SPINAUT-E is used for delivery of drugs that have been indicated for the epidural space. SPINAUT-I is intended for the percutaneous introduction and placement of an epidural catheter.
Device Story
SPINAUT-E and SPINAUT-I are sterile, single-use anesthesia conduction catheters. SPINAUT-E features a flexible catheter, steering handle, and infusion port; internal steering mechanism allows tip deflection (50° or 100°) to guide the catheter through epidural space and soft tissues. SPINAUT-I acts as an introducer to prepare or preserve a path for the catheter. Used by physicians in clinical settings; guided under fluoroscopy. Port connects to syringes for therapeutic agent delivery. Output is the physical placement of the catheter and subsequent drug delivery. Benefits include targeted access to the source of distress for pain management.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included sterilization validation (ISO 11137), shelf-life validation (ASTM F1980-07), biocompatibility (ISO 10993-5, -10, -11, -4, USP <151>), and mechanical performance testing (leakage, tensile strength, fatigue, flexural rigidity, corrosion resistance, and X-ray impermeability).
Technological Characteristics
Materials: SUS 304, PU, ABS, PE, PTFE, Polyurethane. Principle: Steerable catheter with manual handle-actuated tip deflection. Dimensions: 1.7mm O.D., lengths 300-330mm. Connectivity: None. Sterilization: Gamma irradiation. Software: None.
Indications for Use
Indicated for patients requiring epidural drug delivery (SPINAUT-E) or percutaneous placement of an epidural catheter (SPINAUT-I).
Regulatory Classification
Identification
An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.
EBI TargetCath Fluoro-Guided Steerable Catheter System (K032390)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, overlapping each other to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 1, 2016
IMEDICOM Co., Ltd. c/o Ms. Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave. Ste 110 Fullerton CA 92831
Re: K150789 Trade/Device Name: SPINAUT-E, SPINAUT-I Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia conduction catheter Regulatory Class: Class II Product Code: BSO Dated: January 21, 2016
Received: January 27, 2016
Dear Ms. Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K150789
Device Name
SPINAUT-E, SPINAUT-I
#### Indications for Use (Describe)
SPINAUT-E is used for delivery of drugs that have been indicated for the epidural space.
SPINAUT-I is intended for the percutaneous introduction and placement of an epidural catheter.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CER 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CER 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(K150789)
This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Feb 17, 2016
## 1. Applicant / Submitter
IMEDICOM Co., Ltd. #612, 172, LS-ro, Gunpo-si, Gyeonggi-do, 435-824 Republic of Korea
Tel:+82-31-479-1156 Fax:+82-31-479-1157
## 2. Submission Correspondent
LK Consulting Group USA, Inc. 2651 E Chapman Ave. Ste 110, Fullerton CA 92831 Priscilla Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
## 3. Device
- Trade Name: SPINAUT-E, SPINAUT-I י
- י Common Name: Anesthesia conduction catheter
- Classification Name: Anesthesia conduction catheter י
- ' Product Code: BSO
- Classification regulation: 21 CFR 868.5120 י
#### 4. Predicate Device:
I Primary Predicate Device: Myelotec Video Guided Catheter by Myelotec, Inc. (K980734)
Reference Predicate Device: EBI TargetCath Fluoro-Guided Steerable Catheter I System by EBI, L.P. (K032390)
{4}------------------------------------------------
#### 5. Description:
SPINAUT-E consists of a flexible catheter, steering handle and a port for access to the lumen. The catheter has built in steering mechanism that allows for guiding the soft tip through the epidural space and soft tissues for optimal access to the source of distress. The port facilitates the connection of syringes to deliver therapeutic agents by physicians as appropriate to their diagnosis. It is supplied sterile and it is for single use.
There are 9 models in SPINAUT-E: E1000, E2000, E1100, E2100, E3100, S1000, S2000, and S3000. The catheter tip of the model E1000, and E3000 can be bent 50° to the left or to the right (the user can rotate the tip to adjust the direction), but the tip of the model E1100, E2100, and E 3100 can be bent either direction up to 100°. The S1000, S2000, and S3000 have an extra infusion port for drug so that the other main port can be used for other devices. Each model is characterized by the additional port, the lever, and the dimensions.
The SPINAUT-I is intended to be used to prepare or preserve a path for an epidural catheter. When placing SPINAUT-E into the patient, SPINAUT-I can be used as a replacement of the SPINUT-E Catheter Introducer Set. It is supplied sterile and it is for single use.
There are 3 models in SPINAUT-I: ID1000, ID2000, and ID3000 and each model is characterized by the length.
#### 6. Indication for use:
SPINAUT-E is used for delivery of drugs that have been indicated for the epidural space.
SPINAUT-I is intended for the percutaneous introduction and placement of an epidural catheter.
#### 7. Basis for Substantial Equivalence
SPINUAT-E and SPIANTU-I are substantially equivalent to the Myelotec Video Guided Catheter (K980734) made by Myelotec, Inc. and to the EBI TargetCath Fluoro-Guided Steerable Catheter System by EBI, L.P. (K032390).
There are no significant differences between the subject devices and the predicate devices that the subject devices have the similar intended use as the identified predicate devices and they are similar in fundamental scientific technology, design, and size, and they are sterilized via gamma irradiation.
The principles of operation are similar between the subject device and the predicate devices that they consist of a flexible catheter, steering handle and associated port for
{5}------------------------------------------------
access to the lumen. The port facilitates the connection of syringes to deliver physician selected therapeutic agents as appropriate to their diagnosis.
