ICIS® View is a software application used for reference viewing of medical images and associated reports and, as such, fulfills a key role in Agfa HealthCare's Imaging Clinical Information System (ICIS). ICIS® View enables healthcare professionals, including (but not limited to) physicians, surgeons, nurses, and administrators to receive and view patient images and data from multiple departments and organizations within one multi-disciplinary viewer. Users may access the product directly via a web-browser, select mobile devices, healthcare portal or within the Electronic Medical Record (EMR). ICIS® View allows users to perform basic image manipulations and measurements (for example window/level, rotation, zoom, and markups). ICIS® View can optionally be configured for Full Fidelity mode, which is intended for diagnostic use, review and analysis of CR, DX, CT, MR, US images and medical reports. ICIS® View Full Fidelity is not intended to replace full diagnostic workstations and should only be used when there is no access to a workstation. ICIS® View full fidelity is not intended for the display of digital mammography images for diagnosis.
Device Story
ICIS View is a server-based PACS software providing multi-disciplinary viewing of medical images/reports; inputs include DICOM-compliant images from PACS or vendor-neutral archives. Operates via web-browser, mobile devices, or EMR integration. Provides image manipulation tools (window/level, rotation, zoom, markups) and lossless image retrieval in Full Fidelity mode. Used by clinicians in clinical settings to facilitate review and diagnostic decision-making when primary workstations are inaccessible. Benefits include enhanced access to prior studies, time-independent record access, and improved communication between geographically separated providers. Does not replace full diagnostic workstations.
Clinical Evidence
Bench testing only. No clinical studies performed. Diagnostic image quality evaluated by qualified radiologists comparing ICIS View (desktop/mobile) against IMPAX Workstation (K022292) using 6 studies per modality (CR, DX, CT, MR, US). Display performance validated using TG18-QC, TG18-BR, and TG18-LP standards.
Technological Characteristics
PACS software; DICOM compliant; server-based architecture; supports Windows 7 and iOS8; connectivity via web-browser, mobile, or EMR; utilizes standard DICOM transfers; conforms to ACR/NEMA PS3.1-3.20: 2011.
Indications for Use
Indicated for healthcare professionals (physicians, surgeons, nurses, administrators) to view medical images and reports. Full Fidelity mode indicated for diagnostic review/analysis of CR, DX, CT, MR, and US images when full diagnostic workstations are unavailable. Contraindicated for digital mammography diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 1, 2015
AGFA Healthcare N.V. % Ms. Shaeann Cavanagh Regulatory Affairs Specialist NA AGFA Healthcare 10 South Academy Street GREENVILLE SC 29601
Re: K143397
Trade/Device Name: ICIS View Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 24, 2015 Received: April 27, 2015
Dear Ms. Cavanagh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143397
Device Name
ICIS® View
### Indications for Use (Describe)
ICIS® View is a software application used for reference viewing of medical images and associated reports and, as such, fulfills a key role in Agfa HealthCare's Imaging Clinical Information System (ICIS). ICIS® View enables healthcare professionals, including (but not limited to) physicians, surgeons, nurses, and administrators to receive and view patient images and data from multiple departments and organizations within one multi-disciplinary viewer.
Users may access the product directly via a web-browser, select mobile devices, healthcare portal or within the Electronic Medical Record (EMR). ICIS® View allows users to perform basic image manipulations and measurements (for example window/level, rotation, zoom, and markups).
ICIS® View can optionally be configured for Full Fidelity mode, which is intended for diagnostic use, review and analysis of CR, DX, CT, MR, US images and medical reports. ICIS® View Full Fidelity is not intended to replace full diagnostic workstations and should only be used when there is no access to a workstation. ICIS® View full fidelity is not intended for the display of digital mammography images for diagnosis.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY Agfa HealthCare's ICIS® View
Common Name: Picture Archiving and Communications System (PACS) Classification Name: Radiological Image Processing System Regulatory Classification: 21 CFR 892.2050 Product Code: LLZ Proprietary Name: ICIS® View Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Jodi Coleman, Prepared: November 26, 2014 Telephone: 519-746-6210 ext. 2485
#### LEGALLY MARKETED PREDICATE DEVICES A.
This is a 510(k) for Agfa's ICIS® View software, a picture archiving and communications It is substantially equivalent to systems with MIM Software Inc.'s Mobile MIM svstem. (K103785-primary predicate), Agfa HealthCare's IMPAX Workstation (K022292), and Client Outlook Inc.'s eUnity™ (K11164).
