LifeCare PCA Infusion System

K143612 · Hospira, Inc. · MEA · Apr 8, 2016 · General Hospital

Device Facts

Record IDK143612
Device NameLifeCare PCA Infusion System
ApplicantHospira, Inc.
Product CodeMEA · General Hospital
Decision DateApr 8, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The LifeCare PCA™ infusion system is intended for intravenous or epidural administration of analgesic medications that are delivered by a continuous rate of infusion and/or with patient-controlled demand doses.

Device Story

Microprocessor-controlled, pole-mounted, electromechanical infusion pump; delivers analgesic medications intravenously or epidurally. Inputs: physician-programmed parameters; patient-controlled demand doses via pendant. Operation: stepper motor exerts pressure on drug vial (pre-filled or pharmacy-filled) to control infusion rate. Used in clinical settings; operated by healthcare professionals. Features wireless (802.11 a/b/g) and wired interfaces for bi-directional communication with Hospira MedNet™ software for drug library and auto-program transfers; can function standalone. Output: controlled analgesic delivery; visual/audible alarms; diagnostic data. Benefits: precise medication delivery; reduced medication errors via drug libraries; patient-controlled pain management.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including system verification and validation, human factors evaluations, biocompatibility testing per ISO 10993-1, and electrical/electromagnetic compatibility testing per ANSI/AAMI ES60601-1 and IEC standards.

Technological Characteristics

Pole-mounted electromechanical infusion pump; stepper motor actuation. Materials: modified half nut/lead screw. Connectivity: wired/wireless (802.11 a/b/g) to Hospira MedNet™ server. Standards: ANSI/AAMI ES60601-1:2005/(R)2012, IEC 60601-1-2:2007, IEC 60601-1-8:2012, IEC 60601-2-24:2012. Software: microprocessor-controlled with drug library and auto-programming capabilities. Sterilization: administration sets provided sterile (fluid path).

