K150175 · Home Skinovations , Ltd. · ILY · Mar 31, 2015 · Physical Medicine
Device Facts
Record ID
K150175
Device Name
HeatLux Pro I
Applicant
Home Skinovations , Ltd.
Product Code
ILY · Physical Medicine
Decision Date
Mar 31, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HeatLux Pro I is over the counter hand held device intended to emit energy in the visible and near IR spectrum to provide topical heating for the purpose of elevating tissue temperary relief of minor muscular and joint pan and stiffness, minor arthritis pam or muscle spasm, the temporary increase in local and blood carculation, and temporary relaxation of muscles.
Device Story
Hand-held device for OTC topical heating; utilizes array of 24 LEDs (630±10nm) and four temperature-stabilized metal heating plates (41°C). Device connects to AC/DC power adapter for operation. User applies applicator tip directly to skin. Heat and light energy elevate tissue temperature to provide temporary relief of minor muscle/joint pain, stiffness, arthritis, and spasms; increases local blood circulation; promotes muscle relaxation. Device replaces previous battery-powered model with direct AC/DC power for extended operation and increased device lifetime.
Clinical Evidence
Bench testing only. Evaluated thermal profile and temperature stability compared to predicate device. Results demonstrated equivalent performance. No clinical or animal studies performed.
Technological Characteristics
Hand-held applicator with 24 LEDs (630±10nm) and 4 metal heating plates (stabilized at 41°C). Max output 12W; optical power density 10mW/cm². Powered by 100-240 VAC adapter. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-57 (non-laser light source).
Indications for Use
Indicated for OTC use by adults for temporary relief of minor muscular and joint pain, stiffness, minor arthritis pain, muscle spasms, temporary increase in local blood circulation, and temporary muscle relaxation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2015
Home Skinovations Ltd. % Ms. Ahava Stein A. Stein - Regulatory Affairs Consulting Ltd. 20 Hata'as Street, Suite 102 Kfar Saba 4442520 Israel
Re: K150175
Trade/Device Name: HeatLux Pro I Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: ILY Dated: March 1, 2015 Received: March 6, 2015
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known)
#### K150175
Device Name HeatLux Pro I
#### Indications for Use (Describe)
HeatLux Pro I is over the counter hand held device intended to emit energy in the visible and near IR spectrum to provide topical heating for the purpose of elevating tissue temperary relief of minor muscular and joint pan and stiffness, minor arthritis pam or muscle spasm, the temporary increase in local and blood carculation, and temporary relaxation of muscles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
Section 6:
# Special 510(k) Summary
{4}------------------------------------------------
#### 510(K) SUMMARY
## HEATLUX PRO I DEVICE
# 510(k) Number K_
### Applicant Name:
| Company Name: | Home Skinovations Ltd. |
|---------------|----------------------------------------------------------|
| Address: | Tavor Building, POB 533<br>Yokneam Iillit 2069206 Israel |
| | Tel: +972(4)9097440 |
| | Fax: +972(4)9097471 |
| | E-mail: ahava@asteinrac.com |
## Contact Person:
| Official Correspondent: | Ahava Stein |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 4442520 Israel<br>Tel: +972-9-7670002<br>Fax: +972-9-7668534<br>E-mail: ahava@asteinrac.com |
| Date Prepared: | January 22, 2015 |
| Trade Name: | HeatLux Pro I device |
| Classification Name: | CFR Classification section 890.5500; (Product code ILY) |
| Classification: | Class II Medical Device |
## Predicate Device:
The HeatLux Pro I device is substantially equivalent to the previously cleared, HeatLux 1 device, also manufactured by Home Skinovations Ltd.
| Device | Manufacturer | 510(k) No. |
|-----------|------------------------|------------|
| HeatLux 1 | Home Skinovations Ltd. | K120582 |
{5}------------------------------------------------
### Device Description:
The HeatLux Pro I device is a hand held device, using low power light spectrum, array of 24 LEDs, at wavelength of 630±10nm, combined with four metal heating plates (temperature stabilized to 41%). The HeatLux Pro I device consists of an applicator and an AC/DC power adaptor. The applicator is a hand held unit used for treatment, as the treatment surface at the applicator tip comes in direct contact with the skin.
### Device Specifications:
Maximal heating plates output power: 12W Maximal optical power density: 10mWatts/cm2 Package dimensions: 22cm x 22cm x 14cm Weight: 250gr Main Line Frequency (nominal): 50-60 Hz Input Voltage (nominal): 100-240 VAC
#### Intended Use/Indication for Use:
The HeatLux Pro I device is over the counter hand held device intended to emit energy in the visible and near IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffhess, minor arthritis pain or muscle spasm, the temporary increase in local and blood circulation, and temporary relaxation of muscles.
#### Performance Standards:
The HeatLux Pro I device has been tested and complies with the following voluntary recognized standards:
- IEC 60601-1, (Third Edition, 2005 / 2006), Medical Electrical Equipment Part ● 1: General requirements for basic safety and essential performance
- IEC 60601-1-2. (Third Edition, 2007), Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests
- . IEC 60601-2-57, (First Edition, 2011), IEC 60601-2-57 (2011), Medical Electrical Equipment - Part 2-57: Particular Requirements for The Basic Safety And Essential Performance Of Non-Laser Light Source Equipment Intended For Therapeutic, Diagnostic, Monitoring And Cosmetic/Aesthetic Use.
#### Non-Clinical (Bench) Performance Data:
A set of bench tests were performed to evaluate the thermal profile and temperature stability of the HeatLux Pro I device and to compare them to the thermal profile and temperature stability of the HeatLux 1 device.
{6}------------------------------------------------
The results of the bench tests demonstrated that the HeatLux Pro I device has the same thermal profile and temperature stability properties as those reported for the predicate device.
## Pre-Clinical (Animal Study) Performance Data:
Not Applicable
## Clinical Performance Data:
Not Applicable
### Substantial Equivalence:
The indications for use and technological characteristics of the HeatLux Pro I device are substantially equivalent to the indications for use and technological characteristics of the HeatLux 1 device.
The design and components in the HeatLux Pro I device, including the wall adaptor and the applicator are similar to the design and components found in the predicate HeatLux 1 device. The battery was replaced with the direct adaptor charger for longer operation time and in order to increase the device life time. The differences do not negatively affect the safety or efficacy of the device as supported by the results of the performance tests. The performance specifications of the HeatLux Pro I device are substantially equivalent to those in the HeatLux 1 device. The safety features and compliance with safety standards in the HeatLux Pro I device are similar to the safety features and compliance with safety standards found in the HeatLux 1 device. Patient contact materials were partly modified and tested for biocompatibility. Any differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the new HeatLux Pro I device underwent performance testing, including software validation testing and electrical and mechanical safety testing according to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2 and bench tests. These performance tests demonstrated that the differences in the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.
Consequently, it can be concluded that the HeatLux Pro I device is substantially equivalent to the predicate HeatLux 1 device, cleared under 510(k) K120582.
## Conclusions:
Based on the performance testing and comparison to predicate device, the HeatLux Pro I device is substantially equivalent to the HeatLux 1 predicate device for the abovementioned Intended Use/Indication for Use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.