#### DEVICE DESCRIPTION B.
Agfa's ICIS® View system is a picture archiving and communication system (PACS), product code LLZ, intended to provide an interface for the display, annotation, review, printing, storage and distribution of multimodality medical images, reports and demographic information for review and diagnostic purposes within the system and across computer networks.
The new device is substantially equivalent to the predicate devices (K103785, K022292, & K111164). It is a multidisciplinary viewer that allows the user to securely access patient images and reports from any PACS or vendor-neutral archive. Images and reports can be viewed directly via a web-browser, select mobile device, healthcare portal or Electronic Medical Record (EMR). The new device includes some of the clinical tools of the predicate devices specifically the functionality to retrieve original lossless renditions of stored images for diagnostic purposes.
The optional Full Fidelity functionality allows the retrieval of original lossless renditions of stored CR, DX, CT, MR, and US images for diagnostic purposes on select mobile devices or FDA cleared display monitors when there is no access to a full workstation.
Principles of operation and technological characteristics of the new and predicate devices are the same. There is no change to the intended use of the device vs. the predicate devices. Laboratory data including, performance assessments and image quality evaluations conducted with qualified radiologists confirm that performance is equivalent to the predicates.
### C. INTENDED USE
ICIS® View is a software application used for reference viewing of medical images and associated reports and, as such, fulfills a key role in Agfa HealthCare's Imaging Clinical
{4}------------------------------------------------
Information System (ICIS). ICIS® View enables healthcare professionals, including (but not limited to) physicians, surgeons, nurses, and administrators to receive and view patient images and data from multiple departments and organizations within one multi-disciplinary viewer.
Users may access the product directly via a web-browser, select mobile devices, healthcare portal or within the Electronic Medical Record (EMR). ICIS® View allows users to perform basic image manipulations and measurements (for example window/level, rotation, zoom, and markups).
ICIS® View can optionally be configured for Full Fidelity mode, which is intended for diagnostic use, review and analysis of CR, DX, CT, MR, US images and medical reports. ICIS® View Full Fidelity is not intended to replace full diagnostic workstations and should only be used when there is no access to a workstation. ICIS® View full fidelity is not intended for the display of digital mammography images for diagnosis.
Intended use has not changed as a result of any labeling modification(s).
### D. SUBSTANTIAL EQUIVALENCE SUMMARY
Agfa's ICIS® View system has an Indications For Use statement similar to the predicate devices (K103785, K022292, & K111164). The statements have been combined and simplified. There are no changes to the intended use/indications of the device. Intended uses are the same. The devices have the same technological characteristics.
Both ICIS® View and the primary predicate device, Mobile MIM (K103785) are used to display medical images for diagnostic purposes of multiple modalities inlcuding MR and CT. Both provide wireless and portable access to medical images on select Apple products. Furthermore, ICIS® View and Mobile MIM (K103785) contain a statement that it is not intended to replace full diagnostic workstations and should only be used when there is no access to a workstation. ICIS® View and eUnity™ (K111164) are indicated to allow image manipulations and display lossy/lossless images. Both ICIS® View and predicate device, IMPAX Workstation (K02292) are indicated to display, review, transfer, store, and archive medical images. ICIS® View and all three predicate devices (K103785, K022292, & K111164) are not indicated for mammography and all the devices accept standard DICOM transfers.
The only difference of the new device and the primary predicate device, Mobile MIM (K103785), is the new device is server based software that connects to an existing PACS; whereas, Mobile MIM contains an App that connects to the existing PACS. However, the other two predicates, IMPAX Workstation (K022292) and eUnity™ (K111164) are also server based software that connects to the existing PACS. Differences in devices do not alter the intended diagnostic effect.
