The NovaCross Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
Device Story
NovaCross Microcatheter is a sterile, single-use, over-the-wire percutaneous support catheter. It consists of a telescopic shaft, an over-tube, and a proximal handle for manual manipulation and flushing. A key feature is a deployable/retractable distal Nitinol scaffold, visible under fluoroscopy, which expands to the artery width to provide distal anchoring and guidewire support. Used in coronary and peripheral vasculature by physicians; facilitates access for therapeutic devices like atherectomy catheters, PTCA catheters, or stents. The device does not provide therapeutic benefit itself; it serves as a support tool to improve guidewire positioning. Benefits include enhanced support near the treatment site, facilitating subsequent interventional procedures.
Clinical Evidence
Bench testing included biocompatibility (ISO 10993-1), sterilization (ISO 11135-1), and mechanical testing (torque, tensile, pushability, scaffold durability, radial force). A GLP animal study (n=6 pigs) evaluated safety in coronary and peripheral arteries. Endpoints included vessel integrity, gross pathology, and histopathology at 1 and 14 days post-procedure. Results showed no discernible pathological adverse effects and successful achievement of all performance criteria.
Technological Characteristics
Single-lumen, over-the-wire percutaneous catheter. Features a deployable Nitinol scaffold for anchoring. Compatible with 7F guiding catheters and 0.014" guidewires. Working length 135cm. Materials evaluated per ISO 10993-1. Sterilization per ISO 11135-1. No software or electronic components.
Indications for Use
Indicated for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange in patients requiring such access.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three human profiles facing to the right, stacked on top of each other, with a flowing fabric-like element below them.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 4, 2015
Nitiloop Ltd. % Michael Daniel President, Daniel & Daniel Consulting, LLC Daniel & Daniel Consulting 340 Jones Lane Gardnerville. Nevada 89460
Re: K143608
> Trade/Device Name: NovaCross Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: April 10, 2015 Received: April 10, 2015
Dear Michael Daniel,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143608
Device Name NovaCross Microcatheter, 1
Indications for Use (Describe)
The NovaCross Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| <div> <span style="display:inline-block;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="display:inline-block;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services | |
| Food and Drug Administration | |
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional Premarket Notification Submission - 510(k) NovaCross™ Microcatheter 510(k) Number K143608
#### I. SUBMITTER
Nitiloop Ltd. 8 Hamada St. P.O.Box 12892 Herzliya, Israel 46733 Tel: +972-52-396-8925 Fax: +972-72-2501631
#### Contact Person
Michael A. Daniel Daniel & Daniel Consulting, LLC 340 Jones Lane Gardnerville, NV 89460 Tel: 775.392.2970 Cell: 415.407.0223 Fax: 610.545.0799 madaniel@clinregconsult.com
#### II. DEVICE
Name of Device: NovaCross™ Microcatheter Common or Usual Name: NovaCross™ Microcatheter Classification Name: Percutaneous catheter (21 CFR 870.1250) Regulatory Class: II Product Code: DQY
#### III. PREDICATE DEVICE
The primary predicate device is the Roxwood Medical, Inc. MultiCross Support Catheter cleared under K121763, (product code DQY, Regulation No. 870.1250).
A reference device was used in this submission: Corsair Microcatheter (Asahi Intecc Co., Ltd. K083127).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for "Nitiloop CROSSING THE LIMITS". The logo features a red square with a white design on the left side. To the right of the square is the word "Nitiloop" in a bold, sans-serif font. Below "Nitiloop" is the phrase "CROSSING THE LIMITS" in a smaller, lighter font.
## IV. DEVICE DESCRIPTION
The NovaCross™ Microcatheter is a sterile, single-use, single lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange.
The NovaCross™ Microcatheter consists of telescopic shaft. Over Tube, and a proximal Handle Body that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol Scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site.
Subsequent to conventional guidewire placement, therapeutic devices such as atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The NovaCross™ Microcatheter by itself does not provide therapeutic benefit beyond simple facilitation of guidewire support. The NovaCross™ Microcatheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate devices.
### V. INDICATIONS FOR USE
The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both the MultiCross, and the NovaCross™ Microcatheters have a working length of 135cm and are both compatible with a 7F guiding catheter, as well as a 0.014" guidewire. Like the MultiCross Nitinol scaffold, the NovaCross™ Nitinol scaffold is also expanded to the diameter of the artery when deployed.
The two devices use similar technology including a delivery system. The guidewire support with the two devices is accomplished by extending a Nitinol scaffold. The Nitinol element for the NovaCross™, like that of the MultiCross predicate, provides distal anchoring and support, thus exhibiting similar safety questions. These safety questions have been evaluated for the NovaCross™ Microcatheter through extensive design verification and validation testing, including a GLP Animal Study.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for "Nitiloop". The logo consists of a red square with a white curved line inside, followed by the word "Nitiloop" in a bold, sans-serif font. Below the word "Nitiloop" is the phrase "CROSSING THE LIMITS" in a smaller, sans-serif font.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
#### Biocompatibility testing
An evaluation of biocompatibility was performed in compliance with ISO 10993-1. Biocompatibility testing included cytotoxicity, irritation, acute systemic toxicity, sensitization, and hemocompatibility testing. All tests were successfully completed.
### Sterilization, Packaging and Shelf Life Testing
Sterilization validation testing was performed to demonstrate compliance with ISO 11135-1. In addition, shelf life and packaging testing were performed to support the labeled shelf life. All tests were successfully completed.
#### Mechanical Testing
Mechanical bench testing included the following:
- Dimensional
- Simulated use ●
- Tip flexibility ●
- Tip and shaft durability
- Torque response
- Torque strength
- Tensile strength
- Hydrophilic coating integrity
- Corrosion
- Fluid leak
- Air leakage
- Pushability and retractability ●
- Scaffold durability
- Radial force
All tests met the predefined acceptance criteria.
### Animal Study
A GLP animal study, consisting of 6 pigs, was used to test whether the NovaCross™ device could be used safely and effectively as a tool to provide support to the guidewire. The system was used on three arteries (two coronary and one peripheral) in each animal. The safety of the system was evaluated 1 and 14 days post procedure in terms of the integrity of the treated vessels (compared to intact control segments), gross pathology and histopathology. The results demonstrated that no discernible pathological adverse effects were detected, in terms of local and systemic effects
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "Nitiloop". The logo consists of a red square with a white design inside, followed by the word "Nitiloop" in black letters. Below the word "Nitiloop" is the phrase "CROSSING THE LIMITS" in smaller, gray letters.
following the deployment of the Test Item NovaCross™ Microcatheter in coronary and peripheral arteries. The Test Item successfully achieved all performance criteria defined in the study protocol. Pathology and histopathology analyses confirmed the results of the study.
## VIII. CONCLUSIONS
The NovaCross™ Microcatheter was determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.