Defibrillation . Cardioversion ● Noninvasive Pacing . ECG Monitoring . CPR Sensor . For use with ZOLL @ Defibrillators: - R Series ◆ - M Series ● By Trained Personnel only, Including: - Physicians . - Nurses ◆ - Paramedics . - Emergency Medical Technicians . - Cardiovascular Laboratory Technicians . The OneStep Adult Electrodes are not indicated for use on a patient less than 8 years of age or weighs less than 55 Lbs. (25Kg).
Device Story
OneStep Adult Multi-Function Electrode system provides interface between patient and ZOLL R/M Series defibrillators. Inputs: electrical signals for ECG monitoring, CPR sensor data, and energy delivery for therapy. Device consists of disposable solid hydrogel electrodes with pure tin conductive elements and PET backing. Operation: electrodes applied to patient skin; caregiver selects version (Complete, CPR, Pacing, or Basic) based on required therapy (defibrillation, cardioversion, pacing, monitoring). Self-test circuit integrated into electrode to verify continuity. Output: therapy delivery and physiological data to defibrillator. Used in hospital/pre-hospital settings by trained clinicians. Benefits: enables rapid, reliable therapy delivery and monitoring; improved self-test circuit design enhances reliability.
Clinical Evidence
Bench testing only. Design validation (VPN 0730) confirmed integration with ZOLL defibrillators and performance equivalent to predicate devices. No clinical data presented.
Technological Characteristics
Solid hydrogel electrode with pure tin conductive element; PET backing; adhesive perimeter. Connectivity: wired to ZOLL R/M Series defibrillators. Standards: ANSI/AAMI DF80, EN 60601-2-4. Features: integrated self-test continuity circuit, optional CPR sensor, optional pacing (MWP).
Indications for Use
Indicated for adult patients (≥8 years, ≥55 lbs) requiring defibrillation, cardioversion, noninvasive pacing, ECG monitoring, or CPR sensor assistance. For use by trained personnel (physicians, nurses, paramedics, EMTs, cardiovascular lab technicians) with ZOLL R Series and M Series defibrillators.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
## K110742
# NCORPORA
#### APR 1 3 2011
#### 510(k) Summary: Rev. 5
Submitter's Name and Address:
Bio-Detek, Inc. A Division of ZOLL® Medical Corporation 525 Narragansett Park Drive Pawtucket, RI 02861 Tel. (401) 729-1400
Contact Person: Robert Morse Director, Regulatory Affairs Tel. (866) 639-0060 x. 224
Date Summary Prepared: February 23, 2011
Device Name: OneStep ™ Adult Multi-Function Electrode (MFE)
#### Classification Name:
Electrode, Electrocardiograph, Multi-Function with CPR Sensor; Accessory to an Automated External Defibrillator
#### Substantial Equivalence:
The OneStep Electrode Family is substantially equivalent to ZOLL ready-padz OneStep Electrodes that were FDA cleared on 510(k), K060559. In addition, the OneStep Electrode is substantially equivalent to the originally cleared ZOLL stat: padz MWP identified on 510(k), K051076. The intent is to show that the OneStep Electrodes will assist in providing therapy for Defibrillation, Cardioversion, External Noninvasive Pacing, ECG Monitoring and CPR Sensor comparable to the previously cleared devices, for use with ZOLL Defibrillators. All of the cleared devices are in the . Regulatory Class III category.
#### Description of Device:
The OneStep Electrode System is intended for use with ZOLL R Series and M Series for ECG Monitoring, Defibrillation, External Noninvasive Pacing, Cardioversion and CPR Sensor of adult patients in either the hospital or pre-hospital environment.
525 Narragansell Park Drive, Pawtucket, Rhode Island 02861-4323, Tel. (401) 729-1400 or (800) 729-1408
{1}------------------------------------------------
Through part number selection you can choose the full featured Complete model down to the Basic version which offers fewer therapy options. The OneStep Adult Electrodes consist of 5 versions related to the combination of features included with the electrodes. The chart below contains the versions and their features:
| Version | MFE | CPR | MWP | Self Test | ZOLL Part Number |
|----------------------|-----|-----|-----|-----------|------------------|
| OneStep Complete | x | x | x | x | 8900-0214-01 |
| OneStep CPR AA | x | x | | x | 8900-0217-01 |
| OneStep CPR AP | x | x | | x | 8900-0213-01 |
| OneStep Pacing (MWP) | x | | x | x | 8900-0212-01 |
| OneStep Basic | x | | | x | 8900-0211-01 |
OneStep Electrode Versions:
MFE - Multi-Function Electrode
CPR - CPR Sensor
(Cardiopulmonary Resuscitation Aid)
MWP - Imbedded ECG Electrodes in front pad for pacing (MWP = Monitor While Pace)
AA - Optimized for Anterior / Anterior Placement
AP - Anterior / Posterior Placement
The system is comprised of a single use, disposable electrode made of solid hydrogel with a pure tin electrical conductive element having a polyethylene terephtalate (PET) backing with an adhesive perimeter suitable for coupling to patient skin during rescue and/or treatment. By product selection it is determined whether all therapy providing options are available or a reduced choice is made as applicable to caregiver needs.
