The Passport 17M and Passport 12M patient monitors are intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-mvasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sy()//Scy(), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: - . The arrhythmia detection. ST Segment analysis of Mortara algorithm. BIS, RM, CCO, SvOz/Scv02, and PAWP monitoring are intended for adult and pediatric patients only: - ST Segment analysis of Mindray algorithm is intended for adult patients only: ● - C.O. monitoring is restricted to adult patients only: ● - ICG monitoring is only for use on adult patients who meet the following ● requirements: height: 122 to 229cm, weight: 30 to 155kg. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use. The T1 Patient Monitor is intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpOz), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: The arrhythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients; ST Segment analysis of Mindray ECG algorithm is intended for adult patients only; The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
Device Story
Passport Series (17M, 12M, T1) are multiparameter patient monitors; inputs include ECG, respiration, temperature, SpO2, NIBP, IBP, CO2, and other specialized modules (BIS, ICG, AG). Devices process signals to display waveforms, numerics, and trends; provide arrhythmia detection, ST-segment analysis, and resting 12-lead ECG interpretation. Used in healthcare facilities by clinicians; T1 functions as a standalone monitor or a module for 12M/17M. Outputs assist clinicians in patient assessment and clinical decision-making. Modifications include wireless functionality, 12-lead ECG interpretation, and J-point auto-detection.
Clinical Evidence
Bench testing only. Functional and system-level testing performed to verify modifications (wireless modules, 12-lead ECG interpretation, J-point auto-detection). Conformance to consensus standards and FDA Class II Special Controls Guidance for Arrhythmia Detector and Alarm (2003) confirmed.
Technological Characteristics
Multiparameter monitors with integrated touchscreens (17", 12", 5"). Connectivity via wireless modules (ASUS, Silex, Cyberlink). Power: AC/DC or rechargeable Li-ion batteries. Parameters: ECG, SpO2, NIBP, IBP, CO2, etc. Software includes arrhythmia detection, ST-segment analysis (Mortara and Mindray algorithms), and 12-lead ECG interpretation. Data storage via Compact Flash or SD cards.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in healthcare facilities for monitoring physiological parameters (ECG, Resp, Temp, SpO2, PR, NIBP, IBP, PAWP, CO2, O2, AG, ICG, BIS, RM). Specific parameter restrictions apply based on patient age/size (e.g., C.O. adult only; ICG adult height/weight requirements). Not for home, ambulance, or helicopter use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
PASSPORT M SERIES PATIENT MONITORING (INCLUDING MODELS PASSPORT 17M AND PASSPORT 12M) (K132075)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2014
Shenzhen Mindray Bio-Medical Electronics Co., Ltd c/o Ms. Yanhong Bai Manager Regulatory Affairs, Technical Regulation Department Mindray Building, Keji 12th Road South Hi-tech Industrial Park, Nanshan Shenzhen, 518057 P.R. China
Re: K143195
> Trade/Device Name: Passport Series Patient Monitors (including Passport 17M, Passport 12M and T1) Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (With Arrhythmia Detection Or Alarms) Regulatory Class: Class II Product Code: MHX Dated: Undated Received: November 6, 2014
Dear Ms. Yanhong Bai,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K143195
Page 1
510(k) Number (if known): K143195
Device Name : Passport Series Patient Monitors (including Passport 17M, Passport 12M and T1)
#### Passport 12M and 17M Patient Monitors :
The Passport 17M and Passport 12M patient monitors are intended for monitoring. displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-mvasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sy()//Scy(), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
- . The arrhythmia detection. ST Segment analysis of Mortara algorithm. BIS, RM, CCO. SvOz/Scv02, and PAWP monitoring are intended for adult and pediatric patients only:
- ST Segment analysis of Mindray algorithm is intended for adult patients only; ●
- C.O. monitoring is restricted to adult patients only: ●
- ICG monitoring is only for use on adult patients who meet the following ● requirements: height: 122 to 229cm, weight: 30 to 155kg.
The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
### T1 Patient Monitor:
The T1 Patient Monitor is intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpOz), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.
