K021715 · Exactech, Inc. · LOD · Jun 12, 2002 · Orthopedic
Device Facts
Record ID
K021715
Device Name
CEMEX RX, CEMEX ISOPLASTIC & CEMEX XL
Applicant
Exactech, Inc.
Product Code
LOD · Orthopedic
Decision Date
Jun 12, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CEMEX bone cement is intended to be used for the fixation of artificial joint prostheses to the host bone.
Device Story
Bone cement system consisting of liquid monomer (methylmethacrylate, N-N dimethyl p-toluidine, hydroquinone) and dry powder (polymethylmethacrylate, barium sulphate, benzoyl peroxide). Used by surgeons in orthopaedic procedures to fixate metal/plastic joint prostheses to host bone. Powder and liquid mixed in 3:1 ratio; varying formulations (RX, ISOPLASTIC, XL) provide different viscosities for specific surgical application techniques. Liquid packaged in amber glass vials; powder in paper/PE sachets. Sterilized via ethylene oxide (powder) and membrane filtration (liquid). Provides mechanical fixation of implants to bone, aiding joint stability and patient mobility.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on chemical composition, physical properties, and comparison to legally marketed predicate devices.
Technological Characteristics
PMMA bone cement. Components: liquid (methylmethacrylate, N-N dimethyl p-toluidine, hydroquinone) and powder (polymethylmethacrylate, barium sulphate, benzoyl peroxide). 3:1 powder-to-liquid mixing ratio. Viscosity varies by model (RX, ISOPLASTIC, XL). Packaging: amber glass vial (liquid), paper/PE sachet (powder), Tyvek blister pack. Sterilization: ethylene oxide (powder, SAL 10^-6), membrane filtration (liquid, SAL 10^-3).
Indications for Use
Indicated for fixation of prostheses to bone in orthopaedic musculoskeletal procedures for patients with osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, traumatic arthritis, congenital deformities, avascular necrosis, post-traumatic degenerative joint problems, sickle cell anemia, osteoporosis, collagen disease, and revision of previous arthroplasty. Contraindicated in patients with active/incompletely treated infection at the site, loss of musculature or neuromuscular compromise rendering surgery unjustifiable, or allergy to cement components.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
JUN 1 2 2002
# Exactech, Inc. Tecres Cemex® System Bone Cement
page 1 of 2
K021715
# Special 510(k) Summary of Safety and Effectiveness
| Trade Names: | Cemex ISOPLASTIC Bone Cement<br>Cemex RX Bone Cement<br>Cemex XL Bone Cement |
|----------------------|------------------------------------------------------------------------------|
| Common Name: | Bone Cement |
| Classification Name: | Polymethylmethacrylate (PMMA)<br>Bone Cement |
# Legally Marketed Devices for Substantial Equivalence Comparison:
| Model | Manufacturer |
|--------------|--------------------------------------------------------|
| Cemex System | Tecres, S.p.A. (#K000943) |
| CMW3 | CMW Laboratories<br>(Distributed by Dow Corning Wright |
| Simplex P | Howmedica |
| Palacos R | Merck<br>(Distributed by Smith & Nephew) |
| Dough-Type | Zimmer |
## Device Description:
## INTENDED USE
CEMEX RX, CEMEX ISOPLASTIC and CEMEX XL bone cements are intended to be used for the fixation of plastic and metal joint prostheses to host bone.
## INDICATIONS FOR USE
CEMEX bone cement is indicated for the fixation of prostheses to bone in orthopaedic musculoskeletal procedures for osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, tramatic arthritis, congenital deformities, avascular necrosis, post-traumatic degenerative problems of the joint, sickle cell anemia, osteoporosis, collagen disease and for the revision of previous arthroplasty procedures.
