K131649 · Syneron Beauty , Ltd. · ONF · Dec 11, 2013 · General, Plastic Surgery
Device Facts
Record ID
K131649
Device Name
ME
Applicant
Syneron Beauty , Ltd.
Product Code
ONF · General, Plastic Surgery
Decision Date
Dec 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The me is an over-the-counter device intended for the removal of unwanted hair. Me is also intended for permanent reduction in hair growth following an initial treatment regimen with or without maintenance when measured at 6, 9, and 12 months.
Device Story
The me is a compact, over-the-counter hair removal system consisting of a base unit with power supply and a handheld applicator with an integrated air-cooling system. It utilizes Intense Pulse Light (IPL) technology to heat hair roots, inhibiting growth. The device is intended for home use by the patient. Users perform an initial regimen of 7 weekly treatments, with optional monthly maintenance treatments thereafter. Clinical efficacy is determined by hair count reduction measured by independent evaluators comparing images taken before and after treatment. The device provides a non-invasive method for permanent hair reduction, with clinical success defined as >30% hair reduction at 12 months. Healthcare providers may be consulted if severe pain or persistent adverse effects occur.
Clinical Evidence
Prospective clinical study of 87 healthy participants (ages 18-65, skin types I-VI, dark hair). Treatment regimen: 7 weekly sessions. Primary endpoint: hair reduction at 6, 9, and 12 months post-treatment. Results: At 12 months, 45% of subjects achieved >30% hair reduction (no maintenance) and 59% achieved >30% reduction (with maintenance). Body site success rates were 54% (no maintenance) and 72% (with maintenance). No serious adverse events reported; minor effects included redness (0.6%), edema (0.06%), and transient pain.
Technological Characteristics
Compact IPL system; output 2-4 J/cm²; handheld applicator with air-cooling; base unit power supply. No changes to materials, sterilization, or electrical safety from previously cleared versions (K123845, K121598).
Indications for Use
Indicated for removal of unwanted hair and permanent reduction in hair growth in healthy individuals aged 18-65 with dark hair and skin types I-VI.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Shaser Inc.'s IPL Hair Removal System (K120080; K103560)
Tria Beauty, Inc's Tria Laser Hair Removal System (K090820; K120737)
Submission Summary (Full Text)
{0}------------------------------------------------
K131649
DEC 11 2013
#### 510(k) SUMMARY
#### Syneron Beauty Ltd.'s mē
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
### Syneron Beauty Ltd.
Kochav Yokneam Bldg. Yokneam Industrial Zone P.O. Box 14 Yokneam Illit 20692 lsrael
Phone: +972 (4) 9098 700 Facsimile: +972 (4) 9098 701
Contact Person: Omri Hayet
Date Prepared: December 11, 2013
#### Name of Device and Name/Address of Sponsor
m હૈ
# Syneron Beauty Ltd. Kochav Yokneam Bldg. Yokneam Industrial Zone P.O. Box 14 Yokneam Illit 20692
Israel
#### Common or Usual Name
Light based hair removal system
#### Classification Name
ONF- Laser surgical instrument for use in general and plastic surgery and in dermatology, Class II, 21 CFR 878.4810
#### Predicate Devices
Syneron Beauty Ltd.'s me (K123845; K121598) Shaser Inc.'s IPL Hair Removal System (K120080; K103560) Tria Beauty, Inc's Tria Laser Hair Removal System (K090820; K 120737)
{1}------------------------------------------------
#### Intended Use / Indications for Use
The me is an over-the-counter device intended for the removal of unwanted hair. Me is also intended for permanent reduction in hair growth following an initial treatment regimen with or without maintenance when measured at 6, 9, and 12 months.
#### Device Description
The me device is a small over-the-counter, compact system comprised of a base unit assembly with power supply, and connected, via cable, to a handheld applicator with an air-cooling system. The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows.
#### Technological Characteristics
There are no technological differences between the me device presented in this submission and the previously cleared versions of the device (K123845, K121598). The device incorporates Intense Pulse Light (IPL) technology (output 2-4 J/cm²), like other OTC devices for hair removal.
The nonclinical evaluations of the me device were performed in support of clearance of the predicate me devices, and the information regarding sterilization, shelf life, biocompatibility, and electromagnetic compatibility and electrical safety remain unchanged.
