NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
Applicant
Visioneering Technologies, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Apr 30, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
NaturalVue (etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear.
Device Story
NaturalVue contact lenses are daily disposable, visibility-tinted, ionic hydrogel lenses. Manufactured via cast molding using etafilcon A (copolymer of 2-HEMA and MAA) cross-linked with EGDMA and TMPTMA. Lenses incorporate a benzotriazole UV-absorbing monomer to block UV radiation. Supplied sterile in blister packs with isotonic borate buffered saline. Intended for daily wear by patients to correct refractive errors (myopia, hyperopia, astigmatism, presbyopia). Prescribed by eye care professionals. Lenses provide vision correction and UV protection. Single-use design eliminates need for cleaning/disinfection.
Clinical Evidence
Bench testing only. No new clinical studies were conducted. Safety and effectiveness were established by demonstrating equivalence to the predicate device (K140025) and leveraging historical clinical use data for etafilcon A lenses.
Indicated for daily wear for correction of ametropia (myopia, hyperopia, astigmatism) and presbyopia in aphakic and non-aphakic persons with non-diseased eyes. Powers range from -20.00 to +20.00 diopters; astigmatic correction up to -10.00 diopters; reading addition up to +3.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses (K140025)
Aquamax (etafilcon A) Disposable Soft (Hydrophilic) Contact Lenses (K120028)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 30, 2015
Visioneering Technologies, Inc. c/o Ms. Penny Northcutt, RAC, FRAPS, CQA Regulatory Consultant and Official Correspondent 174 Watercolor Way Suite 103-403 Santa Rosa Beach, Florida 32459
Re: K150385
Trade/Device Name: Naturalvue (etafilcon A) Daily Disposable Soft (hydrophilic) Contact Lenses NaturalVue (etafilcon A) SPHERE and ASPHERE NaturalVue (etafilcon A) Toric NaturalVue (etafilcon A) Multifocal NaturalVue (etafilcon A) Multifocal Toric
Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: March 9, 2015 Received: March 10, 2015
Dear Ms. Northcutt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150385
#### Device Name
NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
# Indications for Use (Describe)
#### Spherical and Aspheric
NaturalVue (etafilcon A) Sphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
#### Toric
NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
#### Multifocal Spherical
NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
#### Multifocal Toric
NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.
NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
The lenses are intended for single-use disposable wear.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510k Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| DATE | February 13, 2015 |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | Visioneering Technologies, Inc.<br>Stephen Snowdy, CEO<br>4555 Mansell Road<br>Suite 300<br>Alpharetta, GA 30022<br>770-521-4358 (office phone)<br>404-863-8892 (cell phone)<br>ssnowdy@vtivision.com (email) |
| OFFICIAL<br>CORRESPONDENT | Penny Northcutt, RAC, FRAPS, CQA<br>174 Watercolor Way, Suite 103-403<br>Santa Rosa Beach, FL 32459<br>678-428-6978 (phone)<br>866-630-4082 (fax)<br>pennynorthcutt@theregsolutions.com (email) |
| TRADE NAME | NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic)<br>Contact Lenses |
| COMMON NAME | Soft (hydrophilic) Contact Lenses (daily wear)<br>Soft (hydrophilic) Contact Lenses (disposable) |
| DEVICE<br>CLASSIFICATION<br>AND PRODUCT<br>CODE | Class II<br>Lenses, Soft Contact, Daily Wear, per 21 CFR 886.5925,<br>Product Code: LPL |
| | Class II<br>Lens, Contact, (Disposable), per 21 CFR 886.5925, Product<br>Code: MVN |
| PREDICATE<br>DEVICE | K140025 Pegavision Corporation, NaturalVue (etafilcon A)<br>Daily Disposable Soft (Hydrophilic) Contact Lenses |
# SUBSTANTIAL EQUIVALENCE:
It is Visioneering Technologies, Inc.'s conviction that the data submitted in this 510(k) provides evidence of substantial equivalence for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses with the same established safety profile and effectiveness of the predicate device K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses.
