Reprocessed LigaSure ImpactTM Curved, Large Jaw, Open Sealer/Divider
K150538 · Stryker Sustainability Solutions · NUJ · Jun 25, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150538
Device Name
Reprocessed LigaSure ImpactTM Curved, Large Jaw, Open Sealer/Divider
Applicant
Stryker Sustainability Solutions
Product Code
NUJ · General, Plastic Surgery
Decision Date
Jun 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider LF4318 is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels is desired. The Reprocessed LigaSure Impact™ LF4318 is intended to be used with the ForceTriad™ Energy Platform to cut and seal vessels, and to cut, grasp, and dissect tissue during surgery. The indications for use include open procedures (general, urologic, vascular, thoracic, and gynecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy, etc. The Reprocessed LigaSure Impact™ LF4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7 mm and tissue bundles. The LigaSure™ system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure™ system for these procedures.
Device Story
Hand-held bipolar electrosurgical instrument; used in open surgery; operated by surgeons. Connects to ForceTriad™ Energy Platform via cable. Input: surgeon-actuated jaw closure, activation button/footswitch, and cutter trigger. Device delivers precise energy to tissue for controlled time to achieve permanent fusion; blade divides tissue. Output: sealed and divided vessels/tissue bundles. Benefits: controlled hemostasis and tissue division. Reprocessed device maintains original mechanical design and materials.
Clinical Evidence
Bench and laboratory testing only. Included biocompatibility, sterilization validation, electrical safety (IEC 60601-1, IEC 60601-2-2), electromagnetic compatibility (IEC 60601-1-2), and functional performance (burst pressure, thermal spread, jaw tip temperature, cooling time, reliability). Acute and chronic pre-clinical testing evaluated thermal spread and hemostasis efficacy.
Technological Characteristics
Bipolar electrosurgical instrument; 13.5 mm square shaft; 18 cm length; 36 mm jaw length. Energy source: ForceTriad™ Energy Platform. Manual actuation for jaw closure, rotation, and cutting. Connectivity: proprietary cable connector for generator identification. Sterilization: validated reprocessing cycle.
Indications for Use
Indicated for patients undergoing open surgical procedures (general, urologic, vascular, thoracic, gynecological) requiring ligation and division of vessels (arteries, veins, pulmonary vasculature, lymph) up to 7 mm and tissue bundles. Contraindicated for tubal coagulation or sterilization procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
LigaSure Impact Curved, Large Jaw, Open Sealer/Divider (K123444)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a way that suggests unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 25, 2015
Stryker Sustainability Solutions Mr. Scott English Staff Regulatory Affairs Specialist 1810 West Drake Drive Tempe, Arizona 85283
Re: K150538
Trade/Device Name: Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Dividers Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NUJ Dated: May 20, 2015 Received: May 21, 2015
Dear Mr. English:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Reprocessed Single Use Device Models Included in Clearance:
{2}------------------------------------------------
| Device Model | Device Name | Original<br>Manufacturer |
|--------------|-------------------------------------------------------------|--------------------------|
| LF4318 | LigaSure ImpactTM Curved, Large Jaw, Open<br>Sealer/Divider | Covidien |
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150538
### Device Name
Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Dividers
### Indications for Use (Describe)
The Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider LF4318 is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation of vessels is desired. The Reprocessed LigaSure Impact™ LF4318 is intended to be used with the ForceTriad™ Energy Platform to cut and seal vessels, and to cut, grasp, and dissect tissue during surgery.
The indications for use include open procedures (general, urologic, vascular, thoracic, and gynecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectomies, Nissen fundoplication, colectorny, adhesiolysis, oophorectomy, etc. The Reprocessed LigaSure Impact™ LF4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7 mm and tissue bundles.