{3}------------------------------------------------
#### K143195
Page __ 2__ of __ 2__
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
The arrhythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients;
ST Segment analysis of Mindray ECG algorithm is intended for adult patients only;
The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109)
Over-The-Counter Use
OR
{4}------------------------------------------------
#### 510(K) SUMMARY 1.
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Passport Series Patient Monitors is provided below.
| Device Common Name: | Patient Monitor |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | Passport Series Patient Monitors (including<br>Passport 17M, Passport 12M and T1) |
| Submitter: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680 |
| Contact: | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | November 4, 2014 |
| Classification Regulation: | 21 CFR 870.1025, Class II, Arrhythmia detector<br>and alarm (including ST-segment measurement<br>and alarm) |
| Panel: | Cardiovascular |
# Classification Regulation, Classification Name and Product Codes:
| Product<br>Code | Regulation<br>Number | Panel | Regulation Description | Device Common Name |
|-----------------|----------------------|----------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Primary | | | | |
| Product<br>Code | Regulation<br>Number | Panel | Regulation Description | Device Common Name |
| MHX | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including<br>ST-segment<br>measurement and alarm) | monitor, physiological,<br>patient (with arrhythmia<br>detection or alarms) |
| Secondary | | | | |
| DSI | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including<br>ST-segment<br>measurement and alarm) | detector and alarm,<br>arrhythmia |
| MLD | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including<br>ST-segment<br>measurement and alarm) | monitor, st segment with<br>alarm |
| DRT | 21 CFR<br>870.2300 | Cardiovascular | Cardiac Monitor<br>(including<br>cardiotachometer and<br>rate alarm) | monitor, cardiac (incl.<br>cardiotachometer & rate<br>alarm) |
| DXN | 21 CFR<br>870.1130 | Cardiovascular | Noninvasive blood<br>pressure measurement<br>system | system, measurement,<br>blood-pressure,<br>non-invasive |
| DSK | 21 CFR<br>870.1110 | Cardiovascular | Blood pressure<br>computer | computer, blood-pressure |
| FLL | 21 CFR<br>880.2910 | Cardiovascular | Clinical electronic<br>thermometer | thermometer, electronic,<br>clinical |
| DQA | 21 CFR<br>870.2700 | Cardiovascular | Oximeter | Oximeter |
| DPZ | 21 CFR<br>870.2710 | Cardiovascular | Ear oximeter | oximeter, ear |
| CCK | 21 CFR<br>868.1400 | Anesthesiology | Carbon dioxide gas<br>analyzer | analyzer, gas,<br>carbon-dioxide,<br>gaseous-phase |
| CBQ | 21 CFR<br>868.1500 | Anesthesiology | Enflurane gas analyzer | analyzer, gas, enflurane,<br>gaseous-phase (anesthetic<br>concentration) |
| Product<br>Code | Regulation<br>Number | Panel | Regulation Description | Device Common Name |
| CBS | 21 CFR<br>868.1620 | Anesthesiology | Halothane gas analyzer | analyzer, gas, halothane,<br>gaseous-phase (anesthetic<br>conc.) |
| CBR | 21 CFR<br>868.1700 | Anesthesiology | Nitrous oxide gas<br>analyzer | analyzer, gas,<br>nitrous-oxide, gaseous<br>phase (anesthetic conc.) |
| CCL | 21 CFR<br>868.1720 | Anesthesiology | Oxygen gas analyzer | analyzer, gas, oxygen,<br>gaseous-phase |
| DSB | 21 CFR<br>870.2770 | Cardiovascular | Impedance<br>plethysmograph | plethysmograph,<br>impedance |
| DXG | 21 CFR<br>870.1435 | Cardiovascular | Single-function,<br>preprogrammed<br>diagnostic computer | computer, diagnostic,<br>pre-programmed,<br>single-function |
| OLW | 21 CFR<br>882.1400 | Neurology | Electroencephalograph | index-generating<br>electroencephalograph<br>software |
{5}------------------------------------------------
{6}------------------------------------------------
### Predicate Device:
### K132075 - PASSPORT M SERIES PATIENT MONITORING (INCLUDING MODELS PASSPORT 17M AND PASSPORT 12M), Mindray North America.
### Indications for Use:
### Passport 12M and 17M Patient Monitors:
The Passport 17M and Passport 12M patient monitors are intended for monitoring, displaying, reviewing, alarming, and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (SvO2/ScvO2), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
{7}------------------------------------------------
- The arrhythmia detection, ST Segment analysis of Mortara algorithm, ● BIS, RM, CCO, SvO2/ScvO2, and PAWP monitoring are intended for adult and pediatric patients only;
- ST Segment analysis of Mindray algorithm is intended for adult patients ● only;
- C.O. monitoring is restricted to adult patients only;
- ICG monitoring is only for use on adult patients who meet the following ● requirements: height: 122 to 229cm, weight: 30 to 155kg.
