AQUAMAX (ETAFILCON A) BI-WEEKLY DISPOSABLE SOFT (HYDROPHILIC) CONTACT LENSES

K140025 · Pegavision Corporation · LPL · Oct 15, 2014 · Ophthalmic

Device Facts

Record IDK140025
Device NameAQUAMAX (ETAFILCON A) BI-WEEKLY DISPOSABLE SOFT (HYDROPHILIC) CONTACT LENSES
ApplicantPegavision Corporation
Product CodeLPL · Ophthalmic
Decision DateOct 15, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

Aquamax (Etafilcon A) SPHERE and ASPHERE Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Multifocal Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear. Aquamax (Etafilcon A) SPHERE and ASPHERE Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. Aquamax (Etafilcon A) Multifocal Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Multifocal Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfection and scheduled replacement as prescribed by the eye care professional. NaturalVue (Etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. NaturalVue (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. NaturalVue (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. NaturalVue (Etafilcon A) Multifocal Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear.

Device Story

Ionic hydrogel contact lenses (Etafilcon A) designed for vision correction; available in spherical, aspheric, toric, and multifocal designs. Manufactured via cast molding; tinted with C.I. Reactive Blue 19 for handling; contain benzotriazole UV-absorbing monomer. Supplied sterile in blister packs with isotonic borate buffered saline. Prescribed by eye care professionals for daily wear; frequent replacement lenses require chemical disinfection. Lenses correct refractive errors (myopia, hyperopia, astigmatism, presbyopia) and provide UV protection. Patient benefits include improved visual acuity and convenience of disposable/frequent replacement modalities.

Clinical Evidence

Bench testing only. No clinical studies were required as the USAN name (Etafilcon A) and manufacturing processes are identical to previously cleared predicate devices. Non-clinical performance testing included material properties, biocompatibility (cytotoxicity, maximization, ocular irritation), sterilization validation, and leachability studies.

Technological Characteristics

Material: Etafilcon A (42% polymer, 58% water). Composition: 2-HEMA and MAA copolymer, cross-linked with EGDMA and TMPTMA. Manufacturing: Cast-molded. UV protection: Benzotriazole monomer. Sterilization: Steam. Connectivity: None. Form factor: Spherical, aspheric, toric, multifocal contact lenses.

Indications for Use

Indicated for daily wear correction of ametropia (myopia/hyperopia) and presbyopia in aphakic/non-aphakic persons with non-diseased eyes. Includes options for astigmatism correction up to 10.00D. Available as daily disposable or frequent replacement (2 weeks).

