The iMEC Series Patient Monitors, including iMEC8, iMEC10 and iMEC12, are intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), and carbon dioxide (CO2). All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following: C.O. monitoring is restricted to adult patients only, and PAWP monitoring, arrhythmia detection and ST Segment analysis are not intended for neonatal patients. The iPM Series Patient Monitors, including iPM 8, iPM 10 and iPM 12, are intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead, or 5-lead, or 12-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO2), and anesthetic gas (AG). All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following: C.O. monitoring is restricted to adult patients only, and PAWP monitoring, arrhythmia detection and ST Segment analysis are not intended for neonatal patients. 12-lead monitoring and AG monitoring are available for iPM 10 and iPM 12 Patient Monitors only. The BeneView T1 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead, or 5-lead, or 12-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), and carbon dioxide (CO2). All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following: C.O. monitoring is restricted to adult patients only, and PAWP monitoring, arrhythmia detection and ST Segment analysis are not intended for neonatal patients. The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
Device Story
iMEC, iPM, and BeneView T1 are compact patient monitors derived from the BeneView T Series platform. Devices accept inputs from various sensors (ECG leads, SpO2 probes, blood pressure cuffs, temperature probes, CO2/AG sensors). Signals are processed to display, store, and alarm on physiological parameters. Monitors feature dual ECG algorithm options: the established Mortara algorithm and the new Mindray Monitoring ECG Algorithm for ECG, arrhythmia, and ST segment analysis. Used in healthcare facilities by clinicians to support patient assessment and decision-making. Benefits include continuous monitoring of vital signs to detect clinical changes. Output is presented on integrated displays for real-time review and alarm notification.
Clinical Evidence
Bench testing only. Includes requirements specification review, hardware and software testing, code design/reviews, environmental EMC testing, safety testing, and performance validation. No clinical data presented.
Technological Characteristics
Compact patient monitors; modular parameter configuration. Connectivity includes ECG, SpO2, NIBP/IBP, Temp, CO2, and AG. Software includes Mindray Monitoring ECG Algorithm and Mortara ECG Algorithm. Compliance with recognized safety, performance, and EMC standards.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in healthcare facilities for monitoring physiological parameters including ECG, arrhythmia, ST segment, HR, Resp, Temp, SpO2, PR, NIBP, IBP, C.O., CO2, and AG. C.O. restricted to adults. PAWP, arrhythmia detection, and ST segment analysis not for neonates.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
## k123074
# 510(K) SUMMARY
JUL 0 3 2013
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: __
Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
### Contact Person:
Ms. Susan D. Goldstein-Falk Official Correspondent for Shenzhen Mindray Bio-Medical Electronics Co., Ltd mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 10021 IN CALIFORNIA: TEL: (949) 706-7456 FAX: (949) 706-7457 EMAIL: sgoldstein@mdiconsultants.com
## Alternate Only:
## Mr. Rong Liang
Product Approval Engineer of Technical Regulation Dept. Shenzhen Mindray Bio-Medical Electronics Co., Ltd Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P.R. China TEL: (86)-0755-8188 5637 FAX: (86)-0755-2658 2680 EMAIL: rongliang@mindray.com
EMIRATE. TOURIST, IMMIGRATION
Date Prepared:
Sep 21, 2012
{1}------------------------------------------------
Name of the devices:
●
Trade/Proprietary Name:
iMEC, iPM and BeneView TI Patient Monitors (including models iMEC8, iMEC10, iMEC12, iPM 8, iPM 10, iPM 12, and BeneView T1)
Common Name: Patient Monitor .
- . Classification
| 21 CFR 870.1025 | Arrhythmia detector and alarm<br>(including ST-segment measurement and alarm) | Class II |
|-----------------|-------------------------------------------------------------------------------|----------|
| 21 CFR 870.1025 | Detector and Alarm, Arrhythmia | Class II |
| 21 CFR 870.1025 | Monitor, ST Segment with Alarm | Class II |
| 21 CFR 870.2300 | Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
| 21 CFR 870.1130 | Non-Invasive blood pressure measurement System | Class II |
| 21 CFR 870.1110 | Blood pressure computer | Class II |
| 21 CFR 880.2910 | Clinical Electronic Thermometers –<br>Temperature Monitor with Probe | Class II |
| 21 CFR 870.2700 | Oximeter, Pulse | Class II |
| 21 CFR 870.2710 | Ear Oximeter, Pulse | Class II |
| 21 CFR 868.1400 | Carbon Dioxide Gas Analyzer | Class II |
| 21 CFR 868.1500 | Enflurane gas analyzer | Class II |
| 21 CFR 868.1620 | Halothane gas analyzer | Class II |
| 21 CFR 868.1700 | Nitrous Oxide gas analyzer | Class II |
| 21 CFR 868.1720 | Oxygen gas analyzer | Class II |
| 21 CFR 870.1435 | Single-function, preprogrammed diagnostic computer. | Class II |
## Legally Marketed Predicate Devices:
K092449. BeneView T Series patient monitors (Including Models BeneView T8, BeneView T6 and Bene View TS). Shenzhen Mindray Bio-medical Electronics Co., LTD
### Device Description:
The iMEC Series Patient Monitors (including iMEC10 and iMEC12), iPM Series Patient Monitors (including iPM 8, iPM 10 and iPM 12) and BeneView TI Patient Monitor are the developed new series based on the technical platform of the BeneView T Series patient monitors. The new series, much more compact and flexible with the appearance changes, are the sub-configuration models of the BeneView T Series patient monitors by reducing some parameters monitoring functions such as impedance cardiograph (ICG), bispectral index (BIS) and respiration mechanics (RM) monitoring.
