The STRONG Dental Handpieces, ACL(B)-03C and ACL(B)-03F, are intended for a wide range of dental procedures including: A. Implant placement, including 1. Preparation of the osteotomy site 2. Bone contouring, osteoplasty B. Periodontal surgeries 1. Bone contouring & alveoplasty around living teeth 2. Removal of exostosis C. Bone grafting 1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.) 2. Harvesting autogen living bone 3. Sinus elevation & grafting of alveolar sockets D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions
Device Story
STRONG Dental Handpieces (models ACL(B)-03C, ACL(B)-03F) are gear-driven, handheld dental instruments (1:1 gear ratio). Devices connect to torque-adjustable electrical motors via ISO 3964 coupling. Head clamps accept instruments per ISO 1797-1. Used by dental professionals in clinical settings for surgical procedures including implant placement, periodontal surgery, bone grafting, and tooth extraction. Contra-angle design facilitates access to difficult-to-reach areas. Device transforms motor torque into mechanical rotation for cutting, contouring, or removing bone and tooth structure. Benefits include precise surgical site preparation and efficient tissue removal. Sterilized by end-user via steam autoclave.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 14457:2012 verified via visual inspection, extraction force testing (ACL(B)-03C: 51.2N; ACL(B)-03F: 32.2N), eccentricity (0.014mm/0.012mm), temperature rise (≤10°C), and corrosion resistance. Sterilization validated per ISO 17665-1.
Indicated for dental procedures including implant placement (osteotomy preparation, bone contouring), periodontal surgeries (bone contouring, alveoplasty, exostosis removal), bone grafting (donor site preparation, autogenous bone harvesting, sinus elevation, alveolar socket grafting), and tooth/bone removal and sectioning (impacted third molars, complicated extractions).
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus-like symbol with three human profiles merging into a single form, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The seal is simple, using black lines on a white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2015
Saeshin Precision Co., Ltd. Choi Sae Kwan Quality Assurance Manager 52 Secheon-ro 1-gil, Dasa-eup Dalseong- gun, Daegu 711-814 KOREA
Re: K143418
Trade/Device Name: STRONG Dental Handpieces Regulation Number: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories Regulatory Class: I Product Code: EGS Dated: June 18, 2015 Received: June 22, 2015
Dear Choi Sae Kwan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143418
Device Name STRONG Dental Handpieces
Indications for Use (Describe)
The STRONG Dental Handpieces. ACL(B)-03C and ACL(B)-03F. are intended for a wide range of dental procedures including:
A. Implant placement, including
- 1. Preparation of the osteotomy site
- 2. Bone contouring, osteoplasty
B. Periodontal surgeries
- 1. Bone contouring & alveoplasty around living teeth
- 2. Removal of exostosis
- C. Bone grafting
- 1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)
- 2. Harvesting autogen living bone
- 3. Sinus elevation & grafting of alveolar sockets
D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Saeshin. The logo consists of a blue geometric shape above the company name, "SAESHIN," which is written in a stylized, sans-serif font, also in blue. The geometric shape is composed of several smaller triangles arranged to form a larger, abstract design.
# 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-Tel: +82 53 587 2341 / Fax: +82 53 580 0
# 510(k) Summary
[As required by 21 CFR 807.92] This 510(k) summary is prepared in accordance with 21 CFR807.92
### 1. Date Prepared [21 CFR 807.92(a)(1)]
10/28/2014
### 2. Submitter's Information [21 CFR807.92(a)(1)]
- . Name of Sponsor: Saeshin Precision Co., Ltd.
- -Address: # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Republic of Korea
- Contact Name: Sae Kwan, Choi (Mr.) / Quality Manager ●
- -Telephone No. : +82 53 587 2341
- -Fax No. : +82 53 580 0999
- -Email Address : ksqc@saeshin.com
- Registration Number: ● 3007958831
- Name of Manufacturer: Same as Sponsor
- Same as Sponsor -Address:
### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
- . Trade Name: STRONG Dental Handpieces
- . Common Name: Dental Handpiece and Accessories
- Classification Name: Dental Handpiece and Accessories
- . Classification Panel: Dental
- Classification Regulation: ● 21 CFR 872.4200
- . Product Code: EGS
{5}------------------------------------------------
- . Device Class:
### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
l
The identified predicate devices within this submission are shown as follows:
- 510(k) Number: K100192
- Applicant: . Saeshin Precision Co., Ltd.
- Common Name: Dental Handpieces and Accessories ●
- Device Name: STRONG Dental Handpieces .
There are no significant differences between the proposed models of STRONG Dental Handpieces and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in device design, composition of materials and technical specifications.
