K143757 · Anutra Medical, Inc. · FMF · Jan 8, 2015 · General Hospital
Device Facts
Record ID
K143757
Device Name
ANUTRA Feedback Aspiration Syringe
Applicant
Anutra Medical, Inc.
Product Code
FMF · General Hospital
Decision Date
Jan 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The ANUTRA Feedback Aspiration Syringe is intended for use by healthcare professionals for general purpose fluid aspiration/injection.
Device Story
The ANUTRA Feedback Aspiration Syringe is a 5mL piston-style control syringe used by healthcare professionals for fluid aspiration and injection. It consists of a polypropylene barrel, plunger rod, and synthetic polyisoprene rubber plunger. The device features a finger ring on the plunger rod and tab interfaces with the barrel to provide audible and tactile feedback during operation, allowing for one-handed use. The internal barrel surface is lubricated with polydimethylsiloxane. The device is single-use, non-pyrogenic, and sterilized via gamma irradiation. It connects to other devices via a 6% male Luer Lock. The device is used in clinical settings to facilitate controlled fluid delivery or aspiration, with the tactile/audible feedback mechanism assisting the clinician in monitoring the plunger position during the procedure.
Clinical Evidence
Bench testing only. No clinical data. Testing included ISO 7886-1 (cleanliness, lubricant, capacity, leakage, forces), ISO 594-2 (Luer fitting performance), sterile barrier packaging integrity (ASTM F2096, F88, F1886), and biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, hemocompatibility).
Technological Characteristics
Piston syringe; 5mL nominal volume; polypropylene copolymer barrel/plunger rod; polyisoprene rubber plunger; polydimethylsiloxane lubricant. Features 6% male Luer Lock. Sterilized by gamma irradiation. Meets ISO 7886-1 and ISO 594-2 standards. Mechanical feedback mechanism via plunger rod tabs.
Indications for Use
Indicated for healthcare professionals for general purpose fluid aspiration/injection.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem depicts a stylized human figure with three faces in profile, representing the department's focus on health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 8,2015
Anutra Medical Inc. c/o Cameron Perkins 1000 Perimeter Pike, Suite E Morrisville, NC 27560
Re: K143757
Trade/Device Name: Anutra Feedback Aspiration Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Product Code: FMF Regulatory Class: II Dated: December 18, 2014 Received: December 31, 2014
Dear Cameron Perkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)<br>K143757 | |
| Device Name | |
ANUTRA Feedback Aspiration Syringe
### Indications for Use (Describe)
The ANUTRA Feedback Aspiration Syringe is intended for use by healtheare professionals for general purpose fluid aspiration/injection.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
### This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EFF
{3}------------------------------------------------
### 5.0 510(k) Summary
#### 5.1. Submitter Information
| Company Name: | Anutra Medical, Inc. |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Address: | 1000 Perimeter Pk, Suite E Morrisville, NC 27560 USA |
| Company Phone: | 844-268-8721 |
| Company Fax: | 844-268-8721 |
| Contact Person: | Cameron Perkins<br>Anutra Medical, Inc.<br>Phone: 844-268-8721<br>Fax: 844-268-8721<br>Email: cameron@anutramedical.com |
| Prepared By: | Kristin Benokraitis<br>Director of Quality and Operations<br>Gilero, LLC<br>4022 Stirrup Creek Drive Suite 300<br>Durham, NC 27703<br>Phone: 919-595-8223<br>Fax: 919-595-8221<br>Email: kristinb@gilero.com |
