NP Cements / NP Cements System are intended to be used for the fixation of artificial joint prostheses to the host bone.
Device Story
NP Cements and NP Cements System are radiopaque PMMA bone cements used for fixation of artificial joint prostheses to host bone. NP Cements are supplied as powder and liquid components for manual mixing. NP Cements System integrates powder and liquid storage, mixing, and extrusion into a single closed syringe-like device. Both versions are available in low and high viscosity formulations. Devices are sterile, disposable, and intended for use by surgeons in clinical settings. The cement acts as a mechanical interface between the prosthesis and bone. The system version improves handling and delivery via vacuum-assisted mixing and integrated extrusion. Clinical benefit is the secure fixation of orthopedic implants.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included chemical composition, thermal properties, and mechanical properties (modulus, flexural properties, static compression/bending, fatigue, fracture toughness) conducted per FDA Class II Special Controls Guidance. Biocompatibility evaluated per ISO 10993. Sterilization and shelf-life validated per international standards.
Indicated for fixation of artificial joint prostheses to host bone.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 5, 2015
Cossington Limited % Diane Mandell Horwitz, Ph.D. Mandell Horwitz Consultants, LLC 2995 Steven Martin Drive Fairfax, Virginia 22031
Re: K143134
Trade/Device Name: NP Cements / NP Cements System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: LOD Dated: January 22, 2015 Received: January 22, 2015
Dear Dr. Horwitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Diane Mandell Horwitz, Ph.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K143134
### Device Name NP Cements / NP Cements System
### Indications for Use (Describe)
NP Cements / NP Cements System are intended to be used for the fixation of artificial joint prostheses to the host bone.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### GENERAL INFORMATION 1
#### Submitter and Owner of the 510(k) 1.1
Cossington Ltd 1 Princeton Mews Kingston Upon Thames Surrey KT2 6PT United Kingdom
#### Official Correspondent 1.2
Diane Mandell Horwitz, PhD, RAC Regulatory Affairs Consultants Mandell Horwitz Consultants, LLC 2995 Steven Martin Drive Fairfax, Virginia 20031
Telephone: (703) 307-2921 Fax: (703) 242-1117 E-mail: dmh@mandellhorwitzconsulting.com
#### 1.3 Devices Subject of this 510(k)
NP Cements - Low and High Viscosity NP Cements System - Low and High Viscosity
#### 1.4 Date of Preparation
October 28, 2014
#### NAME OF THE DEVICE AND CLASSIFICATION INFORMATION 2
This traditional 510(k) has been submitted for the following devices.
- NP Cements - Low and High Viscosity
- NP Cements System Low and High Viscosity ●
#### 2.1 Trade/Proprietary Name
As of the date of this application, the Sponsor has not yet developed a trade name for the new bone cements.
{4}------------------------------------------------
#### 2.2 Common/Usual Name
PMMA bone cement for orthopedics PMMA bone cement for orthopedics within a syringe-like device system
#### 2.3 Classification Information
Classification Name: Polymethylmethacrylate (PMMA) Bone Cement
Classification Regulation: 21 CFR § 888.3027
Regulatory Class: Class II
LOD – Bone Cement Product Code:
Panel: Orthopedic
#### PREDICATE DEVICE 3
The predicate device is as follows:
- Tecres Cemex Isoplastic, which was cleared through 510(k) application K021715.
#### DEVICE DESCRIPTION ব
The NP Cements and NP Cements System are PMMA, radiopaque bone cements, containing color additives to impart a green color, designed for the fixation of prosthesis to the living bone. The devices contain the same individual chemical constituents, but are supplied in different ways:
- 1. NP Cements are a traditional bone cement product. The liquid is contained in a vial and the powder in a sachet; these components are packaged in unitary blister with Tyvek lid, which is placed in an aluminum bag. The components are manually mixed immediately prior to use.
- 2. NP Cements System stores the powder and liquid components separately within a closed syringe-like device that serve as both mixing chamber and extrusion device. The device is provided with a vacuum tube and a cannula with the pusher rod. The product is packaged in a double blister, the external blister is sealed with a Tyvek lid, and then placed in an aluminum bag.
The devices are sold disposable and sterile. The NP Cements are available in a low and high viscosity version. The NP Cements System are also available in a low and high viscosity version.
