Post-market analysis of end-user experience and feedback
The sponsor utilized post-market surveillance data and end-user feedback from clinical environments to support the clinical safety and effectiveness of the device, noting that no new clinical tests were performed for this submission.
Post-market analysis of end-user experience; Post-market surveillance / observational analysis
Patients monitored in clinical environments using the Panorama Network
Not applicable for this study
Clinical safety and effectiveness
Indications for Use
The indications for use for the Panorama Patient Monitoring Network include: · Viewing real time patient clinical and demographic data · Graphical and numeric trending of clinical data · Storing and printing of clinical and demographic data · Setting independent alarm limits for data sent by the bedside monitor The clinical data displayed by the Panorama Patient Monitoring Network is obtained from one or more compatible physiological monitors and includes: ECG (Lethal and Non-Lethal Arrhythmia Detection, ST Segment Analysis, QT Analysis and heart rate(HR)), Invasive Blood Pressure (IBP), Non-Invasive Blood Pressure(NIBP), Blood Oxygenation (SpO2), Pulse Rate (PR), Respiration Rate (RESP), Respiration Gases(O2, CO2), Temperature(TEMP), Anesthetic Gases (AG), Cardiac Output(C.O.), Bispectral Index (BIS), and Hemodynamics. The Panorama Patient Monitoring Network is intended for use in a fixed location, in the healthcare facility setting, as a central viewing station. The Panorama Patient Monitoring Network is not intended to the patient at any time or installed in a patient's vicinity. The Panorama Network includes the Panorama Telemetry System which acquires and monitors physiological data for ambulating patients within a defined coverage area. The system processes the physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations. The Panorama Telemetry System is intended for installation in a hospital or clinical environment to provide clinicians with patient physiological data, while allowing for patient mobility. The physiological parameters monitored through telemetry include ECG (Lethal and Non-Lethal Arrhythmia Detection, ST Segment Analysis, OT Analysis and heart rate(HR)), blood oxygenation (SpO2), Pulse Rate (PR), Received data is calculated at the Panorama Central Station, and can be displayed, trended, stored and printed at the Panorama Central Station. The Panorama Patient Monitoring Network is intended for use under the direct supervision of a licensed healthcare practitioner.
Device Story
Panorama Patient Monitoring Network acts as a central viewing station for physiological data; collects inputs from bedside monitors or ambulatory Telepack devices via telemetry; processes data to detect ECG arrhythmias and parameter limit violations; displays, trends, stores, and prints clinical/demographic data. Used in hospital/clinical settings by clinicians; not for direct patient connection. Includes ViewStation (remote display), WorkStation (remote management/alarm adjustment), Web Viewer (intranet access), and Gateway (HL7 translation). Enables remote patient oversight, facilitating clinical decision-making through real-time monitoring and alarm notification.
Clinical Evidence
Bench testing only. No clinical data performed. Safety and effectiveness established through compliance with recognized consensus standards (ANSI/AAMI ES60601-1, IEC60601-1-2, IEC60601-1-6, IEC60601-1-8, IEC60601-2-27, EC57, ISO80601-2-61), risk analysis, and post-market analysis of end-user experience.
Technological Characteristics
Central station/telemetry system; Windows XP embedded OS; touch screen interface. Connectivity via WMTS (telemetry) and intranet (Web Viewer). Software-based arrhythmia, ST-segment, and QT analysis. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 60601-2-27, EC57, and ISO 80601-2-61.
Indications for Use
Indicated for hospital/clinical patients requiring physiological monitoring, including ambulatory patients via telemetry. Monitors ECG (arrhythmia, ST, QT, HR), IBP, NIBP, SpO2, PR, RESP, gases, Temp, C.O., BIS, and hemodynamics. Intended for use under licensed practitioner supervision.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human figures connected by flowing lines, symbolizing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 2, 2015
Shenzhen Mindray Bio-medical Electronics Co., Ltd Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Rd South, Hi-tech Industrial Park Nanshan, Shenzhen 518057, P.R. China
Re: K150082 Trade/Device Name: Panorama Patient Monitoring Network Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DQA, DRG, MSX Dated: July 31, 2015 Received: August 5, 2015
Dear Yanhong Bai,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
K150082 510(k) Number (if known)
Page 1 of 2
Device Name
Panorama Patient Monitoring Network
Indications for Use (Describe)
The indications for use for the Panorama Patient Monitoring Network include:
· Viewing real time patient clinical and demographic data
- · Graphical and numeric trending of clinical data
- · Storing and printing of clinical and demographic data
- · Setting independent alarm limits for data sent by the bedside monitor
The clinical data displayed by the Panorama Patient Monitoring Network is obtained from one or more compatible physiological monitors and includes: ECG (Lethal and Non-Lethal Arrhythmia Detection, ST Segment Analysis, QT Analysis and heart rate(HR)), Invasive Blood Pressure (IBP), Non-Invasive Blood Pressure(NIBP), Blood Oxygenation (SpO2), Pulse Rate (PR), Respiration Rate (RESP), Respiration Gases(O2, CO2), Temperature(TEMP), Anesthetic Gases (AG), Cardiac Output(C.O.), Bispectral Index (BIS), and Hemodynamics.