| Date Summary<br>Prepared: | December 18, 2014 |
### Device Identification 5.2.
| Trade/Proprietary Name: | ANUTRA Feedback Aspiration Syringe |
|--------------------------|--------------------------------------------|
| Common Name: | Control Syringe |
| Classification Name: | Piston Syringe<br>21CFR 880.5860, Class II |
| Classification Panel | General Hospital |
| Product Code: | FMF (Syringe, piston) |
{4}------------------------------------------------
#### 5.3. Predicate Device
The ANUTRA Feedback Aspiration Syringe is substantially equivalent to the following predicate device:
| Device | Manufacturer | 510(k) | Date Cleared |
|--------------------------------------|----------------------------------|---------|--------------|
| BD Single Use,<br>Hypodermic Syringe | Becton, Dickinson and<br>Company | K110771 | May 13, 2011 |
#### 5.4. Device Description
The ANUTRA Feedback Aspiration Syringe is a piston syringe (control syringe) consisting of a plastic barrel which contains a printed graduated scale (5mL nominal volume) and 6% male Luer Lock connection, a plunger, and a plastic plunger rod. The internal surface of the ANUTRA Feedback Aspiration Syringe barrel is lubricated with polydimethylsiloxane (silicone). The plunger rod design contains a finger ring to allow for ease of use, and tab features with interface with the barrel to provide audible and tactile feedback to the user.
The ANUTRA Feedback Aspiration Syringe is single-use only, non-pyrogenic, and sterilized by gamma irradiation.
#### ર.ડ. Intended Use
The ANUTRA Feedback Aspiration Syringe is intended for use by healthcare professionals for general purpose fluid aspiration/injection.
#### 5.6. Predicate Device Comparison - Technical Characteristics
Equivalency of technical characteristics is demonstrated through a direct comparison of the ANUTRA Feedback Aspiration Syringe and the predicate device listed in the table below.
| Technical<br>Characteristic | Subject Device:<br>ANUTRA Feedback<br>Aspiration Syringe | Predicate Device:<br>BD Single Use, Hypodermic<br>Syringe (K110771) |
|-----------------------------|----------------------------------------------------------|---------------------------------------------------------------------|
| Barrel | Yes | Yes |
| Plunger Rod | Yes | Yes |
| Plunger | Yes | Yes |
| Luer Configuration | Luer Lock | 1mL – Luer Slip<br>3mL and 5mL – Luer Lock |
| Nominal Fluid<br>Volume | 5mL | 1mL, 3mL and 5mL |
| Lubrication | Polydimethylsiloxane<br>(silicone) | Unknown, device assumed to<br>be lubricated |
| Sterilization Method | Gamma irradiation | Ethylene oxide or gamma<br>irradiation |
{5}------------------------------------------------
The predicate device is a piston syringe. The ANUTRA Feedback Aspiration Syringe is a type of piston syringe (specifically, a control syringe). The intended use for the ANUTRA Feedback Aspiration Syringe is the same as its predicate device, the BD Single Use, Hypodermic Syringe (K110771).
## Barrel
The volume of fluid contained within the both the ANUTRA Feedback Aspiration Syringe and the predicate device is indicated by graduation marks printed on the outside of the barrel. The ANUTRA Feedback Aspiration Syringe has a plastic barrel with a 5mL nominal fluid volume. A 5mL variant of the predicate device contains the same nominal volume.
# Luer Configuration
Both the subject device and the predicate device are offered in a male Luer Lock configuration. The predicate device is also available in a Slip Luer configuration (1mL only).
## Nominal Fluid Volume
The nominal fluid volume for both the subject device and the predicate device is 5mL. The predicate device is also available in other fluid volumes.
## Lubrication
The internal surface of the ANUTRA Feedback Aspiration Syringe is coated with a polydimethylsiloxane (silicone). The predicate device 510(k) summary does not specify whether or not the predicate device is lubricated, however based on an inspection of the predicate device, it is believed to be lubricated.
# Plunger Rod & Plunger
Both the subject device and the predicate device have a plastic plunger rod and a synthetic rubber plunger; the plunger is attached to the plunger rod with snap-fit retention features. The ANUTRA Feedback Aspiration Syringe plunger rod contains features which provide audible and tactile feedback when the plunger is advanced or retracted, and a finger ring for ease of use and handling the device. The predicate device has a flat push-button on the end of its plunger rod. The mechanism for delivery of fluid is the same for both the subject device and predicate devices.
