K150326 · Candela Corporation · GEX · Apr 16, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150326
Device Name
PicoWay Laser System
Applicant
Candela Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PicoWay Laser System is indicated for the following at the specified wavelength: 532nm; Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. 1064nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. The PicoWay Laser System is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV.
Device Story
Solid-state laser system; delivers 1064nm or 532nm wavelengths; pulse durations ≤900 picoseconds; repetition rates up to 5 Hz. System includes console, articulated arm, and zoom handpiece (3-6mm spot size). Used in clinical settings by trained professionals. User selects fluence and repetition rate via control panel; internal meter verifies beam transmission. Laser energy targets pigment particles via selective photothermolysis; breaks down tattoo ink or pigmented lesions. Provides non-invasive treatment; clinical benefits include clearance of tattoos and pigmented lesions with favorable safety profile.
Clinical Evidence
Prospective, single-arm, self-controlled multicenter study (n=26 subjects, 29 lesions). Primary endpoint: >70% of lesions achieved ≥50% clearance at primary endpoint visit (blinded independent review). Results showed favorable safety/effectiveness; no serious device-related adverse events. Bench testing confirmed compliance with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-22, 60825-1, and 62366.
Technological Characteristics
Solid-state laser; 1064nm/532nm wavelengths; ≤900ps pulse width; up to 5Hz repetition rate. Articulated arm delivery with 3-6mm adjustable zoom handpiece. Electrically powered console with user interface. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62366, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1. Biocompatibility established via predicate.
Indications for Use
Indicated for tattoo removal (532nm: Fitzpatrick I-III, red/yellow/orange; 1064nm: Fitzpatrick I-VI, black/brown/green/blue/purple) and benign pigmented lesion removal (Fitzpatrick I-IV).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 16, 2015
Candela Corporation % Janice Hogan Regulatory Counsel Hogan and Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Massachusetts 19103
Re: K150326
Trade/Device Name: Picoway Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 9, 2015 Received: February 9, 2015
Dear Janice Hogan,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150326
Device Name
PicoWav Laser Svstem
#### Indications for Use (Describe)
The PicoWay Laser System is indicated for the following at the specified wavelength:
532nm; Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.
1064nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple.
The PicoWay Laser System is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV.
Type of Use (Select one or both, as applicable)
× | Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary PicoWay Laser System
| Submitted by: | Candela Corporation<br>530 Boston Post Road<br>Wayland, MA 01778-1886 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ruthie Amir<br>Global Vice President of Clinical, Regulatory, and Education<br>Tel: 508-358-7400 x330<br>Fax: 508-358-5602 |
| Date prepared: | April 3, 2015 |
| Trade Name: | PicoWay Laser System |
| Common Name: | Dermatology Laser System |
| Classification: | Class II<br>Laser surgical instrument for use in general and plastic surgery and in<br>dermatology (21 CFR 878.4810)<br>Product Code GEX |
#### Predicate and Reference Devices:
Cynosure PicoSure™ workstation (K121346) (Primary Predicate) Predicate Devices: PicoWay Laser System (K142372) Syneron Medical Ltd.'s Transcend System (K120510); Candela Corporation's Reference devices: GentleMAX Pro Laser System (K133283); Quanta System's DUOLITE (K103539); Candela's Q-Switched Alexandrite laser (K081324); HOYA ConBio's RevLite Q-switched Nd:YAG Laser System (K103118, K133254); Cutera, Inc.'s CoolGlide Aesthetic Lasers (K132185); Fotona QS Nd:YAG/KTP Laser System (K083889); Cutera's Picosecond Laser System (K140727, K133945).
## Intended Use / Indications for Use:
The PicoWay Laser System is indicated for the following at the specified wavelength:
## 532nm
Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.
#### 1064nm
Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple.
The PicoWay Laser System is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV.
{4}------------------------------------------------
# Description:
The PicoWay Laser System is a solid state laser capable of delivering energy at wavelengths of 1064 nm or 532 nm at short durations less than or equal to 900 picoseconds (ps) at repetition rates up to 5 Hz. The device system is comprised of a system console, an articulated arm, and an attached handpiece. The laser output at each wavelength is generated within the laser chassis and delivered to the skin through an articulated arm delivery system terminated by a zoom handpiece (HP). The light-weight and ergonomic zoom handpiece allows the spot size on the skin to be easily adjusted from 3 mm to 6 mm in steps of 1 mm. The system includes an internal calibration port with an internal meter located on the control panel of the system console, which is used to verify the transmission of the laser beam into the articulated arm. The PicoWay system control panel enables the user to select the desired energy density (fluence) level and repetition rate. The control panel is also used to obtain feedback from the system, such as the number of pulses delivered or spot size selected.
