SIEMENS ACUSON 2000 ULTRASOUND SYSTEM WITH ABVS TRANSDUCER
K081148 · Siemens Medical Solutions USA, Inc., Ultrasound DI · IYO · Dec 22, 2008 · Radiology
Device Facts
Record ID
K081148
Device Name
SIEMENS ACUSON 2000 ULTRASOUND SYSTEM WITH ABVS TRANSDUCER
Applicant
Siemens Medical Solutions USA, Inc., Ultrasound DI
Product Code
IYO · Radiology
Decision Date
Dec 22, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Modified S2000, the S2000 ABVS Ultrasound System, is intended for the following applications: General Radiology, Abdominal, Fetal, Small Parts, Transcranial, OB/GYN, Cephalic, Pediatric, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system supports the 14LSBV for Bmode imaging of a patient's breast using an optional automatic scanning function. The device is not intended to be used as a replacement for screening mammography. The system also provides for the measurement of anatomical structures and analysis as provided in the original S2000 Ultrasound System, which provides information that is used by medical health care professionals for clinical diagnosis purposes.
Device Story
Acuson S2000 ABVS is a diagnostic ultrasound system for clinical imaging and anatomical measurement. It processes acoustic signals from various transducers (e.g., 14LSBV for automated breast scanning) to produce B-mode, M-mode, Doppler, and 3D/4D images. Operated by clinicians in radiology, cardiology, and other specialties, the system provides real-time visualization and quantitative data to assist in clinical diagnosis. Features include tissue harmonic imaging, spatial compounding, tissue equalization, and elasticity imaging. Output is displayed on the system monitor for physician interpretation to guide clinical decision-making. The device is intended for professional use in clinical settings.
Clinical Evidence
Bench testing only. The device modifications were verified and validated according to the manufacturer's design control process to ensure compliance with safety and performance standards, including IEC 60601-1, IEC 60601-2-37, and AIUM/NEMA acoustic output standards.
Technological Characteristics
Diagnostic ultrasound system utilizing pulsed echo and Doppler imaging. Supports multiple transducer types (curved, linear, phased array, mechanical 3D). Complies with UL 60601-1, IEC 60601-2-37, and ISO 10993-1 (biocompatibility). Features include SieClear spatial compounding, 3-Scape real-time 3D, and eSie Touch elasticity imaging. Connectivity via standard ultrasound interfaces.
Indications for Use
Indicated for ultrasound imaging or fluid flow analysis in patients requiring diagnostic evaluation of abdominal, fetal, small parts (breast, testes, thyroid, etc.), transcranial, OB/GYN, pediatric, urological, vascular, and musculoskeletal structures. Includes automatic breast scanning via 14LSBV transducer. Not for screening mammography. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# DEC 22 2011
### 510(k) Summary Prepared August 27, 2008 Revised December 18, 2008
| Sponsor: | Siemens Medical Solutions USA, Inc..<br>Ultrasound Division<br>1230 Shorebird Way<br>P.O. Box 7393<br>Mountain View, California 94039-7393 | |
|-----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Contact Person: | Sheila W. Pickering Ph.D.<br>Telephone:<br>Fax: | (650) 943 7187<br>(650) 943 7053 |
| Submission Date: | April 18, 2008 | |
| Device Name: | Acuson S2000 ABVS Ultrasound System | |
| Common Name: | Diagnostic Ultrasound System with Accessories | |
| Classification:<br>Regulatory Class:<br>Review Category:<br>Classification Panel: | II<br>Tier II<br>Radiology | |
| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN |
|------------------------------------------|---------------|---------------------|
| Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IYC |
| Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-ITX |
## A. Legally Marketed Predicate Devices
The Acuson S2000 ABVS Ultrasound system is substantially equivalent to the Acuson S2000 ultrassund system (K072786)and the U-Systems ultrasound system.
#### B. Device Description:
The Acuson S2000 ABVS Ultrasound System has been designed to meet the following product safety standards:
- UL 60601-1, Safety Requirements for Medical Equipment .
- I IFC 60601-2-37 Diagnostic Ultrasound Safety Standards
- CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment D
- AJUM/NEMA UD-3, 1998 Standard for Real Time Display of Thermal and 이 Mcchanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- AIUM/NEMA UD-2, 1998 Acoustic Output Measurcment Standard for Diagnostic 1 Ultrasound
- l 93/42/EEC Medical Devices Directive
- Safety and EMC Requirements for Medical Equipment L
- . EN/IEC 60601-1
- 8 EN/IEC 60601-1-1
- B EN/IEC 60601-1-2
- IEC 1157 Declaration of Acoustic Power .
- . ISO 10993-1 Biocompatibility
{1}------------------------------------------------
#### C. Intended Use
The Modified S2000, the S2000 ABVS Ultrasound System, is intended for the following The Modified Beeral Radiology, Abdominal, Fetal, Small Parts, Transcranial, OB/GYN, apprications: Contral News Cephalic, Pediatric, Urology, Vascular, Musculoskeletal, Supericial Musculoskeletal, and Peripheral Vascular applications. The system supports the Superilean Massantanteretalters for Use tables, including the 14LSBV for Bmode imaging of a patient's breast using an optional automatic scanning function. The device is not intended to be used as a replacement for screening mammography.