### 5. Description of the Device [21 CFR 807.92(a)(4)]
The STRONG Dental Handpieces; ACL(B)-03C and ACL(B)-03F are gear driven handheld dental handpieces with Gear Ratio of 1:1. They can be driven by torque adjustable electrical motors for surgery treatment. They are attached to drive via ISO 3964 coupling. The head clamp accepts instrument complying with ISO 1797-1. They have contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
The STRONG Dental Handpieces are similar to other commercially available products based on intended use, material, design and use concept. And they also comply with ISO 3964 coupling and ISO 1797-1 shank.
Based on the comparison of intended use and technical features, the STRONG Dental Handpieces are substantially equivalent to the predicate devices
### 6. Intended Use [21 CFR 807.92(a)(5)]
The STRONG Dental Handpieces, ACL(B)-03C and ACL(B)-03F, are intended for a wide range of dental procedures including:
- A. Implant placement, including
- 1. Preparation of the osteotomy site
- 2. Bone contouring, osteoplasty
- B. Periodontal surgeries
- 1. Bone contouring & alveoplasty around living teeth
- 2. Removal of exostosis
{6}------------------------------------------------
- C. Bone grafting
- 1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)
- 2. Harvesting autogen living bone
- 3. Sinus elevation & grafting of alveolar sockets
- D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions
## 7. Technological Characteristics [21 CFR 807.92(a)(6)]
| | ACL(B)-03C | ACL(B)-03F |
|----------------------|-----------------|-----------------|
| Revolution | 0~35,000RPM | 0~35,000RPM |
| Gear Ratio | 1:1 | 1:1 |
| Weight | approx. 45g | approx. 46g |
| Size | Ø19.6x 83.7mm | Ø19.6x 83.7mm |
| Articles | Low speed angle | Low speed angle |
| Standard<br>Coupling | ISO 3964 | ISO 3964 |
3 / 10
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo and address of a company called SAESHIN. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Korea. The phone number is +82 53 587 2341 and the fax number is +82 53 580 0999.
### 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
When compared to the predicate device (K100192), the STRONG Dental Handpieces presented in this submission has the same:
- Intended Use ●
- Device Design ●
- Composition of materials ●
- Technical Specifications .
| Parameter | Proposed | Predicate |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | STRONG Dental Handpieces (Model Nos.:<br>ACL(B)-03C and ACL(B)-03F) | STRONG Dental Handpieces (Model Nos.:<br>AT- II, ACL-01C, ACL-02C and ACL(B)-01C) |
| 510(k) Number | Unknown | K100192 |
| Manufacturer | Saeshin Precision Co., Ltd | Saeshin Precision Co., Ltd. |
| The Indications for Use<br>are the same in the<br>predicate and proposed<br>devices, except that the<br>terms used in the<br>predicate were clarified<br>with plain English in the<br>proposed device | The Strong Dental Handpieces are indicated<br>for wide range of dental procedures.<br>A. Implant placement, including<br>1. Preparation of the osteotomy site<br>2. Bone contouring, osteoplasty<br>B. Periodontal surgeries<br>1. Bone contouring & alveoplasty<br>around living teeth<br>2. Removal of exostosis<br>C. Bone grafting<br>1. Preparation of the donor site (for e.g.<br>symphysis and ascending rames etc.)<br>2. Harvesting autogen living bone<br>3. Sinus elevation & grafting of alveolar<br>sockets | The Strong Dental Handpieces are indicated<br>for wide range of dental procedures.<br>• AT- II for the application in the area<br>of the front teeth, root tip resection,<br>bone removal, osteotomia on the<br>upper and lower jaw, preprosthesis<br>surgical modellation, sequestrotomia,<br>fenestration on the alveolar<br>appendix, apical ventilation, bone<br>modellation, bone smoothing.<br>• ACL-01C, ACL-02C and ACL(B)-01C<br>for the osteotomia on the upper and<br>lower jaw, germectomia,<br>sequestrotomia. |
| Parameter | Proposed | Predicate |