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that forms the hair and neck of each figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 15, 2014 Pegavision Corporation Mr. Tony Hsu, President 2F-1 No. 5, Shing Yeh St. Shan Ding Vil Kwei Shan Hsiang, Taoyuan Hsien 333, Taiwan Re: 510(k) Number: K140025 Trade/Device Name: Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lenses (daily wear) Regulatory Class: Class II Product Code: LPL, MVN Dated: September 15, 2014 Received: September 16, 2014 Dear Mr. Hsu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ Page 2 – Mr. Tony Hsu or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Image /page/1/Picture/8 description: The image shows the text "Deborah L. Falls -S" in a simple, sans-serif font. The text is arranged horizontally, with the name "Deborah L. Falls" appearing prominently, followed by "-S". The text is black against a white background. for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K140025 #### Device Name Aquamax and NaturalVue (Etafilcon A) Soft (Hydrophilic) Contact Lenses #### Indications for Use (Describe) 1. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses #### Spherical and Aspheric Aquamax (Etafilcon A) SPHERE and ASPHERE Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Toric Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. #### Multifocal Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Multifocal Toric Aquamax (Etafilcon A) Multifocal Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear. 2. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement (Replacement Recommended Every 2 Weeks): Spherical and Aspheric Aquamax (Etafilcon A) SPHERE and ASPHERE Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. {3}------------------------------------------------ #### Toric Aquamax (Etafilcon A) Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. # Multifocal Aquamax (Etafilcon A) Multifocal Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. ## Multifocal Toric Aquamax (Etafilcon A) Multifocal Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfection and scheduled replacement as prescribed by the eye care professional. Pegavision recommends up to two week replacement for Etafilcon A lenses, or replacement as recommended by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only. #### 3. NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses Spherical and Aspheric NaturalVue (Etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. ## Toric NaturalVue (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. ## Multifocal NaturalVue (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. ## Multifocal Toric NaturalVue (Etafilcon A) Multifocal Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are {4}------------------------------------------------ indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/5/Picture/1 description: The image contains the text "K140025" at the top. Below that is the logo for PEGA VISION. Underneath the logo is the text "Brighten Your View." # 510k Summary This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | DATE: | September 15, 2014 | |------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Pegavision Corporation<br>Tony Hsu, President<br>2521 Beechwood Drive<br>2F-1 No. 5, Shing Yeh St. Shan Ding Vil Kwei Shan<br>Hsiang, Taoyuan Hsien 333, Taiwan<br>886-3-329-8808 (phone)<br>886-3-329-8897 (fax)<br>tonyhsu@pegavision.com (email) | | OFFICIAL<br>CORRESPONDENT<br>FOR THIS<br>SUBMISSION: | Penny Northcutt, RAC, FRAPS, CQA<br>174 Watercolor Way, Suite 103-403<br>Santa Rosa Beach, FL 32459<br>678-428-6978 (phone)<br>678-513-0937 (fax)<br>pennynorthcutt@theregsolutions.com (email) | | TRADE NAME: | Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent<br>Replacement | | | Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic)<br>Contact Lenses | | | NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic)<br>Contact Lenses | | COMMON NAME: | Soft (hydrophilic) Contact Lenses (daily wear) | | DEVICE<br>CLASSIFICATION and<br>Product Code: | Class II<br>Lenses, Soft Contact, Daily Wear, per 21 CFR 886.5925,<br>Product Code: LPL | | | Class II<br>Lens, Contact, (Disposable), per CFR 886.5925,<br>Product Code: MVN | | PREDICATE<br>DEVICE NAME: | K120228 Pegavision Corporation, Aquamax (Etafilcon A) disposable<br>soft (hydrophilic) contact lenses | | | K013445 Florida Optical Engineering, OSPREY spherical<br>and Multifocal Toric soft daily wear contact lenses | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Pegavision, a company that specializes in vision-related products. The logo features the company name in a stylized font, with the tagline "Brighten Your View" underneath. The image also includes the number K140025, which may be a product or reference number. K962804 Vistakon, Johnson & Johnson Vision Products, ACUVUE (Etafilcon A) Spherical, Toric, and Multifocal, and Toric Multifocal soft (hydrophilic) daily wear contact lenses #### SUBSTANTIAL EQUIVALENCE: It is Pegavision Corporation's conviction that the data submitted in this 510(k) provides evidence of substantial equivalence for the Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses with the same established safety profile and effectiveness of the predicate devices K120028 Aquamax (Etafilcon A) Disposable and Bi-weekly Soft (Hydrophilic) Contact Lenses for spheric lenses for myopia and hyperopia, K013445 Florida Optical Engineering OSPREY spherical multifocal and multifocal toric soft daily wear contact lenses, and K962804 Vistakon, Johnson & Johnson Vision Products ACUVUE (Etafilcon A) spherical, toric, and multifocal soft (hydrophilic) daily wear contact lenses. ## DESCRIPTION OF THE DEVICE: The Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and Natural/ue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are clear or visibility-tinted, containing a UV blocker and are available in a spherical, aspherical multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from Etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% Etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380nm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution. ## INTENDED USE/INDICATIONS FOR USE: ## 1. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses #### Spherical and Aspheric Aquamax (Etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn.by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. ## Toric Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. Traditional 510(k) Pegavision Corporation {7}------------------------------------------------ #### Multifocal Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses mav be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Multifocal Toric Aquamax (Etafilcon A) Multifocal Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses mav be worn by persons who exhibit astigmatism of 10.00 diopters or less. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear. #### 2. