K143003 · Ipulse Limited · OHT · Jun 18, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143003
Device Name
iPulse Smooth Skin Gold Hair Removal System
Applicant
Ipulse Limited
Product Code
OHT · General, Plastic Surgery
Decision Date
Jun 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The iPulse SmoothSkin Gold Hair Removal System is an over the counter device intended for the removal of unwanted hair. The iPulse SmoothSkin Gold Hair Removal System is indicated for the removal of unwanted hair. The iPulse Smoothskin Gold is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
iPulse SmoothSkin Gold is an intense pulsed light (IPL) system for over-the-counter hair removal. Device consists of a handset (capacitor, charger, control electronics, optics, lamp, filter, reflector, light pipe, trigger, and skin tone/proximity sensors) and an external power supply. User operates device by placing handset on skin; optical sensors measure skin tone to determine appropriate energy output. Device delivers xenon arc flashlamp pulses to tissue to target hair follicles. Output settings are adjusted based on measured skin tone. Clinical efficacy is determined by hair count reduction at 6 and 12 months. Benefits include long-term hair regrowth reduction. Device is intended for home use by consumers.
Clinical Evidence
Prospective clinical study of 50 subjects receiving 12 weekly treatments. Primary endpoint: hair reduction at 6 and 12 months post-treatment. Results: 43.9% hair reduction at 6 months (n=50) and 36.0% at 12 months (n=33). 66.7% of subjects achieved >30% hair reduction at 12 months. Analysis used paired sample t-tests on averaged independent assessor counts.
Technological Characteristics
IPL system using xenon arc flashlamp. Wavelength range: 510-1100nm. Pulse duration: 2-10ms. Energy density: 3-6J/cm². Spot size: 3cm². Features integrated optical skin tone sensor for continuous measurement. Powered by external 24V DC supply (110V/230V AC input). Handset includes capacitor, control electronics, and optical assembly.
Indications for Use
Indicated for removal of unwanted hair and permanent reduction in hair regrowth in adults. Permanent reduction defined as long-term, stable reduction in hair count at 6, 9, and 12 months post-treatment regimen.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that forms the shape of a bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2015
iPulse Limited Dr. Michael Kiernan Chief Scientific Officer Technium 2, Kings Road, The Docks Swansea, Wales SA1 8PJ, Wales, United Kingdom
Re: K143003 Trade Name: iPulse SmoothSkin Gold Hair Removal System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT, ONX, GEX Dated: May 15, 2015 Received: May 18, 2015
Dear Dr. Kiernan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143003
Device Name
iPulse SmoothSkin Gold Hair Removal System
#### Indications for Use (Describe)
The iPulse SmoothSkin Gold Hair Removal System is indicated for the removal of unwanted hair. The iPulse Smoothskin Gold is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 3.0 510(K) SUMMARY
| Submission Date: | 15th October 2014 | | | PREDICATE DEVICES | | | DEVICE |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------|---------------------------------------------------------------------------------|
| Submitter Information | | | Device Name | iPulse Hair Removal<br>System | Shaser HRS2 IPL | mē | iPulse Smooth Skin<br>Gold |
| Company Name:<br>Company Address: | iPulse Limited.<br>Technium 2, Kings Road, Swansea, Wales, UK SA1 8PJ | | K Number | K130315 | K120080 | K131649 | K143003 |
| Contact Person: | Dr Mike Kiernan<br>UK +44 1792 485 519<br>mkiernan@cyden.com | | Manufacturer | CyDen Ltd | ShaserInc | Syneron Beauty Ltd | iPulse Ltd |
| Device Information | | | Energy Medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp |
| Trade Name: | iPulse SmoothSkin Gold Hair Removal System | | Wavelength<br>Range | 530-1100nm | 650-1100nm* | 550-1200nm | 510-1100nm |
| Common Name: | Light based over the counter hair removal system | | Pulse Duration | Variable -<br>Single pulse<br>25milliseconds.<br>to<br>Double Pulse 20ms<br>on, 60 ms off. | 30 milliseconds | ~5ms | 2ms to 10ms |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and dermatology | | Energy Density | 7-10J/cm² | 9J/cm² | Up to 9J/cm² | 3-6J/cm² |
| Device Class: | 21 CFR 878.4810 | | Spot Size | 3cm² (1.3cm by 2.4<br>cm) | 2cm² (1cm by 2cm) | 3.3cm2 (3.3cm by<br>1.0cm) | 3cm² |
| Predicate Devices: | iPulse Smoothskin Hair Removal System<br>K130315<br>CyDen Limited | | Delivery Device | Direct Illumination<br>To Tissue | Direct Illumination to<br>Tissue | Direct Illumination to<br>Tissue | Direct Illumination to<br>Tissue |
| | The Shaser HRS2 IPL Hair Removal System<br>K120080<br>ShaserInc | | Pulsing Control | Finger switch | Finger switch | Finger Switch | Finger switch |
| | mē<br>K131649<br>Syneron Beauty Ltd | | Skin Tone Sensor | Optical<br>Measurement<br>Removable from<br>Base Unit.<br><br>Sensor moveable to<br>treatment site | Optical<br>Measurement<br>Integral to Base Unit<br><br>Sensor fixed in base<br>unit, treatment site<br>moved to sensor | Electrical Contact | Optical<br>Measurement<br>Integral to device.<br><br>Continuous<br>measurement. |
