NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System
K150561 · Narang Medical Limited · HRS · Nov 25, 2015 · Orthopedic
Device Facts
Record ID
K150561
Device Name
NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System
Applicant
Narang Medical Limited
Product Code
HRS · Orthopedic
Decision Date
Nov 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NET Brand Small Fragment and Large Fragment Osteosynthesis System is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, distal femur, proximal tibial pilon, fibula, pelvis and acetabulum fractures; periprosthetic fractures; The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine. NET Brand DHS/DCS plating system may be used for fixation of the fractures of proximal femur such as femoral neck, trochanteric, pertrochanteric or intertrochanteric zones, The system is indicated for use in adult patients only. All implants are for single use only.
Device Story
System comprises metallic bone plates and screws for orthopedic trauma and reconstructive surgery. Plates feature various geometric configurations and screw holes for locking/non-locking screws; screws include cortex self-tapping and locking thread designs. DHS/DCS components include plates, lag screws, and compression screws for proximal femur fixation. Devices are used in clinical settings by surgeons for fracture fixation. Implants are provided non-sterile and require sterilization by the user. Output is mechanical stabilization of bone fractures. Benefits include restoration of skeletal alignment and support for bone healing.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing only, including mechanical performance verification (static/dynamic four-point bend tests for plates; torsional, insertion/removal torque, and axial pull-out tests for screws) per ASTM F382, F384, and F543.
Technological Characteristics
Materials: Ti-6Al-4V alloy (ASTM F136) or stainless steel (ASTM F138, F139). Components: Plates (straight/geometric) and screws (cortex/locking). DHS/DCS system includes lag, compression, and cortex screws. Sterilization: Non-sterile, user-sterilized via autoclaving (AAMI ST79, ISO 17665-1). Mechanical performance verified per ASTM F382, F384, F543.
Indications for Use
Indicated for adult patients requiring fixation of small and large bone fractures, arthrodesis, and osteotomy fixation, including clavicle, scapula, humerus, olecranon, radius, ulna, distal femur, proximal tibial pilon, fibula, pelvis, acetabulum, and periprosthetic fractures. DHS/DCS system indicated for proximal femur fractures (femoral neck, trochanteric, pertrochanteric, intertrochanteric). Locking systems suitable for osteopenic bone. Contraindicated for use in the spine.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a simple, flowing line.
November 25, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Narang Medical Limited Mr. Vivek Narang Director 46, Naraina Industrial Area, Phase-1 New Delhi 110028 India
Re: K150561
Trade/Device Name: NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: September 22, 2015 Received: September 29, 2015
Dear Mr. Narang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
Page 2 - Mr. Vivek Narang
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
#### 510(k) Number (if known) K150561
#### Device Name
NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System
#### Indications for Use (Describe)
NET Brand Small Fragment and Large Fragment Osteosynthesis System is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, distal femur, proximal tibial pilon, fibula, pelvis and acetabulum fractures; periprosthetic fractures; The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine.
NET Brand DHS/DCS plating system may be used for fixation of the fractures of proximal femur such as femoral neck, trochanteric, pertrochanteric or intertrochanteric zones, The system is indicated for use in adult patients only. All implants are for single use only.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Premarket Notification 510(k) Summary As required by Section 807.92 General Company Information as required by 807:92 (a) (a.1)
| Submitter's Name | : | Narang Medical Limited |
|---------------------|---|----------------------------------------------------------------------------|
| Address | | |
| Office | : | 46, Community Center, Naraina Industria Area,<br>Phase-1, New Delhi 110028 |
| Factory | : | Plot Number D-4, Sector A-2, Tronica City, Loni,<br>Ghaziabad, 201102 |
| CONTACT PERSON NAME | : | Mr. Vivek Narang |
| TITLE | : | Director |
| PHONE NUMBER | : | +91-45554000 |
| Dated | : | 18-11-2015 |
{4}------------------------------------------------
### This is a bundled submission.
Throughout the submission there is a mention of NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System that represents the range of products covered under this 510(k) submission.