The devices have the same technological characteristics. The new device and the predicate devices (K103785, K022292, & K11164) are picture archiving and communication systems (PACS), Product Code LLZ. Agfa's ICIS® View system is substantially equivalent to the predicate devices (K103785, K022292, & K111164) in that it uses precisely the same technology to capture and display medical data. Descriptive characteristics and performance data are adequate to ensure equivalence.
{5}------------------------------------------------
Performance data including resolution testing and image quality evaluations by qualified radiologists are adequate to ensure equivalence.
| predicate devices. | | | | |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ICIS® View<br>(New Device) | Mobile MIM<br>(Primary<br>PREDICATE-<br>K103785) | IMPAX<br>Workstation<br>(PREDICATE-<br>K022292) | eUnity™<br>(PREDICATE-<br>K111164) |
| Communication | Same as predicates | DICOM | DICOM | HL7, DICOM, IHE |
| Mammographic Use | Same as predicates | No | No | No |
| Modalities for<br>Diagnostic Use | CR, DX, CT, MR, US | SPECT, PET,<br>CT, MR | Radiology<br>(including CR, DX,<br>CT, MR, US) | Radiology<br>(including CR, DX, CT,<br>MR, US) |
| Operating System for<br>Diagnostic Viewing | Windows 7 & iOS8 | iOS | Windows 7 & 8 | Windows & iOS |
| Mobile Device<br>Support for<br>Diagnostic Viewing | iPad® 3, iPad® 4 | Apple iOS<br>handheld devices | | |
| Transfer/Storage/Disp<br>lay of Medical Images | Same as predicates | Yes | Yes | Yes |
| Network Access | Same as K022292 &<br>K111164 | App that<br>connects to<br>existing PACS | Server based<br>software, connects to<br>existing PACS | Server based software,<br>connects to existing<br>PACS |
| User Authentification | Same as predicates | Yes | Yes | Yes |
| Window Level | Same as predicates | Yes | Yes | Yes |
| Rotate/Pan/Zoom | Same as predicates | Yes | Yes | Yes |
| Measurements | Same as predicates | Yes | Yes | Yes |
| Annotations | Same as predicates | Yes | Yes | Yes |
| Indications for Use<br>Statements | ICIS® View is a software<br>application used for<br>reference viewing of<br>medical images and<br>associated reports and, as<br>such, fulfills a key role in<br>Agfa HealthCare's Imaging<br>Clinical Information System<br>(ICIS). ICIS® View enables<br>healthcare professionals,<br>including (but not limited<br>to) physicians, surgeons,<br>nurses, and administrators to<br>receive and view patient<br>images and data from<br>multiple departments and | The Mobile<br>MIM software<br>program is used<br>for the<br>registration,<br>fusion, and/or<br>display for<br>diagnosis of<br>medical images<br>from only the<br>following<br>modalities:<br>SPECT, PET,<br>CT, and MRI. | IMPAX Workstations<br>with MPR, Digital<br>Subtraction and 3D<br>options are intended<br>for use in the<br>acquisition, display,<br>digital processing,<br>review, transfer,<br>storage, archiving<br>and printing of<br>medical images and<br>patient demographic<br>information. | eUnity™ is a software<br>application that displays<br>medical image data and<br>associated clinical reports.<br>eUnity™ performs<br>operations relating to the<br>transfer, storage, display<br>and measurement of<br>image data.<br>eUnity™ allows users to<br>perform image<br>manipulations, including<br>window/level, rotation, |
| Indications for Use<br>Statements continued.. | organizations within one<br>multi-disciplinary viewer.<br>Users may access the<br>product directly via a web-<br>browser, select mobile<br>devices, healthcare portal or<br>within the Electronic<br>Medical Record (EMR).<br><br>ICIS® View allows users to<br>perform basic image<br>manipulations and<br>measurements (for example<br>window/level, rotation,<br>zoom, and markups).<br><br>ICIS® View can optionally<br>be configured for Full<br>Fidelity mode, which is<br>intended for diagnostic use,<br>review and analysis of CR,<br>DX, CT, MR, US images<br>and medical reports. ICIS®<br>View Full Fidelity is not<br>intended to replace full<br>diagnostic workstations and<br>should only be used when<br>there is no access to a<br>workstation.<br><br>ICIS® View full fidelity is<br>not intended for the display<br>of digital mammography<br>images for diagnosis. | Mobile MIM<br>provides wireless<br>and portable<br>access to medical<br>images. This<br>device is not<br>intended to<br>replace full<br>workstations and<br>should be used<br>only when there<br>is no access to a<br>workstation.