K103785 · Mim Software, Inc. · LLZ · Feb 4, 2011 · Radiology
Device Facts
Record ID
K103785
Device Name
MOBILE MIM
Applicant
Mim Software, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Mobile MIM software program is used for the registration, fusion, and/or display for diagnosis of medical images from only the following modalities: SPECT, PET, CT, and MRI. Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation. This device is not to be used for mammography.
Device Story
Mobile MIM is a software application providing wireless, portable access to medical images (SPECT, PET, CT, MRI). It retrieves patient data securely from a MIM workstation or server; images are losslessly compressed for network transfer. The software enables image registration, fusion, multi-planar reconstruction (MPR), maximum intensity projection (MIP), and measurement of distance and intensity values (e.g., Standardized Uptake Value). It provides tools for window/level, zoom/pan, and fusion blending. Used by clinicians in clinical environments when a full workstation is inaccessible. The device operates on off-the-shelf portable hardware (Apple iOS). Users view processed DICOM images to assist in diagnostic decision-making. Benefits include portable access to diagnostic-quality images, facilitating timely clinical review when stationary workstations are unavailable.
Clinical Evidence
Bench testing included display performance evaluation (contrast response, luminosity, resolution, noise) per IEC 62563-1 and TG18 guidelines. Clinical studies were conducted with radiologists using MR, CT, SPECT, and PET images under various environmental conditions to affirm diagnostic viewing capabilities. Software verification and validation testing confirmed performance requirements were met.
Technological Characteristics
Software-based medical imaging viewer for Apple iOS portable devices. Features include image registration, fusion, MPR, MIP, and SUV calculation. Connectivity via secure network to MIM workstation/server. Data transfer uses lossless compression. Calibration procedure required for display performance. Complies with IEC 62563-1 and TG18 guidelines for display quality.
Indications for Use
Indicated for the registration, fusion, and display of medical images (SPECT, PET, CT, MRI) for diagnostic purposes in patients. Intended for use by clinicians when access to a full workstation is unavailable. Contraindicated for mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one light and one dark, with a white circle in the center where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it.
FEB - 4 2011
K103785
Page 1 of 5
## 510(k) Summary of Safety and Effectiveness
(The following information is in conformance with 21 CFR 807.92)
#### Submitter:
MIM Software Inc. 25200 Chagrin Blvd. Suite 200 Cleveland, OH 44122
| Phone: | 216-455-0600 |
|------------------------|--------------------------|
| Fax: | 216-455-0601 |
| Contact Person: | Lynn Hanigan |
| Date Summary Prepared: | Jan 26, 2011 |
| Device Name | |
| Trade Name: | Mobile MIM |
| Common Name: | Medical Imaging Software |
| Trade Name: | Mobile MIM |
|----------------------|------------------------------------------|
| Common Name: | Medical Imaging Software |
| Classification Name: | System, Imaging Processing, Radiological |
#### Predicate Device
MIM Software Inc. (formerly MIMvista Corp.) K062163 MIMviewer 1.0
### Intended Use / Indications for Use
The Mobile MIM software program is used for the registration, fusion, and/or display for diagnosis of medical images from only the following modalities: SPECT, PET, CT, and MRI.
Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.
This device is not to be used for mammography.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one light and one dark, with a circle in the middle where they overlap. To the right of the squares is the text "mim" in lowercase letters, with the word "SOFTWARE" in smaller letters below.
### Device Description
The Mobile MIM software program is used for the registration, fusion, and/or display for diagnosis of medical images from only the following modalities: SPECT, PET, CT, and MRI.
Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.
The software is not to be used for mammography.
It includes the capability to measure distance and image intensity values such as Standardized Uptake Value, displays measurement lines, annotations and regions of interest, and provides window/level, zoom/pan, and fusion blending control functionality.
Mobile MIM retrieves patient image data securely via a network connection with a MIM workstation or server. Processed DICOM images from the workstation or server are losslessly compressed for network transfer and downloaded by Mobile MIM for display.
Mobile MIM operates on "off-the-shelf" portable hardware devices and is therefore subject to factors not typical for reading room workstations (e.g. screen size, environmental variability, network dependencies, etc.). It is therefore required that the user follows the operating instructions properly and utilizes the risk mitigation features in order to make decisions safely and effectively.