{5}------------------------------------------------
# DESCRIPTION OF THE DEVICE:
The NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are visibility-tinted, containing a UV blocker and are available in a spherical, aspheric, toric, multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380mm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution.
# INTENDED USE/INDICATIONS FOR USE:
# NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
## Spherical and Aspheric
NaturalVue (etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
## Toric
NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
## Multifocal
NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eves in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
## Multifocal Toric
NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.
NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
The lenses are intended for single-use disposable wear.
{6}------------------------------------------------
# TECNOLOGICAL CHARACTERISTICS:
# Historical Clinical Use
etafilcon A lenses have been used widely. Their safety and effectiveness have been well established and documented. Their safety and effectiveness can be further exemplified by the following lenses, as well as many others as discussed in the 510(k), cleared by FDA:
- ACUVUE (etafilcon A) Contact lens, clear and visibility tint with UV blocker, K962804 ● submitted by Vistakon USA
- ACUVUE (etafilcon A) contact lens, clear and visibility tint with UV blocker, K991134 ● submitted by Vistakon USA
- Aquamax (etafilcon A) Disposable Soft (Hydrophilic) Contact Lenses, K120028 . submitted by Pegavision Corporation, Taiwan
- NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, . K140025 submitted by Pegavision Corporation, Taiwan
NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses from the previous FDA cleared K140025 did not require clinical studies as the USAN name and manufacturing processes are the same as the above-mentioned predicate (and reference) devices.
# Non-Clinical Study
All tests were conducted in accordance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses.
The non-clinical performance tests have been performed to demonstrate the safety and effectiveness of NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and establish substantial equivalence to the predicate NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses (K140025). The evidence of substantial equivalence to the predicate lenses is described below.
# a) Material Properties Testing
The material properties of NaturalVue (etafilcon A) Soft (Hydrophilic) Contact Lenses were tested as illustrated in the following table.
| | Proposed Device | K140025 Predicate |
|------------------------------------------------|------------------------------------------|------------------------------------------|
| Production Method | Cast-Molded | Cast-Molded |
| USAN Name | etafilcon A | etafilcon A |
| Material Classification | Group 4 high water ionic | Group 4 high water ionic |
| Water Content (%) | 58% | 58% |
| Refractive Index | 1.4023 | 1.4023 |
| Oxygen Permeability<br>(edge corrected) @ 35°C | 19.3 x 10-11<br>(cm²/sec)(ml O2/ml-mmHg) | 19.3 x 10-11<br>(cm²/sec)(ml O2/ml-mmHg) |
| Percent Transmittance | | |
| % T at 593nm | > 95% | > 95% |
| % T at 380-315nm | < 30% | < 30% |
| % T at 315-280nm | < 5% | < 5% |
# b) Biocompatibility
The standard cytotoxicity, maximization, and ocular irritation tests were carried out for the predicate K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and therefore apply to NaturalVue (etafilcon A) Daily Disposable Soft
{7}------------------------------------------------
(Hydrophilic) Contact Lenses. Negative responses were recorded for all tests. The validity of blister packaging for these lenses was demonstrated by passing the standard extraction tests.
# c) Microbiologv
The steam sterilization process has been validated to deliver a minimum SAL of 10-9, thereby complying with the requirement of FDA Group 4. There is shelf-life stability supporting that these lenses remain sterile through the expiration date claimed for the product.
# d) Bacteriostatic Validation
The steam sterilizer was tested for effectiveness by measuring and demonstrating the uniformity of temperature at different locations inside the sterilizer over test period. Tested microorganisms were killed under tested conditions as compared to control.
Lenses remained sterilized and there was no microbial growth for a period of 4 years tested under accelerated conditions. Seal of lens packages remained tight for a period of 4 years as demonstrated by the constant peeling strength tested under accelerated conditions.
## e) Leachability
Studies were conducted to determine the leachable materials from the finished lenses. The results show that, at the levels of the detection reported, there are no leachable monomers and additive residues.
# CONCLUSION:
Based on the performance and manufacturing verification studies, it can be concluded that the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are equivalent to the predicate device K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses with respect to intended use, principles of operation, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.