a.2: The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
#### Proprietary Name:
NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System,
NET Brand of DHS/DCS Plating System
#### Common or Usual Name:
Orthopaedic Bone Plates
Orthopaedic Bone Screws
#### Classification Name:
PLATES, FIXATION, BONE SCREWS, FIXATION, BONE
## Product Code:
HRS, HWC
#### Device Class:
ll
#### Review Panel:
Orthopaedic
#### Regulation Number:
21 CFR 888.3030 and 21 CFR 888.3040
{5}------------------------------------------------
## Variants/Types:
NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System are further subdivided into following categories
| S. No. | Category | Types |
|--------|-------------------------------------------------|---------|
| 01 | Small Fragment Osteosynthesis<br>Plating System | Locking |
| 02 | Large Fragment Osteosynthesis<br>Plating System | Locking |
| 03 | DHS/DCS Plating System | NA |
## Further Description:
NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System consists of plates and screws in a variety of designs and sizes and made from Ti-6Al-4V alloy or stainless steel. Plates are provided in straight designs and in various geometric configurations that are commonly used in trauma and reconstructive surgery. Plates are provided with screw holes to accommodate non-locking and locking screws designs. Screws are provided in, 3.5mm Cortex Self-tapping, 4.5 mm Cortex self-tapping and 2.7mm self-tapping cortex locking, 3.5mm selftapping Cortex Locking, 5.0 mm cortex Locking thread designs in various lengths. This system is not indicated for use in spine.
NET Brand of DHS/DCS Plating System made from Ti-6Al-4V alloy or stainless steel and consist of DHS/DCS Plates, lag Screw, compression screw, and 4.5 Cortex screw Self Tapping, The DHS plates are available with barrel length 25mm (short barrel) and 38mm (Standard barrel) and barrel angels varies in 130° to 150°. The DCS plate is having angle of 95°
The DHS/DCS Screw is available in total length from 50 to 145 mm, thread length 22mm, shaft diameter 7 mm and outer diameter of 12.5. The thread of DHS/DCS Screw has a buttress type.
The DHS/DCS Compression Screw can be used to achieve fracture compression. Its dimension is available with thread length 36mm and outer diameter 4.0 mm.
{6}------------------------------------------------
### a3) IDENTIFICATION OF THE PREDICATE DEVICE:
For the purposes of US FDA's regulation of medical devices, NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System are substantially equivalent in indications and design principles to the following predicate devices.
| S. No. | Device Name | 510(k)<br>Number | Predicate<br>Manufacturer's<br>Name | Subject Device Name |
|--------|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------|-----------------------------------------------------------------------------------|
| 1. | Synthes LCP Proximal<br>Humerus Plate | K011815 | Synthes (USA) | Filos - Proximal Humerus<br>Safety Lock Plate 3.5 -<br>Standard |
| 2. | 3.5 mm LCP Distal Humerus<br>System | K033995 | Synthes (USA) | Distal Humerus Safety Lock<br>Plate 2.7/3.5, Dorsolateral |
| 3. | 3.5 mm LCP Distal Humerus<br>System | K033995 | Synthes (USA) | Distal Humerus Safety Lock<br>Plate 2.7/3.5, Dorsolateral<br>with Lateral Support |
| 4. | 3.5 mm LCP Distal Humerus<br>System | K033995 | Synthes (USA) | Medial Distal Humerus Safety<br>Lock Plate 2.7/3.5mm |
| 5. | Synthes Small Fragment Dynamic<br>Compression Locking (DCL)<br>System | K000684 | Synthes (USA) | LC-DCP Safety Lock Plate 3.5 |
| 6. | Synthes Small Fragment<br>Dynamic Compression<br>Locking (DCL) System | K000684 | Synthes (USA) | Safety Lock 'T' Plate 3.5,<br>Right Angled |
| 7. | Synthes Small Fragment<br>Dynamic Compression<br>Locking (DCL) System | K000684 | Synthes (USA) | Safety Lock 'T' Plate 3.5,<br>Oblique Angled |
| 8. | Synthes Clavicle Hook Plate | K061753 | Synthes (USA) | Clavicle Hook Safety Lock<br>Plate 3.5 |
| 9. | Synthes Small Fragment Dynamic<br>Compression Locking (DCL)<br>System | K000684 | Synthes (USA) | Reconstruction Safety Lock<br>Plate 3.5 - Straight |
| 10. | 3.5 mm LCP Distal Humerus<br>System | (K033995) | Synthes (USA) | Safety Lock Screw Ø 2.7mm -<br>Self Tapping |
| | | | | |