| | STRONG Dental Handpieces (Model Nos.: | STRONG Dental Handpieces (Model Nos.: |
| | ACL(B)-03C and ACL(B)-03F) | AT- II, ACL-01C, ACL-02C and ACL(B)-01C) |
| | D. Removal and sectioning of teeth and | |
| | teeth bone for e.g. impacted third molars | |
| | and complicated extractions | |
| | | |
| Device Design | | |
| Operational Mode | Gear | Gear |
| Gear Ratio | 1:1 | 1:1 |
| Length | 83.7 mm | 84.0, 84.0, 83.6 mm |
| Diameter | Ø 19.6 | Ø 19.6 |
| Dia. of tool | 2.35, 1.60 mm | 2.35 mm |
| Max. Overall Length of | 30 mm | 45, 30, 30 mm |
| Rotary Instrument | | |
| Head Height | 13.7 mm | 13.7 mm |
| Head Diameter | Ø 8 mm | Ø 8, 10 mm |
| Shank | By ISO1797-1 | By ISO1797-1 |
| Length of Shank | 9 – 12 mm | 9 – 12 mm |
| Type of Chuck | Push-Button Locking | Latch or Push-Button Locking |
| Coupling Dimension | By ISO 3964 | By ISO 3964 |
| Type of Connector | Not Applied | Not Applied |
| Accessories | Spanner | Spanner |
| Composition of Materials | | |
| /Surface Treatment | | |
| Gear | SUS420F/Vacuum Heat Treatment | SUS420F/Vacuum Heat Treatment |
| Shank | SUS304 | SUS304 |
| Head and Nozzle | C3604BD-F/Hard Chrome Coated | C3604BD-F/Hard Chrome Coated |
| Chuck | SUS420F/Vacuum Heat Treatment | SUS420F/Vacuum Heat Treatment |
| Handle | AL6061/Anodizing | AL6061/Anodizing |
| Pipe | N/A | C3604BD-F |
| Patient-Contacting | Chuck, Head | Chuck, Head |
| Operator-Contacting | Handle, Head | Handle, Head |
| Technical Specification | | |
| Parameter | Proposed | Predicate |
| | STRONG Dental Handpieces (Model Nos.: | STRONG Dental Handpieces (Model Nos.: |
| | ACL(B)-03C and ACL(B)-03F) | AT- II, ACL-01C, ACL-02C and ACL(B)-01C) |
| Chuck Design | Type1 and Type 3 Push-Button Locking by<br>ISO 1797-1 | Type 2 by ISO 1797-1<br>Type1 Latch or Push-Button Locking by ISO<br>1797-1 |
| Bur Extraction Force(N)<br>Max. Torques<br>Max. Water Pressure<br>Max. Speed in rpm<br>Shank Conformance<br>Coupling Dimension | 45 N, 22N<br>50 Ncm<br>N/A<br>35,000 rpm<br>By ISO 1797-1<br>By ISO 3964 | 55 - 56 N<br>50 Ncm<br>N/A<br>30,000 - 35,000 rpm<br>By ISO 1797-1<br>By ISO 3964 |
| Lubricant<br>Chemical Composition<br>510k#<br>Biocompatible<br>Delivery system | N/A | DO-ALL Dental Handpiece Lubricant<br>K073353<br>N/A. Not intended for patient-contact<br>Spray Nozzel |
| Sterilization | Steam Heat 132 °C/4minutes | Steam Heat 134 °C/4minutes |
| Operating principle | The STRONG Dental Handpieces, ACL(B)- | The STRONG Dental Handpieces, AT- II, |
| | 03C and ACL(B)-03F, are gear driven hand-<br>held dental handpieces with gear ratio of 1:1.<br>It can be driven by torque adjustable<br>electrical motors for surgery treatment. It is<br>attached to drive via ISO 3964 coupling. The<br>head clamp accepts instrument complying<br>with ISO 1797-1. They have contra angle<br>attachment for difficult to reach<br>areas | ACL-01C, ACL-02C and ACL(B)-01C, are<br>gear driven hand-held dental handpieces<br>with gear ratio of 1:1. It can be driven by<br>torque adjustable electrical motors for<br>surgery treatment. It is attached to drive via<br>ISO 3964 coupling. The head clamp accepts<br>instrument complying with ISO 1797-1. They<br>have contra angle attachment for difficult to |
| Parameter | Proposed<br>STRONG Dental Handpieces (Model Nos.:<br>ACL(B)-03C and ACL(B)-03F) | Predicate<br>STRONG Dental Handpieces (Model Nos.:<br>AT- II, ACL-01C, ACL-02C and ACL(B)-01C) |
| | intended to prepare dental cavities for<br>restorations, such as fillings, and for cleaning<br>teeth. | reach areas intended to prepare dental<br>cavities for restorations, such as fillings, and<br>for cleaning teeth. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the address and contact information for Saeshin. The address is #52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Korea. The telephone number is +82 53 587 2341, and the fax number is +82 53 580 0999. The Saeshin logo is at the top of the image.
