K131699 · Mega Electronics , Ltd. · DXH · Nov 26, 2013 · Cardiovascular
Device Facts
Record ID
K131699
Device Name
EMOTION ECG MOBILE
Applicant
Mega Electronics , Ltd.
Product Code
DXH · Cardiovascular
Decision Date
Nov 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The eMotion ECG sensor is a wearable, portable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research. The eMotion ECG Mobile is intended for use in clinical and non-clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters. The eMotion ECG Mobile does not provide any automatic analysis or diagnosis.
Device Story
eMotion ECG Mobile is a telemetry system for remote monitoring of chronic patients. Input: physiological data (ECG, blood pressure, weight, pulse oximetry) from external plug-in devices via Bluetooth. Transformation: mobile application transmits data over mobile networks to a central Health Gateway server using a secured connection. Output: data visualization via Web Monitor on a PC for healthcare professionals. Usage: clinical and non-clinical settings; operated by patients (data collection) and clinicians (data review). Benefit: enables remote monitoring and evaluation of patient health parameters by healthcare providers. Device performs no automatic analysis or diagnosis.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including system A/D conversion, sampling rate, signal frequency, sensitivity, noise, CMRR, and signal range verification. Performance of commercial off-the-shelf (OTS) plug-in devices was also verified.
Technological Characteristics
Wearable, portable, battery-operated (3.7V Li-ion) ECG recorder/transmitter. Connectivity: Bluetooth (to plug-ins) and mobile network (to server). Sampling rates: 100-1000 Hz. 3-lead patient cable. Materials: Latex-free. Standards: ISO 10993-1 (biocompatibility), EN 60601-2-25 (safety/performance). Software: PEMS and web-based platform. Non-sterile, reusable.
Indications for Use
Indicated for chronic patients requiring remote monitoring of health parameters including ECG, blood pressure, weight, and pulse oximetry in clinical or non-clinical settings. No specific age or gender restrictions. Contraindications not specified.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
### EMERGO 9 GROUP
### K131699 b 1/6
### Section 5 - 510(k) Summary
#### 1. Submission Sponsor
Mega Electronics Ltd Pioneerinkatu 6 KUOPIO FI-70800 Finland Phone: +358 17 581 7700 Fax: +358 17 580 0978 Contact: Aki Tiihonen, Operations Director Email: aki.tiihonen@megaemg.com
#### 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: André Kindsvater, Senior Consultant, QA/RA Email: project.management@emergogroup.com
#### 3. Date Prepared
May 20th, 2013
### 4. Device Identification
| Trade/Proprietary Name: | eMotion ECG Mobile |
|----------------------------|-----------------------------------------------------------|
| Common/Usual Name: | Digital Ambulatory Monitor |
| Classification Name: | Transmitters And Receivers, Electrocardiograph, Telephone |
| Classification Regulation: | 21 CFR 870.2920 |
| Product Code: | DXH |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
#### 5. Predicate Devices
Heartrak Smart ECAT, 510(k) Number: K083535 Card Guard PMP4 Medical Web Center, 510(k) Number: K050940
{1}------------------------------------------------
## EMERGO GROUP
KI31699 P 2/6
#### Device Description 6.
The eMotion ECG Mobile is a mobile device, PC and Internet based telemetry solution for the ambulant monitoring of the plug-in device data of chronic patients via a mobile network. Plug-in devices can be ECG devices, blood pressure monitors, weighing scales, etc. The device reads data from the plug-in device via a Bluetooth connection. The application in the mobile device sends the data to a server (Health Gateway) over mobile networks using the secured connection.
Data can be viewed from the Health Gateway server using the Web Monitor. Monitoring is performed using PC application that reads data from the server over the internet using the secured connection.
Figure 5-1. eMotion ECG Mobile concept
#### 7. Intended Use
The eMotion ECG sensor is a wearable, portable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research.
The eMotion ECG Mobile is intended for use in clinical and non-clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters.
{2}------------------------------------------------
·
# EMERGO GEROUP
KI31699 b 3/6
The eMotion ECG Mobile does not provide any automatic analysis or diagnosis.