| ITEM | Mobile MIM | MIMviewer |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications For Use | The Mobile MIM software<br>program is used for the<br>registration, fusion, and/or<br>display for diagnosis of<br>medical images from only the<br>following modalities: SPECT,<br>PET, CT, and MRI.<br>Mobile MIM provides<br>wireless and portable access | MIMviewer is a software<br>package that aids the<br>physician in the diagnosis of<br>patients by means of medical<br>images. MIMviewer is used to<br>display, register and fuse<br>medical images from multiple<br>modalities.<br>The MIMviewer software |
| ITEM | Mobile MIM | MIMviewer |
| | to medical images. This<br>device is not intended to<br>replace full workstations and<br>should be used only when<br>there is no access to a<br>workstation.<br>This device is not to be used<br>for mammography. | program is used for the<br>registration, fusion and<br>display of medical images<br>from multi-modalities, such<br>as SPECT, PET, CT, and<br>MRI. MIMviewer provides<br>tools for image review,<br>manipulation, and analysis<br>that assist physicians both<br>inside and outside the medical<br>environment. |
| Receive, Store,<br>Retrieve, Display,<br>and Process Digital<br>Medical Images | Yes | Yes |
| Display of Clinical<br>Patient Data When<br>No Access to a<br>Workstation | Yes | Yes |
| Image Fusion | Yes | Yes |
| Multi-Planar<br>Reconstruction<br>(MPR) | Yes | Yes |
| Maximum Intensity<br>Projection (MIP) | Yes | Yes |
| Standardized Uptake<br>Value (SUV) | Yes | Yes |
| Distance<br>Measurements | Yes | Yes |
| ITEM | Mobile MIM | MIMviewer |
| Window/Level | Yes | Yes |
| Zoom/Pan | Yes | Yes |
| User Authentication | Yes | Yes |
| Modalities | SPECT, PET, CT, MRI | SPECT, PET, CT, MRI |
| Remote Handheld<br>Viewing Device | Yes | No |
| Operating Platform | Apple® iOS | Windows® 2000/XP<br>MacOS X® 10.4+<br>Linux® |
| Hardware<br>Requirements | Apple® iOS<br>handheld devices | Pentium® 4+<br>G4+ |
Device Comparison Table between new device and predicate:
25200 Chagrin Blvd. Suite 200 Cleveland, OH 44122
{2}------------------------------------------------
K103785
Page 3 of 5
・
25200 Chagrin Blvd. Suite 200
Cleveland, OH 44122
:
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a sequence of alphanumeric characters, specifically 'K163785'. The characters are handwritten and appear to be part of a code or identification number. The number is written in black ink on a white background.
Image /page/3/Picture/10 description: The image shows a logo for MIM Software. The logo consists of two overlapping squares, one light and one dark, with a circle in the center. To the right of the squares is the text "mim" in a bold, sans-serif font. Below the text "mim" is the word "SOFTWARE" in a smaller, sans-serif font.
Page 4 of 5
#### Substantial Equivalence
The comparison chart above provides evidence to facilitate the substantial equivalence determination between Mobile MIM and our chosen predicate device, MIMviewer (K062163)
The differences in the Indication Statements between Mobile MIM and MIMviewer describe specific restrictions on how Mobile MIM is to be used, given the hardware and portability differences between these two devices. Mobile MIM adds the explicate requirement that it should only be used when there is no access to a workstation, and that it is not to be used for mammography.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MIM Software. The logo consists of a square with rounded corners, with a circle inside of it. To the right of the square is the text "mim SOFTWARE" in a sans-serif font.
K103 785
The technological characteristics between Mobile MIM and MIMviewer are different. as the software operates on different hardware. These differences are addressed through the labeling and additional software features of Mobile MIM.
### Performance Data
MIM Software Inc. has conducted display performance testing using Mobile MIM software on various portable devices, both prior to and after utilizing the application's calibration procedure. Testing measured contrast response and evaluated test patterns for luminosity, resolution, and noise according to IEC 62563-1 and TG18 guidelines. All testing passed requirements following the Mobile MIM's calibration procedure.
MIM Software Inc. also performed multiple studies with qualified radiologists using a variety of modalities, specifically MRJ, CT, SPECT, and PET, under different environmental conditions. Results of these studies affirm the diagnostic image viewing capabilities of Mobile MIM when used as indicated.
Furthermore, MIM Software Inc. has conducted verification, validation, and functional testing on the Mobile MIM software. In all cases, the software passed its performance requirements and met specifications.
#### Conclusion
Therefore, from all evidence gathered, it is our belief that Mobile MIM provides a diagnostic viewer of medical images substantially equivalent to the MIMviewer software, with portable device characteristics and functionality.
25200 Chagrin Blvd. Suite 200 Cleveland, OH 44122
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Lynn Hanigan Quality Manager MIM Software 25200 Chargrin Blvd., Suite 200 CLEVELAND OH 44122
FEB - 4 201
Re: K103785
Trade/Device Name: Mobile MIM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 22, 2010 Received: December 27, 2010
Dear Ms. Hanigan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
.
{6}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary S Postel
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): TBD KIO3785
Device Name: Mobile MIM
Indications for Use:
The Mobile MIM software program is used for the registration, fusion, and/or display for diagnosis of medical images from only the following modalities: SPECT, PET, CT, and MRI.
Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.
This device is not to be used for mammography.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K163785
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.