The STRONG Dental Handpieces is indicated for wide range of dental procedures. AT- II for the application in the area of the front teeth, root tip resection, bone removal, osteotomia on the upper and lower jaw, preprosthetic surgical modellation, sequestrotomia, fenestration on the alveolar appendix, apical ventilation, bone modellation, bone smoothing. ACL-01, ACL(B)-01C and ACL-02C for the osteotomia on the upper and lower jaw, germectomia, sequestrotomia.
Device Story
STRONG Dental Handpieces (models AT-II, ACL-01C, ACL(B)-01C, ACL-02C) are mechanical dental instruments; 1:1 transmission ratio; attach to torque-adjustable electrical motors via ISO 3964 coupling; accept ISO 1797-1 compliant instruments. Used by dental professionals for surgical procedures including bone removal, osteotomia, and root tip resection. Device provides mechanical rotation for cutting/shaping bone or teeth; facilitates surgical access in difficult-to-reach areas. Benefits include precise bone/tooth modification during oral surgery.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and intended use comparison to predicate devices.
Technological Characteristics
Mechanical dental handpiece; 1:1 transmission ratio; ISO 3964 coupling; ISO 1797-1 instrument shank compatibility; manual operation via external electrical motor.
Indications for Use
Indicated for dental procedures including root tip resection, bone removal, osteotomia, preprosthetic surgical modellation, sequestrotomia, fenestration, apical ventilation, bone smoothing, and germectomia.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
K100192
019
.
Image /page/0/Picture/2 description: The image shows the logo for SAESHIN. The logo consists of a stylized letter "S" inside of a circle on the left, followed by the word "SAESHIN" in a bold, sans-serif font. The logo is black and white.
Saeshin Precision Co., Ltd.
#93-15, Paho-Dong, Dalseo-Gu, Daegu, 704-220, Republic of Korea Tel 82 53-587-2345 Fax 82 53-587-2347
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).
Date: November 30, 2009
1. Company and Correspondent making the submission:
| | Company |
|-------------------------|---------------------------------------------------------------------|
| Name | Saeshin Precision Co., Ltd. |
| Address | #93-15, Paho-dong, Dalseo-Gu, Daegu, 704-<br>220, Republic of Korea |
| Phone<br>Fax<br>Contact | +82 53-587-2345<br>+82 53-587-2347<br>Y. S. Lee |
- 2. Device:
Proprietary Name - STRONG Dental Handpieces Common Name - Dental Handpieces and Accessories Classification Name - Handpiece, Contra- And Right-Angle Attachment, Dental
- 3. Predicate Device:
- · Surgical Contra-Angle Handpieces Types WS-56E, Surgical straight Handpieces Types, S-11, K011061
- · STRONG Implant Handpieces, K092412
- 4. Classifications Names & Citations: EGS, 872.4200
- 5. Description:
The STRONG Dental Handpieces; AT-II, ACL-01C, ACL(B)-01C and ACL-02C is The STRONG Dental handpieces with transmission ratio of 1:1. It can be driven by torque adjustable electrical motors for surgery treatment. It is attached to drive via ISO 3964 coupline. The head clamp accepts instrument complying with ISO 1797-1. It has contra-angle(AT-II has straight-angle) attachment for difficult to reach 1797-1. It has contra digities for restorations, such as fillings, and for cleaning teeth.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Saeshin. The logo consists of a stylized letter "S" inside of a circle on the left, followed by the word "SAESHIN" in bold, sans-serif font. The logo is black and white.
Saeshin Precision Co., Ltd.
#93-15, Paho-Dong, Dalseo-Gu, Daegu, 704-220, Republic of Korea Tel 82 53-587-2345 Fax 82 53-587-2347
- Indication for use: 6.
The STRONG Dental Handpieces is indicated for wide range of dental procedures.
- AT- II for the application in the area of the front teeth, root tip resection, bone . removal, osteotomia on the upper and lower jaw, preprosthetic surgical modellation, sequestrotomia, fenestration on the alveolar appendix, apical ventilation, bone modellation, bone smoothing.
- ACL-01, ACL(B)-01C and ACL-02C for the osteotomia on the upper and lower . jaw, germectomia, sequestrotomia.
## 7. Review:
The STRONG Dental Handpieces has the same device characteristics as the predicate The Surgical Contra-Angle handpieces Types WS-56E, Surgical Straight Handpieces Types S-11; intended use, material, design and use concept are similar. And they also comply to ISO 3964 coupling and ISO 1797-1 shank.
Based on the comparison of intended use and technical features, the STRONG Dental Handpieces are substantially equivalent to the predicate devices.
- 8. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Saeshin Precision Co., Ltd. concludes that the STRONG Dental Handpieces is safe and effective and substantially equivalent to predicate devices as described herein.
- 9. Saeshin Precision Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA.
END
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Saeshin Precision Company, Limited C/O Mr. Marc M. Mouser Responsible Third Party Official Underwriters Laboratories, Incorporated Laboratory and Testing 2600 Northwest Lake Road Camas, Washington 98607-9526
Re: K100192
Trade/Device Name: STRONG Dental Handpieces Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Acessories Regulatory Class: I Product Code: EGS Dated: January 4, 2010 Received: January 22, 2010
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FEB - 4 2010
{3}------------------------------------------------
## Page 2 - Mr. Mouser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number K K106192
Device Name: STRONG Dental Handpieces
Indication for use:
The STRONG Dental Handpieces is indicated for wide range of dental procedures.
- AT II for the application in the area of the front teeth, root tip resection, bone . removal, osteotomia on the upper and lower jaw, preprosthetic surgical modellation, sequestrotomia, fenestration on the alveolar appendix, apical ventilation, bone modellation, bone smoothing.
- ACL-01, ACL(B)-01C and ACL-02C for the osteotomia on the upper and lower ● jaw, germectomia, sequestrotomia.
Prescription Use (Per 21CFR801 Subpart D)
Over-The-Counter Use -(Per 21CFR801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation
R.Shetz, DNS for Dr. K.P. muley
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Uental Devices
510(k) Number: K/00192
SAESHIN PRECISION Co., Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.