MULTICROSS SUPPORT CATHETER

K121763 · Roxwood Medical, Inc. · DQY · Nov 27, 2012 · Cardiovascular

Device Facts

Record IDK121763
Device NameMULTICROSS SUPPORT CATHETER
ApplicantRoxwood Medical, Inc.
Product CodeDQY · Cardiovascular
Decision DateNov 27, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The MultiCross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Device Story

MultiCross Support Catheter is a sterile, single-use, triple-lumen, over-the-wire percutaneous catheter. Device features inner shaft, outer sheath, proximal handle for manual manipulation, and a temporarily deployable/retractable distal Nitinol scaffold. Scaffold is fluoroscopically visible; expands to arterial width to provide distal anchoring and support for guidewire advancement. Used in coronary and peripheral vasculature; operated by clinicians. Device facilitates guidewire placement to enable subsequent use of therapeutic devices like atherectomy catheters, PTCA catheters, or stents. Does not provide therapeutic benefit itself; serves as support tool for interventional procedures.

Clinical Evidence

No human clinical data. Evidence consists of extensive bench testing (visual inspection, dimensions, particulate, radial force, torque, flexibility, kink resistance, leak, tensile strength, corrosion, coating integrity, sterilization, packaging/shelf-life) and a GLP animal study. Animal study evaluated safety, performance, and handling in coronary and peripheral vasculature; pathology/histopathology results met safety acceptance criteria with no untoward observations.

Technological Characteristics

Triple-lumen, over-the-wire percutaneous catheter. Features a distal Nitinol scaffold for anchoring. 7F guide catheter compatibility; 135 cm working length; 0.014" guidewire compatible. Sterile, single-use. No software or electronic components.