The Panorama Patient Monitoring Network is intended for use in a fixed location, in the healthcare facility setting, as a central viewing station. The Panorama Patient Monitoring Network is not intended to the patient at any time or installed in a patient's vicinity.
The Panorama Network includes the Panorama Telemetry System which acquires and monitors physiological data for ambulating patients within a defined coverage area. The system processes the physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations. The Panorama Telemetry System is intended for installation in a hospital or clinical environment to provide clinicians with patient physiological data, while allowing for patient mobility.
The physiological parameters monitored through telemetry include ECG (Lethal and Non-Lethal Arrhythmia Detection, ST Segment Analysis, OT Analysis and heart rate(HR)), blood oxygenation (SpO2), Pulse Rate (PR), Received data is calculated at the Panorama Central Station, and can be displayed, trended, stored and printed at the Panorama Central Station.
The Panorama Patient Monitoring Network is intended for use under the direct supervision of a licensed healthcare practitioner.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|--|--|--|--|
|-------------------------------------------------|--|--|--|--|
اح Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
K150082 Page 2 of 2
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 5. 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Panorama Patient Monitoring Networks provided below.
| Device Common Name: | Central Station Monitor and Telemetry System |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | Panorama Patient Monitoring Network |
| Submitter: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680 |
| Contact: | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | January 8, 2014 |
| Classification Regulation: | 21 CFR 870.1025, Class II, Arrhythmia detector<br>and alarm (including ST-segment measurement and<br>alarm) |
| Panel: | Cardiovascular |
| Product<br>Code | Regulation<br>Number | Panel | Regulation Description | Device Common Name |
|-----------------|----------------------|----------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Primary | | | | |
| MHX | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including ST-<br>segment measurement<br>and alarm) | monitor, physiological,<br>patient (with arrhythmia<br>detection or alarms) |
| Secondary | | | | |
{5}------------------------------------------------
510(k) Premarket Notification
| Product<br>Code | Regulation<br>Number | Panel | Regulation Description | Device Common Name |
|-----------------|----------------------|----------------|--------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| DSI | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including ST-<br>segment measurement<br>and alarm) | detector and alarm,<br>arrhythmia |
| MLD | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including ST-<br>segment measurement<br>and alarm) | monitor, st segment with<br>alarm |
| DRT | 21 CFR<br>870.2300 | Cardiovascular | Cardiac Monitor<br>(including<br>cardiotachometer and<br>rate alarm) | monitor, cardiac (incl.<br>cardiotachometer & rate<br>alarm) |
| DQA | 21 CFR<br>870.2700 | Cardiovascular | Oximeter | Oximeter |
| DRG | 21 CFR<br>870.2910 | Cardiovascular | Radiofrequency<br>physiological signal<br>transmitter and receiver. | transmitters and receivers,<br>physiological signal,<br>radiofrequency |
| MSX | 21 CFR<br>870.2300 | Cardiovascular | Cardiac monitor<br>(including<br>cardiotachometer and<br>rate alarm). | system,network and<br>communication,physiological<br>monitors |
Predicate Device: Panorama Patient Monitoring Network (K142601)
Hypervisor central monitoring system(including telemetry monitoring system.tms-6016) (K132036)
Philips ST/AR ST and Arrhythmia software (K101521)
#### Indications for Use:
The indications for use for the Panorama Patient Monitoring Network include:
- Viewing real time patient clinical and demographic data ●
- Graphical and numeric trending of clinical data o
- . Storing and printing of clinical and demographic data
- o Setting independent alarm limits for data sent by the bedside monitor
The clinical data displayed by the Panorama Patient Monitoring Network is obtained from one or more compatible physiological monitors and includes: ECG (Lethal and Non-Lethal Arrhythmia Detection, ST Segment Analysis, QT Analysis and heart rate(HR)), Invasive Blood Pressure (IBP), Non-Invasive Blood Pressure(NIBP), Blood Oxygenation (SpO2), Pulse Rate (PR), Respiration Rate (RESP), Respiration Gases(02,CO2), Temperature(TEMP), Anesthetic Gases(AG), Cardiac Output(C.O.), Bispectral Index (BIS), and Hemodynamics.