The LigaSure™ system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure™ system for these procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# SECTION 5: 510(k) SUMMARY
### Submitter:
Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, Arizona 85283
### Contact:
Scott English Staff Regulatory Affairs Specialist 480-763-5333 (o) 480-763-5310 (f) scott.english@stryker.com
#### Date of Preparation: March 2, 2015
### Name of Device:
| Trade/Proprietary Name: | Reprocessed LigaSure Impact™ Curved, Large Jaw, Open<br>Sealer/Divider |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Bipolar Electrosurgical Open and Laparoscopic Instruments |
| Classification Name: | Electrosurgical, Cutting & Coagulation Accessories,<br>Laparoscopic & Endoscopic, Reprocessed<br>(21 CFR§878.4400, Product Code NUJ, Class II) |
# Predicate Devices:
| Model Number | 510(k) Number | 510(k) Title | Original Manufacturer |
|--------------|---------------|------------------------------------------------------------|-----------------------|
| LF4318 | K123444 | LigaSure Impact Curved, Large Jaw, Open Sealer/<br>Divider | Covidien |
# Device Description:
The Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer / Divider is a hand-held bipolar electrosurgical instrument designed exclusively for use with the ForceTriad™ Energy Platform to seal and divide vessels (including pulmonary) up to and including 7 mm in diameter, tissue bundles, and lymphatics during open general surgical procedures. The ForceTriad's tissue-fusion (LigaSure™) mode is designed to deliver precise energy to tissue for a controlled time period to achieve complete and permanent tissue fusion. A blade within the instrument is surgeon actuated to divide tissue.
The instrument has a shaft diameter of 13.5 mm (square), shaft length of 18 cm, and jaw length of 36 mm. The following controls are located on the instrument handle:
{5}------------------------------------------------
- A lever for opening and closing the instrument jaws. The mechanism ● incorporates a latch to hold the jaws in the closed position during vessel sealing and cutting.
- An activation button for generator power to initiate vessel sealing. .
- . A trigger for actuating the cutter.
- A knob to rotate the instrument jaws. .
All controls can be operated with either the right or left hand. Vessel sealing can be initiated using the activation button or utilizing a footswitch connected to the generator. The instrument attaches to the ForceTriad™ Energy Platform via a ten-foot cord with a LiqaSure™ cable connector that identifies the instrument type to the generator.
The instrument is compatible with the Covidien™ ForceTriad™ Energy Platform running software version 3.50 or greater.
The scope of this submission only includes the reprocessed Covidien™ sealer/divider device and not the ForceTriad™ Energy Platform that is used to power the device or the footswitch that connects to the generator. Stryker Sustainability Solutions does not reprocess or market the generator or footswitch.
### Intended Use:
The Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider LF4318 is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels is desired. The Reprocessed LigaSure Impact™ LF4318 is intended to be used with the ForceTriad™ Energy Platform to cut and seal vessels, and to cut, grasp, and dissect tissue during surgery.
The indications for use include open procedures (general, urologic, vascular, thoracic, and avnecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy, etc. The Reprocessed LigaSure Impact™ LF4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7 mm and tissue bundles.
The LigaSure™ system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure™ system for these procedures.
# Summary of Technological Characteristics:
The design, materials, and intended use of Reprocessed LigaSure Impact™ Curved, Large Jaw. Open Sealer/Divider is equivalent to the predicate device. The mechanism of action of the reprocessed device is identical to the predicate device in that the same standard mechanical design, materials, and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Stryker Sustainability Solutions' reprocessing of LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider includes removal of adherent visible soil and decontamination. Each individual device is tested for appropriate function of its components prior to packaging and labeling operations.
{6}------------------------------------------------
### Performance Data:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider. This included the following tests:
- Biocompatibility
- . Validation of Reprocessing
- . Sterilization Validation
- Functional Performance Tests .
- . Electrical Safety Testing
- Electromagnetic Compatibility Testing ●
- Packaging Validation ●
The functional performance testing involved electrical safety and electromagnetic compatibility testing in accordance with IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-2, and verification/comparative testing (to the predicate device). The bench testing involved evaluation of the device's performance and ability to seal and divide vessels up to 7mm, including: burst pressure, thermal spread, max jaw tip temperature, jaw cooling time, device functionality, and device reliability testing.
Acute and chronic pre-clinical testing was conducted to evaluate thermal spread and the ability to achieve hemostasis of vessels and tissues.
The performance testing demonstrates that reprocessed devices are as safe and effective as the predicate and operate as originally intended.
### Conclusion:
Stryker Sustainability Solutions concludes that the Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider is at least as safe and effective to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.