### Sterilization Method
The ANUTRA Feedback Aspiration Syringe is sterilized using gamma irradiation. The predicate device can be sterilized by either gamma irradiation or ethylene oxide.
{6}------------------------------------------------
## Materials
The ANUTRA Feedback Aspiration Syringe is constructed of the following polymeric components:
- . Barrel: Polypropylene copolymer
- Plunger Rod: Polypropylene copolymer ●
- Plunger: Synthetic (polyisoprene) rubber ●
- Lubrication: Polydimethylsiloxane (silicone) ●
The subject device has been tested and meets the biological requirements outlined in ISO 10993-1. A summary of these test results is provided in Section 15 -Biocompatibility.
#### 5.7. Predicate Device Comparison - Performance Characteristics
The performance data supplied with this submission demonstrates that the ANUTRA Feedback Aspiration Syringe meets the specified requirements and is substantially equivalent to the predicate device.
The predicate device (BD Single Use, Hypodermic Syringe) provided an overview of testing completed in the 510(k) Summary (K110771). These tests were used to compare the subject device and the predicate device performance.
Tests Performed on Subject Device and Predicate Device (as indicated in the predicate device 510(k) Summary):
- · ISO 7886-1
- Section 6 (Limits for Acidity or Alkalinity) o
- Section 7 (Limits for Extractable Metals) o
- Annex D Liquid leakage at syringe piston under compression O
- Annex G Forces to operate the plunger о
Tests Performed on Subject Device:
- ISO 7886-1 .
- o Section 5 (Cleanliness)
- Section 8 (Lubricant) o
- Section 9 (Tolerance on graduated capacity) O
- Section 10 (Graduated Scale) O
- Section 11 (Barrel) O
- Section 12 (Piston/plunger assembly); Annex B O
- Section 13 (Nozzle) O
- Section 14.1 (Dead Space) O
- Section 14.2 (Freedom from air and liquid leakage past piston); O Annex B and Annex D
- Section 15 (Packaging) о
- Section 16 (Labeling) O
{7}------------------------------------------------
- ISO 594-2
- o Section 3 (Dimensions and tolerances)
- Section 4.1 (Gauging) O
- Section 5.2 (Liquid leakage from fitting assembly under pressure) O
- Section 5.3 (Air leakage into fitting assembly during aspiration) O
- Section 5.4 (Separation force of fitting assembly) O
- Section 5.5 (Unscrewing torque of fitting assembly) O
- Section 5.6 (Ease of assembly) O
- Section 5.7 (Resistance to overriding) O
- Section 5.8 (Stress cracking) o
- Sterile Barrier Packaging Testing ●
- o ASTM F2096 11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- O ASTM F88-09 Standard Test Method for Seal Strength of Flexible Barrier Materials
- o ASTM F1886 / F1886M 09(2013) Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- . General performance
- Syringe package shall be easy to open with gloved hands o
- Syringe shall be able to aspirate and deliver using one hand O
- Tactile feedback O
- o Finger reach
- Biocompatibility Testing (ISO 10993) ●
- Cytotoxicity by Elution Test (Cytotoxicity) O
- Intracutaneous Reactivity (Irritation or Intracutaneous Reactivity) о
- Maximization Test for Delayed Hypersensivity (Sensitization) O
- Acute Systemic Toxicity (Systemic Toxicity (Acute)) O
- Evaluation of Hemocompatibility: Interaction with Blood O (Hemocompatibility/Hemolysis)
#### 5.8. Conclusion
Test results demonstrated that the ANUTRA Feedback Aspiration Syringe is as safe, as effective and performs as well as or better than the legally marketed predicate device (BD Single Use, Hypodermic Syringe).
Based on comparisons of the device's intended use, technology and performance characteristics, the ANUTRA Feedback Aspiration Syringe is substantially equivalent to the indicated predicate device. Any differences between the ANUTRA Feedback Aspiration Syringe and the predicate device have no significant influence on safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.