The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
### T1 Patient Monitor:
The T1 Patient Monitor is intended for monitoring, displaying, reviewing, storing , alarming, and transferring of multiple physiological parameters including ECG (3-lead, or 5-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
The arrhythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients;
ST Segment analysis of Mindray ECG algorithm is intended for adult patients only;
The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
### Device Description:
The subject Passport Series Patient Monitors includes three monitors:
- . Passport 12M Patient Monitor
- Passport 17M Patient Monitor ●
- T1 Patient Monitor
All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters. The Passport 12M and 17M Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor. The T1
{8}------------------------------------------------
patient monitor is one of the available modules that can be plugged into the Passport 17M or 12M monitor. The T1 can also be used as a standalone monitor and when used as a standalone monitor, it provides a subset of the functions that are provided by Passport 17M and 12M.
Each of the three monitors that are the subject of this 510(k) are modified versions of previously cleared Mindray devices. The Passport 17M and Passport 12M were previously cleared in K132075, on April 18, 2014 and the T1 was previously cleared in K123074 on July 3, 2013.
### Performance Data:
- . To establish the substantial equivalence of the Passport Series Patient Monitors (including Passport 12M, Passport 17M and T1), Mindray conducted functional and system level testing on the subject devices. The testing provided an evaluation of the performance of the device relevant to each of the modifications to the subject devices since their previous clearance. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate.
- . In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
- Mindray has also followed the FDA Class II Special Controls Guidance . Document: Arrhythmia Detector and Alarm issues on October 28, 2003.
- . The T1 has been modified to provide wireless functionality. The existing wireless functionality of Passport 12M Monitor has been modified to include a new wireless module. The wireless capabilities of the Passport 17M remain unchanged. Mindray conducted Wireless functionality testing to ensure the performance of the new wireless modules meet specifications and are equivalent to the predicate device.
### Substantial Equivalence:
Comparison of Indications - Both the predicate device and the subject family of devices are multiparameter patient monitors intended to be used in healthcare facilities under the direction of clinical professionals. The indications for use of the subject T1 device are a subset of those for the predicate device, as the T1 device includes a subset of the functionality. The indications for use of the Passport 12M, Passport 17M and T1 also include the new feature of 12-lead ECG interpretation. Although this feature is not present in the predicate device, it is present in the cleared Mindray V Series Monitoring System (K132026), and thus does not constitute a new intended use for a multi-parameter monitor.
Comparison of Technological Characteristics - The table below compares the key technological feature of the subject devices to the primary predicate device (the Mindray Passport Series M K132075). The features in grey are the features that have been modified since their previous clearances and that are the subject of this 510(k).
{9}------------------------------------------------
| | Predicate Device (K132075) | | Subject Devices | | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-----------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Feature | 17M | 12M | 17M | 12M | T1 |
| Integrated<br>display and<br>touchscreen | 17"<br>1280*1024<br>pixels | 12" 800*600<br>pixels | Same | Same | 5", 480*272<br>pixels<br>(same as<br>K123074) |
| Secondary<br>display | Independent<br>control and<br>display | Display is<br>linked to<br>integrated<br>display | Same | Same | Independent<br>display and<br>control via a<br>VGA port in the<br>T1 docking<br>station |