Indications for Use

Indicated for patients requiring access to discrete regions of the coronary and peripheral vasculature for guidewire exchange, used in conjunction with a steerable guidewire.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY K121763 # NOV 2 7 2012 This summary of the 510(k) premarket notification for the MultiCross Support Catheter is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR8807.92. # GENERAL INFORMATION Applicant: Roxwood Medical, Inc. 150 Executive Park Boulevard, Suite 3700 San Francisco, CA 94134 U.S.A. Phone: 408-373-4751 Fax: 415-467-1253 # Contact Person: Mehrdad Farhangnia Chief Executive Officer Roxwood Medical, Inc. 150 Executive Park Boulevard, Suite 3700 San Francisco, CA 94134 U.S.A. Phone: 408-373-4751 Fax: 415-467-1253 # Date Prepared: June 13, 2012 #### DEVICE INFORMATION The MultiCross Support Catheter is a percutaneous catheter for use in the coronary and peripheral vasculature. Trade Name: MultiCross Support Catheter Generic/Common Name: Percutaneous Catheter Classification: 21 CFR 870.1250, Class II Product Code: DQY {1}------------------------------------------------ MULTICROSS SUPPORT CAT 510(k) PREMARKET NOTIFICATION # PREDICATE DEVICE(S) - Vascular Solutions Twin-Pass™ OTW Dual Access Catheter (K072618) � - EndoCross ENABLER-P Catheter System (K092174) � - BridgePoint Medical Stingray Orienting Balloon Catheter (K080987) . # INDICATIONS FOR USE The MultiCross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange. ## PRODUCT DESCRIPTION The MultiCross Support Catheter is a sterile, single-use, triple lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange. The MultiCross Support Catheter consists of an inner shaft, outer sheath, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site. Subsequent to conventional guidewire placement, therapeutic devices such as atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The MultiCross Support Catheter by itself does not provide therapeutic benefit beyond simple facilitation of guidewire support. The MultiCross Support Catheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate devices. # TECHNOLOGICAL CHARACTERISTICS The technological characteristics of the MultiCross Support Catheter are similar to the predicate devices. Performance data were provided to support the determination of substantial equivalence. | Feature | MultiCross Support<br>Catheter<br>(Subject Device) | Vascular Solutions<br>Twin-PassTM OTW Dual<br>Access Catheter | EndoCross<br>ENABLER-P<br>Catheter System | BridgePoint Medical<br>Stingray Orienting<br>Balloon Catheter | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K_____ | K052257 | K092174 | K080987 | | Indications for<br>Use | The MultiCross<br>Support Catheter is<br>intended to be used in<br>conjunction with a<br>steerable guidewire to<br>access discrete regions<br>of the coronary and<br>peripheral vasculature<br>and for guidewire | The Twin-Pass Dual<br>Access Catheter is to be<br>used in conjunction with<br>steerable guidewires in<br>order to access discrete<br>regions of the coronary<br>and peripheral arterial<br>vasculature, to facilitate<br>placement of guidewires<br>and other interventional | The ENABLER-P<br>Catheter System is<br>intended to be used<br>in conjunction with a<br>steerable guidewire<br>to access discrete<br>regions of the<br>peripheral<br>vasculature and for | The BridgePoint<br>Medical StingrayTM<br>Orienting Balloon<br>Catheters are indicated<br>for directing, steering,<br>controlling, and<br>supporting a<br>guidewire in order to<br>access discrete regions | {2}------------------------------------------------ ROXWOOD MEDICAL, INC. # MULTICROSS SUPPORT CATHETER 510(k) PREMARKET NOTIFICATION | Feature | MultiCross Support<br>Catheter<br>(Subject Device)<br>exchange. | Vascular Solutions<br>Twin-Pass™ OTW Dual<br>Access Catheter<br>devices, and for use<br>during two guidewire<br>procedures. | EndoCross<br>ENABLER-P<br>Catheter System<br>guidewire exchange. | BridgePoint Medical<br>Stingray Orienting<br>Balloon Catheter<br>of the coronary and<br>peripheral vasculature. | |---------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | Guide Catheter<br>Compatibility | 7F | 5F | 7F | 6F | | Working<br>Length | 135 cm | 140 cm | 107 cm | 135cm | | Guidewire<br>Compatibility | 0.014" Guidewire<br>compatible;<br>Up to three guidewires<br>may be exchanged | 0.014" Guidewire<br>compatible;<br>Up to two guidewires<br>may be exchanged | 0.035" Guidewire<br>Compatible | 0.014" Guidewire<br>compatible; | | Lumen | Triple | Dual | Dual | Triple | | Distal Shaft<br>Capability | The Nitinol scaffold<br>provides distal<br>anchoring and<br>supports the<br>advancement of the<br>guidewire. | Standard Percutaneous<br>Catheter Distal Shaft<br>Structure (no support<br>mechanism) | The balloon provides<br>distal anchoring and<br>supports the<br>advancement of the<br>guidewire. | The balloon provides<br>distal anchoring and<br>supports the<br>advancement of the<br>guidewire. | #### SUBSTANTIAL EQUIVALENCE The MultiCross Support Catheter is substantially equivalent to Vascular Solutions Twin-Pass OTW Dual Access Catheter (K072618), the EndoCross ENABLER-P Catheter System (K092174, and the BridgePoint Medical Stingray Orienting Balloon Catheter (K080987). The subject device and the predicate devices are percutaneous catheters. The proposed indications for use for the MultiCross Support Catheter are substantially equivalent to the indications for use for the predicate devices. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the MultiCross Support Catheter is substantially equivalent to the predicate devices. # TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION #### Performance Testing - Bench Extensive bench testing was conducted on the MultiCross Support Catheter to evaluate the performance of the device and to support a determination of substantial equivalence to the predicate devices. Non-clinical testing assessed the following aspects of the device: - Catheter Inspection (Visual Inspection/Physical Dimensions) � - Particulate Testing - Radial Outward Force . - Contrast Flush Injection . - Simulated Use � - Torque Strength - Flexibility and Kink Resistance - Leak: Liquid - Air {3}------------------------------------------------ K121763 Page 4 of 4 MULTICROSS SUPPORT CATHETER 510(k) PREMARKET NOTIFICATION - Tensile Strength . - Corrosion Resistance Test . - Coating Integrity . - In vivo Animal Validation Study . - . Biocompatibility - Sterilization Validation . - Packaging and Shelf-life ♥ All testing was performed in accordance with recognized standards. # Safety and Performance Testing - Animal A GLP animal study was performed to evaluate the safety, performance and handling of the MultiCross Support Catheter in the coronary and peripheral vasculature as compared to a control device. Based on pathology and histopathology results, the safety acceptance criteria for the study were met. Performance and handling observations were made based on detailed characteristics of the device. No untoward observations were found by the clinician. #### Conclusion of Performance Testing The collective results of the bench testing and the GLP animal study demonstrate that the MultiCross Support Catheter meets the established specifications necessary for consistent performance for its intended use and is substantially equivalent to the predicate devices. #### CONCLUSION The MultiCross Support Catheter is a percutaneous vascular support catheter and shares its design and mechanism of action with the identified predicate devices. The results of the bench testing and GLP animal study confirm that the MultiCross Support Catheter functions to its specifications, performs as intended and exhibits the appropriate characteristics of a percutaneous support catheter. The MultiCross Support Catheter is substantially equivalent to the predicate devices in terms of technological characteristics, intended use and performance. No new issues of safety or effectiveness are raised by the MultiCross Support Catheter. #### SUMMARY The MultiCross Support Catheter is substantially equivalent to the predicate devices. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of an eagle or bird with outstretched wings, rendered in a bold, black color. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Roxwood Medical, Inc. Mr. Mehdrad Farhangnia Chief Executive Officer 150 Executive Park Blvd, Suite 3700 San Francisco, CA 94134 Re: K121763 Trade/Device Name: MultiCross Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 16, 2012 Received: November 19, 2012 Dear Mr. Farhangnia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set NOV 2 7 2012 {5}------------------------------------------------ # Page 2 - Mr. Mehdrad Farhangnia forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic in product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. > Sincerely yours, Matthew G Hillebrenne for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ROXWOOD MEDICAL, INC. # 510(k) Number (if known): KI21 76 3 Device Name: MultiCross Support Catheter # Indications For Use: The MultiCross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) MS Kelli Division of Cardiovascular Devices 510(k) Number K121763
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...