LIGASURE IMPACT CURVED, LARGE JAW, OPEN SEALER / DIVIDER
K123444 · Covidien, Formerly Valleylab, A Division of Tyco H · GEI · Jan 30, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123444
Device Name
LIGASURE IMPACT CURVED, LARGE JAW, OPEN SEALER / DIVIDER
Applicant
Covidien, Formerly Valleylab, A Division of Tyco H
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels is desired. The LigaSure Impact LF4318 is intended to be used with the ForceTriad™ energy platform to cut and seal vessels, and to cut, grasp and dissect tissue during surgery. The indications for use include open procedures (general, urologic, vascular, thoracic, and gynecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectornies, Nissen fundoplication, colectorny, adhesiolysis, oophorectorny, etc. The LigaSure Impact LF4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7mm and tissue bundles. The LigaSure™ system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Device Story
Sterile, single-use, hand-held bipolar electrosurgical instrument; used with ForceTriad Energy Platform. Features long, curved jaw and pistol-grip handle with integrated hand switch and cutter. Operates by delivering precise energy to tissue for controlled time to achieve permanent fusion. Used in open surgical procedures by surgeons for vessel ligation, division, grasping, and dissection. Connects to generator via 10-foot cord; generator uses unique product identifier for device recognition. Benefits include improved access/visibility to critical structures and ergonomic handling for left/right-handed users.
Clinical Evidence
Bench testing and pre-clinical testing performed. Bench testing evaluated performance and ability to seal/divide vessels up to 7mm. Pre-clinical testing (acute and chronic) evaluated thermal spread and hemostasis in simulated clinical settings. No human clinical data provided.
Technological Characteristics
Bipolar electrosurgical instrument; sterile, single-use. Features curved jaw, shaft-based mechanism, and pistol-grip handle. Connects to ForceTriad energy platform. Complies with IEC 60601-1 and IEC 60601-2-2 safety standards.
Indications for Use
Indicated for open surgical procedures (general, urologic, vascular, thoracic, gynecological) requiring ligation and division of vessels (arteries, veins, pulmonary vasculature, lymph) up to 7mm and tissue bundles. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
K1 23 444
Image /page/0/Picture/1 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black square with a white rectangle in the middle and two smaller white rectangles above and below it. The word is written in a simple, sans-serif font and is the main focus of the image.
:7
## 510(k) Summary
Date summary prepared: November 7, 2012
## 510(k) Submitter/Holder
Covidien 5920 Longbow Drive Boulder, CO 80301
#### Contact
Dawn D. Tindall Sr. Associate, Regulatory Affairs Telephone: 303-581-7034 Fax: 303-530-6313 Email: dawn.tindall@covidien.com
## Name of Device
Trade Name: Catalog Number: Common Name: Classification Name: LigaSure Impact™ Curved, Large Jaw, Open Sealer / Divider LF4318 Bipolar electrosurgical instrument Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, class II, GEI).
## Predicate Devices
The LigaSure Impact™ Curved, Large Jaw, Open Sealer / Divider was compared and found to be substantially equivalent to the following products of comparable type in commercial distribution:
Device Common Name: Bipolar electrosurgical instrument Trade Name: Enseal G2 Tissue Sealer Catalog Number: NSEALX22L 510(k) Number: K112033 Manufacturer: Ethicon Endo-Surgery, LLC
Device Common Name: Bipolar electrosurgical instrument Trade Name: LigaSure Atlas, Tissue Fusion Open Instrument Catalog Number: LS1020 510(k) Number: K010013 Manufacturer: Covidien
Device Common Name: ForceTriad Electrosurgical Generator; LigaSure Instruments Trade Name: ForceTriad generator 510(k) Number: K070162 Manufacturer: Covidien
#### Device Description
The LigaSure Impact™, Curved, Large Jaw, Open Sealer/Divider (LF4318) is a sterile, single-use, hand-held bipolar electrosurgical instrument designed exclusively for use with the ForceTriad™ Energy Platform (cleared under K051644, K070162) to seal and divide vessels (including pulmonary) up to and including 7mm in diameter, tissue bundles, and lymphatics during open general surgical procedures. The ForceTriad's tissue-fusion
JAN 3 0 2013
22
510(k): LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider - LF4318
{1}------------------------------------------------
1 2 3 4 4
Page 2 of 2
(LigaSure) mode is designed to deliver precise energy to tissue for a controlled time period to achieve complete and permanent tissue fusion.
The device combines the benefits of a long, curved jaw and a shaft-based jaw mechanism for improved access and visibility to critical structures. The pistol grip style handle was designed for improved comfort and usability, with an integrated hand switch and cutter for the subsequent sealing and dividing of tissue. Hand controls have been symmetrically placed to facilitate handling by both left and right-handed users. The LF4318 device connects to the ForceTriad energy platform with a 10 foot cord containing a LigaSure cable connector. This connector functions as a unique product identifier for device-specific recognition by the generator.
#### Intended Use
The LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels is desired. The LigaSure Impact LF4318 is intended to be used with the ForceTriad energy platform to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.
The indications for use include open procedures (general, urologic, vascular, thoracic, and gynecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectornies, Nissen fundoplication, colectorny, adhesiolysis, oophorectorny, etc. The LigaSure Impact LF4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7mm and tissue bundles.
The LigaSure™ system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
#### Performance
The testing of the LF4318 involved extensive functional, bench and pre-clinical testing. The functional testing involved safety and performance testing in accordance with IEC 60601-1 and IEC 60601-2-2 and verification/comparative testing (to the predicate devices). The bench testing involved evaluation of the devices performance and ability to seal and divide vessels up to 7mm. The pre-clinical testing involved both acute and chronic testing evaluating the thermal spread and ability to achieve hemostasis of tissue and vessels in a variety of simulated clinical setting confirmed that the LF4318 is as safe and effective as the predicate devices and operated as intended.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is in black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Covidien, Formerly ValleyLab, A Division of Tyco Healthcare % Ms. Dawn Tindall Senior Associate, Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301
January 30, 2013
Re: K123444
Trade/Device Name: LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider (LF4318) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: November 07, 2012 Received: November 08, 2012
Dear Ms. Tindall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Ms. Dawn Tindall
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K123444
#### Device Name: LigaSure Impact™ Curved, Large Jaw, Open Sealer / Divider (LF4318)
Indications for Use:
The LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider is a dedicated bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels is desired. The LigaSure Impact LF4318 is intended to be used with the ForceTriad™ energy platform to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.
The indications for use include open procedures (general, urologic, vascular, thoracic, and gynecological) where ligation and division of vessels is performed. These procedures include vaginal hysterectomies. Nissen fundoplication, colectorny, adhesiolysis, oophorectorny, etc. The LigaSure Impact LF4318 can be used on vessels (arteries, veins, pulmonary vasculature, and lymph) up to and including 7mm and tissue bundles.
The LigaSure™ system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Long H. Chen
(Division Sign-Off) Division of Surgical Devices 510(k) Number K123444
20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.