In terms of features, the major difference between the subject device and the primary predicate device is that the predicate device has one more designated port allowing the use of a flexible fiber optic endoscope. This is an extra feature that the predicate device offers but the common epidural catheters in the market are guided under fluoroscope as the subject device is, so we believe this difference would not render the device NSE. To support this, we have identified a reference predicate device which has the similar intended use and also used under fluoroscope as the subject device.
The materials used in the subject devices might be different from the predicate devices; however, the biocompatibility testing results of the subject device support that the subject devices are biocompatible. The size and the internal design also might be different between the devices but the performance testing results show that the subject device would perform as well as the predicate devices.
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SPINAUT-E<br>SPINAUT-I | Myelotec Video Guided Catheter | EBI TargetCath Fluoro-Guided<br>Steerable Catheter System |
| 510(k) Number | - | K980734 | K032390 |
| Product Code | BSO | BSO | BSO |
| Manufacturer | IMEDICOM Co., Ltd. | Myelotec, Inc. | EBI, L.P. |
| Intended Use | SPINAUT-E is used for delivery of<br>drugs that have been indicated for<br>the epidural space.<br>SPINAUT-I is intended for the<br>percutaneous introduction and<br>placement of epidural catheters. | When used with a fiberoptic<br>endoscope, this device can be used<br>for observing epidural anatomy.<br>pathology and delivery of drugs<br>approved for epidural indications. | When used with a fluoroscope, the<br>EBI® TargetCath™ Fluoro-Guided<br>Steerable Catheter System can be<br>used in the lumbar and sacral spine<br>for delivery of drugs approved for<br>epidural indications. The System<br>may also be used for the purpose of<br>assisting in the diagnosis and<br>treatment of disease utilizing a<br>caudal approach via the sacral<br>hiatus. |
| Diameter(O.D.) of<br>Catheter | 1.7mm | 2.7mm, 3.0mm, 3.3mm | 2.4mm |
| Length of Catheter | 300mm/ 315mm/ 330mm | 300mm | 280mm |
| Infusion port | One port | Dual port | Dual port |
| Material | SUS 304, PU, ABS, PE, PTFE,<br>ABS, Polyurethane | - | - |
| Steering | Steerable | Steerable | Steerable |
| Single use | Yes | Yes | Yes |
{6}------------------------------------------------
| Components | Epidural catheter, guide wire,<br>catheter introducer, introducer<br>needle, and needle cap | Epidural catheter, guide wire,<br>catheter introducer, introducer<br>needle, and needle cap | Epidural catheter and various<br>accessories |
|-------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------|
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible |
| Endoscope capable | Incapable | Capable | Incapable |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization |
#### 8. Non-Clinical Testing
- Sterilization validating testing has been performed in accordance with ISO 11137-1, 2, . 3 and ISO 11737-1. 2. 3. and the test results met the pre-set criteria.
- The tests to validate the shelf life of the device through the proposed shelf life were conducted using the accelerated aging method in accordance to ASTM F1980-07 and the test results validated 3 year shelf life. The following tests were conducted: - SPINAUT-E: Visual Inspection (Appearance), Size Measurement, Leakage Test (Catheter), Tensile strength (Catheter), Curved strength (Catheter), Fatigue test (Catheter), X-ray impermeable test, Corrosion resistance test, Flexural Rigidity (Catheter introducer and the needle), Pulling (Drawing) Test (Catheter introducer and the needle), Package Tensile Test, Package Peeling Test, Package Burst Test, Package Dye Penetration Test, Package Bubble Test, and Sterility Test. - SPINAUT-I: Visual Inspection (Appearance), Size Measurement, Leakage Test, Tensile strength, Flexural Rigidity, Connecting Part (Conical) Fitting, Package Tensile Test, Package Peeling Test, Package Burst Test, Package Dye Penetration Test, Package Bubble Test. and Sterility Test.
- · The subject device falls into the category below based on the ISO 10993, Part I. Category: External communicating device (tissue/bone/dentin communicating), Limited exposure (<24 hours) The following biocompatibility tests were performed and the test results supported that the subject devices are biocompatible. - Extraction Test, Cytotoxicity (ISO 10993-5), Skin Sensitization (ISO 10993-10), Intracutaneous Reactivity (ISO 10993-10), Acute Systemic Toxicity Test (ISO 10993-11), Pyrogen (USP 37 <151>), Hoemolysis (ISO 10993-4), Endotoxin-specific LAL (Limulus Amebocyte Lysate) test (USP 38 NF33)
- Various bench tests and sterility test were performed to evaluate the performance and . the safety of the subject devices and the test results met the pre-set criteria. The following tests have been conducted. - SPINUAT-E: Visual Inspection (Appearance), Leakage Test (Catheter), Tensile strength (Catheter), Curved strength (Catheter), Fatigue test (Catheter), X-ray impermeable test. Corrosion resistance test. Flexural Rigidity (Catheter introducer and the needle), Pulling (Drawing) Test (Catheter introducer and the needle), and Sterility Test.
{7}------------------------------------------------
- SPINAUT-I: Visual Inspection (Appearance), Leakage Test, Tensile strength, Flexural Rigidity, Connecting Part (Conical) Fitting, and Sterility Test.
The test results supported that the subject device is substantially equivalent to the predicate devices in safety and effectiveness.
## 9. Conclusion
Based on the similarities, we conclude that the SPINAUT-E and SPINAUT-I are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.