| 11. | Synthes Small Fragment<br>Dynamic Compression<br>Locking (DCL) System | K000684 | Synthes (USA) | Safety Lock Screw Ø 3.5mm -<br>Self Tapping |
| 12. | Synthes (USA) 3.5 mm and 4.5<br>mm Locking Compression Plate<br>(LCP) System with Expanded<br>Indications | K082807 | Synthes (USA) | Broad LC-DCP Safety Lock<br>Plate 4.5 /5.0 |
| 13. | Synthes (USA) 3.5 mm and 4.5<br>mm Locking Compression Plate<br>(LCP) System with Expanded<br>Indications | K082807 | Synthes (USA) | Narrow LC-DCP Safety Lock<br>Plate 4.5/5.0 |
| 14. | Synthes (USA) LCP Proximal<br>Femur Plate and Screws | K030858 | Synthes (USA) | Proximal Femoral Safety Lock<br>Plate 4.5/5.0/7.3 |
| 15. | Synthes LCP Distal Femur Plates | K062564 | Synthes (USA) | Distal Femoral Safety Lock<br>Plate 4.5/5.0 |
| 16. | Synthes LCP Proximal Tibia Plate<br>Synthes 4.5 mm Titanium LCP<br>Proximal Tibia Plating System | For Stainless<br>Steel Plate<br>(K011978)<br>For<br>Titanium<br>Plate<br>K023802 | Synthes (USA) | Proximal Lateral Tibial Safety<br>Lock Plate 4.5/5.0 |
| 17. | Synthes Small Fragment<br>Dynamic Compression<br>Locking (DCL) System | K000684 | Synthes (USA) | Safety Lock 'T' Plate 4.5/5.0 |
| 18. | Synthes Small Fragment<br>Dynamic Compression<br>Locking (DCL) System | K000684 | Synthes (USA) | L Buttress Safety Lock Plate<br>4.5/5.0 |
| 19. | Synthes Small Fragment<br>Dynamic Compression<br>Locking (DCL) System | K000684 | Synthes (USA) | T Buttress Safety Lock Plate<br>4.5/5.0 |
| 20. | Synthes 4.5 mm LCP<br>Straight Reconstruction<br>Plates | K051986 | Synthes (USA) | Reconstruction Safety Lock<br>Plate 4.5/5.0 |
| 21. | Synthes Large Fragment Dynamic<br>Compression Locking (DCL)<br>System | K000682 | Synthes (USA) | Safety Lock Screw Ø 5.0mm |
| | | | | |
| 22. | Synthes Limited Contact-<br>Dynamic Hip Screw Implant (LC-<br>DHS)" for K923613 | K 923613 | Synthes (USA) | Dynamic Hip Compression<br>Plate (with Dynamic<br>Compression Holes) |
| | Synthes Titanium Limited<br>Contact-Dynamic Hip Screw<br>Implant (Ti LC-DHS)" for<br>K953607 | K953607 | | |
| | Synthes Limited Contact-<br>Dynamic Hip Screw Implant (LC-<br>DHS)" for K923613 | K 923613 | | Dynamic Hip Compression |
| 23. | Synthes Titanium Limited<br>Contact-Dynamic Hip Screw<br>Implant (Ti LC-DHS)" for<br>K953607 | K953607 | Synthes (USA) | Plate - Short Barrel (with<br>Dynamic Compression Holes) |
| 24. | DYNAMIC CONDYLAR<br>SCREW OR D.C.S | K840954 | Synthes (USA) | DCS Plate 95° with Dynamic<br>Compression Holes |
| | Synthes Dynamic Hip Plate | K 923613<br>K953607 | Synthes (USA) | Standard Lag Screw, Ø<br>12.5mm |
| 25. | Synthes Limited Contact-<br>Dynamic Hip Screw Implant (LC-<br>DHS)" for K923613 | K 923613 | | |
| | Synthes Titanium Limited<br>Contact-Dynamic Hip Screw<br>Implant (Ti LC-DHS)" for<br>K953607 | K953607 | Synthes (USA) | Compression Screw |
| 26. | Synthes Cortical Screws | K112583 | Synthes (USA) | Cortex Screw Ø 3.5mm, Self<br>Tapping |
| 27. | Synthes Cortical Screws | K112583 | Synthes (USA) | Cortex Screw Ø 4.5mm, Self<br>Tapping |
{7}------------------------------------------------
{8}------------------------------------------------
## These implants are sold non-sterile, the products have to be sterilized prior to use.
#### A5). (5) A statement of the intended use of the device
#### Indications for Use:
NET Brand Small Fragment and Large Fragment Osteosynthesis System is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, distal femur, proximal tibial pilon, fibula, pelvis and acetabulum fractures; periprosthetic fractures; The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine.
{9}------------------------------------------------
NET Brand DHS/DCS plating system may be used for fixation of the fractures of proximal femur such as femoral neck, trochanteric, pertrochanteric or intertrochanteric zones.
The system is indicated for use in adult patients only. All implants are for single use only.
### a6). Summary of Technological Characteristics as compared to the predicate devices:
SUBSTANTIAL EQUIVALENCE INCLUDING COMPARISON WITH PREDICATE DEVICES
The subject device and the predicate devices have the same intended use and have the same technological characteristics. The subject and predicate devices are all fabricated from the same or similar materials and share similar design characteristics.