Indications for Use

Indicated for patients requiring intravenous or epidural administration of analgesic medications via continuous infusion or patient-controlled demand doses.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other, with a wing-like shape extending above them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 8, 2016 Hospira, Inc Ms. Yuliya Matlin Associate Director, Global Regulatory Affairs 275 North Field Drive Lake Forest, Illinois 60045 Re: K143612 Trade/Device Name: Lifecare PCATM Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEA, FPA Dated: March 4, 2016 Received: March 10, 2016 Dear Ms. Matlin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang - for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K143612 Device Name LifeCare™ PCA Infusion System Indications for Use (Describe) The LifeCare PCA™ infusion system is intended for intravenous or epidural administration of analgesic medications that are delivered by a continuous rate of infusion and/or with patient-controlled demand doses. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Hospira. The logo consists of the word "Hospira" in a sans-serif font, with a stylized graphic above it. The graphic features three curved lines that converge at the top, resembling a stylized flower or abstract design. The logo is presented in grayscale. ## 510(k) SUMMARY A summary of 510(k) information in accordance with the requirements of 21 CFR 807.92 is provided below for K143612. | Submitter Information | | |-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | Hospira, Inc. | | Address | 275 North Field Dr., Lake Forest, IL 60045 | | Phone number | 858-391-1142 | | Fax number | 224-212-5401 | | Establishment<br>Registration Number | 3005579246 | | Name of contact<br>person | Yuliya Matlin | | Date prepared | April 5, 2016 | | 510(k) Number | K143612 | | Name of device | | | Trade or proprietary<br>name | LifeCare PCA™ Infusion System | | Common or usual<br>name | Patient Controlled Analgesia (PCA) Infusion Pump and<br>Administration Sets | | Classification name | Infusion Pump | | Classification panel | General Hospital | | Regulation | 21 CFR 880.5725 (Infusion Pump) | | Product Code(s) | MEA - Patient Controlled Analgesia (PCA) Infusion Pump<br>FPA - Administration Sets | | Legally marketed<br>device(s) to which<br>equivalence is<br>claimed | Hospira LifeCare™ PCA3 Infusion System, cleared under K043256<br>Hospira LifeCare PCA™ Infusion Pump System with Hospira<br>MedNet™ Software cleared under K042800<br>No Reference Devices were used in this submission | | Device description | The Hospira LifeCare PCA™ Infusion System is a microprocessor controlled, pole mounted, electromechanical infusion pump that allows a patient to self-administer analgesic using an attached patient pendant, within physician prescribed, programmed parameters. The Hospira LifeCare PCA™ Infusion System also includes a sterile empty vial that can be pharmacy-filled with appropriate medications, and dedicated administration sets to administer the medication.<br>The infuser is powered from an AC power source and has an internal battery to maintain operation for short periods when an AC power source is not available. A stepper motor exerts pressure on an inserted drug vial to control the infusion of analgesic into a patient.<br>The infuser will accept either a pre-filled drug vial manufactured by Hospira or a sterile empty vial that can be filled by a hospital's pharmacy. Administration Sets compatible for use in the LifeCare PCA™ Infuser are provided sterile (fluid path) and intended for single-use application.<br>The infuser is equipped with wireless (802.11 a/b/g) and wired interfaces to Hospira MedNet™ Software. Drug Libraries and Auto-Programs are transferred using these interfaces. The subject device bi-directionally communicates to the hospital information system through the Hospira MedNet™ Server, which is optional software, and the device can also be programmed and used without the Hospira MedNet™ software. | | Indications for Use | The LifeCare PCA™ infusion system is intended for intravenous or epidural administration of analgesic medications that are delivered by a continuous rate of infusion and/or with patient-controlled demand doses. | | Summary of Intended<br>Use Compared to<br>Predicate Device | The indication for use is nearly identical to the predicate. The difference in the intended use statements of the predicate and subject device clarifies that the subject device may deliver continuous or patient controlled administration of analgesic medications via intravenous or epidural routes of administration. This change in wording does not alter the function of the device in comparison to the predicate and is intended to provide clarity to the device's functionality. | | Summary of<br>Technological<br>Characteristics<br>Compared to Predicate<br>Device | The LifeCare PCA™ Infusion System with Hospira MedNet™ employs the same fundamental scientific technology, and principles of operation and intended/indications for use as are used by the predicate device(s). In addition, the LifeCare PCA™ Infusion System has the following technological characteristics comparable to predicate device(s) which previously received 510(k) clearance:<br>• Utilize the same pumping technology; deliver fluids over the same flow rates with same accuracy.<br>• Have similar drug library features and wired and wireless communication options using a facility's network infrastructure.<br>• Incorporates the same basic Infusion System design with the exception of the design updates described in this 510(k).<br>• Incorporate the same basic software driven electronic control system<br>• Use comparable dedicated administration sets. | | Summary of<br>Technological<br>Characteristics<br>Compared to Predicate<br>Device (Continued) | Modifications to the software, hardware, labeling, and PCA consumables as<br>compared to the predicate device include:<br><br>Software: Changes to the user interface to enhance auto-programming<br>workflow. Alarm risk prioritization (low, medium, high) and alarm visual<br>indicator updates to comply with IEC 60601-1 Edition 3. Addition of wireless<br>strength and battery capacity indicators. Updates to the Biomed capabilities<br>to include more diagnostic information and usage counters. Improvements<br>to the wireless connectivity to include global region support, enhanced<br>roaming, and to update wireless system components. Implement<br>cybersecurity enhancements.<br><br>Hardware: Miscellaneous changes to hardware to comply with IEC60601-1<br>Edition 3. Strengthened door and latch design. Material change to half nut/lead<br>screw. Added seal around barcode reader window to prevent fluid ingress.<br><br>Labeling: Updates to the System Operating Manual, Technical Service<br>Manual and on pump label to align with IEC 60601-1 Edition 3 and to<br>incorporate updated workflow for auto-programming. Updates to Sterile<br>Empty vial providing guidelines for barcode label placement after filling with<br>medication.<br><br>PCA Consumables: modified distal male luer adaptor design on all<br>administration sets. Increased activation pressure for the Pressure Activated<br>Valve (PAV). Changed slide clamp color to orange. | | Performance Data - Non<br>Clinical Testing | An assurance case was provided for the LifeCare PCA™ Infusion System as<br>recommended in the FDA guidance document, Infusion Pumps Total Product<br>Life Cycle.<br><br>The stated goal of the assurance case is:<br>• The LifeCare PCA™ infusion system design is adequately safe for its<br>intended use.<br><br>The assurance case defined the device system, including the indications for<br>use, system definition, operational description, patient populations, and use<br>environments. The supporting assurance arguments covered the following<br>attributes:<br>• Residual risks are analyzed and determined to be acceptably low<br>• Design is verified and valid for its indications for use as it relates to<br>safety<br>• Device design is reliable | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "Hospira" in a serif font, with the "H" capitalized and the rest of the letters in lowercase. Above the word, there is a logo consisting of three curved lines that resemble an abstract representation of a medical device or a stylized wave. A horizontal line is present below the word "Hospira". {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the Hospira logo. The logo features the company name in a serif font, with the "H" slightly larger than the other letters. Above the name is an abstract graphic consisting of three curved lines that converge at the top, resembling a stylized arch or a stylized representation of motion. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a simple, sans-serif font, with the letters in gray. Above the name is an abstract graphic consisting of three curved lines that converge at a point above the "i" in Hospira, creating a sense of movement or connectivity. | Performance Data – Non<br>Clinical Testing<br>(Continued) | The following evidence was included in the assurance case: | | |-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------| | | Performance Test Summary<br>System verification and validation activities for LifeCare PCA™ Infusion<br>System confirmed that the system meets user needs and design<br>inputs. All the testing met the acceptance criteria.<br><br>Human factors evaluations have been conducted to validate the<br>effectiveness of use error related mitigations<br><br>Software documentation as recommended in the following FDA<br>guidance documents:<br>Infusion Pumps Total Product Life Cycle Content of Premarket Submissions for Software Contained in<br>Medical Devices Content of Premarket Submissions for Management of<br>Cybersecurity in Medical Devices Biocompatibility evaluations of consumable products were conducted in<br>accordance with ISO 10993-1:2009 (Corrigendum 2010) FDA Draft Guidance for Industry and Food and Drug Administration<br>Staff Use of International Standard ISO-10993, "Biological<br>Evaluation of Medical Devices Part 1: Evaluation and Testing"<br>issued April 2013 Electrical and Electromagnetic Compatibility testing were conducted. The<br>Infusion System complies with following standards: | | | | Basic Safety and Essential<br>Performance | ANSI/AAMI ES60601-1:2005/(R) 2012 and<br>A1:2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012, | | | Electromagnetic<br>Compatibility | IEC Standard 60601-1-2:2007 3rd Ed. | | | Alarm Systems | IEC 60601-1-8, Edition 2.1 2012-11 | | | Infusion Pump Basic<br>Safety and Essential<br>Performance | IEC 60601-2-24 Medical Electrical Equipment – Part-<br>Edition 2.0 2012-10 | | Performance Data –<br>Clinical Testing | Clinical evaluation is not required for this submission to support substantial<br>equivalence. Human Factors studies have been conducted on subject<br>devices, demonstrating passing results. | | ## CONCLUSIONS DRAWN FROM CLINICAL AND NON-CLINICAL DATA The conclusions drawn from the nonclinical tests demonstrate that the device is substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...