<br><br>The device is not<br>to be used for<br>mammography. | They are indicated<br>for use by the<br>physician to aid in<br>diagnosis, and by<br>medical professionals<br>whenever they would<br>require or desire<br>access to medical<br>images and patient<br>demographic<br>information.<br><br>MPR and 3D<br>functions allow the<br>user to view 3D<br>image data from<br>perspectives different<br>from that in which it<br>was acquired. Digital<br>subtraction allows a<br>user viewing time<br>series studies to<br>remove densities not<br>enhanced by a<br>contrast medium,<br>highlighting vascular<br>structures. | zoom, measurement and<br>markup.<br><br>eUnity™ displays both<br>lossless and lossy<br>compressed images. For<br>lossy images, the medical<br>professional user must<br>determine if the level of<br>loss is acceptable for their<br>purposes.<br><br>Display monitors used for<br>reading medical images<br>for diagnostic purposes<br>must comply with<br>applicable regulatory<br>approvals and with<br>quality control<br>requirements for their use<br>and maintenance.<br><br>Mammographic images<br>and digitized film screen<br>images must not be<br>reviewed for primary<br>image interpretations.. |
Table 1 on the below summarizes the similarities and differences between the new device and predicate devices.
{6}------------------------------------------------
## Table 1: Device Predicate Comparison
### TECHNOLOGICAL CHARACTERISTICS E.
Agfa's ICIS® View system is a picture archiving and communication system (PACS), product code LLZ, intended to provide an interface for the display, annotation, review, printing, storage and distribution of multimodality medical images, reports and demographic information for review and diagnostic purposes within the system and across computer networks.
It is a software device running on commercially available computer equipment that allows users to view and modify DICOM compliant medical images and patient information.
Principles of operation and technological characteristics of the new and predicate devices are the same. There is no change to the intended use of the device vs. the predicate devices. The new device is substantially equivalent to the predicate devices (K103785, K022292, & K111164). It is a multidisciplinary viewer that allows the user to securely access patient images and reports from any PACS or vendor-neutral archive. Images and reports can be viewed directly via a webbrowser, select mobile device, healthcare portal or Electronic Medical Record (EMR). The new
{7}------------------------------------------------
device includes some of the clinical tools of the predicate devices specifically the functionality to retrieve original lossless renditions of stored images for diagnostic purposes. There are no differences between the device and the predicates (K103785, K022292, & K111164) that impact safety and effectiveness.
### F. TESTING
Laboratory testing and software testing (for a moderate level of concern device) using equivalent test protocols as used for Medical Image Management Devices as part of verification and validation under design controls (according to 21 CFR 820.30).
No animal or clinical studies were performed in the development of the new device. No patient treatment was provided or withheld.
1. ICIS® View vs. Agfa IMPAX (K02292) - Requirements & Performance of the full fidelity view were evaluated:
To determine substantial equivalence, diagnostic image quality was evaluated by qualified medical professionals comparing images across 3 platforms:
- ICIS® View Desktop - using an FDA clearaed diagnostic monitor
- ICIS® View Mobile using a calibrated iPad® 3 and iPad® 4 ●
- Agfa's full diagnostic PACS workstation IMPAX 6.6.1 (K022292 predicate) ●
A sample set of an average of 6 imaging studies per modality (CR, DX, CT, MR, US) were evaluated. Qualified radiologists were asked to provide an acceptable or unacceptable score when comparing the diagnostic quality, including evaluations of contrast, sharpness, artifacts and overall image quality in ICIS® View Full Fidelity mode (both desktop and mobile) to the IMPAX predicate.