| Additional<br>display<br>features | The minitrends diagram,<br>OxyCRG diagram, other<br>monitor view, and calculation<br>can be viewed when using an<br>external LCD screen | | Same | Same | Added |
| Wireless<br>module<br>(ASUS) | The ASUS module is used for<br>connecting to a network<br>wirelessly, constructing a<br>monitoring network with a<br>central monitoring system<br>(CMS). | | Same | Added support<br>for Silex<br>wireless module | Added wireless<br>function using<br>the Cyberlink<br>wireless<br>module |
| Module rack | Independent of the patient<br>monitor, provides 8 standard<br>module slots to extend the<br>measurement capabilities of<br>the system | | Same | Same | Not supported |
| Power supply | Two<br>rechargeable<br>Lithium-ion<br>battery or<br>AC power<br>supply | One<br>rechargeable<br>Lithium-ion<br>battery or AC<br>power supply | Same | Same | One<br>rechargeable<br>Lithium-ion<br>battery or DC<br>power supply or<br>AC power<br>supply<br>(same as<br>K123074) |
| Feature | Predicate Device (K132075) | | Subject Devices | | |
| | 17M | 12M | 17M | 12M | T1 |
| Battery | Chargeable Lithium-Ion, 11.1 VDC, 4500 mAh, 350 g | Same | Same | Same | Chargeable Lithium-Ion, 7.4 VDC, 2500 mAh (same as K123074) |
| External memory card | Compact Flash | Same | Same | Same | Secure Digital (same as K123074) |
| Data Recorder | The thermal recorder records patient information, measurement numerics, up to three waveforms, etc. | Same | Same | Same | Not supported |
| Speaker | Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and multi-level tone modulation | Same | Same | Same | Same |
| ECG | 3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, heart rate | Same | Same | Same | Same |
| 12-lead ECG interpretation | Not supported | Supported | Supported | Supported | Supported |
| ECG J-point auto detection | Supported | Same | Same | Same | Added |
| Respiration rate (Resp) | Measurement range: Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm. Accuracy: 7 to 150 rpm: $\pm$ 2 rpm or $\pm$ 2%, whichever is greater; 0 to 6 rpm: Not specified. | Same | Same | Same | Same |
| Temperature (Temp) | Measurement range: 0 to 50°C (32 to 122°F) Accuracy: $\pm$ 0.1°C or $\pm$ 0.2°F (without probe) | Same | Same | Same | Same |
| Feature | Predicate Device (K132075) | | Subject Devices | | |
| | 17M | 12M | 17M | 12M | T1 |
| Pulse oxygen<br>saturation<br>(SpO2) | Mindray SpO2 Module<br>Measurement range: 0 to 100%<br>Accuracy:70 to 100%: ±2%<br>(adult/pediatric mode)<br>70 to 100%: ±3% (neonate<br>mode)<br>0% to 69%: Not specified.<br>Masimo SpO2 Module<br>Measurement range: 1 to 100%<br>Accuracy:70 to 100%: ±2%<br>(measured without motion in<br>adult/pediatric mode)<br>70 to 100%: ±3% (measured<br>without motion in neonate<br>mode)<br>70 to 100%: ±3% (measured<br>with motion)<br>1% to 69%: Not specified.<br>Nellcor SpO2 Module<br>Measurement range: 0 to 100%<br>Accuracy:70 to 100%: ±2%<br>(adult/pediatric)<br>70 to 100%: ±3% (neonate)<br>0% to 69% Not specified | | Same | Same | Same |
## Device Comparison Table
{10}------------------------------------------------
{11}------------------------------------------------
K143195
Page 8 of 13
{12}------------------------------------------------
| | Predicate Device (K132075) | | Subject Devices | | |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Feature | 17M | 12M | 17M | 12M | T1 |
| Pulse rate<br>(PR) | PR from Mindray SpO2<br>Module<br>Measurement range: 20 to 254 bpm<br>Accuracy: $\pm$ 3 bpm<br><br>PR from Masimo SpO2<br>Module<br>Measurement range: 25 to 240 bpm<br>Accuracy: $\pm$ 3 bpm (measured without motion)<br>$\pm$ 5 bpm (measured with motion)<br><br>PR from Nellcor SpO2<br>Module<br>Measurement range: 20 to 300 bpm<br>Accuracy:20 to 250 bpm: $\pm$ 3 bpm<br>251 to 300 bpm, not specified<br><br>PR from IBP Module<br>Measurement range: 25 to 350 bpm<br>Accuracy: $\pm$ 1 bpm or $\pm$ 1%, whichever is greater" | | Same | Same | Same |
| | Predicate Device (K132075) | | Subject Devices | | |
| Feature | 17M | 12M | 17M | 12M | T1 |
| Non-invasive<br>blood<br>pressure<br>(NIBP) | Measurement range:<br>Adult<br>Pediatric<br>Systolic: 40 to 270<br>200<br>Diastolic: 10 to 210<br>150<br>Mean: 20 to 230<br>to 165<br>Accuracy:<br>Max mean error: ±5 mmHg<br>Max standard deviation: 8<br>mmHg | Measurement range:<br>Adult<br>Pediatric<br>Neonate<br>40 to<br>40 to 135<br>10 to<br>10 to 100<br>20<br>20 to 110 | Same | Same | Same |
