HOSPIRA LIFECARE INFUSION SYSTEM, MODEL 12384-04-07; LIFECARE PCA INFUSION SYSTEM STERILE EMPY VIALS, MODEL 6021-04-03
K043256 · Hospira, Inc. · MEA · Dec 6, 2004 · General Hospital
Device Facts
Record ID
K043256
Device Name
HOSPIRA LIFECARE INFUSION SYSTEM, MODEL 12384-04-07; LIFECARE PCA INFUSION SYSTEM STERILE EMPY VIALS, MODEL 6021-04-03
Applicant
Hospira, Inc.
Product Code
MEA · General Hospital
Decision Date
Dec 6, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hospira LifeCare® PCA3 Infusion System is intended for accurate, volumetric, infusion of analgesic drugs by continuous or patient-demanded (PCA) intravenous administration. The LifeCare PCA® Infusion System is also indicated for short-term (less than 96 hours) continuous epidural administration of analgesic drugs.
Device Story
Hospira LifeCare® PCA3 Infusion System is a pole-mounted, microprocessor-controlled, electromechanical infusion pump. Device uses a stepper motor to drive a plunger against a pre-filled or pharmacy-filled drug vial, delivering analgesic medication into a sterile, dedicated administration set. Operated in hospital settings by clinicians; patients use a pendant to self-administer doses within physician-programmed parameters. Powered by AC source with internal battery backup. Output is controlled volumetric infusion of analgesics. Device lacks wireless or network capabilities and is not compatible with MedNet™ software. System provides controlled delivery to manage patient pain, ensuring accuracy through programmed limits.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics and performance specifications to predicate devices.
Technological Characteristics
Microprocessor-controlled electromechanical infusion pump; stepper motor actuation; pole-mounted form factor; AC power with internal battery backup; uses sterile, dedicated, non-pyrogenic, latex-free administration sets. No wireless or network connectivity. Software is embedded firmware.
Indications for Use
Indicated for patients requiring continuous or patient-controlled analgesic (PCA) intravenous infusion, or short-term (less than 96 hours) continuous epidural analgesic administration in hospital clinical settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
LifeCare PCA® Infusion System with Hospira MedNet™ Software (K042800)
{0}------------------------------------------------
DEC - 6 2004
1043256
Confidential
# 510(K) Summary
1. Name of Submitter:
Hospira, Incorporated 275 North Field Drive Lake Forest, Illinois 60045
Owner/Operator # 9063339
### 2. Manufacturer and Establishment Registration Number:
Hospira, Inc. - Morgan Hill 755 Jarvis Drive Morgan Hill, CA 95037
Establishment Registration # 2921482
- 3. Proprietary or Trade Name of Proposed Device: Hospira LifeCare® PCA3 Infusion System
- 4. Common Name: Infusion Pump
5. Device Classification, Pancode and ProCode: Class II, 80-FRN (Infusion Pump) Class II, 80-FPA (Administration Sets)
- 6. Performance Standards: No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for intravenous infusion pumps. Infusion pumps are listed in 21 CFR 880.5725.
- 7. Intended Use: The Hospira LifeCare® PCA3 Infusion System is intended for accurate, volumetric, infusion of analgesic drugs by continuous or patient-demanded (PCA) intravenous administration.
#### 8. Indications for Use:
The LifeCare® PCA3 Infusion System is also indicated for short-term (less than 96 hours) continuous epidural administration of analgesic drugs. The epidural route can be used to provide anestbessa or analgesia. The pump must be used with sterile, dedicated, LifeCare PCA administrations sufer
#### 9. Proposed Device Description:
The Hospira LifeCare® PCA3 Infusion System is a microprocessor controlled, pole mounted, standalone, electromechanical infusion pump that allows a patient to self administer, analgeric usines, a patient pendant, within physician prescribed, programmed parameters. A stepper motor exerts pressure on an inserted drug vial to control the infusion of analgesic into a patient. The infuser is powered from an AC power source and has an internal battery to maintain operation for short periods when an AC power source is not available. The infuser will accept a pre-filled drug vial manufactured by Haspire and includes one sterile empty vial that can be filled by a hospital's pharmacy.
As of May 03, 2004, both the infuser and the dedicated LifeCare® PCA administration sets are manufactured and distributed by Hospira Incorporated, formerly the Hospital Products Division of Abbott Laboratories.
