Detection | Contour tracking | Contour detection applied to all<br>phases individually |
| Global Cardiac<br>Function Analysis | Ejection Fraction, Cardiac<br>Output, Volume Curves are<br>calculated from shapes of<br>myocardial contours. | Ejection Fraction, Cardiac<br>Output, Volume Curves are<br>calculated from shapes of<br>myocardial contours. |
| Regional Wall<br>Motion Analysis | Wall motion, Wall thickness,<br>Wall thickening are calculated | Wall motion, Wall thickness,<br>Wall thickening are calculated |
| Strain Analysis | Strain is calculated | None |
{6}------------------------------------------------
| Result Display | | |
|----------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| | Cine View,<br>Polar Map Of Measurement<br>Results<br>Time-Curve Of Regional<br>Measurement | Cine View,<br>Polar Map Of Measurement<br>Results<br>Time-Curve Of Regional<br>Measurement |
# 16. SAFETY:
The device is designed and manufactured under the Quality System Requlations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC62304 and IEC62366.
# 17. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrates that the requirements for the applications have been met. Bench studies were conducted using numerical phantoms to analyze the accuracy of extraction/display of aneurysm shaped regions as well as measurement calculations and to analyze cardiac function and strain. Additionally, clinical evaluations were conducted to demonstrate that the CT/XA Cerebral Artery Morphological Analysis and MR Wall Motion Tracking applications performed as intended. The results confirmed that CT/XA Cerebral Artery Morphological Analysis was comparable to manual measurements and/or segmentations and that the contour tracking process of the MR Wall Motion Tracking application met the required success ratio.
Software Documentation for a Moderate Level of Concern, per the FDA quidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
## 18. CONCLUSION
The software applications included in the Vitrea Software Toshiba Package, VSTP-001A, perform in a manner similar to and are intended for the same use as the predicate devices. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.