#### Indications for Use
Device Name: OneStep™ Adult Multi-Function Electrode
Intended Use:
- Defibrillation .
- Cardioversion ●
- Noninvasive Pacing �
- . ECG Monitoring
- CPR Sensor .
For use with ZOLL @ Defibrillators:
- R Series ◆
- M Series ●
{2}------------------------------------------------
By Trained Personnel only, Including:
- Physicians .
- Nurses ◆
- Paramedics .
- Emergency Medical Technicians .
- Cardiovascular Laboratory Technicians .
The OneStep Adult Electrodes are not indicated for use on a patient less than 8 years of age or weighs less than 55 Lbs. (25Kg).
#### Comparison of Technological Characteristics:
The intended use of the OneStep Multi-Function Electrodes as described in the indications for use, and labeling has not changed as a result of this submission. For the connection of the OneStep Family of Electrodes in this submission to the ZOLL defibrillator cables are similar to the predicate devices with the addition of the M Series utilizing an adaptor for CPR Monitoring functionality which can be procured under ZOLL part number 8000-0599. The OneStep MFE's within this submission as compared to the two ZOLL predicate electrode devices using the same hydrogel, acrylic adhesive closed cell foam and have a conductive area of the same dimensional size that meet/exceed National and International Standards ANSI/AAMI DF80 and EN 60601-2-4 respectively.
#### Nonclinical Testing & Clinical Service:
The OneStop Electrode System has been subjected to extensive performance testing to ensure the device meets all of its functional requirements and performance specifications as defined in applicable National/International recognized standards. The ZOLL OneStep Electrode Family has been in clinical service since the latter part of October, 2006 through a substantial equivalence determination by FDA for clearance to distribute through a Traditional 510(k) process. This Special 510(k) has been compiled to address the relocation of the self test continuity circuit from the opening process of the pouch to two disconnect points on each electrode. Upon electrode removal from the release liner material, exposed when the pouch is opened, the self test circuit will disengage allowing therapy to be carried-out. It is our intent to present a safer and more reliable circuit disconnects to enhance the effectiveness of the self test process.
Validation Project Number (VPN) 0730 addresses design validation to ensure the safety and effectiveness considerations have been successful in integrating the OneStep electrode with previously stated ZOLL defibrillators and performs as well and/or better than the legally marketed predicate devices. Applicable standards are defined in Form FDA 3514, VPN 0730 and Section 2. Device Description, within this premarket submission.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three abstract, curved shapes that resemble a bird in flight or flowing ribbons.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zoll Medical Corporation c/o Mr. Robert Morse Bio-Detek, Inc. 525 Narragansett Park Drive Pawtucket, RI 02861
APR 1 3 2011
Re: K110742
Trade/Device Name: Onestep Adult Multi-function Electrode Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, DRO, DQA, LDD, LIX Dated: March 1.7, 2011 Received: March 18, 2011
Dear Mr. Morse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{4}------------------------------------------------
Page 2 - Mr. Robert Morse
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours?
signature
Bram D. Zuckerman, M.D. for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) number (if known): _ 长 ١١٥٦٧ 2
Device Name: OneStep™ Adult Multi-Function Electrode
Intended Use:
- Defibrillation .
- Cardioversion .
- Noninvasive Pacing �
- ECG Monitoring .
- CPR Sensor, (Optional) �
For use with ZOLL وه Defibrillators:
- R Series .
- M Series .
By Trained Personnel only, Including:
- Physicians ●
- Nurses ●
- Paramedics .
- Emergency Medical Technicians .
- Cardiovascular Laboratory Technicians ●
The OneStep Adult Electrodes are not indicated for use on a patient less than 8 years of age or weighs less than 55 Lbs. (25Kg).
(Please do not write below this line - Continue on another page if necessary)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | | | |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (Division Sign-Off) | | | | |
| for Division of Cardiovascular Devices | | | | |
| 510(k) Number | K110 792 | | | |
| Prescription Use | <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span> </div> | OR | Over the Counter Use | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> </div> |
| (Per 21 CFR 801.109) | Rev. 4 | | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.