#### Clinical Performance Data
In a medically supervised clinical study, the me was effective in the removal of unwanted hair, when used as directed. 87 healthy, dark haired participants of all skin types (I-VI) and between the ages of 18 to 65 years old, who were willing to refrain from any type of hair removal methods beyond shaving during the study duration, were enrolled. A treatment area (5x5 cm2) in the underarms, arms and legs were treated once a week for 7 treatments, over a period of 6 weeks, and were evaluated at 3 months after treatment also known as the "3-Month Group". Thereafter, 58 (54 females and 4 males with an average age of 34 years old) of these participants continued in long term clinical studies to evaluate permanent hair reduction at 6, 9 and 12 months with and without monthly maintenance treatments after the initial 7 treatments had been completed.
Table 1 presents the study data for subjects that did not receive monthly maintenance treatments after the initial regimen of 7 treatments ("No Maintenance" group), and Table 2 presents the data for subjects who received monthly treatment after the initial treatment period ("Initial and Maintenance Treatment" group). 72% of subjects were of light skin tone (Skin type I-IV) while the remainder were of darker skin tone (Skin Type V and VI). The "6-month Group" consisted of 40 participants. Each participant had a non-maintenance side and a maintenance side per body area (arms, legs and axilla) and monthly treatments were performed only on the maintenance side. Similarly, the "9-Month Group" consisted of 26 subjects and lastly the "12-Month Group" consisted of 44 subjects.
{2}------------------------------------------------
The end point for the long term clinical studies was the overall hair reduction achieved at 6, 9 and 12 months following a basic treatment regimen (7 weekly treatments) as measured by two independent evaluators through hair count on images taken from the treatment area that were subsequently averaged. These images were taken just before the first treatment and at subsequent follow ups within the study period of up to 1 year. The effects of performing monthly maintenance following the Basic Treatment regimen of 7 weekly treatments were evaluated in participants of the various study groups at 6, 9 and 12 months from the start of the study and are summarized in the table below. A subject was considered a success if all treated body parts achieved greater than 30% hair reduction at that time point. A body site is considered a success if the treated site achieved greater than 30% hair reduction at that time point.
There were no serious adverse events observed during the entire course of all study evaluations conducted and any minor transient effects such as skin irritation or redness were documented by medical personnel and are described below.
*Table 1 below represents subjects who were evaluated in at least one follow up time point. Non maintenance body sites received only the initial (basic) treatment (7 weekly treatments) with the me device and no additional treatment during follow-up assessments at up to 12 months after end of basic treatment.
Note: There were 40 participants at the 6 month visit, 26 participants at the 9 month visit and 44 participants at the 12 month visit. A total of 58 participants were evaluated in at least one follow up time point 6, 9 or 12 months. Subjects were evaluated at 2, 3, or all follow up time points.
| | Average Results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Initial (Basic) Treatment (7 weekly<br>treatments) | 54% reduction observed in 87 participants (143 sites) |
| 3-Month Group | 44% reduction observed in 87 participants (139 sites) |
| | Long Term (≥6 months) Follow-Up |
| 6-Month Group | 48% reduction (40 participants, 65 sites) |
| 9-Month Group | 45% reduction (26 participants, 44 sites) |
| 12-Month Group | 37% reduction (44 participants, 72 sites) |
| % subjects met success (>30% hair reduction)<br>on all body sites at 12 months post-treatment<br>Subject Success is defined as greater than<br>30% hair reduction at all treatment sites at 12<br>months. | 45% success observed in the 44 subjects with<br>long-term follow-up at 12 months |
| % body sites met success (>30% hair<br>reduction) at 12 months post-treatment<br>Body site success is defined as greater than<br>30% hair reduction on arm or leg or axilla at<br>12 months | 54% success observed in the 72 sites with long-<br>term follow-up at 12 months |
# Table 1: Hair Reduction Data with Initial Treatment Regimen (No Maintenance*) (Return for Follow Up at Variable Time Points 6 or 9 or 12 Months)
{3}------------------------------------------------
**Table 2 represents body sites where maintenance was performed throughout the study. Like Table 1 these subjects were evaluated in at least one follow up time point. For maintenance body sites, the subjects were assigned to perform maintenance treatments monthly after completing the initial (basic) treatment regimen.