{6}------------------------------------------------
The Panorama Patient Monitoring Network is intended for use in a fixed location, in the healthcare facility setting, as a central viewing station. The Panorama Patient Monitoring Network is not intended to be directly connected to the patient at any time or installed in a patient's vicinity.
The Panorama Network includes the Panorama Telemetry System which acquires and monitors physiological data for ambulating patients within a defined coverage area. The system processes the physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations. The Panorama Telemetry System is intended for installation in a hospital or clinical environment to provide clinicians with patient physiological data, while allowing for patient mobility.
The physiological parameters monitored through telemetry include ECG (Lethal and Non-Lethal Arrhythmia Detection, ST Segment Analysis, QT Analysis and heart rate(HR)), blood oxygenation (SpO2), Pulse Rate (PR). Received data is calculated at the Panorama Central Station, and can be displayed, trended, stored and printed at the Panorama Central Station.
The Panorama Patient Monitoring Network is intended for use under the direct supervision of a licensed healthcare practitioner.
## Device Description:
The subject Panorama Patient Monitoring Network includes eight Components:
- o Panorama Central Station
- Panorama Telemetry System
- Panorama Telepack ●
- Compatible Physiological Monitor(s) ●
- o ViewStation
- WorkStation ●
- Web Viewer ●
- Gateway ●
- 1、The Panorama Central Station is a device that monitors patients, collects and stores data.and performs alarm notification .The clinical data displayed at the Central Station is derived from compatible physiological monitor(s) or the Panorama Telepack.
- 2、The Panorama Telemetry System is designed to acquire and monitor physiological data for ambulating patients within a defined coverage area.
- 3、The Panorama Telepack is a battery powered ambulatory monitoring device intended for use in a hospital/clinical environment to provide clinicians with patient
{7}------------------------------------------------
physiological data, while allowing for patient mobility. Data received from the Telepack is transmitted to the Panorama Telemetry System .
- 4、Mindray's cleared Compatible Physiological Monitors that transmit patient data to the Central Station include:
- · Passport 12M/17M Monitor(K143195)
- · Passport 8/12 Monitor(K132662)
- · T1(K143195)
- · Passport V Monitor(K091834)
- · V Series Monitor(K132026)
- · DPM 6/7 Monitor(K092449)
- · Spectrum/ Spectrum OR Bedside Monitor(K062098)
- · Passport 2 Bedside Monitor (K020550)
- 5、The Panorama ViewStation is a software application that allows the user to remotely display and print patient information from the Central Station.
- 6、The Panorama WorkStation is a software application that allows the user to remotely display and print patient information, enter or change patient demographic information, change alarm limits and responses, place patients in standby or discharge patients.
- 7、The Panorama Web Viewer is a software application that provides the user to remotely view patient information from the Central Station through a facilities intranet using a Microsoft Internet Explorer web browser.
- 8、The Panorama Gateway is an embedded Windows XP software application that allows translation of the Mindray proprietary network format into standard formats such as HL7.
## Summary of performance testing
- The Panorama Network has been tested and found to be in compliance with ● recognized safety, performance and electromagnetic compatibility standards.
- A risk analysis has been developed to identify potential hazards and document the o mitigation of the hazards. The device's software has been verified and validated in accordance with the appropriate test requirements.
- The Panorama Network has been tested and complies with the following recognized ● consensus standards:
- V ANSI/AAMI ES60601-1
- > IEC60601-1-2
- > IEC60601-1-6
- > IEC60601-1-8
- A IEC60601-2-27
- > EC57
{8}------------------------------------------------
A ISO80601-2-61
- o The Panorama Network was tested and complies with established parameter performance specifications.