> Volume 1 of 1
{1}------------------------------------------------
## 10. Predicate Device Information:
Infusion pumps cleared for commercial distribution and determined to be appropriate for use as predicates are summarized in the following table.
| 510(k) # | Product Name | Clearance Date |
|----------|----------------------------------------------------------------|----------------|
| K042800 | LifeCare PCA® Infusion System with Hospira MedNet™<br>Software | 10/28/2004 |
| K022203 | Abbott LifeCare® PCA3 infusion System | 08/01/2002 |
## 11. Comparison to Legally Marketed Device(s)
| Factors | Subject Device(s)<br>Hospira LifeCare® PCA3 Infusion System | Predicate Device(s)<br>Abbott LifeCare® PCA3<br>Infusion System<br>and<br>LifeCare® PCA Infusion<br>System with Hospira™<br>MedNet Software |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended for volumetric, infusion of analgesic drugs by<br>continuous or patient-demanded (PCA) intravenous<br>administration. It is indicated for short-term (less than<br>96 hours) continuous epidural administration of<br>analgesic drugs. | Same |
| Indications for Use | Hospital clinical settings using sterile, dedicated,<br>LifeCare® PCA administration sets. | Same |
| Operating Principle | Advancement of a syringe-like drug vial by a stepper<br>motor driven plunger to expel drug from the vial into an<br>attached sterile intravenous administration set designed to<br>be used exclusively with LifeCare® PCA infusers. | Same |
| Administration Sets and<br>Fluid Contact Materials | Sterile, dedicated, non-pyrogenic, latex-free LifeCare®<br>PCA administration sets. | Same |
| Physical Features | Materials, Size, Weight, Power Sources, Battery Type,<br>Power Cord | Same |
| Environmental Features | Operating Temperature, Storage Temperature, Relative<br>Humidity, Pressure | Same |
| Performance Features | Delivery Rates, Dose Units, Delivery Accuracy,<br>Delivery Modes, Occlusion Pressure Limits, Alarm<br>Types and Conditions | Same |
| Wireless MedNet™<br>Software Compatibility | No wireless/network capability.<br>Not compatible with MedNet™ Software. | Same as Abbott<br>LifeCare® PCA3<br>Infusion System |
{2}------------------------------------------------
### 12. Statement of Safety and Effectiveness
The Hospira LifeCare® PCA3 Infusion System meets the functional claims and intended use as described in the product labeling. The proposed modifications do not raise new issues of safety and/or effectiveness.
### 13. Statement of Substantial Equivalence:
The Hospira LifeCare® PCA3 infusion System is substantially equivalent in terms of safety and effectiveness to the predicate Abbott LifeCare® PCA3 Infusion System and the LifeCare PCA® Infusion System with Hospira MedNet™ Software based on the following characteristics.
- 1) Same intended use,
- 2) Same indication for use,
- 3) Same fundamental technology and operating principle.
- 4) Same physical, operational, environmental and performance attributes, and
- 5) Same materials of construction for infuser components and administration sets.
The claim for substantial equivalence is supported by the information provided in this 510(k) submission.
Prepared and submitted by:
Patricia Meleski 11/23/04
Patricia Melerski Manager, Global Device Regulatory Affairs Hospira, Inc. 275 North Field Drive Lake Forest, IL 60045 Phone: 224/212-4880 Fax: 224/212-5401
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 6 2004
Ms. Patricia Melerski Manager, Global Device Regulatory Affairs Hospira, Incorporated 275 North Field Drive, Dept. 389, Bldg. H2 Lake Forest, Illinois 60045
Re: K043256
Trade/Device Name: Hospira LifeCare® PCA3 Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEA Dated: November 23, 2004 Received: November 24, 2004
Dear Ms. Melerski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Melerski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Suxen Runner
(Chin Lin, Ph.D.
(Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
:
:
:
| 510(k) Number (if known) | K043256 |
|--------------------------|----------------------------------------|
| Device Name: | Hospira LifeCare® PCA3 Infusion System |
Indications for Use
The Hospira LifeCare® PCA3 Infusion System is indicated for accurate, volumetric, infusion of analgesic drugs by continuous or patient-demanded (PCA) intravenous administration. The LifeCare PCA® Infusion System is also indicated for short-term (less than 96 hours) continuous epidural administration of analgesic drugs.
| Prescription Use _X_<br>(Part 21 801 Subpart D) | AND/OR | Over-The_Counter Use<br>(Part 21 CFR 807 Subpart C) |
|-------------------------------------------------|--------|-----------------------------------------------------|
|-------------------------------------------------|--------|-----------------------------------------------------|
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
| 510(k) Number | K043256 |
|---------------|---------|
|---------------|---------|
Volume 1 of 1
·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.