Note: There were 40 participants at the 6 month visit, 26 participants at the 9 month visit and 44 participants at the 12 month visit. A total of 58 participants were evaluated in at least one follow up time point 6, 9 or 12 months. Subjects were evaluated at 2, 3, or all follow up time points.
| | Average Results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Initial (Basic) Treatment (7 weekly<br>treatments) | 57% reduction observed in 87 participants (143<br>sites) |
| 3-Month Group | 58% reduction observed in 87 participants (142<br>sites) |
| Long Term (≥6 months) Follow-Up | |
| 6-Month Group | 55% reduction (40 participants, 69 sites) |
| 9-Month Group | 56% reduction (26 participants, 46 sites) |
| 12-Month Group | 52% reduction (44 participants, 75 sites) |
| % subjects met success (>30% hair reduction)<br>on all body sites at 12 months post-treatment<br>Subject Success is defined as greater than<br>30% hair reduction at all treatment sites at 12<br>months. | 59% success observed in the 44 subjects with<br>long-term follow-up at 12 months |
| % body sites met success (>30% hair<br>reduction) at 12 months post-treatment<br>Body site success is defined as greater than<br>30% hair reduction on arm or leg or axilla at<br>12 months. | 72% success observed in the 75 sites with long-<br>term follow-up at 12 months |
| Table 2: Hair Reduction Data with Initial and Maintenance** Treatment (Return for | | | |
|-----------------------------------------------------------------------------------|--|--------------------------------------------------------|--|
| | | Follow Up at Variable Time Points 6 or 9 or 12 Months) | |
Tables 3 and 4 show data from the 21 of the 87 participants who completed non-maintenance (Table 3) and monthly maintenance (Table 4) treatments for up to 1 year continuously following their initial weekly basic treatments and returned for follow-up at 6, 9 and 12 months. As shown in the tables below, this subgroup of participants had slightly higher hair reduction at 12 months. Unlike Tables 1 and 2, the subjects in Tables 3 and 4 were evaluated at all follow up time points of 6, 9, and 12 months.
| Table 3: Hair Reduction Data with Initial Treatment (No Maintenance*) for Participants | |
|----------------------------------------------------------------------------------------|--|
| Returned for Follow Up at 6, 9 and 12 months | |
| | Average Results |
|---------------------------------------|------------------------------------------------------------|
| Basic Treatment (7 weekly treatments) | 54% reduction observed in 87 participants (143 body sites) |
| 3 Months after Initial Treatment | 44% reduction observed in 87 participants (139 body sites) |
{4}------------------------------------------------
| | Long Term (≥6 months) Follow-up |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| 6 Months after Initial Treatment | 44% reduction observed in 21 participants (37<br>body sites) |
| 9 Months after Initial Treatment | 45% reduction observed in 21 participants (37<br>body sites) |
| 12 Months after Initial Treatment | 42% reduction observed in 21 participants (37<br>body sites) |
| % subjects met success (>30% hair reduction)<br>on all body sites at 12 months post-treatment<br>Subject Success is defined as greater than<br>30% hair reduction at all treatment sites at 12<br>months. | 52% success observed in the 21 subjects with<br>long-term follow-ups to 12 months |
| % body sites met success (>30% hair<br>reduction) at 12 months post-treatment<br>Body site success is defined as greater than<br>30% hair reduction on arm or leg or axilla at<br>12 months. | 68% success observed in the 37 sites with long-<br>term follow-up at 12 months |
* Non maintenance body sites received only the initial (basic) treatment (7 weekly treatments) with the me device and no additional treatment during follow-up assessments at up to 12 months after end of basic treatment.
# Table 4: Hair Reduction Data with Initial and Maintenance Treatment** for Participants Returned for Follow Up at 6, 9 and 12 months
| | Average Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Basic Treatment (7 weekly treatments) | 57% reduction observed in 87 participants (143<br>body sites) |
| 3 Months after Initial Treatment | 58% reduction observed in 87 participants (142<br>body sites) |
| | Long Term (≥6 months) Follow-up |
| 6 Months after Initial Treatment | 55% reduction observed in 21 participants (40<br>body sites) |
| 9 Months after Initial Treatment | 59% reduction observed in 21 participants (40<br>body sites) |
| 12 Months after Initial Treatment | 60% reduction observed in 21 participants (40<br>body sites) |
| % subjects met success (>30% hair reduction) on<br>all body sites at 12 months post-treatment<br>Subject Success is defined as greater than 30%<br>hair reduction at all treatment sites at 12 months. | 62% success observed in the 21 subjects with<br>long-term follow-ups to 12 months |
| % body sites met success (>30% hair reduction)<br>at 12 months post-treatment<br>Body site success is defined as greater than 30%<br>hair reduction on arm or leg or axilla at 12 | 80% success observed in the 40 sites with long<br>term follow-up at 12 months |
{5}------------------------------------------------
| months. | |
|---------|--|
|---------|--|
**For maintenance body sites, the subjects were assigned to perform maintenance treatments monthly after completing the initial (basic) treatment regimen.