| Invasive<br>blood<br>pressure<br>(IBP) | Measurement range: -50 to<br>300 mmHg<br>Accuracy:±2% or ±1 mmHg,<br>whichever is greater (without<br>sensor) | | Same | Same | Same |
| Pulse<br>Pressure<br>Variation<br>(PPV) | Supported feature of IBP | | Same | Same | Added |
| Cardiac<br>output (C.O.) | The temperature change is<br>displayed as a curve in the<br>C.O. split screen, and the<br>monitor calculates the C.O.<br>value from this curve. The<br>monitor is capable of storing 6<br>measurements. | | Same | Same | Not supported |
| Continuous<br>cardiac<br>output (CCO) | CCO/SvO2 interface module<br>is used to interface with<br>Edwards Vigilance II monitor<br>/ Vigileo Monitor which<br>measures continuous cardiac<br>output (CCO) and mixed<br>venous oxygen saturation<br>(SvO2). | | Same | Same | Not supported |
| Feature | Predicate Device (K132075) | | Subject Devices | | |
| | 17M | 12M | 17M | 12M | T1 |
| Mixed/central<br>venous<br>oxygen<br>saturation<br>(SvO2/ScvO2) | CCO/SvO2 interface module<br>is used to interface with<br>Edwards Vigilance II monitor<br>/ Vigileo Monitor which<br>measures mixed venous<br>oxygen saturation (SvO2) and<br>central venous oxygen<br>saturation (ScvO2). | | Same | Same | Not supported |
| Carbon<br>dioxide (CO2) | Sidestream CO2 Module:<br>Measurement range: 0 to 99<br>mmHg<br>Accuracy:<br>0 to 40 mmHg: ±2 mmHg<br>41 to 76 mmHg: ±5% of the<br>reading<br>77 to 99 mmHg: ±10% of the<br>reading | | Specifications<br>unchanged<br><br>Modules have<br>been<br>repackaged to<br>occupy a single<br>slot | Specifications<br>unchanged<br><br>Modules have<br>been<br>repackaged to<br>occupy a single<br>slot | Specifications<br>unchanged<br><br>Modules have<br>been<br>repackaged to<br>occupy a single<br>slot |
| | Microstream CO2 Module:<br>Measurement range: 0 to 99<br>mmHg<br>Accuracy:<br>0 to 38 mmHg: ±2 mmHg<br>39 to 99 mmHg: ±5% of the<br>reading ±5% of the<br>reading+0.08% of (the<br>reading-38) | | | | |
| | Mainstream CO2 Module:<br>Measurement range: 0 to 150<br>mmHg<br>Accuracy:<br>0 to 40 mmHg: ±2 mmHg<br>41 to 70 mmHg: ±5% of the<br>reading<br>71 to 100 mmHg: ±8% of the<br>reading<br>101 to 150 mmHg: ±10% of<br>the reading" | | | | |
| Feature | Predicate Device (K132075) | | Subject Devices | | |
| | 17M | 12M | 17M | 12M | T1 |
| Oxygen (O2) | Oxygen values are calculated based on user input and not directly measured. | | Same | Same | Not supported |
| Anesthetic gas (AG) | The AG module can identify two anesthetic gases in a mixture automatically and distinguish between them according to their contributions to the MAC value for display as the primary and secondary anesthetic agents. | | Same | Same | Not supported |
| Impedance cardiograph (ICG) | Measurement range:<br>SV: 5 to 250 ml<br>HR: 44 to 2m<br>C.O. 1.4 to 15 L/min<br><br>Accuracy:<br>SV: Not specified.<br>HR: ±2 bpm<br>C.O. Not specified." | | Same | Same | Not supported |
| Bispectral index (BIS) | Measured parameters: EEG, BIS, BIS L, BIS R | | Same | Same | Not supported |
| Respiration mechanics (RM) | Measurement range: Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm.<br>Accuracy: 7 to 150 rpm:±2 rpm or ±2%, whichever is greater;<br>0 to 6 rpm: Not specified." | | Same | Same | Not supported |
| 3 Additional SpO2 Sensors | None | | The model names of the three SPO2 sensors are LNCS NeoPt, LNCS Neo, LNCS Inf. These three SPO2 sensors are cleared in K132037. | | |
| 4 New NIBP cuffs | None | | The model names of the four NIBP cuffs are CM1306, CM1307, CM1506, CM1507. All this four NIBP cuffs are substantial equivalent to the NIBP cuffs which are cleared in K132075 and K092449. | | |
| Supports T1 as a module | Not supported | | Supported | Supported | N/A |
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
### Substantial Equivalence Conclusion:
Based on the detailed comparison of specifications for each of the modifications to the previously cleared Passport M Series devices, the performance testing and conformance with applicable standards, the Passport Series Patient Monitors (including Passport 12M, Passport 17M and T1) can be found substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.