The CrossCountry™ Transbronchial Access Tool is to be utilized through a flexible endoscope with an extended working channel by physicians who are trained in endoscopic techniques to puncture the tracheobronchial wall and facilitate access of additional endobronchial tools for patients with endobronchial lesions, peripheral lung nodules, or lung masses.
Device Story
CrossCountry Transbronchial Access Tool is a flexible endobronchial device consisting of a wire and a catheter; used through a flexible endoscope with an extended working channel. Physician-operated in clinical settings to access lung lesions located outside the airways (without a bronchus sign). The wire features a sharp distal tip to puncture the tracheobronchial wall; the catheter is then advanced to mechanically dilate the puncture site. This creates a channel for subsequent insertion of other endoscopic tools to sample tissue. Benefits include enabling access to peripheral lung nodules or masses that are otherwise difficult to reach. The device is single-use and sterile.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (tensile, shelf life, simulated use, distribution, compatibility, dimensional) and a preclinical animal study comparing the CrossCountry tool to a control device. All acceptance criteria were met, confirming the device performs as intended.
Technological Characteristics
Components: wire and catheter. Materials: Copolyester, elastomer, 304 stainless steel braid, titanium, and nitinol. Dimensions: Catheter OD 2.01mm, Wire OD 0.66mm. Principle: Mechanical puncture and dilation. Connectivity: None. Sterilization: Ethylene oxide. Single-use.
Indications for Use
Indicated for patients with endobronchial lesions, peripheral lung nodules, or lung masses requiring transbronchial access for diagnostic or therapeutic tool placement.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
LungPoint Tools (LungPoint Sheath and LungPoint Dilation Balloon) (K131234)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2015
Covidien Ilc Deborah Fleetham Regulatory Affairs Manager 161 Cheshire Lane, Suite 100 Plymouth, MN 55441
Re: K142934
Trade/Device Name: CrossCountryTM Transbronchial Access Tool Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: II Product Code: EOO Dated: May 15, 2015 Received: May 18, 2015
Dear Ms. Fleetham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Teiashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
### 510(k) Number (if known): K142934
Device Name: The CrossCountry™ Transbronchial Access Tool
### Indications for Use:
The CrossCountry™ Transbronchial Access Tool is to be utilized through a flexible endoscope with an extended working channel by physicians who are trained in endoscopic techniques to puncture the tracheobronchial wall and facilitate access of additional endobronchial tools for patients with endobronchial lesions, peripheral lung nodules, or lung masses.
Prescription Use X AND / OR (Part 21 CFR 801 Subpart D)
Over-the-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
{3}------------------------------------------------
# Section 7 510(k) Summary
The 510(k) Summary for the CrossCountry™ transbronchial access tool is included on the following page.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a white cross in the center, followed by the word "COVIDIEN" in blue, sans-serif font. The logo is simple and modern, and the colors are clean and professional.
### 510(k) Summary Covidien llc. Traditional 510(k) CrossCountry™ Transbronchial Access Tool
#### I. Submitter
Covidien llc 161 Cheshire Lane Suite 100 Plymouth, MN 55441 U.S.A.
Contact Person: Deborah Fleetham, Manager Regulatory Affairs Ph: 763-210-4091 Fax: 763-210-4098 Email : deborah.fleetham@covidien.com Date Prepared : June 12, 2015:
#### II. DEVICE
| Trade Name : | CrossCountry™ Transbronchial Access Tool |
|----------------------|--------------------------------------------------|
| Common Name: | Endoscopic Tool |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number: | 21 CFR 874.4680 |
| Product code: | EOQ |
#### III. PREDICATE DEVICE(S)
| Primary Predicate: | LungPoint Tools (LungPoint Sheath and LungPoint Dilation<br>Balloon) K131234 | | |
|----------------------|------------------------------------------------------------------------------|--|--|
| Common Name: | Sheath and Dilation Balloon | | |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories | | |
| Regulation Number: | 21 CFR 874.4680 | | |
| Product code: | EOQ | | |
| Secondary Predicate: | Wang Transbronchial Aspiration Needle – K914181 | | |
| Common Name: | Transbronchial Needle | | |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories | | |
| Regulation Number: | 21 CFR 874.4680 | | |
| Product code: | EOQ | | |
Primary Predicate: LungPoint Tools (LungPoint Sheath and LungPoint Dilation Balloon) - K131234
{5}------------------------------------------------
Secondary Predicate: Wang Transbronchial Aspiration Needle - K914181
No additional reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The CrossCountry™ transbronchial access tool (CrossCountry Tool) is a flexible endobronchial tool made up of two components: a wire and a catheter. The wire is used inside of the catheter. The product is used through a flexible endoscope with an extended working channel. The CrossCountry tool is designed to puncture a hole in the tracheobronchial wall then, using the catheter to dilate the channel, allows for subsequent endoscopic tool placement. This allows for access to lesions without a bronchus sign (outside the airways).