| Specific<br>Indications for<br>Use | The iPulse Hair<br>Removal System is<br>an Over-the-counter<br>device intended for<br>the removal of<br>unwanted hair. | The Shaser HRS2 is<br>an Over-the-counter<br>device intended for<br>the removal of<br>unwanted hair. It is<br>also intended for<br>permanent reduction<br>in unwanted hair.<br>Permanent hair<br>reduction is defined<br>as the long-term<br>stable reduction in<br>the numbers of hairs<br>regrowing when<br>measured at 6,9 and<br>12 months after the<br>completion of a<br>treatment regimen. | The me is an over-<br>the-counter device<br>indicated for the<br>removal of unwanted<br>hair. Me is also<br>intended for<br>permanent in hair<br>growth following an<br>initial treatment<br>regimen with or<br>without maintenance<br>when measured at<br>6,9 and 12 months. | The iPulse<br>SmoothSkin Gold<br>Hair Removal System<br>is indicated for the<br>removal of unwanted<br>hair. The iPulse<br>Smoothskin Gold is<br>also indicated for the<br>permanent reduction<br>in hair regrowth,<br>defined as the long-<br>term, stable<br>reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9 and<br>12 months after the<br>completion of a<br>treatment regime. | | | |
| Table 1: Predicate Device Comparison | | | | | | | |
A comparison of the key technological characteristics of the iPulse Smoothskin Gold System to the predicate devices is provided in Table 1.
# CONFIDENTIAL
{4}------------------------------------------------
{5}------------------------------------------------
#### Device Description
The iPulse SmoothSkin Hair Removal System is an intense pulsed light (IPL) system consisting of:
Handset - contained within the handset is the Capacitor, Capacitor Charger, Control electronics, Optics (Lamp, Filter, Reflector, Light Pipe), Trigger mechanism and Skin Tone / proximity Sensors (STS);
External Power Supply - used to convert the electricity from the mains supply (either 110V or 230V, 50/60Hz) to a lower DC value, typically 24V. This power supply unit is an "off-the-shelf" component which meets all the relevant electrical safety standards.
# Intended Use
The iPulse SmoothSkin Gold Hair Removal System is an over the counter device intended for the removal of unwanted hair.
#### Indications for Use
The iPulse SmoothSkin Gold Hair Removal System is indicated for the removal of unwanted hair. The iPulse Smoothskin Gold is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
# Performance Data - Non Clinical
Nonclinical, clinical and usability testing has been completed on the iPulse SmoothSkin Hair Removal System. Nonclinical testing included biocompatibility, electrical safety and software testing.
Usability testing was completed in 24 subjects to evaluate device human factors and label comprehension.
# Performance Data - Clinical
Clinical testing was conducted in 50 subjects at a single treatment center. Each subject received 12 weekly treatments to the selected body location. Prior to treatment the site was shaved and any hair trimmings removed. The treatment was performed by a suitable operator to ensure consistency of coverage thus enabling accurate efficacy determination. The device output setting for each subject was determined by the measured skin tone at the actual treatment site.
{6}------------------------------------------------
Hair counts were performed on a fixed area located at the center of the treatment site using templates and skin landmarks. All treatment sites were photographed prior to each treatment and at the review point. All subjects were shaved one week prior to both baseline and the 6 and 12 month review points thus ensuring hair count consistency.
Hair counts in the selected area on each subject were determined by 2 independent assessors and the results averaged. The percentage difference in hair count pre and post treatment was calculated and a Paired Sample t-test used to compare means. Only cases where the pre and post treatment hair count areas could be clearly identified on the clinical photographs were analysed.
The efficacy results are provided in Table 2.
| | SmoothSkin Gold |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Number of Subjects at 6 Months<br>Post Treatment | 50 |
| Hair Reduction at 6 Months Post<br>Treatment | 43.9% |
| Number of Subjects at 12 Months<br>Post Treatment | 33 |
| Hair Reduction at 12 Months Post<br>Treatment | 36.0% |
| % of Subjects met success (>30%<br>hair reduction) on all body areas at<br>12 months post-treatment. Subject<br>Success is defined as greater than<br>30% hair reduction at all treatment<br>sites at 12 months.* | 66.7% |
| * The SmoothSkin Gold clinical study only used single treatment sites on<br>individual subjects. | |
| <b>Table 2: Clinical Efficacy Results</b> | |
# Substantial Equivalence
The iPulse SmoothSkin Gold Hair Removal System has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the iPulse Smoothskin Gold Hair Removal System and the listed predicate devices do not raise any issues of safety or efficacy. Performance data supports that the device is as safe and as effective as the predicate devices for its intended use. Therefore, the iPulse SmoothSkin Gold Hair Removal System may be found substantially equivalent to its predicate devices.
# CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.