# Technological Characteristics:
The PicoWay Laser System has the same intended use and similar indications for use, technological characteristics, and operating principles as the Cynosure PicoSure™ workstation (K121346) and the PicoWay Laser System (K142372). The PicoWay device design and components are the same as the previously cleared PicoWay System, and very similar to those of the PicoSure predicate. For each of these device systems, the treatment handpiece is attached to an articulating arm that is connected to the main system console. For each system, the user interface is located at the front/top of the console. For the PicoWay and predicate devices, the laser output at each wavelength is generated within the laser chassis and delivered to the skin through an articulated arm delivery system with a handpiece attached to the end. The handpiece allows the spot size on the skin to be adjusted according to device specifications. Each system thus consists of the articulating arm (and attached handpiece), as well as an electrically powered system console that houses the software, user interface, and produces the laser energy. The PicoWay provides similar key design aspects, including the same or similar spot sizes, laser wavelengths, pulse width, and laser types, as its predicate devices. The frequency (repetition rate) of the PicoWay System is the same as or within the frequency range of the predicates. Further, each of the devices presents a range of spot sizes to allow the user to choose the most appropriate spot size for each patient. Minor differences do not raise any new types of safety or effectiveness questions because the PicoWay parameters are the same as the PicoWay predicate and within the range of the PicoSure predicate.
# Performance Data:
Electrical Safety and Electromagnetic Compatibility: Electrical safety and electromagnetic compatibility (EMC) testing for the PicoWay Laser System was conducted by an independent test laboratory in accordance with IEC 60601-1, Medical equipment, Part 1: General requirements for basic safety and essential performance and with IEC 60601-1-2, Collateral Standard: Electromagnetic Compatibility - Requirements and Tests, 3rd ed. The PicoWay System was determined to be in conformance with applicable IEC standards (IEC 62366, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1).
Biocompatibility: The biocompatibility of the PicoWay device is also established based on the predicate.
{5}------------------------------------------------
Software: Software verification and validation testing was conducted and results demonstrated that testing results were found acceptable for software release.
Clinical Data: A single arm, prospective, self-controlled multicenter study was conducted to evaluate the safety and effectiveness of the PicoWay System for the removal of benign pigmented lesions. The clinical evaluation of 26 subjects (29 lesions) at 2 investigational sites demonstrated that the PicoWay performs as intended and presents a favorable for its indications. The majority of subjects were female, Caucasian, with a mean age of 48 years.
The results of the clinical study demonstrated the favorable safety and effectiveness profile for the PicoWay for the indicated use for benign pigmented lesions removal. As discussed above, the PicoWay device successfully met the protocol-defined primary hypothesis for pigmented lesions clearance, where more than 70% of the treated pigmented lesions achieved 50% or more clearance at the primary endpoint visit based on blinded, independent review. Additional statistical analyses support the robustness of the primary endpoint results.
Consistent with the results observed for the primary endpoint, the study results of the investigator assessments similarly demonstrated a substantial degree of pigmented lesions clearance. Subject assessments further confirmed that the device treatment was generally considered low pain, and available satisfaction data trended towards overall satisfaction with treatment.
Treatment with the PicoWay device also demonstrated a positive safety profile, with no device related serious adverse events, no deaths, and no withdrawals due to adverse events. Of the 107 treatments performed in the study, only 3 adverse events were reported for 2 subjects. These events were not severe and resolved or improved during the study. Anticipated treatmentassociated responses with the PicoWay laser did not present any new types of safety questions as compared to the predicate devices. Therefore, the study demonstrated strong safety results of the PicoWay device for its indicated use.
Based on the clinical performance as documented in the pivotal clinical study, the PicoWay System was found to have a safety and effectiveness profile that is similar to the predicate devices.
All performance testing demonstrated that the PicoWay Laser System performs according to specifications and functions as intended.
## Summary of Substantial Equivalence:
The PicoWay and the predicate devices have the same intended use with the same or similar indications for use. The PicoWay Laser System presents the same or similar technological characteristics as its predicate devices, including the laser type, wavelengths, device design, pulse width, frequency, spot sizes and system components. Any minor differences do not present any new types of safety or effectiveness questions since the PicoWay parameters are the same as the PicoWay predicate. Further, PicoWay performance has been demonstrated in a clinical investigation, and results confirm the safety and effectiveness profile of the PicoWay device and its predicates all operate with the same mechanism of action based on selective photothermolysis of pigment particles using laser energy. Therefore, the PicoWay has the same intended use and the same or similar indications for use, technological characteristics, and principles of operation as the predicate devices. The PicoWay is substantially equivalent to the predicate devices.
{6}------------------------------------------------
# Conclusions:
Clinical testing of the PicoWay device demonstrated that the device performs as intended with a favorable safety profile. Results in the study were comparable to those reported for the predicate device, in support of substantial equivalence. The non-clinical data further support the safety of the device, and software verification and validation testing demonstrates that the PicoWay device is expected to perform as intended in the specified use conditions. The PicoWay System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.