The product is packaged in a Tyvek™1 pouch and sterilized with ethylene oxide.
#### V. INDICATION FOR USE
The CrossCountry™ transbronchial access tool is to be utilized through a flexible endoscope with an extended working channel by physicians who are trained in endoscopic techniques to puncture the tracheobronchial wall and facilitate access of additional endobronchial tools for patients with endobronchial lesions, peripheral lung nodules, or lung masses.
#### VI. SUMMARY OF CHARACTERISTICS COMPARED TO PREDICATE DEVICE
Transbronchial endoscopic techniques are the foundation for the CrossCountry tool and both the primary and secondary predicate. All of these tools are intended to access lung lesions. At a high level, the subject and predicate devices are based on the following technological elements:
- . All are introduced endoscopically and used to reach the target lung tissue
- Each device is inserted through a channel which can be a bronchoscope or other . working channel.
- . All are transient devices intended for short-term introduction through a naturally occurring orifice.
- The CrossCountry tool and the Wang Transbronchial Aspiration Needle have . sharp distal ends to puncture the tracheobronchial wall.
See table below for a detailed summary of the characteristics compared to the predicate device.
T Trademark of its respective owner
{6}------------------------------------------------
| Characteristic | CrossCountry Tool | Broncus Medical Inc.<br>Lungpoint Tools<br>(Primary Predicate)<br>K131234 | Wang Transbronchial<br>Aspiration Needle<br>(Secondary Predicate)<br>K914181 |
|-------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Device Classification | Class II | Class II | Class II |
| FDA Product Code | EOQ | EOQ | EOQ |
| Technological Characteristics | | | |
| Components | Wire and Catheter | LungPoint Sheath and<br>LungPoint Balloon | Transbronchial Aspiration<br>Needle |
| Single Use | Yes | Yes | Yes |
| Anatomical Location | Lung | Lung | Lung |
| Introduction into the<br>body | Endobronchial | Endobronchial | Endobronchial |
| Sharp Distal Tip | Yes - Wire | Not known | Yes - Needle |
| Radiopaque distal end | Yes | Yes | Yes |
| Dilation Ability | Yes | Yes | No |
| Dilation Mechanism | Mechanical | Pneumatic | None |
| Working Outer<br>Diameter | Catheter OD: 2.01mm<br>Wire OD: 0.66mm | Sheath - 2.65 mm<br>Balloon: 1 mm<br>(uninflated)<br>4 mm (inflated) | Tip OD: 1.9 mm |
| Working Length | Catheter: 109.9 cm<br>Wire: 116.1 cm | Sheath: 90 cm<br>Balloon: 97.5 cm | 130 cm |
| Tip Diameter | Catheter: Tapers down to<br>slightly larger than the<br>wire - Wire: 0.41mm | Not applicable | 22 gauge needle<br>(0.028 in / 0.71mm) |
| Material | | | |
| Working Outer<br>Material | Polymer – Copolyester<br>Elastomer<br>Braid - 304 Stainless<br>Steel | Not Known | PTFE |
| Wire Tip<br>Wire | Titanium<br>Nitinol | Not Known | Stainless Steel |
The following technological differences exist between the subject and predicate devices:
- The CrossCountry tool dilates using mechanical pressure while the LungPoint . Tools dilate using pneumatic pressure.
{7}------------------------------------------------
- . The CrossCountry tool incorporates a sharp wire tip to puncture the tracheobronchial wall while the LungPoint Tool must rely on a separate device to puncture the wall.
- The CrossCountry tool provides access for other endoscopic tools to sample tissue while the Wang transbronchial aspiration needle is able to sample tissue directly.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility Data:
Biocompatibility testing was completed in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993 "Biological Evaluation fo Medical Devices Part 1: Evaluation and Testing" May 1, 1995 and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" as recognized by FDA. The following tests were conducted:
- Cytotoxicity
- Sensitization ●
- . Intracutaneous Reactivity
# Design Verification and Validation Testing Data:
- Tensile Testing
- Shelf Life Testing
- Simulated Use Testing ●
- Distribution Testing ●
- Compatibility Testing ●
- . Dimensional Testing
# Animal Study Data:
A preclinical study was conducted using the CrossCountry access tool on one side of the lung and a control device on the opposite side of the lung. All of the acceptance criteria were met. This study showed the CrossCountry access tool design meets its intended use.
#### SUBSTANTIAL EQUIVALENCE CONCLUSION VIII.
Covidien Ilc has demonstrated that the proposed CrossCountry access tool is substantially equivalent to the LungPoint Tools and the Wang Transbronchial Aspiration Needle which are legally marketed for the same intended use. The non-clinical data and the invivo animal study provide adequate justification to demonstrate that the CrossCountry access tool should perform as intended in the specified use condition.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.