Actual results will vary from person to person and depending on body site being treated (arms, legs, or underarms). Hair re-growth may fluctuate (increase or decrease) over time, depending on the person, body area and whether the instructions were followed correctly. While permanent reduction of hair results can be achieved without monthly maintenance, it is recommended that user's perform monthly treatments following their basic treatments.
As noted above, there were no serious adverse events in the clinical study. In the clinical study, the majority (85%) of participants did not experience side effects with treatment. The most common side effect is redness after treatment that disappears within several hours without medical treatment. In addition, folliculitis may be observed in some cases.
The user may feel warmth, tingling, or itching. It is expected that the user may feel up to a moderate level of pain. If the pain is severe/intense or the pain persists for more than 24 hours after a treatment, consultation with a physician is recommended.
| Anticipated Effects on Skin<br>Appearance | Number of<br>Reports | Incidence (%) / of<br>3,122 treatments | % Subjects (n/102)* |
|-------------------------------------------------|----------------------|----------------------------------------|---------------------|
| Redness | 20 | 0.6% | 4% (4/102) |
| Pruritis/itching | 2 | 0.06% | 1% (1/102) |
| Edema | 2 | 0.06% | 1% (1/102) |
| Inflammation | 2 | 0.06% | 1% (1/102) |
| Acne | 2 | 0.06% | 2% (2/102) |
| Total Anticipated | 28 | 0.9% | 9% (9/102)* |
| Device-related<br>Adverse Events | | | |
| Infection | 0 | 0% | 0% (0/87) |
| Blister | 2 | 0.06% | 1% (1/102) |
| Hyper/hypopigmentation | 0 | 0% | 0% (0/87) |
| Scarring | 0 | 0% | 0% (0/87) |
| Severe pain | 11 | 0.3% | 2% (5/102) |
| Total Adverse Events | 13 | 0.4% | 6% (6/102) |
| Overall anticipated effects +<br>adverse events | 41 | 1% | 15% (15/102) |
In the clinical study, the following side effects of treatment were reported:
{6}------------------------------------------------
* Safety evaluations were reported for a total of 102 subjects including the 87 subjects included in the efficacy data for up to 3 months post-treatment and an additional 15 subjects included in the long-term studies for >6 months follow-up after treatment.
#### Substantial Equivalence
The me is substantially equivalent to the Company's own previously cleared me device (K123845; K121598), Shaser's IPL System (K120080; K103560), and Tria Beauty, Inc's Tria Laser Hair Removal System (K090820; K120737).
The me has the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate devices. As noted above, the only principal change implemented in this submission compared to the previously cleared device version is to expand the indications for use of the device to include permanent hair reduction, based on longer-term clinical data through 12 months. Clinical data substantiates permanent hair reduction of at least 30% measured at 12 months post-treatment compared to baseline. This indication encompasses the same long term data (12 months) as other previously cleared predicates devices such as the Shaser's IPL System (K120080; K103560) and the Tria Laser Hair Removal System (K090820; K 120737). No changes to the device specifications were required to support the expanded indications for use. Thus, there are no technological differences between the me device presented in this submission and the previously cleared versions of the device (K 123845, K121598). Clinical data supports expansion of the indications for use to include permanent hair reduction. Thus, the me is substantially equivalent.
#### Conclusion
Therefore, the nonclinical and clinical evaluations of the me device demonstrate that the device performs as intended, and is substantially equivalent to the predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or extensions reaching upwards. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 11, 2013
Syneron Beauty Ltd. c/o Hogan Lovells US LLP Janice M. Hogan 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K131649
Trade/Device Name: më Hair Removal System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONF Dated: November 07, 2013 Received: November 07, 2013
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{8}------------------------------------------------
Page 2 - Janice M. Hogan
.. . . . . . . . . . .
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Binita S. Asharaكرة
2013.12.11 17:34:24-05'00
Binita Ashar, MD, MBA, FACS Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131649
Device Name: Syneron me
Indications For Use: The mẽ Is an over-the-counter device intended for the removal of unwanted hair. Me is also intended for permanent reduction in hair growth following an initial treatment regimen with or without maintenance when measured at 6, 9, and 12 months.
| Prescription Use | AND/OR | Over-The-Counter Use | X |
|-----------------------------|--------|------------------------|---|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden
2013.12.11 15:00:46 -05'00'
| (Division Sign-Off) for BSA | |
|------------------------------|---------|
| Division of Surgical Devices | |
| 510(k) Number | K131649 |
Page 1 of __1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.