> Section 4 Page 2 of 3
{1}------------------------------------------------
# Exactech, Inc. Tecres Cemex® System Bone Cement
page 2 of 2
K021715
# Special 510(k) Summary of Safety and Effectiveness
# CONTRAINDICATIONS
CEMEX bone cement is contraindicated in the presence of active or incompletely treated infection which could involve the site where the cement is to be applied. CEMEX bone cement is contraindicated where the loss of musculature or neuromuscular compromise in the affected limb would render the surgical procedure unjustifiable. CEMEX bone cement is contraindicated in patients who are allergic to any of its components.
# GENERAL DESCRIPTION - Substantial Equivalency Information
The chemical constituents in CEMEX RX, ISOPLASTIC, and XL bone cements are identical to those in the predicate CEMEX SYSTEM cement (#K000943). The liquid component contains methylmethacrylate, N-N dimethyl p-toluidine, and hydroquinone. The dry powder component contains polymethylmethacrylate, barium sulphate and benzoyl peroxide. Like the predicate CEMEX SYSTEM Bone Cement (#K000943), the ratio between the powder and liquid components of the proposed cements is 3:1.
The proportion of each ingredient in the ISOPLASTIC and RX models varies slightly from the predicate model. The XL bone cement formulation is identical to that of CEMEX SYSTEM (#K000943). The differences result in a range of viscosities to accommodate various application techniques.
Cemex XL. RX and ISO bone cements are also similar to other competitive cement products including "Simplex P" by Howmedica, "CMW 3" by Dow Corning Wright, "Dough" by Zimmer, and "Palacos R" by Smith & Nephew. Several of the predicate devices are indicated for the fixation of pathological fractures but Cemex bone cements are not indicated for this application.
## PACKAGING
The liquid monomer component is contained in a trumpet shaped amber glass vial. The powder component is contained in a paper polyethylene (PE) film sachet. The product is packaged in unitary blister pack with Tyvek® lids. The outer packaging is a heavy weight cardboard box.
## STERILITY ASSURANCE
The powdered component is sterilized by ethylene oxide (EO) to a Sterility Assurance Level (SAL) of 10°. The liquid component is sterilized by a membrane filtration technique to a SAL of 10-3.
> Section 4 Page 3 of 3
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 1 2 2002
Ms. Lisa Simpson Regulatory Representative Exactech® 2320 NW 66th Court Gainesville, Fl 32653
Re: K021715
Trade Name: Tecres Bone Cement RX, ISOPLASTIC, XL Regulation Number: 21 CFR 888.3027 Regulation Name: PolymethyImethacrylate (PMMA) bone cement Regulatory Class: II Product Code: LOD Dated: May 22, 2002 Received: May 23, 2002
Dear Ms. Simpson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Lisa Simpson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Mach N Millkern
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Tecres Cemex Bone Cement
Indications for Use
page iofl
ON1715 510(k) Number:
. Device Names:
Cemex Rx Bone Cement Cemex XL Bone Cement Cemex Isoplastic Bone Cement
## INTENDED USE
CEMEX bone cement is intended to be used for the fixation of artificial joint prostheses to the host bone.
#### INDICATIONS FOR USE
CEMEX bone cement is indicated for the fixation of protheses to bone in orthopaedic musculoskeletal procedures for osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, tramatic arthritis, congenital deformities, avascular necrosis, post-traumatic degenerative problems of the joint, sickle cell anemia, osteoporosis, collagen disease and for the revision of previous arthroplasty procedures.
## CONTRAINDICATIONS
CEMEX bone cement is contraindicated in the presence of active or incompletely treated infection which could involve the site where the cement is to be applied.
CEMEX bone cement is contraindicated where the loss of musculature or neuromuscular compromise in the affected limb would render the surgical procedure unjustifiable.
CEMEX bone cement is contraindicated in patients who are allergic to any of its components
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | or | Over the Counter Use |
|------------------|----|----------------------|
|------------------|----|----------------------|
for Mark A Millbern
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
| 510(k) Number | K021715 |
|---------------|--------------------------|
| | Section 3<br>Page 1 of 1 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.