### 9. Summary of Non-Clinical Performance Data
#### ● Biocompatibility
The materials contacting patients of chuck and the head were totally same and have been previously cleared by FDA through K100192 of predicate device, STRONG Dental Handpiece. The test report has been issued by Korea Testing & Research Institute, 7-6, Gomak-Ri, Wolgot-Myeon, Gimpo-Si, Gyunggi-Do, 415-871, Korea.
The categorization of contact was established under ISO 10993-1:2009-10-15.
According to the recommendations in this standard and FDA Blue Book Memo #G95-
- 1, the following tests are applicable and were performed:
- Cytotoxicity according to ISO 10993-5 -
- -Sensitization according to ISO 10993-10
- -Irritation or intracutaneous reactivity according to ISO 10993-10
#### ● Bench Testing
Bench testing was performed to ensure the performance of the STRONG Dental Handpieces. ACL(B)-03C and ACL(B)-03F, verify conformity to ISO 14457 and demonstrate substantial equivalence to the predicates.
ACL(B)-03C and ACL(B)-03F samples were compliant with ISO 14457: 2012 Dentistry - Handbieces And Motors and demonstrated substantial equivalence to the predicates.
#### -Visual inspection of general design
This test was performed with normal visual and profile projector. All the articles comply with the acceptance criteria of Section 7.2 of the handpiece standard.
#### Extraction force -
The STRONG Dental Handpieces, ACL(B)-03C, for extraction test mandrel type 3 from the locking chuck system shall be at least 45N. In addition, STRONG Dental Handpieces, ACL(B)-03F, for extraction test mandrel type 5 from the locking chuck system shall be at least 22N. The results are reported in the table below:
| Article No. | Mean (N) |
|-------------|----------|
| ACL(B)-03C | 51.2 N |
| ACL(B)-03F | 32.2 N |
#### Eccentricity -
The eccentricity of the test mandrel in rotation and without applied load shall not be exceeded the total indicated run-out of 0.08 mm. The results are reported in the table below
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Saeshin. The logo consists of a geometric shape made up of several triangles in shades of blue, positioned above the company name, "SAESHIN," written in a bold, sans-serif font, also in blue. The geometric shape resembles a stylized cube or diamond, giving the logo a modern and technological feel.
| Article No. | Mean (mm) |
|-------------|-----------|
| ACL(B)-03C | 0.014 |
| ACL(B)-03F | 0.012 |
#### Resistance to sterilizing procedure -
After 250 cycles of 132 ° for 4 minutes in the autoclave, there was no sign of deterioration.
| Article No. | Deterioration<br>detected | Extraction<br>> 45 N, > 22<br>N | Max. speed<br>±10% rpm | Noise <<br>70dB |
|-------------|---------------------------|---------------------------------|------------------------|-----------------|
| ACL(B)-03C | No | 51.2 N | 35,120 | 55 |
| ACL(B)-03F | No | 32.2 N | 35,090 | 54 |
#### Temperature rise -
Rise of maximum temperature at the touchable surface of the housing under rated running conditions shall not exceed 20℃ compared to the temperature of the environment. The results are reported in the table below;
| Article No. | Mean (°C) |
|-------------|-----------|
| ACL(B)-03C | 10 |
| ACL(B)-03F | 6 |
#### -Resistance to corrosion
Handpieces are corrosion resistant. There was no signs of corrosion after having autoclave procedure 10 times at 132 ℃ for 4 minutes at 22kPa.
#### ● Clinical and non-clinical tests
A non-clinical evaluation, based on literature research, has been done. The evaluation of the applicable market data showed that STRONG Dental handpieces, ACL(B)-03C and ACL(B)-03F, do not pose known or new clinical risks than similar medical devices currently on the market. Based on those results clinical test have not been executed.
#### . Sterilization
The STRONG Dental handpieces, ACL(B)-03C and ACL(B)-03F, is provided nonsterile and labeled for sterilization by the end user, the Instructions for use include the following parameters:
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo and contact information for Saeshin. The address is #52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Korea. The telephone number is +82 53 580 0925 and the fax number is +82 53 580 0939.
| Sterilization Type | Temperature | Time | Load<br>Characteristics | Dry Time |
|---------------------------------------------|-------------|--------|-------------------------|----------|
| Steam Sterilization<br>(Pre-vacuum<br>Type) | 132 °C | 4 min. | Sterilization Bag | 30 min. |
The sterilization method was validated per ISO 17665-1: 2006 Sterilization of health care products – Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
### 10.Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic, 21 CFR Part 807, and based on the information provided in this premarket notification Saeshin Presicion Co., Ltd. Concludes that the STRONG Dental Handpieces are substantially equivalent to predicate devices as described herein.
In all the respects, the STRONG Dental Handpieces is the equivalent to currently marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.