CPR Dura-Padz Reusable Defibrillation Electrode with Dura-padz Gel
K150198 · Bio-Detek, Inc. · MKJ · May 2, 2015 · Cardiovascular
Device Facts
Record ID
K150198
Device Name
CPR Dura-Padz Reusable Defibrillation Electrode with Dura-padz Gel
Applicant
Bio-Detek, Inc.
Product Code
MKJ · Cardiovascular
Decision Date
May 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
Defibrillation ECG Monitoring Cardioversion CPR Feedback The CPR Dura-padz Reusable Defibrillation Electrode is used in conjunction with Dura-padz Gel, and for use with the following ZOLL Biphasic-only defibrillators with max. energy setting of 200 Joules: AED Pro M Series E Series R Series The device will be used in pre-hospital, alternate care and hospital settings by trained personnel only, including: Physicians Nurses Paramedics Emergency Medical Technicians Cardiovascular Laboratory Technicians First Responders The CPR Dura-padz Reusable Defibrillation Electrodes are not for use with ZOLL AED Plus and/or any Public Access Defibrillators. The CPR Dura-padz Reusable Defibrillation Electrodes are not indicated for use on a patient less than 8 years of age or weighing less than 55 lbs (25kg).
Device Story
Reusable defibrillation electrode system; used with Dura-padz Gel and compatible ZOLL biphasic defibrillators. Inputs: electrical signals for ECG monitoring and CPR compression data via integrated sensor. Device transforms inputs into real-time CPR feedback and cardiac monitoring data for display on host defibrillator. Used by trained clinicians (physicians, nurses, paramedics, EMTs) in clinical/pre-hospital settings. Output enables providers to monitor cardiac rhythm and optimize CPR compression quality. Benefits: facilitates effective resuscitation and monitoring. Reusable up to 100 patient uses.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and verification/validation of functional requirements against recognized standards.
Technological Characteristics
Reusable multi-function electrode; includes integrated CPR sensor. Compatible with ZOLL biphasic defibrillators (max 200J). Materials: conductive electrode components and reusable gel. Connectivity: wired interface to host defibrillator. Reusable up to 100 patient uses.
Indications for Use
Indicated for adult patients (>= 8 years old, >= 55 lbs/25 kg) requiring defibrillation, cardioversion, ECG monitoring, or CPR feedback. Used with specific ZOLL biphasic defibrillators (AED Pro, M, E, R Series) by trained medical personnel in hospital, pre-hospital, or alternate care settings. Contraindicated for use with ZOLL AED Plus, public access defibrillators, or pediatric patients under 8 years/55 lbs.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
Image /page/0/Picture/12 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 2, 2015
Bio-Detek, Inc. % Shannon Duhamel Regulatory Affairs Specialist Zoll Medical Corporation 269 Mill Road Chelmsford, Massachusetts 01824-4105
Re: K150198
Trade/Device Name: CPR Dura-padz Reusable Defibrillation Electrode with Dura-padz Gel Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, LIX Dated: April 13, 2015 Received: April 14, 2015
Dear Shannon Duhamel,
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# INDICATIONS FOR USE
# 510(k) Number (if known): K150198
# Device Name: CPR Dura-padz Reusable Defibrillation Electrode with Dura-padz Gel
Intended Use:
- Defibrillation
- ECG Monitoring
- Cardioversion ●
- CPR Feedback ●
The CPR Dura-padz Reusable Defibrillation Electrode is used in conjunction with Dura-padz Gel, and for use with the following ZOLL Biphasic-only defibrillators with max. energy setting of 200 Joules:
- AED Pro
- M Series ●
- E Series ●
- R Series ●
The device will be used in pre-hospital, alternate care and hospital settings by trained personnel only, including:
- . Physicians
- Nurses
- Paramedics
- Emergency Medical Technicians ●
- Cardiovascular Laboratory Technicians
- First Responders
The CPR Dura-padz Reusable Defibrillation Electrodes are not for use with ZOLL AED Plus and/or any Public Access Defibrillators.