### 2. Display Device Assessment - TG18 Testing
The executed validation plan for image quality required calibrated display devices for all assessments.
A direct comparison of all images discussed was conducted such that envrionmental conditions, the evaluator, infrastructure etc. were all identical to make the best comparison assessment possible.
To assess and compare the overall image quality of the display devices used for validation the "Assessment of Display Performance for Medical Imaging Systems" was used (TG18-QC, TG18-BR, TG18-LP).
All results met acceptance criteria.
Performance data including resolution testing and image quality evaluations by qualified radiologists are adequate to ensure equivalence to the predicates.
{8}------------------------------------------------
The product, manufacturing and development processes conform to product safety and medical imaging standards including:
#### PRODUCT STANDARD G.
- . ACR/NEMA PS3.1-3.20: 2011 Digital Imaging and Communications in Medicine (DICOM).
## OUALITY MANAGEMENT STANDARDS
- ISO 14971:2012 Application of Risk Management to Medical Devices
- ISO 13485:2003 Medical Devices Quality Management Systems Requirements For ● Regulatory purposes
### RISK ASSESSMENT AND MANAGEMENT SUMMARY H.
During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with a conventional PACS system previously released to the field.
For ICIS® View there are a total of 20 risks in the broadly acceptable region and two risks in the ALARP region. Zero risks were identified in the Not Acceptable Region. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk.
The benefits of ICIS® View are:
- Image and information review capabilities are enhanced over what is offered by traditional film: ●
- Immediate access to all available prior patient studies. о
- Time independent access to patient records and information. o
- Simultaneous access to multimodality, multi-departmental image review (a chest x-ray, o an, MRI, and CT for example).
- Improved resource utilization of both equipment and staff. ●
- Improved communication between local and geographically separated healthcare o providers promoting collaborative care.
- Authorized Users are able to share same patient record and contribute to that record. O
- Distribution of clinical data either through digital access, distribution or hardcopy print. .
The residual risk of each hazardous situation was analyzed and it was determined that the overall benefits to the patient outweigh the residual risks.
### CONCLUSIONS I.
Agfa's ICIS® View system has an Indications For Use statement similar to the predicate devices (K103785, K022292, & K111164). The statements have been combined and simplified. There are no changes to the intended use/indications of the device. Intended uses are the same. The devices have the same technological characteristics.
{9}------------------------------------------------
Both ICIS® View and the primary predicate device. Mobile MIM (K103785) are used to display medical images for diagnostic purposes of multiple modalities inlcuding MR and CT. Both provide wireless and portable access to medical images on select Apple products. Furthermore, ICIS® View and Mobile MIM (K103785) contain a statement that it is not intended to replace full diagnostic workstations and should only be used when there is no access to a workstation. ICIS® View and eUnity™ (K111164) are indicated to allow image manipulations and display lossy/lossless images. Both ICIS® View and predicate device, IMPAX Workstation (K02292) are indicated to display, review, transfer, store, and archive medical images. ICIS® View and all three predicate devices (K103785, K022292, & K111164) are not indicated for mammography and all the devices accept standard DICOM transfers.
The only difference of the new device and the primary predicate device, Mobile MIM (K103785), is the new device is server based software that connects to an existing PACS; whereas, Mobile MIM contains an App that connects to the existing PACS. However, the other two predicates, IMPAX Workstation (K022292) and eUnity™ (K111164) are also server based software that connects to the existing PACS. Differences in devices do not alter the intended diagnostic effect.
The devices have the same technological characteristics. The new device and the predicate devices (K103785, K022292, & K11164) are picture archiving and communication systems (PACS), Product Code LLZ. Agfa's ICIS® View system is substantially equivalent to the predicate devices (K103785, K022292, & K111164) in that it uses precisely the same technology to capture and display medical data. Descriptive characteristics and performance data are adequate to ensure equivalence.
Principles of operation and technological characteristics of the new and predicate devices are the same. There is no change to the intended use of the device vs. the predicate devices.
This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.