#### 8. Comparison of Technological Characteristics
The following table compares the eMotion ECG Mobile to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | Mega Electronics Ltd | Universal Medical Inc. | CARD GUARD SCIENTIFIC<br>SURVIVAL, LTD |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | eMotion ECG | Heartrak Smart ECAT | Card Guard PMP4<br>Medical Web Center |
| 510(k) Number | TBD | K083535 | K050940 |
| Product Code | DXH | DXH | DXH |
| Regulation<br>Number | 21 CFR 870.2920 | 21 CFR 870.2920 | 21 CFR 870.2920 |
| Regulation<br>Name | Transmitters And<br>Receivers,<br>Electrocardiograph,<br>Telephone | Transmitters And<br>Receivers,<br>Electrocardiograph,<br>Telephone | Transmitters And<br>Receivers,<br>Electrocardiograph,<br>Telephone |
| Indications for<br>Use | The eMotion ECG sensor is<br>a wearable, portable,<br>externally applied,<br>electrocardiograph<br>recorder and transmitter<br>for the purpose of health<br>monitoring, biofeedback<br>and scientific research.<br>The eMotion ECG Mobile is<br>intended for use in clinical<br>and non-clinical settings to<br>collect and transmit health<br>parameters to healthcare<br>professionals for<br>monitoring and evaluation.<br>Health parameters are<br>collected from a variety of<br>commercially available,<br>external plug-in devices<br>such as ECG sensors,<br>Weight Scales, Blood | Heartrak Smart ECAT is a<br>wireless ambulatory,<br>multi-channel, continuous<br>ECG event recorder with<br>embedded arrhythmia<br>detection algorithms.<br>Heartrak Smart ECAT<br>registers symptomatic and<br>asymptomatic cardiac<br>event triggered by a<br>patient manually or auto-<br>triggered by embedded<br>arrhythmia detection<br>algorithms. Using wireless<br>technology, Heartrak<br>Smart ECAT, when placed<br>with range of a compatible<br>RF receiver, uploads<br>recorded ECG parameter<br>data to receiver. When<br>data upload is complete | The PMP4 Medical Web<br>Center is a Software<br>application intended for<br>supporting remote<br>monitoring of<br>Electrocardiographic<br>(ECG), Spirometric,<br>Fetal/Maternal, Blood<br>Pressure, Heart Rate,<br>Blood Glucose, Blood<br>Oxygen Saturation, Body<br>Weight and optionally<br>other patients' vital signs<br>and parameters.<br>The data is received from<br>transducers/monitors,<br>which are external to the<br>system. |
#### Table SA - Comparison of Characteristics
{3}------------------------------------------------
#### Mega Electronics LTD Traditional 510(k) Premarket Submission eMotion ECG Mobile
# EMERGO GROUP
| Manufacturer | Mega Electronics Ltd | Universal Medical Inc. | CARD GUARD SCIENTIFIC<br>SURVIVAL, LTD |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Trade Name | eMotion ECG<br>Pressure Meters and Pulse<br>Oximeters.<br>The eMotion ECG Mobile<br>does not provide any<br>automatic analysis or<br>diagnosis. | Heartrak Smart ECAT<br>data can be reviewed and<br>analyzed at a physician's<br>office, clinic or monitoring<br>center. | Card Guard PMP4<br>Medical Web Center |
| Overall Design | PEMS and Software | PEMS | Software |
| Sterile | non-sterile | non-sterile | N/A |
| Single-Use | no | no | No |
| Battery<br>Operated | Re-chargeable 3.7 V Li-ion<br>battery | Internal Li-Ion<br>rechargeable battery 3.6V | N/A |
| Data<br>Transmission | Bluetooth & Mobile Net | Bluetooth & Mobile Net | N/A |
| Sampling rate | Selectable 100, 125, 250,<br>500 or 1000 Hz | 205 Hz | N/A |
| AC Powered | no | no | N/A |
| Patient Cable | 3-lead | 3-lead | N/A |
| Latex Free | yes | yes | N/A |
| Data<br>transmission | via mobile device | RF within 10 m | via PDA, Cellular Phone,<br>iTV, PC |
| Web-based<br>Medical Center<br>Platform | Yes | N/A | Yes |
| Complies with<br>ISO 10993-1 | Yes | Yes | N/A |
| Electrical Safety<br>Testing Passed | Yes | Yes | N/A |
#### 9. Non-Clinical Performance Data
The eMotion ECG Mobile has been fully verified and validated following written test protocols to demonstrate that the design meets the requirements and performs as intended. The test results including pass/fail determination are documented in the corresponding test reports.