{2}------------------------------------------------
And comparing with the cleared BeneView T Series Patient Monitors using Mortara ECG algorithm, the subject patient monitors adopt Mindray Monitoring ECG Algorithm as an optional ECG algorithm to implement ECG monitoring, arrhythmia detection, and ST Segment Analysis.
#### Statement of intended Use:
The iMEC Series Patient Monitors, including iMEC8, iMEC10 and iMEC12, are intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), and carbon dioxide (CO2). All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following: C.O. monitoring is restricted to adult patients only, and PAWP monitoring, arrhythmia detection and ST Segment analysis are not intended for neonatal patients.
The iPM Series Patient Monitors, including iPM 8, iPM 10 and iPM 12, are intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead, or 5-lead, or 12-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO2), and anesthetic gas (AG). All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following: C.O. monitoring is restricted to adult patients only, and PAWP monitoring, arrhythmia detection and ST Segment analysis are not intended for neonatal patients. 12-lead monitoring and AG monitoring are available for iPM 10 and iPM 12 Patient Monitors only.
The BeneView T1 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead, or 5-lead, or 12-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), and carbon dioxide (CO2). All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following: C.O. monitoring is restricted to adult patients only, and PAWP monitoring, arrhythmia detection and ST Segment analysis are not intended for neonatal patients.
**6-4**
{3}------------------------------------------------
## k123074
The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
### Technology:
The iMEC Series Patient Monitors (including iMEC10 and iMEC12), iPM Series Patient Monitors (including iPM 8, iPM 10 and iPM 12) and BeneView TI Patient Monitor are substantially equivalent to the predicate devices BeneView T Series Patient Monitors (K092449).
#### Test Summary:
The iMEC Series Patient Monitors (including iMEC8, iMEC10 and iMEC12), iPM Series Patient Monitors (including iPM 8, iPM 10 and iPM 12) and BeneView TI Patient Monitor comply with the recognized safety, performance and electromagnetic compatibility standards. A risk analysis has been developed to identify potential hazards and document . the mitigation of the hazards. Mindray's product development process required that the following activities be completed during the development of those patient monitors:
- Requirements specification review Hardware and Software testing Code design and code reviews Environmental EMC testing Safety testing Performance testing Hardware and Software validation
#### Conclusion:
The results of all testing demonstrate that the iMEC Series Patient Monitors, iPM Series Patient Monitors and BeneView T1 Patient Monitor are as safe, as effective, and perform as well as the predicate devices.
MINDRAY PROPRIETARY Patient Monitors
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center . WO66-G609 Silver Spring, MD 20993-002
Image /page/4/Picture/3 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### July 3, 2013
Shenzhen Mindray Bio-Medical Electronics Co., Ltd Susan Goldstein-Falk 55 Northern Blvd., Suite 200 Great Neck, NY 11021 US
Re:
Trade/Device Name: iMEC, iPM and BeneView T1 Patient Monitors (including models iMEC8 iMEC10,iMEC12, iPM 8, iPM 10, iPM 12, and BeneView T1)
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class II Product Code: MHX Dated: May 24, 2013 Received: June 6, 2013
#### Dear Susan Goldstein-Falk:
K123074
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRHs) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MA Hilleman
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123074
Device Name: iMEC, iPM and BeneView T1 Patient Monitors
Indications for Use:
#### iMEC .
The iMEC Series Patient Monitors, including iMEC8, iMEC10 and iMEC12, are intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO>), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), and carbon dioxide (CO2).
All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following:
- C.O. monitoring is restricted to adult patients only;
- PAWP monitoring is not intended for neonatal patients;
- The Mindray ECG Algorithm arrhythmia detection is intended for adult and pediatric patients, and the Mindray ECG Algorithm ST Segment analysis is intended for adult patients only.
## • iPM
The iPM Series Patient Monitors, including iPM 8, iPM 10 and iPM 12, are intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead, or 12-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO₂), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO2), and anesthetic gas (AG).
All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following:
- C.O. monitoring is restricted to adult patients only;
- PAWP monitoring is not intended for neonatal patients;
- The Mortara ECG Algorithm arrhythmia detection and ST Segment analysis is intended for adult and pediatric patients. The Mindray ECG Algorithm arrhythmia detection is intended for adult and pediatric patients, and the Mindray ECG Algorithm ST Segment analysis is intended for adult patients only.
- 12-lead monitoring and AG monitoring are available for iPM 10 and iPM 12 Patient Monitors only.
## Page 1 of 2
{7}------------------------------------------------
#### BeneView T1 .
The BeneView T1 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead, or 5-lead, or 12-lead selectable), arrhythmia detection and ST Segment analysis, heart rate (HR), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), and carbon dioxide (CO2).
All the parameters can be applied to single adult, pediatric and neonatal patients with the exception of the following:
- C.O. monitoring is restricted to adult patients only;
- PAWP monitoring is not intended for neonatal patients;
- The Mortara ECG Algorithm arrhythmia detection and ST Segment analysis is intended for adult and pediatric patients. The Mindray ECG Algorithm arrhythmia detection is intended for Adult and Pediatric patients, and the Mindray ECG Algorithm ST Seqment analysis is intended for adult patients only.
The monitors are to be used in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
Prescription Use X 1 Fosonption 0601 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.