The system also provides for the measurement of anatomical structures and analysis as provided in the original S2000 Ultrasound System, which provides information that is used by medical heulth care professionals for clinical diagnosis purposes.
#### I). Substantial Equivalence
The submission device is substantially equivalent to the predicate devices with regard to both intended use and technological characteristics.
### E. Performance Data
The S2000 modifications are verified and validated according to the company's design control process.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of the human form.
Public Health Service
# DEC 16 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Ms. Shelly Pearce Regulatory Affairs Siemens Medical Solutions USA, Inc. Ultrasound Division 1230 Shorebird Way MOUNTAIN VIEW WAY CA 94043-1344
Re: K081148
Trade/Device Name: Acuson S2000 ABVS Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: November 21, 2008 Received: November 26, 2008
Dear Ms. Pearce:
This letter corrects our substantially equivalent letter of December 22, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Acuson S2000 ABVS Ultrasound System, as described in your premarket notification:
Transducer Model Number
CW2 Probe CW5 Probe EC9-4 Curved Array 9L4 Linear Array 14L5 Multi-D Array 4P1 Phased Array 6C2 Curved Array 4C1 Curved Arrav 4V1 Phased Array
{3}------------------------------------------------
10V4 Phased Array 14L5 SP Linear Array 7CF2 Curved Array Mechanical 3D 9EVF4 Curved Array V5Ms Multiplane TEE 18L6HD Linear Array 8V3 Phased Array 14L5BV Multi-D Array
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours,
Jope Mr Why
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
Special 510(k) Notification
Siemens Ultrasound Division
CONFIDENTIAL
Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
K081148
Device Name:
ACUSON S2000 ABVS Ultrasound System
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|--------------------------|------------------------------|---------------------------|-----------------------------------|
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitud<br>e<br>Doppler | Color<br>Velocity<br>Imaging | Combin<br>ed<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | P | P | P | P | P | P | | | | |
| Fetal | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10, 11,<br>13 |
| Abdominal | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10, 11,<br>13 |
| Intraoperative<br>(Note 9) | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10, 11,<br>14 |
| Intraoperative<br>Neurological | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,10, 11,<br>14 |
| Pediatric | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10, 11 |
| Small Organ<br>(Note 1) | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10,<br>11,14 |
| Neonatal<br>Cephalic | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10 |
| Adult Cephalic | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10 |
| Cardiac | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,6,7,8,10 |
| Trans-<br>esophageal | | P | P | P | P | P | P | | BMDC | |
| Transrectal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,10,<br>11,14 |
| Transvaginal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,10, 11 |
| Transurethral<br>Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC | Note2,3,4,5,6,7,8,<br>10, 11,14 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,10,<br>11,14 |
{5}------------------------------------------------
#### Special 510(k) Notification Siemens Ultrasound Division CONFIDENTIAL
| Musculo-skeletal<br>Superficial | P | P | P | P | P | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 |
|-------------------------------------|---|---|---|---|---|------|-------------------------------|
| Other (specify)<br>Neonatal Cardiac | P | P | P | P | P | BMDC | Note 3,4,6 |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Additional Comments:
Note 1 For example: breast, testes, thyroid, penis, prostate, etc.
Note 2 Ensemble tissue harmonic imaging
Note 3 SieClear multi-view spatial compounding
Note 4 Tissue Equalization Technology
Note 5 3-Scape real-time 3D imaging
Note 6 Cadence contrast agent imaging
Note 7 B&W SieScape panoramic imaging
Note 8 Power SieScape panoramic imaging
Note 9 For example: vascular, abdominal
Note 10Clarify VE vascular enhancement technology
Note 11 Advanced Sieclear multi-view spatial compounding
Note 13 STIC
Note 14 eSie™ Touch elasticity imaging/FTI
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Aogui m Why
Prescription Use (Per 21 CFR 801.109) 1
(Division/Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K081148
{6}------------------------------------------------
Special 510(k) Notification Siemens Ultrasound Division CONFIDENTIAL
#### Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name:
Intended Use:
# 14LSBV Multi-D Array Transducer for use with ACUSON S2000 ABVS
Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|---------------------------|-----------------------|-----------------------------------|
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Note 1) | | N | | | | | | | | Note<br>2,3,4,5,7,8,10,<br>11, 14 |
| Neonatal<br>Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063085; E = added under Appendix E
Additional Comments:
Note 1 For example: breast, testes, thyroid, penis, prostate, etc.
Note 2 Ensemble tissue harmonic imaging
Note 3 SieClear multi-view spatial compounding
Note 4 Tissue Equalization Technology
Note 5 3-Scape real-time 3D imaging
Note 6 Cadence contrast agent imaging
Note 7 B&W SieScape panoramic imaging
{7}------------------------------------------------
Special S10(k) Notification Siemens Ultrasound Division CONFIDENTIAL
Note 8 Power SieScape panoramic imaging Note 10Clarify VE vascular enhancement technology Note 11 Advanced Sieclear multi-view spatial compounding Note 14 eSie™ Touch elasticily imaging/FTI
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Hozan M. Wly
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices, 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.