The device was also tested against the recognized consensus standard EN 60601-2-25:1999 Basic Safety and Essential Performance of Electrocardiographs.
The proper functioning of the applicable plug in devices was also verified.
The device and the applicable plug-in's passed all tests successfully.
{4}------------------------------------------------
EMERGO 9 GROUP
K131699 b 5/6
All required software testing was completed as part of the software verification and validation. Please refer to Section 016 Software.
#### Summary of Performance Testing Result
All the specified performance tests have been passed successfully.
| Test | Pass / fail criteria | Results | |
|------|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1 | System A/D<br>Conversion | 14 bit | Passed |
| 2 | Sampling rate | 1000 Hz | Passed |
| 3 | IP Class | IP20 | Passed |
| 4 | Signal frequency<br>band | ECG: 1 Hz ... 30 Hz | Passed |
| 5 | System sensitivity | 1µV / bit (peak to peak) or 0.2µV / bit<br>(peak to peak) switchable<br>ECG: 1.33 µV / bit (peak to peak) | Passed |
| 6 | System noise | 1 µV RMS | Passed |
| 7 | CMRR | 90 dB minimum<br>104 dB typical (type tested) | Passed |
| 8 | Signal range | 14 bit: +/- 8192µV (peak to peak) | Passed |
| 9 | Accelerometer | 10-bit +/- 8g mode, typical sensitivity 64 bits/g<br>Output data rate: 250 Hz<br>Accuracy: +/- 5% at 1g | Passed |
| 10 | ECG Waveform<br>shape | ECG Complex - Recognizable<br>Ventricular Fibrillation - Recognizable<br>30, 60, 120, 180 and 240 BPM +/- 2 %<br>Sine wave 10Hz and 40 Hz +/- 2 %, shape a<br>clear sine wave | Passed |
| 11 | Sensor power | Re-chargeable Li-Po or Li-ion 140mAh, 3.7V | Passed |
| 12 | Battery life | 300 full re-charge cycles (80%) | Passed |
| 13 | Operating time | Online with Bluetooth: ca. 4h | Passed |
#### Table 5B – Performance Testing Summary - ECG Sensor
The performance of the commercial OTS plug-ins has been tested and all tests passed successfully.
{5}------------------------------------------------
Mega Electronics LTD Traditional 510(k) Premarket Submission eMotion ECG Mobile
# EMERGO GROUP
K131699 P 6/6
| K-Number | Device | Results |
|----------|--------------------------------------------|---------|
| K043217 | UA-767PBT Digital Blood Pressure Monitor | Passed |
| K061822 | HEM-780N3 Automatic Blood Pressure Monitor | Passed |
| K102350 | 3150 WristOX2 | Passed |
| none | UC-321 PBT C40 Weight Scale | Passed |
| none | HBF-206IT Weight Scale | Passed |
#### Table 5C - Performance Testing - OTS Plug-ins
The eMotion ECG Mobile meets all the requirements for overall design, biocompatibility, and electrical safety confirms that the output meets the design inputs and specifications. The eMotion ECG Mobile passed all testing and supports the claims of substantial equivalence and safe operation.
The eMotion ECG Mobile complicable voluntary standards for biocompatibility. The device passed all the testing in accordance with national and international standards.
#### 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the eMotion ECG Mobile and the predicate devices do not raise any questions regarding its safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the eMotion ECG Mobile is substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, sterilization, biocompatibility, performance characteristics, and intended use.
The eMotion ECG Mobile, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized representation of an eagle with its wings spread. The eagle is facing to the right and has three wavy lines emanating from its body.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 26, 2013
Mega Electronics Ltd. c/o Mr. Andre Kindsvater Senior Consultant QA/RA Prinsessegracht 20 2514 AP, The Hague Netherlands
Re: K 13 l 699
> Trade/Device Name: Emotion ECG Mobile Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter & Receiver Regulatory Class: II (two) Product Code: DXH Dated: October 22, 2013 Received: October 24, 2013
Dear Mr. Andre Kindsvater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
OwenP.Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### K131699
### Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: eMotion ECG Mobile •
Indications for Use:
The eMotion ECG sensor is a wearable, portable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research.
The eMotion ECG Mobile is intended for use in clinical and non-clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters.
The eMotion ECG Mobile does not provide any automatic analysis or diagnosis.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by
Owen P. Faris -S
Date: 2013.11.26
13:18:19-05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.