The CPR Dura-padz Reusable Defibrillation Electrodes are not indicated for use on a patient less than 8 years of age or weighing less than 55 lbs (25kg).
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{3}------------------------------------------------
# 510(k) Summary:
| Submitter's Name and Address: | Bio-Detek, Inc.<br>525 Narragansett Park Drive<br>Pawtucket, RI 02861 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Application Correspondent: | Shannon Duhamel<br>RA Specialist<br>(978) 421-9574<br>sduhamel@zoll.com |
| Date Summary Prepared: | July 9, 2014 |
| Classification: | Class III |
| Device Name | CPR Dura-padz Reusable Defibrillation Electrode with Dura-padz Gel |
| Common Name | Automated External Defibrillator Multi-Function Electrodes |
| Classification Name | Automated External Defibrillators (MKJ;<br>21 CFR 870.5310)<br>Cardiopulmonary Resuscitation Aid (LIX;<br>21 CFR 870.5200) |
| Predicate Devices | K100565 - Dura-padz Reusable Defibrillation Electrode used in conjunction with Dura-padz Gel<br>K110742 - ZOLL OneStep Adult Multi-Function Electrode |
#### Substantial Equivalence - Non-Clinical Evidence:
The existing features and functions (defibrillation, cardioversion and ECG monitoring) of the CPR Dura-padz Reusable Defibrillation Electrode have been cleared per K100565. As with the cleared predicate (K100565), the subject device is intended for use on adult patients with the following ZOLL biphasic-only defibrillators: AED Pro, E Series, R Series and M Series. The CPR Feedback monitoring function is the same technology used in the predicate device cleared per K110742. Safety, efficacy and substantial equivalence was shown through verification and validation testing.
{4}------------------------------------------------
# Substantial Equivalence - Clinical Evidence:
N/A - Clinical evidence was not necessary to show substantial equivalence.
# Description:
As with the cleared predicate device (K100565), the CPR Dura-padz Reusable Defibrillation Electrode is intended for use with the following ZOLL biphasic-only defibrillators: AED Pro, E Series, R Series and M Series for ECG monitoring, defibrillation and cardioversion. The addition of a CPR sensor (cleared per K110742) to the subject device will enable CPR feedback. As with the currently marketed predicate device (K100565), the CPR Dura-padz electrode is intended for use in conjunction with Dura-padz Gel on adult patients, and the electrode is reusable up to 100 patient uses. The device Indications For Use are as follows:
- Defibrillation ●
- ECG Monitoring
- Cardioversion ●
- CPR Feedback ●
The CPR Dura-padz Reusable Defibrillation Electrode is used in conjunction with Dura-padz Gel, and for use with the following ZOLL Biphasic-only defibrillators with max. energy setting of 200 Joules:
- AED Pro
- M Series ●
- E Series ●
- R Series
The device will be used in pre-hospital, alternate care and hospital settings by trained personnel only, including:
- Physicians ●
- Nurses ●
- Paramedics ●
- Emergency Medical Technicians
- Cardiovascular Laboratory Technicians
- First Responders
The CPR Dura-padz Reusable Defibrillation Electrodes are not for use with ZOLL AED Plus and/or any Public Access Defibrillators.
The CPR Dura-padz Reusable Defibrillation Electrodes are not indicated for use on a patient less than 8 years of age or weighing less than 55 lbs (25kg).
{5}------------------------------------------------
# Comparison of Technological Characteristics:
The CPR Feedback, ECG monitoring, defibrillation and cardioversion functions of the CPR Dura-padz electrode cleared with the predicate devices K110742 and K100565 have remained unchanged in the proposed version of the device.
### Performance Testing:
The CPR Dura-padz Reusable Defibrillation Electrode with Dura-padz Gel has been subjected to extensive performance testing to ensure the device meets all of its functional requirements and performance specifications as defined in applicable National/International recognized standards. Performance testing is provided in Section 18 of this submission.
## Conclusion:
The information provided in this 510(k) demonstrates that the CPR Dura-padz Reusable Defibrillation Electrode with Dura-padz Gel features and functions are substantially equivalent to those of the indicated commercially distributed devices with regard to performance, safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.