{5}------------------------------------------------
# 5 INDICATIONS FOR USE
Below is the indication for use.
NP Cements /NP Cements System are intended to be used for the fixation of artificial joint prostheses to the host bone.
# 6 COMPARISON OF THE INTENDED USE WITH THE PREDICATE DEVICE
The NP Cements and the NP Cements System have the same indication for use statements (other than the identification of the trade name) as the predicate device, and thus have the same intended use, as shown in Table 1.
| Device | Indication for Use |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Cossington<br>NP Cement /<br>NP Cement System | NP Cements /NP Cements System are intended to be used for the fixation of artificial joint prostheses to the host bone. |
| Tecres Cemex Isoplastic<br>(K021715) | Cemex bone cement is intended to be used for the fixation of artificial joint prostheses to the host bone. |
| Table 1: Comparison of the Indication Statement with the Predicate Device | |
|---------------------------------------------------------------------------|--|
|---------------------------------------------------------------------------|--|
### COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE 7 PREDICATE DEVICE
The NP Cements and NP Cements System share many of the same technological characteristics compared to the predicate Cemex Isoplastic, including important considerations such as the same materials, and mechanical and chemical-physical performances. There are, however, some differences in the shelf life, method of sterilization of the powder component, the presence of colored pigments and how the device is supplied to the surgeon for the system version. These comparisons are summarized in Table 2.
| Characteristics | Cossington<br>NP Cement HV/LV<br>NP Cement HV/LV System | Tecres<br>Cemex Isoplastic<br>K021715 | Comparison |
|--------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------|
| Main Components | Polymethylmethacrylate<br>(PMMA)<br>Methylmethacrylate (MMA)<br>Barium Sulphate | Polymethylmethacrylate<br>(PMMA)<br>Methylmethacrylate (MMA)<br>Barium Sulphate | Same |
| Other Components | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Same |
| Color Additives | Pigments (FD&C Blue No.1 and<br>FD&C Yellow No.5 ) | None | Different |
| Mixing/Application | Manual<br>Syringe-like device (System<br>version) | Manual | Same for Manual:<br>Different for<br>system |
Table 2: Comparison of the Technological Characteristics with the Predicate Device
{6}------------------------------------------------
| Characteristics | Cossington<br>NP Cement HV/LV<br>NP Cement HV/LV System | Tecres<br>Cemex Isoplastic<br>K021715 | Comparison |
|------------------------------------------------|-----------------------------------------------------------|---------------------------------------|---------------------------------------------|
| Powder Sterilization<br>Method | Gamma-ray irradiation (manual)<br>Ethylene Oxide (system) | Ethylene Oxide | Same for manual;<br>different for<br>system |
| Sterility Assurance<br>Level (SAL) -<br>Powder | 10-6 | 10-6 | Same |
| Liquid Sterilization<br>Method | Filtration | Filtration | Same |
| SAL – Liquid | 10-3 | 10-3 | Same |
| Shelf Life | 2 years | 5 years | Different |
#### PERFORMANCE DATA 8
This 510(k) submission provided performance data to establish the substantial equivalence of the new bone cements to the predicate bone cement. Performance testing was conducted in accordance with the "FDA Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement; Guidance for Industry and FDA" dated July 17, 2002. The following is a summary of the performance data.
Sterilization and Shelf Life: The devices are sterilized using standard methods and the sterilization cycles have been validated following international standards. The shelf life of the devices has been established in ongoing stability studies.
Biocompatibility: Biocompatibility testing of the NP Cements and NP Cements System has been performed to show the device materials are safe, biocompatible and suitable for their intended use. Both ISO 10993 and FDA Draft Guidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" have been taken into account to evaluate the biocompatibility of the device materials.
Performance Testing (Chemical, Material and Mechanical): Performance testing was performed to characterize the bone cements in accordance with special controls guidance document. This testing included the following:
- Mixing and application characteristics (e.g., dough time, setting time) ●
- Chemical composition (e.g., residuals, molecular weight and polymer structure, glass transition temperature)
- Thermal properties (e.g., polymerization temperature)
- Mechanical properties (e.g., modulus and flexural properties, static compression and ● bending, fatigue testing, fracture toughness and viscoelasticity)
The performance data demonstrate that the new devices are substantially equivalent to the predicate device, and meet the requirements of the Special Controls Guidance document.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.