The subject and predicate devices encompass the same range of physical dimensions, are packaged using the same materials, and are to be sterilized by the same methods. Any differences in the technological characteristics do not raise new issues of safety or efficacy. Performance data provided to demonstrate substantial equivalence included engineering analysis and mechanical testing according to ASIM F382, ASTM F 384 and ASTM F543.
Overall, Small Fragment and Large Fragment Osteosynthesis System of Narang Medical Limited have the following similarities to the predicate devices:
- * has the same intended use,
- * uses the same operating principle,
- * incorporates the same basic design,
- * incorporates the same or very similar materials, and
* has similar packaging and can be sterilized using the same materials and processes.
#### Following is the summary of parameters in which the comparison has been verified:
| S. No. | Characteristics | Predicate Device Versus New<br>Device (Auxein Brand) | Remarks |
|--------|--------------------------|--------------------------------------------------------------------------------------|------------|
| 01 | Indications for use | Similar intended use in New Device<br>and Predicate device | Equivalent |
| 02 | Material | Same material used in New Device<br>and Predicate device | Equivalent |
| 03 | Performance<br>Standards | Same performance standards used in<br>both New Device as well as predicate<br>device | Equivalent |
{10}------------------------------------------------
| 04 | Sterilization | Same method of sterilization used in<br>both New Device as well as<br>Predicate device | Equivalent |
|----|-----------------------------|----------------------------------------------------------------------------------------|------------|
| 05 | Dimensional<br>Verification | Same dimensions found in both New<br>Device as well as Predicate device | Equivalent |
#### b1). Discussion on the non-clinical testing performed
Following are the applicable product standards considered for non-clinical standards
- A: Material Standards
- B: Performance Standards
A: Material Standards: The raw material standards are the first standards to be complied, as it ensures compliance to the materials to be used for manufacturing of metallic surgical implants.
following material standards have been adopted and complied:
ASTM F 136: Standard specification for wrought Titanium-6Aluminium-1. 4Vanadium ELI (Extra low interstitial) Alloy for surgical implant applications.
ட Standard Specification for Wrought 2. ASTM 138: 18 chromium-14Nickel-2.5Molybdenum stainless steel bar and wire for surgical implants.
ட 3. ASTM 139: Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Sheet and Strip for Surgical Implants
#### B: Performance Standards:
The performance of NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System has been verified as per the following standards
- ASTM F 382,
- ASTM F 384 and
- ASTM F 543
- For Bone Plates:
As per ASTM F 382 and ASTM F 384 Static Four Point Bend Test: Conforms, Dynamic Four Point Bend Test: Conforms
- For Bone . Screws:
SECTION 05
{11}------------------------------------------------
- As per ASTM F 543 :Torsional Properties: Conforms, Driving Torque : . Conforms, Pull-out Test: Conforms
#### b2). Discussion on the clinical evaluation referenced and relied upon:
NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System,
NET Brand of DHS/DCS Plating System are of similar design and pattern as well as similar indications for use. Therefore Clinical information was not necessary to demonstrate substantial equivalence.
## CONCLUSION:
#### General, Safety and Performance conclusion:
| S. No. | Parameter of<br>Conclusion | Proposed Device | Predicate<br>Device |
|--------|----------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------|
| 01 | Product Code | For Bone Plates: HRS<br>For Bone Screws: HWC | Same |
| 02 | Regulation<br>Number | For Bone Plates: 21CFR 888.3030<br>For Bone Screws: 21CFR 888.3040 | Same |
| 03 | Regulatory Class | Class II | Class II |
| 04 | Indications For<br>use | Same Indications For Use | Similar |
| 05 | Sterilization | Provided Non-Sterile and to be sterilized using<br>Autoclaving Method to achieve SAL of 10-6 AAMI<br>ST79, ISO 17665-1 | Similar |
{12}------------------------------------------------
| 06 | Mechanical Test<br>Performance | For Bone Plates: | Same |
|----|--------------------------------|-----------------------------------------|------|
| | | As per ASTM F 382 and ASTM F 384 | |
| | | - Static Four Point Bend Test Conforms | |
| | | - Dynamic Four Point Bend Test Conforms | |
| | | For Bone Screws: | |
| | | As per ASTM F 543 | |
| | | - Torsional Properties Conforms | |
| | | - Insertion Torque Conforms | |
| | | - Removal Torque Conforms | |
| | | - Axial Pull-out Test Conforms | |
| 07 | Material<br>Standards | ASTM F 136, ASTM F 138 and ASTM F 139 | Same |
General, Safety and Performance Conclusion:
From the available data available we can justify that the NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System are as safe, as effective and perform as same indications for use as that of already marketed predicate devices identified in a3. Of 510(k) summary.
Therefore, our devices can be considered safe and effective for their intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.