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement (Replacement Recommended Every 2 Weeks): #### Spherical and Aspheric Aquamax (Etafilcon A) SPHERE and ASPHERE Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Toric Aquamax (Etafilcon A) Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. #### Multifocal Aquamax (Etafilcon A) Multifocal Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Multifocal Toric Aquamax (Etafilcon A) Multifocal Toric Soft (Hydrophilic) Contact Lenses for Frequent Replacement are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased Traditional 510(k) Pegavision Corporation {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Pegavision, with the text "PEGA" in blue and "VISION" in gray. Below the logo is the text "Brighten Your View" in a smaller font size. At the top right corner of the image, the text "K140025" is visible. eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfection and scheduled replacement as prescribed by the eye care professional. Pegavision recommends up to two week replacement for Etafilcon A lenses, or replacement as recommended by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only. ## 3. NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses #### Spherical and Aspheric NaturalVue (Etafilcon A) Sphere and Asphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Toric NaturalVue (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. #### Multifocal NaturalVue (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. #### Multifocal Toric NaturalVue (Etafilcon A)Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less. NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are intended for single-use disposable wear. # TECNOLOGICAL CHARACTERISTICS: {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for Pegavision, with the tagline "Brighten Your View" underneath. The logo is in blue and gray. The image also contains the text "K140025" in the upper right corner. #### Historical Clinical Use Etafilcon A lenses have been used widely. Their safety and effectiveness have been well established and documented for more than eighteen years. Their safety and effectiveness can be further exemplified by three lenses, as well as many others as discussed in the 510(k), cleared by FDA: · ACUVUE (Etafilcon A) Contact lens, clear and visibility tint with UV blocker, K962804 submitted by Vistakon USA • ACUVUE (Etafilcon A) contact lens, clear and visibility tint with UV blocker, K991134 submitted by Vistakon USA · Aquamax (Etafilcon A) Disposable Soft (Hydrophilic) Contact Lenses, K120028 submitted by Pegavision Corporation, Taiwan Aquamax (Etaflicon A) Soft (Hydrophilic) Contact Lenses from the previous FDA cleared K120028 did not require clinical studies as the USAN name and manufacturing processes are the same as the above-mentioned ACUVUE predicate (and reference) devices. #### Non-Clinical Study All non-clinical tests were conducted in accordance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses. The non-clinical performance tests have been performed to demonstrate the safety and effectiveness of Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for Frequent Replacement, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and establish substantial equivalence to predicate lenses Aquamax Disposable Soft (Hydrophilic) Contact Lenses (K120028) and OSPREY spherical multifocal and multifocal toric soft daily wear contact lenses (K013445) and ACUVUE (Etafilcon A) spherical, toric, and multifocal soft (hydrophilic) daily wear contact lenses (K962804). The evidence of substantial equivalence to the predicate lenses is described below. ## a) Material Properties Testing The material properties of Aquamax and NaturalVue (Etafilcon A) Soft (Hydrophilic) Contact Lenses were tested as illustrated in the following table: | | Proposed Device | K120028 Predicate | K013445 Predicate | K962804 Predicate | |------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------|--------------------------------------------------|----------------------------------------------------| | Production Method | Cast-Molded | Cast-Molded | Lathe-Cut | Cast-Molded | | USAN Name | Etafilcon A | Etafilcon A | Hioxifilcon B | Etafilcon A | | Material Classification | Group 4 high water ionic | Group 4 high water ionic | Group 1 low water non-ionic | Group 4 high water ionic | | Water Content (%) | 58% | 58% | 48% | 58% | | Refractive Index | 1.4023 | 1.4023 | 1.425 | 1.4000 | | Oxygen Permeability<br>(edge corrected) @ 35°C | $19.73 \times 10^{-11}$<br>(cm²/sec)(ml O₂/ml-mmHg) | $19.73 \times 10^{-11}$<br>(cm²/sec)(ml O₂/ml-mmHg) | $15 \times 10^{-11}$<br>(cm²/sec)(ml O₂/ml-mmHg) | $21.4 \times 10^{-11}$<br>(cm²/sec)(ml O₂/ml-mmHg) | | Percent Transmittance<br>% T at 593nm | > 95% | > 95% | >95% | > 85% | | % T at 380-315nm | < 30% | < 30% | UNK | < 30% | | % T at 315-280nm | < 5% | < 5% | UNK | < 5% | #### b) Biocompatibility The standard cytotoxicity, maximization, and ocular irritation tests were carried out for Aquamax (Etafilcon A) Bi-Weekly Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and therefore Traditional 510(k) Pegavision Corporation {10}------------------------------------------------ Image /page/10/Picture/1 description: The image contains the text "K140025" at the top. Below that is the logo for PEGA VISION. Underneath the logo are the words "Brighten Your View". apply to NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. Negative responses were recorded for all tests. The validity of blister packaging for these lenses was demonstrated by passing the standard extraction tests. #### c) Microbiology The steam sterilization process has been validated to deliver a minimum SAL of 10-6, thereby complying with the requirement of FDA Group 4. There is shelf-life stability supporting that these lenses remain sterile through the expiration date claimed for the product. #### d) Bacteriostatic Validation The steam sterilizer was tested for effectiveness by measuring and demonstrating the uniformity of temperature at different locations inside the sterilizer over test period. Tested microorganisms were killed under tested conditions as compared to control. Lenses remained sterilized and there was no microbial growth for a period of 5 years tested under accelerated conditions. Seal of lens packages remained tight for a period of 5 years as demonstrated by the constant peeling strength tested under accelerated conditions. #### e) Leachability Studies were conducted to determine the leachable materials from the finished lenses. The results show that, at the levels of the detection reported, there are no leachable monomers and additive residues. #### CONCLUSION: Based on the verification performance, it can be concluded that the Aquamax (Etafilcon A) Bi-Weekly Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, and NaturalVue (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are equivalent to the predicate devices -K120028 Aquamax (Etafilcon A) Disposable Soft (Hydrophilic) Contact Lenses, K013445 Florida Optical Engineering OSPREY spherical and Multifocal Toric soft daily wear contact lenses, and K962804 Vistakon, Johnson & Johnson Vision Products ACUVUE (Etafilcon A) spherical, toric, and multifocal soft (hydrophilic) daily wear contact lenses with respect to intended use, principles of operation, and technological characteristics.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...