- o No clinical tests were performed on the Panorama Network. The clinical safety and effectiveness of the central monitoring networks has been long established through analysis of end-user experience and feedback gained through post-market analysis.
### Substantial Equivalence:
### Comparison of Indications -
The Panorama Patient Monitoring Network/K142601 shall serve as the primary predicate device respecting Indications for Use, performance specifications and technology.
The Hypervisor Central Monitoring System /K132036 is provided as a predicate device supporting bispectral index (BIS) that has been added to the subject Panorama. Section 12.4.4 includes additional details related to this new parameter.
The Philips ST/AR ST and Arrhythmia software/K101521 is provided as a predicate device supporting QT analysis that has been added to the subject Panorama. Section 12.4.1 includes additional details related to this new parameter.
The subject Panorama considered substantially equivalent to the predicate devices respecting the indications for use.
## Comparison of Technological Characteristics –
The table below compares the key technological feature of the subject devices to the primary predicate device 。
| Technology | Subject /<br>Panorama Patient<br>Monitoring Network | Predicate /<br>Panorama Patient<br>Monitoring Network<br>(K142601) |
|--------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Central Station<br>Telemetry Server<br>Keyboard<br>Mouse | 2U (Horizontal) or Vertical<br>2U (Horizontal) Tower<br>Yes<br>Yes | 2U (Horizontal) or Vertical<br>2U (Horizontal) Tower<br>Yes<br>Yes |
| Display<br>Operating System<br>Longview-VGA extender | Touch screen<br>Windows XP (embedded)<br>Yes | Touch screen<br>Windows XP (embedded)<br>Yes |
| Telemetry System<br>TIM Transceiver<br>Frequency<br>Repeater | 2U (horizontal rack mount)<br>4U (horizontal rack mount)<br>WMTS<br>Yes | 2U (horizontal rack mount)<br>4U (horizontal rack mount)<br>WMTS<br>Yes |
{9}------------------------------------------------
| Antenna | Radio Frequency | Radio Frequency |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telepack | Telepack-608(WMTS) | Telepack-608(WMTS) |
| Compatible Physiological<br>Monitors | Passport 12M/17M<br>Monitor(K143195)<br>Passport 8/12<br>Monitor(K132662)<br>T1(K143195)<br>Passport V<br>Monitor(K091834)<br>V Series Monitor(K132026)<br>DPM 6/7<br>Monitor(K092449)<br>Spectrum/ Spectrum OR<br>Bedside Monitor(K062098)<br>Passport 2 Bedside Monitor<br>(K020550) | Passport 12M/17M<br>Monitor(K143195)<br>Passport 8/12<br>Monitor(K132662)<br>T1(K143195)<br>Passport V<br>Monitor(K091834)<br>V Series Monitor(K132026)<br>DPM 6/7<br>Monitor(K092449)<br>Spectrum/ Spectrum OR<br>Bedside Monitor(K062098)<br>Passport 2 Bedside<br>Monitor (K020550) |
| ViewStation | Yes | Yes |
| WorkStation | Yes | Yes |
| Web Viewer | Yes | Yes |
| Gateway | Yes | Yes |
| Monitoring Parameters | ECG (Lethal and Non-<br>Lethal Arrhythmia<br>Detection, ST Segment<br>Analysis, QT Analysis and<br>heart rate(HR)), Invasive<br>Blood Pressure (IBP), Non-<br>Invasive Blood<br>Pressure(NIBP), Blood<br>Oxygenation (SpO2), Pulse<br>Rate (PR), Respiration Rate<br>(RESP), Respiration<br>Gases(02,CO2),<br>Temperature(TEMP),<br>Anesthetic Gases(AG),<br>Cardiac Output(C.O.),<br>Bispectral Index (BIS), and<br>Hemodynamics. | ECG (Lethal and Non-<br>Lethal Arrhythmia<br>Detection, ST Segment<br>Analysis, and heart<br>rate(HR)), Invasive Blood<br>Pressure (IBP), Non-<br>Invasive Blood<br>Pressure(NIBP), Blood<br>Oxygenation (SpO2), Pulse<br>Rate (PR), Respiration<br>Rate (RESP), Respiration<br>Gases(02,CO2),<br>Temperature(TEMP),<br>Anesthetic Gases(AG),<br>Cardiac Output(C.O.), and<br>Hemodynamics. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.